K Number
K100228

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-02-24

(29 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The VersaPulse P20 Laser System is intended for use in surgical procedures requiring open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue.

The VersaPulse P20 Ho:YAG (2.1 μm wavelength) is indicated for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: Arthroscopy, Urinary lithotripsy, Gynecological surgery, Gastroenterological surgery, Urology, ENT surgery, General surgery

Device Description

The VersaPulse P20 Laser System is a desktop 20 Watt surgical Holmium laser and is an improved model configuration of the 20 Watt VersaPulse PowerSuite Holmium Surgical Laser, [510 (k) K011703].

Main functional components :

  • A Laser console .
  • . Control and display panel
  • A fiber port for Laser delivery systems
  • . A covered foot switch
  • Operating software
  • A variety of fiber optic delivery devices with accessories .

The system uses the same type of laser delivery devices as the predicate.

AI/ML Overview

1. A table of acceptance criteria and the reported device performance

No explicit acceptance criteria or reported device performance metrics (e.g., sensitivity, specificity, accuracy) are provided in the document. The submission focuses on demonstrating substantial equivalence to a predicate device.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

Not applicable. This device is a laser system; the evaluation focuses on engineering, safety, and functional testing, not on clinical performance metrics derived from patient data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

Not applicable. Ground truth as typically understood in medical image analysis (e.g., expert consensus on diagnoses) is not relevant for this type of device.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

Not applicable.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. The device is a surgical laser, not an AI-powered diagnostic or assistive tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable. The "ground truth" for this device would be its adherence to engineering specifications and safety standards, confirmed through various types of technical and functional testing, rather than clinical ground truth from patient data.

8. The sample size for the training set

Not applicable. The device is a hardware system, not a machine learning model requiring a training set.

9. How the ground truth for the training set was established

Not applicable.


Study that proves the device meets the acceptance criteria:

The document states: "The appropriate testing, including safety, performance and functional testing, to determine substantial equivalence of the VersaPulse P20 Laser System has been conducted."

This statement indicates that a series of internal tests were performed by the manufacturer (Lumenis, Inc.) to ensure the VerasPulse P20 Laser System met the necessary safety, performance, and functional requirements. The goal of these tests was to demonstrate "substantial equivalence" to a predicate device (Holmium 20 Watt VersaPulse PowerSuite and Dual Wavelength Surgical Lasers, K011703).

The summary does not provide specific details on the methodologies, sample sizes, or results of these "safety, performance, and functional testing" procedures. Instead, the FDA's decision to clear the device (K100228) signifies their acceptance of the manufacturer's presented data and conclusion of substantial equivalence. This type of submission relies on the manufacturer's internal verification and validation activities rather than external clinical trials with patient data for performance evaluation in the context of diagnostic or AI devices.

{0}------------------------------------------------

K 10028

FEB 2 4 2010

Attachment 6 - 510(k) Summary of Safety and Effectiveness - VersaPulse P20 Laser System

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

SUBMITTER INFORMATION: 6.1.

Submitter's Name:Lumenis, Inc.
Address:3959 West 1820 SouthSalt Lake City, UT 84104
Contact:Douglas Stante
Phone:801-656-2620
Fax:801-565-2627
Date of Preparation:January 20, 2010

6.2. DEVICE NAME:

Trade Name(s):VersaPulse P20 Laser System
Common/Usual Name:powered laser surgical instrument
Classification Names:79 GEX
CFR Reference:21 CFR 878.4810 Laser surgical instrument for use ingeneral and plastic surgery and in dermatology

6.3. PREDICATE DEVICE:

The predicate device is the Holmium 20 Watt VersaPulse PowerSuite (Ho:Yag) and Dual Wavelength (Ho: Yag/ Nd: Yag) Surgical Lasers, K011703, August 29, 2001.

{1}------------------------------------------------

6.4. DEVICE DESCRIPTION:

The VersaPulse P20 Laser System is a desktop 20 Watt surgical Holmium laser and is an improved model configuration of the 20 Watt VersaPulse PowerSuite Holmium Surgical Laser, [510 (k) K011703].

Main functional components :

  • A Laser console .
  • . Control and display panel
  • A fiber port for Laser delivery systems
  • . A covered foot switch
  • Operating software
  • A variety of fiber optic delivery devices with accessories .

The system uses the same type of laser delivery devices as the predicate.

6.5. INTENDED USE:

The VersaPulse P20 Laser System is intended for use in surgical procedures requiring open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue.

TECHNOLOGICAL CHARACTERISTICS SUMMARY & SUBSTANTIAL EQUIVALENCE 6.6. STATEMENT:

The subject device, the VersaPulse P20 Laser System, has the same intended use, general design and same fundamental scientific technology as the Lumenis predicate device.

6.7. PERFORMANCE DATA SUMMARY:

The appropriate testing, including safety, performance and functional testing, to determine substantial equivalence of the VersaPulse P20 Laser System has been conducted.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of two main elements: a circular seal on the left and a stylized symbol on the right. The circular seal contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. The symbol on the right is a stylized representation of a human figure, possibly meant to evoke a sense of care and well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

FEB 2 4 2010

Lumenis, Inc. % Mr. Douglas Stante Global Director Regulatory and Quality Systems Compliance 3959 West 1820 South Salt Lake City, Utah 84104

Re: K100228

Trade/Device Name: VersaPulse P20 Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: January 25, 2010 Received: February 01, 2010

Dear Mr. Stante:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the iddications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contractuality warranties. We remind you; however, that device labeling must be truthful and not mist leading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device it the found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements aft the Act

{3}------------------------------------------------

Page 2 - Mr. Douglas Stante

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 53.1-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21.CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment 2 - Indications for Use Statement

510(k) Number (if known)K 100 228
Device NameVersaPulse P20
Indications for UseThe VersaPulse P20 Ho:YAG (2.1 μm wavelength) is indicated for use in surgical procedures involving open, laparoscopic and endoscopic ablation, vaporization, excision, incision, and coagulation of soft tissue in medical specialties including: ArthroscopyUrinary lithotripsyGynecological surgeryGastroenterological surgeryUrologyENT surgeryGeneral surgery
Prescription Use (Per 21 CFR 801. 109)X OR Over-The-Counter Use_________________

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K100228

ﻳﺔ، ﻓ

·

ﻟﺴﻴﺔ

・・・

:28

:: :

። .

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.