K Number
K100090

Validate with FDA (Live)

Device Name
IMPLANTABLE CLIP
Manufacturer
Date Cleared
2010-02-02

(21 days)

Product Code
Regulation Number
878.4300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GENICON implantable clips are intended for use during procedures for the purpose of approximating soft tissue, closing off vessels and other structures in order to stop bleeding or to connect internal tissues to aid in healing.

Device Description

The ligation clips are composed exclusively of titanium, and are supplied as sterile products in various sizes (mini-micro, micro, smallmedium, medium, medium-large, large and extra large) with six clips per disposable holder. The titiaum used meets all requirements for A.S.T.M. specifications F-67 95 "Unalloyed Titanium for Surgical Implant Applications", Grade I and ISO 5832-2-93 and 10993 "Implants for Surgery - Metallic Materials - Part 2: Unalloyed Titanium".

AI/ML Overview

The provided text is a 510(k) summary for the GENICON Titanium Ligation Clip. It describes the device, its intended use, and substantial equivalence to a predicate device, but it does not contain information about a study that proves the device meets specific acceptance criteria in the way a diagnostic AI/ML device would.

Instead, the summary focuses on:

  • Material Composition: Stating the device is made of titanium meeting specific ASTM and ISO standards.
  • Substantial Equivalence: Claiming the device is substantially equivalent in material and function to a predicate device (Ethicon K830503).
  • Safety and Efficacy: Asserting that titanium is "well recognized as being safe and effective for long term implantation" and that millions of titanium clips applied yearly "attest to the wide acceptance of this method of hemostatic and ligation control."
  • Compliance with Standards: Stating that devices are manufactured according to GMP, AAMI, ASTM, and ISO 13485 requirements.

Therefore, many of the requested items related to an AI/ML device performance study cannot be answered from this document.

Here's an analysis based on the information available in the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Implicit)Reported Device Performance
Material meets specified biocompatibility/strength standards.Titanium used meets ASTM F-67 95, Grade I, and ISO 5832-2-93 and 10993.
Substantially equivalent to predicate device (Ethicon K830503)Material composition and disposable holder material are "substantially equivalent." Functionally, clips are "substantially equivalent."
Safe and effective for its intended use (ligation, hemostasis)."Titanium is well recognized as being safe and effective for long term implantation. Millions of titanium clips are applied yearly and since inception in the 1960's attest to the wide acceptance of this method of hemostatic and ligation control."
Manufacturing qualityDevices are manufactured according to Good Manufacturing Practices (G.M.P.), Association for Advancement of Medical Instrumentation (A.A.M.I.) and American Society for Testing and Materials (A.S.T.M.) requirements and applicable Harmonized Standards ISO 13485 (This refers to the manufacturing process, not clinical performance data from a specific study).

2. Sample size used for the test set and the data provenance

Not applicable. This document describes a medical device (titanium ligation clip), not an AI/ML diagnostic or predictive model that would use a "test set" in this context. The "safety and efficacy information" refers to general acceptance of titanium clips over time, not a specific clinical study with a test set.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. This is not an AI/ML device's performance study.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. This is not an AI/ML device's performance study.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI/ML device, nor is an MRMC study described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

The document implicitly relies on:

  • Material Standards: ASTM and ISO specifications for titanium.
  • Historical Clinical Use/Outcomes: "Millions of titanium clips are applied yearly...attest to the wide acceptance of this method of hemostatic and ligation control."
  • Substantial Equivalence: Ground truth for equivalence is based on comparison to a legally marketed predicate device, likely involving material properties, design, and intended use.

8. The sample size for the training set

Not applicable. This is not an AI/ML device.

9. How the ground truth for the training set was established

Not applicable. This is not an AI/ML device.

{0}------------------------------------------------

K100090

FEB - 2 2010

510(k) Summary Statement

Submitters Name: GENICON 6869 Sta Point Court Suite #114 Winter Park, FL 32792 Ph: 1.321.297.1189 Fx: 1.407.679.0880

Contact Name: Melanie Murphy

Name of Device: Titanium Ligation Clip

Safety & Effectiveness Data Summary:

Classification Name: Clip, Implantable Common/Usual Name: Ligation Clip Proprietary Name: n/a at this time

Classification: Class II

Implantable Clip:#79 FZP
Hemostatic Clip:#79 MCH Reg. #878.4300 - 36-2 - 1. 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 7 " " " " " " " " " " " " " " " " " " " " " " " " " "

Performance Standards: Devices are manufactured according to Good Manufacturing Practices (G.M.P.), Association for Advancement of Medical Instrumentation (A.A.M.I.) and American Society for Testing and Materials (A.S.T.M.) requirements and applicable Harmonized Standards ISO 13485.

Material Composition: A.S.T.M F-67 95, Grade I Titanium I.S.O. #5832-2-93, Grade I Titanium

Intended Use: The GENICON implantable clips are intended for use during procedures for the purpose of approximating soft tissue, closing off vessels and other structures in order to stop bleeding or to connect internal tissues to aid in healing.

Device Description: The ligation clips are composed exclusively of titanium, and are supplied as sterile products in various sizes (mini-micro, micro, smallmedium, medium, medium-large, large and extra large) with six clips per disposable holder. The titiaum used meets all requirements for A.S.T.M. specifications F-67 95 "Unalloyed Titanium for Surgical Implant Applications", Grade I and ISO 5832-2-93 and 10993 "Implants for Surgery - Metallic Materials - Part 2: Unalloyed Titanium".

Predicate Devices: Ethicon K830503

{1}------------------------------------------------

Comparison of Technological Characteristics: The titanium clip material and disposable holder material substantially equivalent to the predicate device. In function, the clips are the substantially equivalent to the predicate device as well.

ﻣﺎ ﺟ

(

Safety and Efficacy Information: Titanium is well recognized as being safe and effective for long term implantation. Millions of titanium clips are applied yearly and since inception in the 1960's attest to the wide acceptance of this method of hemostatic and ligation control.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". Inside the circle is an abstract emblem resembling an eagle or a stylized human figure, composed of three curved lines. The emblem is positioned to the right of the text.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

FEB - 2 2010

Genicon % Mr. Gary W. Haberland President 6869 Stapoint Court #114 Winter Park, Florida 32792

Re: K100090

Trade/Device Name: Genicon Ligation Clip Regulation Number: 21 CFR 878,4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP, MCH Dated: January 08, 2010 Received: January 12, 2010

Dear Mr. Haberland:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Mr. Gary W. Haberland

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.lyfm

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

12,100090 510(k):

Section 4 - Indications for Use

·Device Name: GENICON Ligation Clip

Indications for Use:

The GENICON implantable clips are intended for use during procedures for the purpose of approximating soft tissue, closing off vessels and other structures in order to stop bleeding or to connect internal tissues to aid in healing. -

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

Page 1 of

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

00090

Daniel Keene for MXM
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number

§ 878.4300 Implantable clip.

(a)
Identification. An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.(b)
Classification. Class II.