SUREFIT DUAL DISPERSIVE ELECTRODE

K100047 · Conmed Corporation · GEI · May 14, 2010 · General, Plastic Surgery

Device Facts

Record IDK100047
Device NameSUREFIT DUAL DISPERSIVE ELECTRODE
ApplicantConmed Corporation
Product CodeGEI · General, Plastic Surgery
Decision DateMay 14, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2

Intended Use

The CONMED® Surefit™ Dual Dispersive Electrodes is intended for use in surgical procedures in which Electrosurgery equipment with contact quality monitoring system is used.

Device Story

The CONMED® Surefit™ Dual Dispersive Electrode is a single-use, non-sterile dispersive return electrode with a pre-attached cord. It functions as a passive component in an electrosurgical circuit, providing a return path for RF energy from the patient back to the electrosurgical generator. The device features a conductive area covered with a soft, conformable hydro-gel adhesive and a non-conductive border adhesive to isolate the conductive area from surgical fluids. It is used in surgical settings where electrosurgical equipment with contact quality monitoring systems is employed. The electrode is applied to the patient's skin by clinical staff; the generator monitors contact quality to ensure safety. By providing a return path, the device enables the safe delivery of electrosurgical energy for cutting and coagulation, reducing the risk of patient burns at the return site.

Clinical Evidence

Bench testing only. Conformance to IEC 60601-2-2:2009 (Medical Electrical Equipment) and ISO 10993 (Biological Evaluation of Medical Devices, including cytotoxicity, irritation, and delayed-type hypersensitivity) was demonstrated.

Technological Characteristics

Single-use, non-sterile dispersive electrode with pre-attached cord. Conductive hydro-gel adhesive surface with non-conductive border. Complies with IEC 60601-2-2:2009 and ISO 10993-1, 5, and 10. Passive device; no internal software or active electronics.

Indications for Use

Indicated for use with electrosurgical generators equipped with contact quality monitoring systems (e.g., REM, ARM, NESSY) during electrosurgery for all patient populations, provided there is sufficient surface area to ensure full skin contact.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Conmed Electrosurgery. The logo consists of a stylized graphic to the left of the word "CONMED" in all caps. Below the word "CONMED" is the word "ELECTROSURGERY" in all caps. The graphic is a square with a curved shape inside of it. Klooch # 510k Summary of Safety and Effectiveness CONMED® Surefit™ Dual Dispersive Electrode This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92. ConMed Electrosurgery is hereby submitting the 510(s) Summary of Safety and Effectiveness for the 510(k) Number _______________________________________________________________________________________________________________________________ as of January 6, 2010. ## A. Submitter ConMed Electrosurgery 14603 E. Fremont Ave. Centennial, Colorado 80112 MAY 1 4 2010 ## B. Company Contact Shawn Riedel Vice President, Quality Assurance and Regulatory Affairs 14603 E. Fremont Ave. Centennial, Colorado 80112 Phone: 303-269-8224 Fax: 303-699-9854 e-mail: shawn riedel@es.conmed.com ## C. Device Name | Trade Name: | CONMED® SurefitTM Dual Dispersive Electrode | | | |----------------------|----------------------------------------------------------------|----------|--------| | Common Name: | Electrosurgical unit and accessories | | | | Classification Name: | Electrosurgical Cutting and Coagulation Device and accessories | | | | Regulation Number: | 21 CFR 878.4400 | Class II | 79 GEI | | Panel: | General and Plastic Surgery | | | ## D. Predicate Device Name CONMED® Surefit™ Dual Dispersive Electrode ConMed Electrosurgery 510(k) K002002 ## E. Device Description The CONMED® Surefit™ Dual Dispersive Electrode is a single use, non-sterile dispersive electrode with a pre-attached cord. The purpose of the return electrode is to complete the electrosurgical circuit between the generator, the active electrode, and the patient. The CONMED® Surefit™ Dual Dispersive Electrode is to be used on any patient where full skin {1}------------------------------------------------ contact and a suitable placement site can be obtained. Use of this product for unintended applications could lead to an unsafe condition. The surface of the conductive area is covered with a soft, conformable hydro-gel conductive adhesive. The pad also has a non-conductive border adhesive surrounding the entire conductive area to isolated the conductive area from surgical fluids. #### F. Intended Use The CONMED® Surefit™ Dual Dispersive Electrodes is intended for use in surgical procedures in which Electrosurgery equipment with contact quality monitoring system is used. ## G. Indications for Use CONMED® Surefit™ Dual Dispersive Electrodes are designed for use with only those electrosurgical generators equipped with contact quality monitoring systems (i.e. REM™, ARM™, and NESSY, ™etc.) during electrosurgery and provides a path for, RF energy produced at the active electrode to return to the generator. The Surefit™ Dual Dispersive Electrode is for use with all patient populations providing there is sufficient surface area to ensure full contact of the electrode with the patient's skin #### H. Technological Characteristics The technological characteristics of the proposed device are identical to the predicate device. #### Safety Information ـــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ Questions of safety and effectiveness are the same for this device as they are for the predicated devices and other patient return electrodes on the market. There are no new technologies incorporated into the device. #### J. Biocompatibility The biological safety of selected components of the CONMED® Surefit™ Dual Dispersive Electrode has been assured through the selection of materials which demonstrate appropriate levels of biocompatibility. These were selected on the basis of Part 1 of ISO 10993-1, "Biological Evaluation of Medical Devices." ## K. Performance Testing Where applicable, specific testing performed was in conformance with the requirements of the following standards: IEC 60601-2-2:2009 Edition 5.0 2009-02, Medical Electrical Equipment Part 2-2 ISO 10993-5: 1999 Biological Evaluation of Medical Devices - Part 5 Tests for cytotoxicity, in vitro methods ISO 10993-10: 2002/Amd 1:2006 Biological Evaluation of Medical Devices - Tests for Irritation and delayed-type hypersensitivity {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The seal is black and white. MAY 1 4 2010 Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002 Conmed Corporation % Mr. Shawn Riedel Vice President, Quality Assurance and Regulatory Affairs 14603 E. Fermont Avenue Centennial, Colorado 80112 Re: K100047 Trade/Device Name: CONMED® Surefit™ Dual Dispersive Electrode Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI, JOS Dated: April 15, 2010 Received: April 16, 2010 Dear Mr. Riedel: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {3}------------------------------------------------ Page 2 - Mr. Shawn Riedel forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use 510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________ Device Name: CONMED® Surefit™ Dual Dispersive Electrode Indications for Use: CONMED® Surefit™ Dual Dispersive Electrodes are designed for use with only those electrosurgical generators equipped with contact quality monitoring systems (i.e. REM™, ARM™, and NESSY, ™etc.) during electrosurgery and provides a path for, RF energy produced at the active electrode to return to the generator. The Surefit™ Dual Dispersive Electrode is for use with all patient populations providing there is sufficient surface area to ensure full contact of the electrode with the patient's skin. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orthopedic, and Pestorative Devices Page *_ of _* 510(k) Number K100047
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%