K Number
K093981

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-07-22

(210 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The TensorTip - Non-invasive Hemodynamic blood pressure monitor trending device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.

Device Description

The TensorTip - Non-invasive Hemodynamic Blood Pressure Monitor measurement chamber is designed to measure spot-check of heartbeat and blood pressure trending levels by simply inserting the finger into the device "chamber". It contains a dedicated finger compartment having soft gel on the upper compartment lead door suiting the finger to be placed inside. The device incorporates LEDs (Light Emitting Diode); a sensor array enables to sense the spectrum from near UV to near IR and a rechargeable battery for power supply. The device has large display for displaying temporal Pulse Per Minute (PPM) and Systolic and Diastolic Blood Pressure trending. In addition it shows the capillary blood waveform and the blood pressure variation during the measurement period. It is utilizing software that collects and presents the measured data.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Cnoga Medical TensorTip, based on the provided 510(k) summary:

1. Table of Acceptance Criteria and Reported Device Performance

The 510(k) summary primarily focuses on demonstrating substantial equivalence to predicate devices and adherence to relevant standards. Specific numerical acceptance criteria for blood pressure accuracy (e.g., mean difference and standard deviation compared to a reference method) are not explicitly stated in the provided text. However, the summary indicates that clinical validation met the requirements of ISO 81060-2, which is the standard for non-invasive sphygmomanometers. This standard typically defines accuracy requirements.

Since the exact numerical acceptance criteria are not specified, the table will reflect the general statement of conformity to the standard.

Acceptance Criteria (Aligned with ISO 81060-2)Reported Device Performance
Clinical validation requirements of ISO 81060-2Met the requirements

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated in the provided text. The document only mentions "human testing" for clinical validation.
  • Data Provenance: Not explicitly stated. The document indicates "human" testing was performed, but does not specify the country of origin of the data or whether it was retrospective or prospective.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided in the given 510(k) summary. For blood pressure monitors, ground truth is typically established by trained medical professionals using a reference device (e.g., a mercury sphygmomanometer or an oscillometric device validated against an invasive arterial line).

4. Adjudication Method for the Test Set

This information is not provided in the given 510(k) summary.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

This information is not applicable or not provided. The device described is a non-invasive hemodynamic blood pressure monitor, not an AI-assisted diagnostic imaging system that would typically involve human "readers" or an MRMC study in this context. The 510(k) describes a standalone device.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, a standalone performance study was done. The 510(k) describes the TensorTip as a device that directly measures and displays blood pressure and pulse rate. The "Performance Data" section states that "Cnoga Medical's TensorTip spot-check pulse rate and non-invasive blood pressure trending device has been successfully tested with bench, human and safety testing to support the determination of substantial equivalence with predicate devices." This implies testing the device's inherent accuracy in measuring these physiological parameters.

7. The Type of Ground Truth Used

The ground truth used for clinical validation of blood pressure devices typically involves measurements from a validated reference device (e.g., a mercury sphygmomanometer or an oscillometric device that has itself been validated against more invasive methods like intra-arterial pressure). While not explicitly detailed, the statement "Clinical validation has met the requirement of ISO 81060-2 standard" strongly implies that such a reference standard was used to establish ground truth for the test set.

8. The Sample Size for the Training Set

This information is not provided in the given 510(k) summary. The document describes a medical device, not a machine learning model that typically involves a distinct "training set" in the conventional sense of AI development. While the device certainly uses algorithms, the summary doesn't parse out data into training/test sets in that manner.

9. How the Ground Truth for the Training Set Was Established

This information is not provided as a separate "training set" and its ground truth establishment are not directly discussed in the context of the device's development as presented in this 510(k) summary. The focus is on the device's overall performance validation against standards.

{0}------------------------------------------------

K093981
P1/3

Image /page/0/Picture/1 description: The image shows the text "CNOGA MEDICAL Seeing Beyond Imagination". The text is arranged in three lines, with "CNOGA" on the top line, "MEDICAL" on the second line, and "Seeing Beyond Imagination" on the third line. There is a graphic between the N and G in the first line.

JUL 22 2010

510(K) SUMMARY FOR CNOGA MEDICAL'S TENSORTIP™ - NON-INVASIVE HEMODYNAMIC BLOOD PRESSURE MONITOR

DATE PREPARED: JULY 10, 2010

1. 510(K) OWNER NAME

Cnoga Medical ACAL North Industrial Park, Ha-Ilan 83 St., Or Akiva, Israel Tel: +972 4 6361080, Fax: +972 4 6361321, E mail: yosef@cnoga.com (Mailing address: POB 79, Zichron Yaacov, 30900, Israel).

Contact person name: Dr. Yosef Segman, Cnoga Medical's CEO. Telephone: 972-4-6361080 ext. 105, E-mail: yosef@cnoga.com

Submitter: Ms. Tali Hazan - R.A Consultant Telephone: 972-50-5292304, Fax: 972-151-508963806, E-mail: tali.hazan@012.net.il

DEVICE NAME 2.

Common/Usual Name: Non-Invasive Measurement Chamber. Proprietary/Trade name: TensorTip™ - Non-invasive Hemodynamic Blood Pressure Monitor

Classification: Cnoga Medical's TensorTip- Non-invasive Hemodynamic Blood Pressure Monitor has been classified as Class II devices under the following classification names:

Classification NameProduct Code21 CFR Ref.Panel
System, measurement, blood-pressure, non-invasiveDXN870.1130Cardiovascular
Impedance plethysmographDSB870.2770Cardiovascular

{1}------------------------------------------------

C N & G 1. D 1 1 A 11 Seeing Beyond Imagination

3. PREDICATE DEVICES

Cnoga Medical's TensorTip device is substantially equivalent to the following Predicates:

  • 3.1 A&D Engineering's A&D Medical Ua-767pbt Digital Blood Pressure Monitor (non-invasive), cleared under 510(k) number K043217.
  • 3.2 Finapres Medical Systems' Finometer Noninvasive Hemodynamic Monitor, cleared under 510(k) number K023723.
  • 3.3 Finapress Medical Systems' Portapress Ambulatory Continuous noninvasive blood pressure monitor, cleared under 510(k) number K023338.

DEVICE DESCRIPTION 4.

The TensorTip - Non-invasive Hemodynamic Blood Pressure Monitor measurement chamber is designed to measure spot-check of heartbeat and blood pressure trending levels by simply inserting the finger into the device "chamber".

It contains a dedicated finger compartment having soft gel on the upper compartment lead door suiting the finger to be placed inside. The device incorporates LEDs (Light Emitting Diode); a sensor array enables to sense the spectrum from near UV to near IR and a rechargeable battery for power supply. The device has large display for displaying temporal Pulse Per Minute (PPM) and Systolic and Diastolic Blood Pressure trending. In addition it shows the capillary blood waveform and the blood pressure variation during the measurement period. It is utilizing software that collects and presents the measured data. Body contact materials were evaluated for biocompatibility with accordance to FDA's Memorandum - #G95 1, May 1, 1995 and ISO 10993-1:2003 - Biological evaluation of medical devices - Part 1: Evaluation and testing with acceptable results.

{2}------------------------------------------------

CN S G A I. D I CAL 11 Sceing Beyond Imagination

5. INTENDED USE

The TensorTip - Non-invasive blood pressure trending device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory.

K093981

p3/3

The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.

TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE 6.

Cnoga Medical's TensorTip-Non-invasive Hemodynamic Blood Pressure Monitor device combines two features (Pulse rate spot-check and blood pressure trending) into one device. Both features are substantially equivalent to the predicate devices already commercially available in the market, as identified in this 510(k) summary, section 3. These predicate devices have the same intended use and share a similar technology as our TensorTip device.

Certain differences in the technology utilized for the TensorTip did not raise new safety and effectiveness questions and found to be substantial equivalence for the device intended use.

7. PERFORMANCE DATA

Cnoga Medical's TensorTip spot-check pulse rate and non-invasive blood pressure trending device has been successfully tested with bench, human and safety testing to support the determination of substantial equivalence with predicate devices. Safety and EMC tests were conducted with accordance to EN 60601-1 and EN/IEC 60601-1-2 standards. Clinical validation has met the requirement of ISO 81060-2 standard.

Tests results are supporting all labeling claims and substantial equivalency.

CONCLUSIONS 8.

The evaluation of our device performances demonstrates that it is as safe and as effective as predicate devices.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the United States Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of a bird-like figure, possibly representing an eagle, with its wings spread.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

JUL 2-2 2010

Cnoga Medical, Ltd. c/o Ms. Tali Hazan Regulatory Affairs Advisor for Medical Devices Ramot Naftali M.P. Upper Galilee ISRAEL 13830

Re: K093981

Trade/Device Name: TensorTip - Noninvasive Hemodynamic Blood Pressure Monitor Regulatory Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: 74 DXN Dated: June 12, 2010 Received: June 16, 2010

Dear Ms. Hazan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 -- Ms. Tali Hazan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
W.M.d.

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

C D が پائي Seeing Bevond Imagination

Indications for Use

510(k) Number (if known):

Device Name: TensorTip - Non-invasive Hemodynamic Blood Pressure Monitor

Indications for Use:

The TensorTip - Non-invasive Hemodynamic blood pressure monitor trending device is a small, lightweight, handheld, device intended for measuring and display of Blood Pressure trending (systolic and diastolic) and spot-check of Peripheral pulse rate (PPR) and Peripheral pulse wave (PPW). Measurement is performing on capillary finger tip tissue (other than the thumb). The ring finger is the recommended site. The results of each measurement are stored in the system memory. The device is intended for use in the home environment. It is intended to be used by any person aged above 18 years old.

Prescription Use > (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1

(Posted November 13, 2003)

ivision Sian-Off) Division of Cardiovascular Devices

510(k) Number K093981

Cnoga 510(k) AI TensorTip - Non-invasive Hemodynamic Blood Pressure Monitor July 10, 2010 Page 1 of 1

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).