K Number
K093800

Validate with FDA (Live)

Date Cleared
2010-06-16

(188 days)

Product Code
Regulation Number
880.6850
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Robinson MedSurg QuickScrew Rack RCK24-143 is intended to organize, hold, and protect the QuickScrew screwdrivers for transport, storage, sterilization, and use during surgery. The subject instrument racks are suitable for use in prevacuum steam. The subject instrument trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed devices.
Validated Sterilization Parameters:
Method: Prevacuum Steam Autoclave
Temperature: 270°F (132°C)
Exposure Time: 4 minutes
Drying Time: 60 minutes

Device Description

The Robinson MedSurg QuickScrew Rack RCK24-143 is an instrument rack for organizing, holding, and protecting reusable surgical instruments during transportation, storage, and sterilization. It is designed for adequate exposure of the trays' contents to sterilant during the sterilization process.
The Rack is made of anodized aluminum with polytherimide tube holders for the instruments and to allow for air and sterilant to flow through and around the rack.

AI/ML Overview

The request is about a medical device called the "Robinson MedSurg QuickScrew Rack RCK24-143," a sterilization rack. The provided documentation (K093800) is a 510(k) premarket notification summary and related FDA correspondence. This device is not an AI/ML powered device, therefore the vast majority of the requested information is not available and does not apply to this device.

Here's an assessment based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance CriteriaReported Device Performance
Conformance to AAMI ST77:2006 for Containment Devices for reusable medical device sterilization"The rack conforms to this standard and performed as intended."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. The device is a physical sterilization rack, not a software or AI/ML device that would typically have a "test set" in the context of data. The testing mentioned is performance testing for a medical device's function, not a data-driven model.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No "ground truth" establishment by experts in the context of data analysis is described, as this is a physical device.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable. There is no mention of adjudication methods for a "test set" as this is not an AI/ML or diagnostic device.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device, so MRMC studies and AI assistance are irrelevant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. The device is a physical product, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable. The "truth" for this device relates to its ability to perform its function (organizing, holding, protecting instruments, and allowing sterilant flow) as validated by performance testing against established standards (AAMI ST77:2006).

8. The sample size for the training set

Not applicable. There is no "training set" as this is a physical device, not an AI/ML model.

9. How the ground truth for the training set was established

Not applicable. As above, there is no "training set" or corresponding ground truth.

Summary of Device-Specific Information:

The Robinson MedSurg QuickScrew Rack RCK24-143 is a sterilization rack. Its performance testing was conducted according to AAMI ST77:2006 Containment Devices for reusable medical device sterilization. The reported result is that "The rack conforms to this standard and performed as intended." This implies that the acceptance criteria for the device were defined by the requirements outlined in AAMI ST77:2006, and the device successfully met those requirements through the conducted performance tests. The specific details of these tests (e.g., number of test cycles, specific measurements taken) are not provided in the summary.

{0}------------------------------------------------

TAB 1

K093800

Revised 2

510(k) Summary

As required by the Safe Medical Device Act of 1990, and codified in 21 CFR 807.92 upon which the substantial equivalence is based.

Robinson MedSurg QuickScrew Racks

Date Prepared: December 1, 2009

JUN 1 3 2010

  • A. Submitter's Name
    Robinson MedSurg, LLC 7430 E. Park Meadows Drive, Suite 300 Lone Tree, CO 80124

Establishment Registration Number: #3006119027

  • B. Company Contact
    Peggy Henline Operations Manager T. 303-706-1100 F. 303-662-8484 phenline(@robinsonmedsurg.com

C. Device Name

Trade Name:Robinson MedSurg QuickScrew Rack RCK24-143
Common Name:Sterilization Rack
Classification:Sterilization Wrap Containers, Trays, Cassettes and Other Accessories
Regulatory Class:II
Product Code:KCT
Classification Number:21 CFR §880.6850

D. Predicate Devices

The subject Robinson MedSurg QuickScrew Rack RCK24-143 is substantially equivalent in Intended Use and Fundamental Scientific Technology to the following legally marketed device in commercial distribution: Smith and Nephew Instrument Tray cleared under K073551, Smith and Nephew Instrument Trays cleared under K090562, and the ContainMed Versipod Instrument Trays cleared under K071783.

{1}------------------------------------------------

E. Description of Device

  • DEVICE DESCRIPTION: The Robinson MedSurg QuickScrew Rack RCK24-143 is an instrument rack for organizing, holding, and protecting reusable surgical instruments during transportation, storage, and sterilization. It is designed for adequate exposure of the trays' contents to sterilant during the sterilization process.
    The Rack is made of anodized aluminum with polytherimide tube holders for the instruments and to allow for air and sterilant to flow through and around the rack.

Intended Use F.

The Robinson MedSurg QuickScrew Rack RCK24-143 is intended to organize, hold, and protect the QuickScrew screwdrivers for transport, storage, sterilization, and use during surgery. The subject instrument racks are suitable for use in prevacuum steam. The subject instrument trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed devices.

Validated Sterilization Parameters:

MethodTemperatureExposure TimeDrying Time
Prevacuum SteamAutoclave270°F (132°C)4 minutes60 minutes

G. Comparison of Technological Characteristics

The subject Robinson MedSurg QuickScrew Racks RCK24-143 have the same fundamental technological characteristics as the predicate devices. The subject racks are substantially equivalent in design, materials, and intended use to the predicate devices. There are no significant differences between the proposed and predicate devices that raise new questions of safety or efficacy and is, therefore, substantially equivalent to the predicate devices.

H. Summary of Performance Data

Performance testing was conducted according to AAMI ST77:2006 Containment Devices for reusable medical device sterilization. The rack conforms to this standard and performed as intended.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three intertwined strands representing health, services, and people.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Randolph C. Robinson President Robinson Medsurg LLC 7430 East Park Meadows Drive, Suite 300 Lone Tree, Colorado 80124

JUN 1 3 2010

Re: K093800

Trade/Device Name: Robinson MedSurg QuickScrew Rack Model RCK 24-143 Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: May 16, 2010 Received: May 19, 2010

Dear Mr. Robinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Robinson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Susan Punno

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

TAB 3

Indications for Use Statement

510(k) Number:

Device Name: Robinson MedSurg QuickScrew Rack Model RCK 24-143 V

Indications for Use:

The Robinson MedSurg QuickScrew Rack RCK24-143 is intended to organize, hold, and protect the QuickScrew screwdrivers for transport, storage, sterilization, and use during surgery. The subject instrument racks are suitable for use in prevacuum steam. The subject instrument trays are not intended to maintain sterility; they are intended to be used in conjunction with a validated FDA cleared sterilization wrap in order to maintain sterility of the enclosed devices.

Validated Sterilization Parameters:

MethodTemperatureExposure TimeDrying Time
Prevacuum SteamAutoclave270°F (132°C)4 minutes60 minutes

Prescription Use (Part 21 CFR 801 Subpart D) and/or

Over-The-Counter Use XX (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

sion Sign-Off)
usion of Anesthesiology, General Hospital
ection Control, Dental Devices

Page 1 of 1

(k) Number: K093800

§ 880.6850 Sterilization wrap.

(a)
Identification. A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.(b)
Classification. Class II (performance standards).