K Number
K093895

Validate with FDA (Live)

Date Cleared
2010-05-20

(153 days)

Product Code
Regulation Number
870.2910
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ALLY™ Platform System is a multi-platform medical device that is an accessory to approved 510k devices and is intended to be used as a means of receiving, storing, and transmitting patient biometric data including vital signs measurements. The ALLY™ Platform System is to be used upon prescription and under the supervision of an authorized healthcare provider.

The ALLY " Platform System will send information to a remote database to be accessed by authorized clients, such as healthcare professionals or patient family members. Electronic data consisting of patient vital sign information, educational material, reminders, questionnaires, and patient responses are exchanged over a secure network connection. Information may be communicated via standard telephone lines, high speed internet connections, serial port interfaces, or Bluetooth® radio transceivers.

The ALLY™ Platform is a multi-platform application that is not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts. The ALLY™ Platform is not intended to act as an emergency response system. All patient medical diagnosis and treatment are to be performed by an appropriate health care professional.

Device Description

ALLY™ is comprised of a home unit, legally marketed diagnostic peripherals, and a server. The home unit is a communications device that collects data from the diagnostic peripherals and. The patient in the home and transmits the data to the server over the home's existing telephone line or other identified method (existing high-speed connection in patients' home). The diagnostic peripherals comprise a wide range of legally marketed medical electronic devices capable of measuring patient parameters such as blood pressure, pulse rate, weight, etc. The server application responds to calls from the home units, collects the data, and then compares the data to preset thresholds. The healthcare practitioners and privileged users access the server through a secure website.

ALLYTM Care Portal: The ALLY TM Care Portal is intended to be used as an accessory to communication tools, such as the ALLYTM Home Appliance (RTX 337X). The ALLY TM Care Portal enables healthcare providers to request, receive, and review historical patient information. It is intended to be used in combination with a variety of external devices.

ALLYTM Home Appliance (RTX 337X): The ALLYTM Home Appliance (RTX 337X) is for use in non-clinical settings (such as the home), as an accessory device that is intended to be a communication tool to enable healthcare providers to receive historical patient information. It is intended to be used in combination with a variety of external devices.

The ALLY™ Home Appliance (RTX 337X) serves as the remote communication link between compatible external devices, and the compatible healthcare facility at another location. The healthcare facility could be at a disease management center or with the healthcare/wellness provider or other out of hospital caregivers.

AI/ML Overview

This 510(k) summary is for the ALLY™ Platform System, a telehealth system. Based on the provided document, here's an analysis of the acceptance criteria and study information:

The document does not contain specific details about acceptance criteria or a dedicated study proving the device meets them in the way modern AI/ML device submissions typically do (e.g., performance metrics, ground truth, expert review). This is largely because the ALLY™ Platform System is a data transmission and storage platform, not a diagnostic or AI-powered analytical device. Its primary function is to securely transmit biometric data.

Here's an attempt to answer your questions based on the available information, noting where information is not present:


1. Table of Acceptance Criteria and Reported Device Performance

Given the nature of the ALLY™ Platform System, acceptance criteria would likely revolve around technical functionalities like successful data transmission, data integrity, security, and compatibility with peripherals, rather than diagnostic accuracy metrics. The document does not provide a table of performance metrics.

Acceptance Criteria (Inferred from device description)Reported Device Performance (Inferred)
Successful Data Transmission: Ability to collect data from diagnostic peripherals and transmit to a server.The device is intended to collect data and transmit it over various methods (telephone, high-speed internet, serial port, Bluetooth). The server application responds to calls from home units and collects data.
Data Integrity: Ensured preservation of patient biometric data during transmission and storage.The system is designed to receive, store, and transmit patient biometric data. Implicitly, this requires data integrity.
Security: Secure transmission and access for authorized users.Information is exchanged over a secure network connection. Healthcare practitioners and privileged users access the server through a secure website.
Compatibility: Ability to interface with a wide range of legally marketed medical electronic devices (diagnostic peripherals).Intended to be used in combination with a variety of external devices and compatible external devices.
Reliability/Functionality: Device does not provide automated treatment decisions, perform data manipulation, diagnostics, or real-time alerts.Explicitly stated: "not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts."
Level of Concern: No expected injury or death due to failure or latent design flaw.Concluded device is "Medium Concern" (Moderate), implying satisfactory safety assessment.

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a test set or data provenance in the context of evaluating a diagnostic algorithm's performance. The ALLY™ Platform System is a communication and data management system, not one that generates diagnostic insights. Therefore, there wouldn't typically be a "test set" of patient data for performance evaluation in the same way an AI-powered image analysis tool would have.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

N/A. Ground truth establishment with experts is not relevant for this type of device, as it does not perform diagnostic tasks.

4. Adjudication Method for the Test Set

N/A. Adjudication methods like 2+1 or 3+1 are used for establishing ground truth in diagnostic studies, which is not applicable here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

N/A. This device is not an AI-powered diagnostic tool, and therefore, an MRMC study comparing human reader performance with and without AI assistance is not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

N/A. The ALLY™ Platform System is a data transmission platform, not a standalone diagnostic algorithm. It explicitly states it is "not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts."

7. The Type of Ground Truth Used

N/A. Ground truth (e.g., expert consensus, pathology, outcomes data) is used for validating diagnostic devices. For the ALLY™ Platform System, validation would likely involve demonstrating technical functionality, data integrity, security, and compliance with communication standards.

8. The Sample Size for the Training Set

N/A. This device does not use machine learning or AI models that require a training set of data.

9. How the Ground Truth for the Training Set was Established

N/A. Not applicable, as there is no training set or ground truth in the context of an AI/ML model for this device.


Summary:

The 510(k) submission for the ALLY™ Platform System focuses on its function as a secure communication and data storage platform for biometric data, acting as an accessory to other approved medical devices. It explicitly states that it does not provide automated treatment decisions, perform data manipulation, diagnostics, or real-time alerts.

Therefore, the typical acceptance criteria and study designs for AI/ML-powered diagnostic devices (involving test sets, ground truth, expert adjudication, MRMC studies, and training sets) are not relevant to this submission. Instead, the "study" demonstrating the device meets its acceptance criteria likely involved testing its technical capabilities, interoperability, security features, and compliance with electrical safety and communication standards to ensure it performs its intended function of reliably receiving, storing, and transmitting patient biometric data. The provided document heavily emphasizes substantial equivalence to predicate devices based on similar intended use and technological characteristics, suggesting that the primary "proof" of meeting criteria was demonstrating these similarities and outlining safety considerations for a "Moderate" level of concern device.

{0}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS (21CFR 807.92)

I. DATE PREPARED: 12/14/2009

MAY 2 0 2010

  • II. SUBMITTER:
    Continuity Health Solutions 9304 Forest Lane, Suite N272 Dallas, TX 75243 Telephone: 866.517.3135 Fax: 214.553.8946

III. CONTACT PERSON:

Bryan Poteet, CEO Telephone: 866.517.3135

IV. DEVICE NAME:

Trade/Proprietary Name: Common/Usual Name: Classification Name:

ALLYTM Platform System Telehealth System Physiological Signal Transmitters and Receivers

V. DEVICE CLASSIFICATION

Class II CFR section: 870.2910 Physiological Signal Transmitters and Receivers Product code: DRG Panel: Cardiology

VI. PREDICATE DEVICES:

K090801 Electronic House Call System Class II Decision Date: 04/08/2009

K063612 Health hero Network Class II Decision Date: 12/29/2006

{1}------------------------------------------------

K093895
p2/3

VII. SUBSTANTIAL EQUIVALENCE CONCLUSIONS:

Continuity Health Solutions concludes that the intended use for the Continuity Health Solutions ALLYTM Platform System is the same as that of the predicate devices, and that the technological characteristics demonstrate that they are equivalent to the predicate device. A comparison of the technological characteristics of the predicate and legally marketed devices available has been performed.

Thus, this premarket notification has demonstrated substantial equivalence.

VIII. DEVICE DESCRIPTION AND INTENDED USE:

Device Description:

ALLY™ is comprised of a home unit, legally marketed diagnostic peripherals, and a server. The home unit is a communications device that collects data from the diagnostic peripherals and. The patient in the home and transmits the data to the server over the home's existing telephone line or other identified method (existing high-speed connection in patients' home). The diagnostic peripherals comprise a wide range of legally marketed medical electronic devices capable of measuring patient parameters such as blood pressure, pulse rate, weight, etc. The server application responds to calls from the home units, collects the data, and then compares the data to preset thresholds. The healthcare practitioners and privileged users access the server through a secure website.

ALLYTM Care Portal

The ALLY TM Care Portal is intended to be used as an accessory to communication tools, such as the ALLYTM Home Appliance (RTX 337X). The ALLY TM Care Portal enables healthcare providers to request, receive, and review historical patient information. It is intended to be used in combination with a variety of external devices.

ALLYTM Home Appliance (RTX 337X)

The ALLYTM Home Appliance (RTX 337X) is for use in non-clinical settings (such as the home), as an accessory device that is intended to be a communication tool to enable healthcare providers to receive historical patient information. It is intended to be used in combination with a variety of external devices.

The ALLY™ Home Appliance (RTX 337X) serves as the remote communication link between compatible external devices, and the compatible healthcare facility at another location. The healthcare facility could be at a disease management center or with the healthcare/wellness provider or other out of hospital caregivers.

{2}------------------------------------------------

Intended Use:

The ALLY™ Platform System is a multi-platform medical device that is an accessory to approved 510k devices and is intended to be used as a means of receiving, storing, and transmitting patient biometric data including vital signs measurements. The ALLY™ Platform System is to be used upon prescription and under the supervision of an authorized healthcare provider.

The ALL YTM Platform System will send information to a remote database to be accessed by authorized clients, such as healthcare professionals or patient family members. Electronic data consisting of patient vital sign information, educational material, reminders, questionnaires, and patient responses are exchanged over a secure network connection. Information may be communicated via standard telephone lines, high speed internet connections, serial port interfaces, or Bluetooth® radio transceivers.

The ALLYM Platform is a multi-platform application that is not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts. The ALLY™ Platform is not intended to act as an emergency response system. All patient medical diagnosis and treatment are to be performed by an appropriate health care professional.

IX. SAFETY INFORMATION:

The ALLYTM Platform System has no patient contact and is utilized only by trained professionals. Trained professionals allow sufficient review to afford identification and intervention in the event of a malfunction have evaluated the output of the device. This will allow sufficient review to afford identification and intervention in the event of a malfunction.

X. CONCLUSION:

Continuity Health Solutions believes sufficient information is included to reach a determination of substantial equivalence. We conclude that the subject device is as safe and effective including the component and accessory devices.

As a result of this analysis and the risk reduction activities, and has reviewed the guidance document for the content of premarket submissions for software contained in medical device, we conclude that there is no software or hardware component in the ALLY TM Platform System that would be expected to result in injury or death to a patient due to a failure or latent design flaw. Thus, the "Level of Concern" of the ALLY™ Platform System product is "Moderate".

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is relatively small compared to the caduceus symbol. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Continuity Health Solutions c/o Mr. Bryan Poteet CEO 9304 Forest Lane, Suite N272 Dallas. TX 75243

MAY 0 2010

K093895 Re:

Trade/Device Name: ALLYTM Platform System Regulatory Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitter and Receiver Regulatory Class: II (two) Product Code: 74 DRG Dated: April 13, 2010 Received: April 19, 2010

Dear Mr. Poteet:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

{4}------------------------------------------------

Page 2 - Mr. Bryan Poteet

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): N/A

Device Name: ALLY™ Platform System

Intended Use:

The ALLY™ Platform System is a multi-platform medical device that is an accessory to approved 510k devices and is intended to be used as a means of receiving, storing, and transmitting patient biometric data including vital signs measurements. The ALLY™ Platform System is to be used upon prescription and under the supervision of an authorized healthcare provider.

The ALLY " Platform System will send information to a remote database to be accessed by authorized clients, such as healthcare professionals or patient family members. Electronic data consisting of patient vital sign information, educational material, reminders, questionnaires, and patient responses are exchanged over a secure network connection. Information may be communicated via standard telephone lines, high speed internet connections, serial port interfaces, or Bluetooth® radio transceivers.

The ALLY™ Platform is a multi-platform application that is not intended to provide automated treatment decisions or perform data manipulation, diagnostics, or real-time alerts. The ALLY™ Platform is not intended to act as an emergency response system. All patient medical diagnosis and treatment are to be performed by an appropriate health care professional.

Prescription Use (Per 21 CFR 801 Subpart D) AND/OR

Over the Counter Use (Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE)

W.M. Wood.

510(k) Number K0613095

§ 870.2910 Radiofrequency physiological signal transmitter and receiver.

(a)
Identification. A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.(b)
Classification. Class II (performance standards).