K Number
K093801

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-07-08

(210 days)

Product Code
Regulation Number
870.4360
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The IMPELLA Controller is an extracorporeal bypass control unit intended to be used to provide circulatory support for periods up to 6 hours. It is also intended to be used to provide circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass. The Impella Controller is intended to be used by trained healthcare professionals in healthcare facilities and medical transport (i.e. ambulance, helicopter, or fixed-wing aircraft) environments.

The IMPELLA Controller also displays pressure measurement readings, which are useful in determining intravascular pressure.

Device Description

The IMPELLA Controller is a microprocessor-based pump motor driver and purge system. It is designed to operate on AC current (110-240 VAC, 47-63 Hz) or on an internal, rechargeable battery. The IMPELLA Controller generates the signals required to power the drive motor of one of ABIOMED's IMPELLA Percutaneous Support Catheters (the IMPELLA RECOVER LP 2.5 (cleared under K063723), the IMPELLA 5.0 LP and the IMPELLA 5.0 LD (cleared under K083111). The IMPELLA Controller also serves to deliver an infusate of the catheter's drive motor, and to provide useful information regarding the catheter's performance. It is intended to be used by trained healthcare professionals in hospital and medical transport environments. It is lightweight, portable and has an accompanying cart for ease of transport within the hospital.

AI/ML Overview

The provided text describes the IMPELLA Controller's acceptance criteria and the study that proves it meets these criteria.

1. Table of Acceptance Criteria and Reported Device Performance:

The primary acceptance criteria for the IMPELLA Controller were its conformity to established international and internal performance standards, and its equivalence to the predicate device.

Acceptance CriteriaReported Device Performance
International Standards Compliance
IEC 60601-1-2 (Electromagnetic Compatibility - EMC/EMI)Conformed to IEC 60601-1-2, including pertinent IEC 61000-3-X and IEC 61000-4-X standards, and EN 55011.
IEC 60601-1-Part 1 (Electrical Safety)Conformed to IEC 60601-1 - Part 1.
ISTA 2A and EN 868 (Packaging and Shipping)Conformed to ISTA 2A and EN 868.
RTCA/DO-160C (Altitude and Vibration during Operation)Tested per RTCA/DO-160C.
ISO 1135-4 (Transfusion equipment)Applied.
ISO 8536-4 (Infusion equipment)Applied.
EN ISO 10993-1:2003 and parts (Biocompatibility)Conformed to EN ISO 10993-1:2003 and its parts, including EN ISO 10993-7.
EN ISO 11135 (Sterility testing)Conformed to EN ISO 11135.
FDA Guidance for Software in Medical DevicesComplied with FDA 2005 document "Guidance for Industry and FDA Staff- Guidance for the Content of Pre-market Submissions for Software Contained in Medical Devices".
Internal Performance Specifications
System DurabilityMet.
System PerformanceMet.
System CharacterizationMet.
System Flow CharacterizationMet.
Sensor System ResponseMet.
Equivalence to Predicate DeviceDemonstrated equivalent performance characteristics to the predicate system (K063723) through in vitro testing.

2. Sample size used for the test set and the data provenance:

The text does not specify a sample size for a "test set" in the context of patient data or clinical trials. The performance testing was conducted in vitro, meaning in a controlled laboratory environment, not on human subjects. Therefore, there is no direct patient data or provenance mentioned. The testing focused on the device's technical specifications and compliance with standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

This information is not provided in the document section related to performance data. As the tests were in vitro, the "ground truth" would be established by technical specifications and reference standards, rather than expert consensus on patient data.

4. Adjudication method for the test set:

An adjudication method for a "test set" in the context of reviewing clinical cases or images is not applicable here, as the study was in vitro and focused on engineering and functional performance, not clinical scenarios requiring expert interpretation.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

An MRMC comparative effectiveness study was not performed. The IMPELLA Controller is a medical device for circulatory support and not an AI-assisted diagnostic tool that would typically involve human readers interpreting cases.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The performance testing described was primarily "standalone" in the sense that it evaluated the device's technical capabilities and compliance with standards without a human operator's variable input being a primary focus of the performance assessment (though it is intended for use by trained healthcare professionals). The tests focused on the device's inherent design and functionality, such as system durability, flow characterization, and sensor response.

7. The type of ground truth used:

The ground truth used for the in vitro testing was established based on:

  • Established engineering design specifications (e.g., specific flow rates, pressure readings, durability benchmarks that the device was designed to meet).
  • International and national standards for medical device safety, electromagnetic compatibility, electrical safety, biocompatibility, sterility, software design, packaging, and transport (e.g., IEC 60601 series, ISO 10993 series, FDA guidance documents).
  • Performance characteristics of the predicate device, which served as a benchmark for equivalence in internal performance testing.

8. The sample size for the training set:

The concept of a "training set" is typically associated with machine learning or AI algorithm development. Since the IMPELLA Controller is a hardware-based medical device with microprocessor control, and the provided text does not mention any machine learning components, there is no training set in that context.

9. How the ground truth for the training set was established:

As there is no mention of a training set for machine learning in the provided document, the mechanism for establishing its ground truth is not applicable. The "ground truth" for the device's development and verification would stem from engineering principles, medical requirements for circulatory support, and regulatory standards.

{0}------------------------------------------------

KOS3801

JUL - 8 2010

IMPELLA CONTROLLER 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 of the Code of Federal Regulations, Part 807.92.

A. Application Information: Date Prepared:

July 1, 2010

ABIOMED, Inc.

Submitter's Name & Address: 22 Cherry Hill Drive Danvers, MA 01923

Contact Person:

Dr. Robert Stewart Manager, FDA Programs Ph: 978-646-1567 Fax: 978-777-5692 E-mail: rstewart@abiomed.com

B. Device Information:

Trade or Proprietary Name:IMPELLA Controller
Common or Usual Name:Non-roller type Cardiopulmonary Blood Pump
Classification Name:Class III, KFM, 21 CFR - 870.4360
Performance Standard:Performance standards do not currently exist for these devices
(i.e. none established under section 514 of the F D & C Act)

C. Predicate Device:

The Impella Power Supply, the Impella Mobile Pump Console, and the Braun Vista Basic Purge System, which were cleared for use together as an extracorporeal bypass control unit for the RECOVER LP 2.5 PERCUTANEOUS CARDIAC SUPPORT SYSTEM, K063723.

D. Device Description:

The IMPELLA Controller is a microprocessor-based pump motor driver and purge system. It is designed to operate on AC current (110-240 VAC, 47-63 Hz) or on an internal, rechargeable battery. The IMPELLA Controller generates the signals required to power the drive motor of one of ABIOMED's IMPELLA Percutaneous Support Catheters (the IMPELLA RECOVER LP 2.5 (cleared under K063723), the IMPELLA 5.0 LP and the IMPELLA 5.0 LD (cleared under K083111). The IMPELLA Controller also serves to deliver an infusate of the catheter's drive motor, and to provide useful information regarding the catheter's performance. It is intended to be used by trained healthcare professionals in hospital and medical transport environments. It is lightweight, portable and has an accompanying cart for ease of transport within the hospital.

E. Intended Use:

The IMPELLA Controller is an extracorporeal bypass control unit intended to be used to provide circulatory support for periods up to 6 hours. It is also intended to be used to provide circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass. The Impella Controller is intended to be used by trained healthcare professionals in healthcare facilities and medical transport (i.e. ambulance, helicopter, or fixed-wing aircraft) environments.

The IMPELLA Controller also displays pressure measurement readings, which are useful in determining intravascular pressure.

{1}------------------------------------------------

F. Technological Characteristics:

The IMPELLA Controller employs the same basic, functional scientific technology as its predicate device cleared under K063723.

G. Comparison to Predicate Device:

The IMPELLA Controller has the same intended use as its predicate system, which is made up of a primary, AC powered Power Supply, a Mobile Pump Console, and a Braun VISTA Purge System. The primary function of the IMPELLA Controller remains the same as the predicate system, which is supplying motor power to control and monitor an IMPELLA Percutaneous Support Catheter. The IMPELLA Controller has also been designed to provide ease of patient portability using the same technology as its predicate system. Both the IMPELLA Controller and its predicate system are software driven, microprocessor-based consoles. No modifications to any of the IMPELLA Percutaneous Support Catheters are needed prior to their use with the IMPELLA Controller.

Differences between the IMPELLA Controller and the predicate extracorporeal bypass control unit are the integration of the separate predicate systems (the pump drive hardware, the primary power supply and purge system for the pump motor bearings) into a single control unit, resulting in equivalent performance characteristics to the predicate system. In addition, whereas the predicate console, the Mobile Pump Console utilized a monochromatic display for the display of operational information, the IMPELLA Controller incorporates a Color LCD, for better visibility. The equivalency of the indications for use, the design features and the functional characteristics of the IMPELLA Controller raise no new safety or effectiveness issues.

H. Summary of Performance Data:

The performance characteristics of the IMPELLA Controller were tested in vitro to verify that it met its performance specifications (established by ABIOMED), along with FDA or international standards for software Verification and Validation, and safety testing for EMC, Electrical Safety, and Transport.

The following international standards were met:

  • Electromagnetic compatibility testing was in conformance with IEC 60601-1-2, including all . pertinent IEC 61000-3-X and IEC 61000-4 --X standards for EMC/EMI along with EN 55011.
  • Electrical safety testing was in conformance with IEC 60601-1- Part 1. .
  • Packaging and shipping testing was in conformance with ISTA 2A and EN 868. .
  • Altitude and vibration testing during operation was tested per RTCA/DO-160C.

A number of additional standards were applied:

  • ISO 1135-4, Transfusion equipment for medical use -- Part 4. �
  • ISO 8536-4, Infusion equipment for medical use -- Part 4. .
  • Biocompatibility testing was in conformance with EN ISO 10993-1:2003 and its parts, including . EN ISO 10993-7.
  • Sterility testing is in conformance with EN ISO 11135.

Software design and testing was in compliance with:

  • FDA 2005 document titled "Guidance for Industry and FDA Staff- Guidance for the Content of . Pre-market Submissions for Software Contained in Medical Devices".
    Internal performance characteristics testing protocols were also used to demonstrate that the IMPELLA Controller performed equivalently to the predicates (given in K063723). The main in vitro systems tests completed were:

  • System Durability Testing. .

  • System Performance Test. .

  • System Characterization Test. .

{2}------------------------------------------------

  • System Flow Characterization Test. .
  • . Sensor System Response Test.

Additional specialized performance testing was completed to verify that the IMPELLA Controller met its design specifications.

The performance testing verified that the IMPELLA Controller, met its design specifications, and passed all of the standards testing. Overall, the in vitro performance testing results were equivalent to those demonstrated for the predicate (provided in K063723).

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle. The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JUL -- 8 2010

Abiomed, Inc. c/o Robert Stewart, Ph.D. Manager, FDA Programs 22 Cherry Hill Drive Danvers, MA 01923

Re: K093801

IMPELLA® 2.5, 5.0 and 5.0/LD with Mobile Console (IMC) Regulation Number: 21 CFR 870.4360 Regulation Name: Non-roller type cardiopulmonary bypass blood pump Regulatory Class: Class III Product Code: KFM. DWA Dated: June 18, 2010 Received: June 21, 2010

Dear Dr. Stewart:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be

{4}------------------------------------------------

Page 2 - Robert Stewart, Ph.D.

found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

R.v.chner

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

5.0 INDICATIONS FOR USE

510(k) Number (if known): (09380-

Device Name: IMPELLA Controller

Indications for Use:

The IMPELLA Controller is an extracorporeal bypass control unit intended to be used to provide circulatory support for periods up to 6 hours. It is also intended to be used to provide circulatory support (for periods up to 6 hours) during procedures not requiring cardiopulmonary bypass. The Impella Controller is intended to be used by trained healthcare professionals in healthcare facilities and medical transport (i.e. ambulance, helicopter, or fixed-wing aircraft) environments.

The IMPELLA Controller also displays pressure measurement readings, which are useful in determining intravascular pressure.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Muna R. bilal

(Division Sian-Off) Division of Cardiovascular Devices

510(k) Number Ko9380 |

§ 870.4360 Nonroller-type blood pump.

(a)
Nonroller-type cardiopulmonary and circulatory bypass blood pump —(1)Identification. A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (
i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
(2)
Classification —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
Nonroller-type temporary ventricular support blood pump —(1)Identification. A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
Classification. Class III (premarket approval).(c)
Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required. A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.