K Number
K093794

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-05-14

(155 days)

Product Code
Regulation Number
880.2800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Smart-Read EZTest – Steam self-contained biological indicators (SCBI) are for monitoring the efficacy of saturated steam sterilization processes. Performance characteristics are established in accordance with USP 31 for the pre-vacuum 121°C steam process. Additional pre-vacuum saturated steam sterilization temperatures are also included in the Certificate of Analysis. Smart-Read EZTest – steam SCBIs are also appropriate for use in saturated steam processes of 132°C, 134°C and 135°C.

Device Description

The biological indicator consists of a self-contained unit which includes: bacterial spores inoculated onto a paper carrier, a small glass ampoule containing sterile culture medium and color indicator, and a plastic vial that serves as the culture tube. This unit complies to the performance characteristics described in USP 31 and ANSI/AAMI/ISO 11138-1.

AI/ML Overview

Acceptance Criteria and Device Performance Study for Smart-Read™ EZTest® - Steam Biological Indicator

This report describes the acceptance criteria and the study that proves the Smart-Read™ EZTest® - Steam biological indicator meets those criteria for reduced incubation time.

1. Table of Acceptance Criteria and Reported Device Performance

The core acceptance criterion for the reduced incubation time (RIT) study is that 97% of growth results observed at 7 days should be available at the RIT of 10 hours for the biological indicator (BI) that have been exposed to a partial sterilization cycle.

Acceptance CriterionReported Device Performance (Growth at 10 hours vs. Growth at 7 days)
Reduced incubation time (10 hours) results must show a minimum of 97% concordance with the results obtained after a 7-day incubation period for BIs exposed to partial sterilization cycles. This is based on sets of 100 BIs exposed to a steam process that reduced the spore population to approximately 1 spore per BI, where 30-80 units demonstrated growth (indicating a partial kill).Partial Cycle #1: 97.7% at 10 hours, 100% at 7 days (43/44 at 10h vs. 44/44 at 7d) Partial Cycle #2: 97.8% at 10 hours, 100% at 7 days (44/45 at 10h vs. 45/45 at 7d) Partial Cycle #3: 97.3% at 10 hours, 100% at 7 days (71/73 at 10h vs. 73/73 at 7d) Partial Cycle #4: 97.5% at 10 hours, 100% at 7 days (78/80 at 10h vs. 80/80 at 7d) Partial Cycle #5: 100% at 10 hours, 100% at 7 days (52/52 at 10h vs. 52/52 at 7d) Partial Cycle #6: 100% at 10 hours, 100% at 7 days (23/23 at 10h vs. 53/53 at 7d when looking at the last full growth %)Partial Cycle #7: 100% at 10 hours, 100% at 7 days (73/73 at 10h vs. 73/73 at 7d) Partial Cycle #8: 100% at 10 hours, 100% at 7 days (48/48 at 10h vs. 48/48 at 7d) Partial Cycle #9: 97.8% at 10 hours, 100% at 7 days (45/46 at 10h vs. 46/46 at 7d) Partial Cycle #10: 98.3% at 10 hours, 100% at 7 days (59/60 at 10h vs. 60/60 at 7d) Partial Cycle #11: 98.4% at 10 hours, 100% at 7 days (60/61 at 10h vs. 61/61 at 7d) Partial Cycle #12: 97.5% at 10 hours, 100% at 7 days (78/80 at 10h vs. 80/80 at 7d) Overall, all partial cycles meet or exceed the 97% concordance requirement at 10 hours compared to 7 days, demonstrating the device's ability to achieve reduced incubation time.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The study involved multiple partial sterilization cycles, each using a varying number of biological indicators (BIs).

    • Partial Cycle #1: 44 BIs
    • Partial Cycle #2: 45 BIs
    • Partial Cycle #3: 73 BIs
    • Partial Cycle #4: 80 BIs
    • Partial Cycle #5: 52 BIs
    • Partial Cycle #6: 53 BIs
    • Partial Cycle #7: 73 BIs
    • Partial Cycle #8: 48 BIs
    • Partial Cycle #9: 46 BIs
    • Partial Cycle #10: 60 BIs
    • Partial Cycle #11: 61 BIs
    • Partial Cycle #12: 80 BIs
      The total number of BIs tested across all partial cycles is 715.
  • Data Provenance: The document does not explicitly state the country of origin but implies laboratory testing conducted by SGM Biotech, Inc., located in Bozeman, MT, USA. The study design is prospective as it involves controlled exposure of BIs to partial sterilization cycles. The report indicates that the test was performed with "four different spore crops and multiple lots of product from each crop", suggesting multiple batches of the device were evaluated.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The ground truth for this device is established through microbiological growth assessment. This does not typically involve interpretation by human "experts" in the same way clinical imaging studies do. The assessment of growth (yellow color change) or no growth (clear and purple) is an objective, binary outcome. Therefore, there were no "experts" with specific qualifications like radiologists establishing the ground truth in this context. The results are based on direct observation of the color change in the culture medium, which is an intrinsic characteristic of the biological indicator's function.

4. Adjudication Method for the Test Set

Adjudication methods like 2+1 or 3+1 are typically used for subjective clinical interpretations. For the Smart-Read EZTest, the outcome (growth or no growth as indicated by color change) is a direct, objective result. There is no indication of an adjudication method being used because the results are determined by the biological reaction rather than human interpretation requiring consensus.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices where human readers interpret results, and the AI's role is to assist in that interpretation. The Smart-Read EZTest is a biological indicator designed for a direct chemical and biological reaction, not requiring human interpretation beyond observing a color change. Therefore, assessing how human readers improve with AI assistance is not applicable.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, a standalone study was performed. The data presented demonstrates the performance of the biological indicator itself in detecting sterilization failure (survival of spores). The "device performance" refers to the BI's ability to accurately show growth within the reduced incubation time, independent of any human intervention beyond activating and incubating the device and observing the final color change.

7. Type of Ground Truth Used

The ground truth used is a biological outcome (microbial growth of Geobacillus stearothermophilus spores) directly observed through a color change in the culture medium. For the purpose of establishing reduced incubation time, the 7-day incubation period serves as the gold standard ground truth, against which the 10-hour incubation results are compared. The 7-day result is considered the definitive indicator of spore survival or death.

8. Sample Size for the Training Set

The document does not explicitly state a separate "training set" for the reduced incubation time study. The study describes the methodology for establishing the reduced incubation time, which implies a development and testing process (i.e., the presented data acts as the validation that the 10-hour time is effective). The "test was performed with four different spore crops and multiple lots of product from each crop," suggesting robust testing across various manufacturing iterations and spore batches.

9. How the Ground Truth for the Training Set Was Established

As noted above, a distinct "training set" in the context of an AI/algorithm is not directly applicable here. However, the process for establishing the ground truth (7-day incubation) for the development and validation of the 10-hour reduced incubation time is as follows:

The ground truth for spore survival or death (growth or no growth) is established by incubation for 7 days at 60 ± 2°C. This extended incubation period ensures the maximum opportunity for any surviving spores to proliferate and cause a color change. This 7-day incubation serves as the reference standard against which the performance of the shorter (10-hour) incubation time is evaluated. The principle is that if spores survive, they will show growth within 7 days. The study then aimed to demonstrate that the same growth would be observable reliably at 10 hours.

{0}------------------------------------------------

K093794
p185

. .

Tab 11 510(k) Summary

Section 21 CFR §807.92:

MAY 1 4 2010

510(k) Notification Smart-Read 11 EZTest® - Steam biological indicator monitor for steam sterilizers

Submitted by:

SGM Biotech, Inc. 10 Evergreen Drive Bozeman, MT 59715

Contact:

John R. Gillis, Ph.D. President Tel: 406-585-9535 Fax: 406-585-9219 Email: jgillis@sgmbiotech.com

This summary was prepared on 26 Apr 2010 [as required by Section 807.92(a)(1)]

DEVICE:

Classification Name: Indicator, Biological Sterilization Process

Trade name: Smart-Read™ EZTest® - Steam biological indicator Common name: Self-contained biological indicator for steam [as required by Section 807.92(a)(2)]

CLASSIFICATION: Class II (General Hospital/General Controls)

SUBSTANTIALLY EQUIVALENT OT THE FOLLOWING LEGALLY MARKETED DEVICES:

EZTest® - steam - SGM Biotech, Inc. (K930682) EZTest® - steam - SGM Biotech, Inc. (K963841) [as required by Section 807.92(a)(3)]

FDA Question 5 Page 2 K093794

{1}------------------------------------------------

DESCRIPTION OF DEVICE: Smart-Read EZTest – steam [as required by Section 807.92 (a)(4)]

The biological indicator consists of a self-contained unit which includes: bacterial spores inoculated onto a paper carrier, a small glass ampoule containing sterile culture medium and color indicator, and a plastic vial that serves as the culture tube. This unit complies to the performance characteristics described in USP 31 and ANSI/AAMI/ISO 11138-1.

Spores

Bacterial endospores of Geobacillus stearothermophilus, ATCC 7953 or equivalent, are inoculated onto a 591 Whatman filter paper carrier cut into ¼" x ¾" strips to be placed inside the plastic body of the device. The population of spores on each individual piece of paper will be between 1.0 x 10(x) and 5 x 10(x).

The spore concentration will be labeled as a 10- or 10 concentration.

The population data is based on the heat shock population assay. The heat shock procedure is 95°C to 100°C for 15 minutes as outlined in USP 31.

Culture Tube

The culture tube is made of a polypropylene. The plastic culture tube is 1.75" in length and 0.34" in OD. The 0.02″ wall allows the culture tube to be flexible so that the culture media glass ampoule can be broken to activate the test.

Culture Medium

The culture medium, consisting of a formulated soybean casein digest medium containing bromocresol purple, this medium is filled into ampoules of a Type I borosilicate glass, flame sealed and sterilized. The sealed ampoules are placed into the plastic body with the spore paper.

Filter Paper Vent

The plastic body is closed with a ½" square autoclavable filter paper vent material.

Cap

A polypropylene plastic cap is used to hold the vent material in place. This cap has three small holes in it to allow steam and air to freely move in and out of the body.

FDA Question 5 Page 3 K093794

- Reference USP 21

{2}------------------------------------------------

INCUBATION:

Following exposure to the sterilization process the exposed Smart-Read EZTest is incubated at 60 + 2°C for 10 hours. If spores have survived (sterilization failure) the unit will turn yellow due to the growth of the surviving spores.

PHYSICAL PERFORMANCE:

Reduced incubation time was established by exposing sets of 100 Bls to a steam process that reduced the spore population to approximately 1 spore per Bl. The exposure was acceptable if at least 30 units demonstrated growth and no more than 80 units demonstrated growth. The units were incubated at 60 ± 2°C for 7 days. The reduced incubation time was identified when 97% of the results of the 7 day results were available.

This test was performed with four different spore crops and multiple lots of product from each crop. The data appears below.

Fume and 1 1 Hrs. 1 - 8 Hrs. 3 5 9 Hrs. 3 10 Hrs. 2 10 Hrs. 2 Hrs. 2 Hrs. 2 Hrs. 1 1 2 Hrs.
Partial Cycle #1 *************************************************************************************************************************************************************
Positives0 / 4422 / 4436 / 4443 / 4444 / 4444 / 4444 / 44
Percent Growth04650.0%81.89697.7%100%100%100%
Partial Cycle a2 2000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000 1000
Positives30 / 4541 / 454414544 / 4545 / 4545 / 4545 / 45
Percent Growth66.75691.1%97.8%97.8%100%100%100%
Partlat Cycle #3 20 22 10 12 10 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2yr - Sitter Art Start Time2017 2017334 - 15
Positives8/734917365 / 7371 / 7573/7373 / 7373/73
Percent Growth11.09667.1%89.0%97.3%100%100%100%
Partial Cycle #4 11 12 18 11 12 12 12 12 12 12 12 12 12 12 12 12 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10 10
Positives58 / 8071 / 807618078 / 8080 / 8080 / 8080 / 80
Percent Growth72.5%88.8%95.0%97.5%100%100%100%
Time 20 1 1 2 3 1 1 7 Hrs 1 8 Hrs : 10 Hrs 1 10 Hrs. 1 - 24 Hrs. 1 2 Hrs. 1 2 Hrs. 1 2 168 Hrs.
Partlal Cycle #5 1 = 1 = 1 = 1 = 1 = 1 = 1 = = = = =and and and the same of the comments of the mail of the mail of the
Positives32/5151 / 5252 / 5252 / 5252 / 5252 / 5252 / 52
Percent Growth61.5%98.1%100%100%100%100%100%
Partial Cycle #6 " " 130 " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " " "中国有意義 (1) 1977年 1997年 文章書 (1) 文字 (1) 文字
Positives44 / 5353 / 5323 / 5323 / 5323 / 2323 1 2353 / 53
Percent Growth83.096100%100%100%100%100%100%
Partial Cycle #7 . 3 . 0 . 0 . 0 . 0 . 0 . 1 . 1 . 1 . 1 . 1 . 1 . 1 . 12017 - 12:55
Positives36/7366 / 737217373 / 7373 / 7373 / 7373 / 73
Percent Growth49.3%90.4%98.646100%100%100%100%
Time 3 1 1 1 2 0 1 7 Hrs. 1 8 Hrs. 3 10 Hrs. 3 2 24 Hrs. 2 24 Hrs. 1 2 72 Hrs. 1 2 188 Hrs.
Partial Cycle #8 : - fr. - fr. - fr. - 1. 2 . 1. 1. 1. 1. 1.1 * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * *
Positives26 / 4843 / 4848 / 4848 / 4848 / 4848 / 4848 / 48
Percent Growth52.4%89.696100%100%100%100%100%
Partial Cycle #9- F.: .
Positives30 / 4644 / 464514645 / 46વર્ણવર્ણવર્સ
Percent Growth65.2%95.7%97.8%97.8%100%100%100%
Partial Cycle #10: '- પ્રતિ。. : : *!- -12 - 11 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 - 12 -
Positivesરવ / 60રેક / સ્વિ59 / 60રત દરટવે છે.રત ભ્વસ્ત્ર
Percent Growth90.09696.7%88.39698.34698.36698.3%100%
Partial Cycle #11::.. .
46 / 61-----55/6160 / 61λί-50 / 61
Positives58161୧୦ । ୧।રા । રા
Percent Growth75.4%90. 29695.19698.4%98.4%98.496100%
Time Time Time7 Hrs ! 8 Hrs. 8 9 Hrs. 1 10 Hrs.24 Hrs. 72 Hrs. 1 3 168 Hrs.
Partial Cycle 12 "1 - 1983 - 1ייר.DA Question 5 Page 4
Positives28 / 8071 / 8076 / 8078 / 8080 / 8080 / 8080 / 80
Percent Growth72.5%88.894વે 5 જુ97.5%100%100%100%K093794

Biological Data for Smart-Read Reduced Incubation Time

{3}------------------------------------------------

K093794
ρ⁴δ5

INTENDED USE OF DEVICE: [as required by Section 807.92(a)(5)]

Smart-Read EZTest – Steam self-contained biological indicators (SCBI) are for monitoring the efficacy of pre-vacuum saturated steam sterilization processes. Performance characteristics are established in accordance with USP 31 for the 121°C steam process. Additional pre-vacuum saturated steam sterilization temperatures are also included in the Certificate of Analysis. Smart-Read EZTest – steam SCBIs are also appropriate for use in processes of 132°C, 134°C and 135°C.

The Smart-Read EZTest - steam SCBI is placed into a sterilizer load in a horizontal position in a location that is judged to be the most difficult to sterilize. A sterilization cycle appropriate for the particular type of load is run. Following the sterilization cycle, the operator removes the SCBI from the load and waits approximately 10 minutes until the device has cooled so it can be comfortably handled. The operator then activates the self-contained ampoule by placing it into the activation well in the Smart-Well incubator and gently pulling forward. This action mechanically flexes the side wall of the plastic vial to break the glass ampoule containing the culture medium. The activated Smart-Read EZTest – steam SCBI is then placed in an incubator at 60 ± 2°C for 10 hours which provides conditions conducive to growth of any bacterial spores which might have survived the sterilization cycle. Microbial growth of the indicator organisms will produce acid in the microbiological culture medium containing bromcresol purple, thus lowering the pH of the solution and invoking a yellow color change of the media components. If sterilization conditions are met the spores have been killed and the media will remain clear and purple at the conclusion of the 10 hour incubation time.

SUMMARY OF THE TECHNOLOGICAL CHARACTERISTICS OF Smart-Read EZTest – steam COMPARED TO THE CHARACTERISTICS OF THE PREDICATE DEVICE: [as required by Section 807.92(a)(6)]

Smart-Read EZTest – steam, is similar in type to other products presently in the marketplace whose useful intent is the monitoring of sterilizer efficacy, has been determined to be safe and effective.

Active components of the indicator have been in routine use more than 30 years. They predate the medical device amendment of the Food and Drug Cosmetic Act and are recognized by the United States Pharmacopoeia 31.

The incubation time of 10 hours at 60 + 2°C meets the requirements of the CDRH guidance "Validating Biological Indicator Incubation Time".

FDA Question 5 Page 5 K093794

{4}------------------------------------------------

K093799
p5 of 5

DETERMINATION OF SUBSTANTIALLY EQUIVALENT BASED ON PHYSICAL CHARACTERISTICS: [as required by Section 807.92 (b)(1)]

A comparison is provided of the Smart-Read EZTest – steam to the legally marketed predicate device describing similarities and differences such as technology and other important characteristics.

  • A. EZTest - steam (K930682):
    ls manufactured in precisely the same manner as the Smart-Read EZTest device. The only difference is the extension of the label claims for an incubation temperature of 60 + 2°C and a reduced incubation time (RIT) of 10 hours.

  • B. EZTest steam (K963841):
    This device is exactly the same as the Smart-Read EZTest - steam. The difference is the extension of the label claims for incubation temperature of 60 ± 2°C and the RIT of 10 hours.

DETERMINATION OF SUBSTANTIALLY EQUIVALENCE BASED ON IN USE PERFORMANCE: [as required by Section 807.92 (b)(2)]

The resistance performance of Smart-Read EZTest - steam is exactly the same as EZTest - steam K930682 and K963841.

CONCLUSION: fas required by Section 807.92 (b) (3)]

Smart-Read EZTest – steam is appropriate for monitoring pre-vacuum steam sterilization processes of 121°C, 132°C, 134°C and 135°C and meets the USP 31, ANSI/AAMI/ISO 11138-1, and ANSI/AAMI ST 59:1999 requirements.

FDA Question 5 Page 6 K093794

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol resembling three overlapping lines or strokes, possibly representing a stylized human figure or a symbol of health and well-being.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

John R. Gillis, Ph.D. President SGM Biotech, Incorporated 10 Evergreen Drive, Suite E Bozeman, Montana 59715

MAY 1 4 2010

Re: K093794

Trade/Device Name: Self-Contained Biological Indicator Smart-Read® EZTest® - Steam Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: FRC Dated: April 26, 2010 Received: April 27, 2010

Dear Dr. Gillis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2- Dr. Gillis

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

h for.

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

TAB 9 Indications for Use

KO93794 510(k) Number (if known):

Smart-Read EZTest - Steam Device Name: Self-contained Biological Indicator

Indications for Use:

Smart-Read EZTest – Steam self-contained biological indicators (SCBI) are for monitoring the efficacy of saturated steam sterilization processes. Performance characteristics are established in accordance with USP 31 for the pre-vacuum 121°C steam process. Additional pre-vacuum saturated steam sterilization temperatures are also included in the Certificate of Analysis. Smart-Read EZTest – steam SCBIs are also appropriate for use in saturated steam processes of 132°C, 134°C and 135°C.

The Smart-Read EZTest – steam SCBI is placed into a sterilizer load in a horizontal position in a location that is judged to be the most difficult to sterilize. A sterilization cycle appropriate for the particular type of load is run. Following the sterilization cycle, the operator removes the SCBI from the load and waits approximately 10 minutes until the device has cooled so it can be comfortably handled: The operator then activates the self-contained ampoule by placing it into the activation well in the Smart-Well incubator and gently pulling forward. This action mechanically flexes the side wall of the plastic vial to break the glass ampoule containing the culture medium. The activated Smart-Read EZTest steam SCBI is then placed in an incubator at 60 ± 2°C for 10 hours which provides conditions conducive to growth of any bacterial spores which might have survived the sterilization cycle. Microbial growth of the indicator organisms will produce acid in the microbiological culture medium containing bromcresol purple, thus lowering the pH of the solution and invoking a yellow color change of the media components. If sterilization conditions are met the spores have been killed and the media will remain clear and purple at the conclusion of the 10 hour incubation time.

signature

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

14093794 510(k) Number:

FDA Question 1 Page 2 K093794

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).