GC AADVA IMPLANT SYSTEM

K093749 · GC America, Inc. · DZE · Dec 6, 2010 · Dental

Device Facts

Record IDK093749
Device NameGC AADVA IMPLANT SYSTEM
ApplicantGC America, Inc.
Product CodeDZE · Dental
Decision DateDec 6, 2010
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Intended Use

The GC Aadva Implant System is a titanium alloy screw-type endosseous dental implant and endosseous dental implant abutment, which is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for a prosthetic device, such as an artificial tooth, in order to restore a patient's chewing function. The GC Aadva Implant System is intended for immediate loading only when good primary stability is achieved and with appropriate occlusal loading.

Device Story

GC Aadva Implant System is a screw-type endosseous dental implant system; composed of Ti-6Al-4V ELI titanium alloy. Components include implants and abutments connected via tapered-joint. Implants feature sandblasted and acid-etched surfaces. Used by dental professionals in clinical settings for surgical placement into jaw bone to support prosthetic devices (artificial teeth). Restores patient chewing function. Immediate loading permitted only with sufficient primary stability and controlled occlusal forces.

Clinical Evidence

No clinical test data submitted. Bench testing only. Nonclinical data includes solubility, corrosion/electrochemical properties, cytotoxicity, and fatigue testing of Ti-6Al-4V ELI alloy. Animal testing confirmed suitability for surgical use and compatibility with bone tissue. Packaging validated per ISO 11607-1.

Technological Characteristics

Material: Ti-6Al-4V ELI titanium alloy. Design: Screw-type endosseous implant with tapered-joint abutment connection. Surface: Sandblasted and acid-etched. Sterilization: Validated packaging per ISO 11607-1. Mechanical properties conform to FDA Special Controls Guidance for Root-form Endosseous Dental Implants.

Indications for Use

Indicated for patients requiring dental restoration via endosseous implants in upper or lower jaw arches to support prosthetic devices and restore chewing function. Intended for immediate loading contingent upon achievement of good primary stability and appropriate occlusal loading.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## 5. 510(k) Summary | Submitter/Contact<br>Person | H. Carl Jenkins<br>The Wood Burditt Group<br>DEC - 6 2010<br>FDA Regulatory Counseling<br>10 E. Scranton Avenue, Suite 201<br>Lake Bluff, IL 60044<br>(ph) (847) 234-7500 x 205<br>(fax) (847) 574-0728<br>(email) hcjenkins@woodburditt.com | |-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Applicant | GC America, Inc.<br>3737 W. 127th Street<br>Alsip, IL 60803<br>800.323.3386 x4042<br>708.897.4042<br>708.897.4031 (fax) | | Manufacturer | GC CORPORATION.<br>76-1 HASUNUMA-CHO,<br>ITABASHI-KU<br>TOKYO 174-8585 JAPAN | | Sterilization Facility | RADIA INDUSTRY CO., LTD<br>168 OOYAGI, TAKASAKI<br>GUNNMA 370-0072 JAPAN | | Device Name | GC Aadva Implant System | | Common Name | Root-form Endosseous Dental Implant and Endosseous<br>Dental Implant Abutment | | Classification | Class II<br>Procode DZE<br>Regulation: 21 CFR 872.3640 | 5. 510(k) Summary -0 {1}------------------------------------------------ | Identification of<br>Predicates and Summary<br>of Substantial<br>Equivalence | The GC Aadva Implant System is substantially equivalent<br>with respect to the intended use, design, risks, device<br>characteristics and performance aspects to numerous cleared<br>devices, including: | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | 510(k) / Product / Manufacturer | | | - K072425 / JNE IMPLANT SYSTEM / GC<br>AMERICA, INC. | | | - K971196 / ENDOPORE ENDOSSEOUS DENTAL<br>IMPLANT SYSTEM / INNOVA CORP. | | | - K002513 / ASTRA TECH IMPLANTS - DENTAL<br>SYSTEM / ASTRA TECH, INC. | | | - K033984 / STRAUMANN DENTAL IMPLANT<br>SYSTEM / INSTITUT STRAUMANN AG | | | The GC Aadva Implant System is comparable to the GC JNE<br>Implant System (K072425) in technological characteristics.<br>More specifically, the GC Aadva Implant System and the GC<br>JNE Implant System have the same component material,<br>chemical composition, body type, design shape, engaging<br>method, implant surface treatment and dimensions (lengths<br>and diameters). The GC Aadva Implant System and GC JNE<br>Implant System also have the same indications for use. | | Device Description | The GC Aadva Implant System is an endosseous dental<br>implant made of Ti-6Al-4V ELI alloy and consists of several<br>components. Geometrically, the implant is screw-type. An<br>abutment is connected to the implant through a tapered-joint.<br>Implants are treated with sandblast and acid etching using<br>scanning electron microscopy (SEM). | | | The device functions by being surgically implanted in the<br>bone of the upper or lower jaw arches in order to provide<br>support for a prosthetic device, such as an artificial tooth, in<br>order to restore a patient's chewing function. | | | With regard to the scientific concepts that form the basis for<br>the device, root-form endosseous dental implant devices are<br>characterized by four geometrically distinct types: basket,<br>screw, solid cylinder, and hollow cylinder. The GC Aadva | | | Implant System is a “screw” endosseous dental implant. With<br>regard to the physical and performance characteristics of the<br>GC Aadva Implant System, the design shape, engaging<br>method, implant surface treatment and dimensions (lengths<br>and diameters) are the same as the lawfully marketed<br>predicate GC JNE Implant System. | | Intended Use and<br>Indications | The GC Aadva Implant System is a titanium alloy screw-type<br>endosseous dental implant and endosseous dental implant<br>abutment, which is intended to be surgically placed in the<br>bone of the upper or lower jaw arches to provide support for a<br>prosthetic device, such as an artificial tooth, in order to restore<br>a patient's chewing function.<br>The GC Aadva Implant System is intended for immediate<br>loading only when good primary stability is achieved and with<br>appropriate occlusal loading. | | Performance Testing | <b>I. Nonclinical Data</b><br>Bench testing that studied the solubility, the corrosion and<br>electrochemical properties of the metals, the cytotoxicity, and<br>the fatigue of the GC Aadva Implant System was conducted.<br>Testing indicated that Ti-6Al-4V alloy is estimated to be a<br>corrosion resistant material and not cytotoxic. Furthermore,<br>testing indicated that Ti-6Al-4V ELI alloy has acceptable<br>strength according to the fatigue test results.<br><br>Additionally, the packaging system for sterilized products and<br>non-sterile components of GC Aadva Implant System is the<br>same as that of JNE Implant System (K072425) in constituent<br>materials and the structure, which has been validated<br>according to ISO 11607-1.<br><br>Animal testing that studies titanium alloy's suitability for<br>surgical implant use was conducted. Results indicate that Ti-<br>6Al-4V ELI alloy is suitable for surgical implant use.<br><br>Also, animal testing was conducted to determine the<br>compatibility of the GC Aadva Implant System to bone tissue.<br>Results indicate that Ti-6Al-4V ELI alloy is compatible to<br>bone tissue.<br>Bench testing was conducted in accordance with the | | | | | procedures outlined in the Guidance for Industry and FDA | | | Staff Class II -- Special Controls Guidance Document: Root- | | | form Endosseous Dental Implants. | | | | | | II. Clinical Data | | | No clinical test data is being submitted with this 510(k). | | | III. Conclusions drawn from clinical and nonclinical<br>data indicating that the new device is safe and<br>effective for its intended use and performs as<br>well or better than predicate device. | | | Performance testing data indicates that the new device is safe<br>and effective for its intended use and performs as well or<br>better than predicate device. Specifically, testing demonstrates<br>that component material in the subject device is the same as<br>the component material in a predicate device, and conforms to<br>relevant ISO and ASTM standards. Bench testing conducted<br>in accordance with the procedures outlined in the Guidance<br>for Industry and FDA Staff Class II -- Special Controls<br>Guidance Document: Root-form Endosseous Dental Implants<br>demonstrates that this device meets the mechanical properties<br>recommendations by FDA. | | 5. 510(k) Summary Page 5.2 of 5.4 {2}------------------------------------------------ {3}------------------------------------------------ {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the bird symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 GC America, Incorporated C/O Mr. Carl Jenkins Wood Burditt Group 10 E. Scranton Avenue, Suite 201 Lake Bluff, Illinois 60044 Re: K093749 DEC - 6 2010 Trade/Device Name: GC Aadva Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: November 24, 2010 Received: November 26, 2010 > and the comments of the count . . : Dear Mr. Jenkins: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {5}------------------------------------------------ ## Page 2- Mr. Jenkins Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, th fis Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## 4. Indications for Use ## DEC - 6 2010 510(k) Number (if known): K093749 Device Name: GC Aadva Implant System Indications for Use: The GC Aadva Implant System is a titanium alloy screw-type endosseous dental implant and endosseous dental implant abutment, which is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for a prosthetic device, such as an artificial tooth, in order to restore a patient's chewing function. The GC Aadva Implant System is intended for immediate loading only when good primary stability is achieved and with appropriate occlusal loading. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) > Concurrence of CDRH, Office of Device Evaluation (ODE) times (current > > (Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: k093741 Page 4.1 of 4.1 4. Indications For Use
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%