K Number
K093736

Validate with FDA (Live)

Date Cleared
2011-03-10

(461 days)

Product Code
Regulation Number
888.3353
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DePuy Corail Revision Stem is indicated for cementless use in the treatment of:

  1. A severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, rheumatoid arthritis or congenital hip dysplasia;
  2. Avascular necrosis of the femoral head;
  3. Acute traumatic fracture of the femoral head or neck;
  4. Failed previous surgery, including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthoplasty, or other total hip replacement; and
  5. Certain cases of ankylosis.

The subject Corail Revision Stem is intended for cementless use as part of the femoral component in a total hip arthroplasty.

Device Description

The Corail Revision Stem is part of a modular prosthesis system for use in total hip replacement. It mates with DePuy femoral heads which articulate with DePuy acetabular systems with either a metal or UHMPE liner. The Corail Hip System, Revision Stem is a line addition to the Corail hip stem family. The stems are available in 9 lengths in sizes 10-20 in standard and high-off versions. The subject titanium alloy hip stems are monolithic, grit blasted stems plasma sprayed with hydroxyapatite powders on the entire stem length. The stems feature a 12/14 AMT (Articul/eze Mini Taper) Taper trunnion which are compatible with previously cleared metal and ceramic DePuy femoral heads with corresponding 12/14 tapers and offsets up to +13 mm.

AI/ML Overview

This 510(k) summary (K093736) describes the DePuy Corail® Hip System, Revision Stem.

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria in a table format. Instead, it indicates that "All of the observed results indicate that the Corail Revision Hip System is substantially equivalent to devices currently marketed." This implies that the performance matched or exceeded what is considered acceptable for the predicate devices.

Acceptance Criteria (Implied)Reported Device Performance
Substantial equivalence to predicate devices in:Substantially equivalent to predicate devices in intended use, indications for use, materials, geometry, design, and performance.
Non-clinical testing results for:Bench testing (Range of Motion, proximal fatigue, ceramic head compatibility) and coating characterization were provided.
Coating thickness and applicationNonporous, calcium phosphate coating applied by plasma spraying to a nominal thickness of 155 ± 35 microns from hydroxyapatite powders.

(Note: The 510(k) process for this type of device typically relies on demonstrating substantial equivalence to a legally marketed predicate device rather than setting specific numerical performance targets that need to be met independently for a novel technology.)

2. Sample Size and Data Provenance for Test Set

  • Sample Size: Not applicable. This submission relies on substantial equivalence and non-clinical testing, not a clinical test set with patient data.
  • Data Provenance: Not applicable. The testing was non-clinical bench testing and coating characterization.

3. Number of Experts and Qualifications for Ground Truth (Test Set)
Not applicable. No clinical test set with ground truth established by experts was used.

4. Adjudication Method (Test Set)
Not applicable. No clinical test set with an adjudication method was used.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. The submission states, "No clinical testing was required to demonstrate substantial equivalence." This type of study would involve human readers and clinical data, which was not part of this submission. Since no such study was performed, there is no effect size reported for human readers improving with or without AI assistance.

6. Standalone (Algorithm Only) Performance Study
Not applicable. This device is a medical implant (hip stem), not an AI algorithm. Therefore, a standalone performance study of an algorithm was not conducted.

7. Type of Ground Truth Used
Not applicable. For this device, ground truth as typically understood in AI/imaging studies (e.g., expert consensus, pathology, outcomes data) was not established from clinical data. The demonstration of safety and effectiveness was based on non-clinical testing and substantial equivalence to predicate devices.

8. Sample Size for Training Set
Not applicable. This device is a medical implant, not an AI algorithm. There is no concept of a "training set" in this context.

9. How Ground Truth for Training Set Was Established
Not applicable. As this is not an AI algorithm, there was no training set or ground truth to establish for training purposes.

{0}------------------------------------------------

长 093736

510 (k) Summary

:

:

MAR 1 0 2011

(As required by 21 CFR 807.92 and 21 CFR 807.93)

NAME OF SPONSOR:DePuy Orthopaedics, Inc.700 Orthopaedic DriveWarsaw, Indiana 46582Establishment Registration Number: 1818910
510(K) CONTACT:Rhonda MyerSenior Regulatory Affairs AssociateTelephone: (574) 371-4927Facsimile: (574) 371-4987Electronic Mail: Rmyer7@its.jnj.com
DATE PREPARED:March 1, 2011
PROPRIETARY NAME:DePuy Corail® Hip System, Revision Stem
COMMON NAME:Hydroxyapatite-coated hip prosthesis
CLASSIFICATION:Class III per 21 CFR 888.3330: Hip jointmetal/metal semi-constrained, with anuncemented acetabular component, prosthesis(KWA)Class II per 21 CFR 888.3353: Hip jointmetal/ceramic/polymer semi-constrainedcemented or nonporous uncemented prosthesis(LZO)Class II per 21 CFR 888.3353: Hip jointmetal/ceramic/polymer semi-constrainedcemented or nonporous uncemented prosthesis(MEH)Class II per 21 CFR 888.3360: Hip jointfemoral (hemi-hip) metallic cemented oruncemented prosthesis (LWJ)
DEVICE PRODUCT CODE ANDDESCRIPTION:KWA: prosthesis, hip, semi-constrained (metaluncemented acetabular component)LZO: prosthesis, hip, semi-constrained,metal/ceramic/polymer, cemented or non-porous,uncementedMEH: prosthesis, hip, semi-constrained,uncemented, metal/polymer, non-porous,calcium-phosphateLWJ: prosthesis, hip, semi-constrained,metal/polymer, uncemented

{1}------------------------------------------------

SUBSTANTIALLY EQUIVALENT DEVICES:

DePuy Corail Hip Stem, K042992 DePuy KAR™ Hip Stem, K961732

DEVICE DESCRIPTION:

The Corail Revision Stem is part of a modular prosthesis system for use in total hip replacement. It mates with DePuy femoral heads which articulate with DePuy acetabular systems with either a metal or UHMPE liner. The Corail Hip System, Revision Stem is a line addition to the Corail hip stem family. The stems are available in 9 lengths in sizes 10-20 in standard and high-off versions. The subject titanium alloy hip stems are monolithic, grit blasted stems plasma sprayed with hydroxyapatite powders on the entire stem length. The stems feature a 12/14 AMT (Articul/eze Mini Taper) Taper trunnion which are compatible with previously cleared metal and ceramic DePuy femoral heads with corresponding 12/14 tapers and offsets up to +13 mm.

INDICATIONS AND INTENDED USE:

Indications:

The DePuy Corail Revision Stem is indicated for cementless use in the treatment of:

  • I. A severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, rheumatoid arthritis or congenital hip dysplasia;
    1. Avascular necrosis of the femoral head;
    1. Acute traumatic fracture of the femoral head or neck;
    1. Failed previous surgery, including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthoplasty, or other total hip replacement; and
    1. Certain cases of ankylosis.

Intended Use:

The subject Corail Revision Stem is intended for cementless use as part of the femoral component in a total hip arthroplasty.

Summary of Technologies/Substantial Equivalence:

The substantial equivalence of the subject Corail Revision Stem is demonstrated by similarities in intended use, indications for use, materials, geometry, design and performance as compared to the predicate devices. The changes presented in this 510(k) do not present new issues of safety or effectiveness as the technological characteristics are basically the same as the predicates. Based on the materials, geometry, mechanical testing and indications for use, the DePuy Corail Revision Hip System is considered to be substantially equivalent to the predicate devices.

Non-clinical Testing:

{2}------------------------------------------------

Non-clinical testing and analysis were provided, including bench testing and coating characterization. Bench testing included Range of Motion, proximal fatigue and ceramic head compatibility testing. The nonporous, calcium phosphate coating was applied by plasma spraying to a nominal thickness of 155 ± 35 microns from hydroxyapatite powders. The calcium phosphate coating is applied by a different vendor and was characterized per FDA's "510(k) Information needed for Hydroxyapatite Coated Orthopedic Implants." All of the observed results indicate that the Corail Revision Hip System is substantially equivalent to devices currently marketed. Therefore, the subject device is as safe, as effective, and performs at least as safely and effectively as legally marketed predicates.

Clinical Testing:

No clinical testing was required to demonstrate substantial equivalence.

Conclusion:

The subject Corail Revision Stem is substantially equivalent to the predicate devices identified in this premarket notification.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with three curved lines representing its wings or body.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DePuy Orthopaedics. Inc. % Ms. Rhonda Myer Senior Regulatory Affairs Associate 700 Orthopaedic Drive Warsaw, Indiana 46581-0988

MAR 1 0 2011

Re: K093736

Trade/Device Name: DePuy Corail Hip System, Revision Stem Regulation Number: 21 CFR 888.3330 Regulation Name: Hip joint metal/metal semi-constrained, with an uncemented acetabular component, prosthesis Regulatory Class: Class III Product Code: KWA, LZO, MEH, LWJ Dated: February 17, 2011 Received: February 18, 2011

Dear Ms. Myer:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

Page 2 -- Ms. Rhonda Myer

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

A.S. B. r2 h
for

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use Statement

093736 510 (k) Number (if known):

Device Name: DePuy Corail® Hip System, Revision Stem

Indications for Use:

The DePuy Corail Revision Stem is indicated for cementless use in the treatment of:

    1. A severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis, or congenital hip dysplasia;
    1. Avascular necrosis of the femoral head;
    1. Acute traumatic fracture of the femoral head or neck;
    1. Failed previous hip surgery, including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthoplasty, or other total hip replacement; and
    1. Certain cases of ankylosis.
Prescription UseX
(Part 21 CFR 801 Subpart D)AND/OR
Over-The-Counter Use
(21 CFR 807 Subpart C)

(Please do not write below this line. Continue on another page if needed.)

(Division Sign Off) for M. Melkerson

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K093736

DePuy Orthopaedics, Inc.

Corail Hip System, Revision Stem Traditional 510(k)

Page 31 of 364

§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.

(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.