K Number
K093678

Validate with FDA (Live)

Manufacturer
Date Cleared
2011-01-13

(412 days)

Product Code
Regulation Number
866.3040
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Platelia™ Aspergillus EIA is an immunoenzymatic sandwich microplate assay for the detection of Aspergillus galactomannan antigen in adult and pediatric serum and Bronchoalveolar Lavage (BAL) fluid samples.

The Platelia™ Aspergillus EIA is a test which, when used in conjunction with other diagnostic procedures such as microbiological culture, histological examination of biopsy samples and radiographic evidence, can be used as an aid in the diagnosis of Invasive Aspergillosis.

Device Description

The Platelia™ Aspergillus EIA is a one-stage immunoenzymatic sandwich microplate assay which detects galactomannan in human serum and BAL fluid samples. The assay uses rat EBA-2 monoclonal antibodies, which are directed against Aspergillus galactomannan, and have been characterized in previous studies. The monoclonal antibodies are used, (1) to coat the wells of the microplate and bind the antigen, and (2) to detect the antigen bound to the sensitized microplate reagent: peroxidase-linked monoclonal antibodies).

Serum or BAL fluid samples are heat-treated in the presence of EDTA in order to dissociate immune complexes and to precipitate proteins that could possibly interfere with the test. The treated samples and conjugate are added to the wells coated with monoclonal antibodies, and incubated. A monoclonal antibody - galactomannan monoclonal antibody / peroxidase complex is formed in the presence of galactomannan antigen.

The strips are washed to remove any unbound material. Next, the substrate solution is added, which will react with the complexes bound to the well to form a blue color reaction. The enzyme reaction is stopped by the addition of acid, which changes the blue color to yellow. The absorbance (optical density) of specimens and controls is determined with a spectrophotometer set at 450 and 620/630 nm wavelength.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

Device: Platelia™ Aspergillus EIA (Catalog 62793)
Intended Use: Immunoenzymatic sandwich microplate assay for the detection of Aspergillus galactomannan antigen in adult and pediatric serum and Bronchoalveolar Lavage (BAL) fluid samples, as an aid in the diagnosis of Invasive Aspergillosis when used with other diagnostic procedures.


1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity and specificity. Instead, it presents the performance results. Assuming the "reported device performance" are the de facto acceptance criteria for this 510(k) submission, here are the key performance metrics:

Metric (Sample Type)Acceptance Criteria (from study results)Reported Device Performance
Serum Samples
Sensitivity (Pediatric Patients)31.0-73.8% (95% CI for combined Proven & Probable)52.9% (9/17)
Sensitivity (Adult Patients)61.6-90.2% (95% CI for combined Proven & Probable)79.3% (23/29)
Specificity (Pediatric Patients)79.4-92.1% (95% CI for combined sites)87.0% (94/108)
Specificity (Pediatric Samples)97.7-99.0% (95% CI for combined sites)98.5% (1600/1625)
Specificity (Adult Patients)82.6-93.0% (95% CI for combined sites)88.8% (127/143)
Specificity (Adult Samples)97.7-99.0% (95% CI for combined sites)98.5% (1243/1262)
BAL Fluid Samples
Sensitivity (SOT Recipients)56.5-100% (95% CI for combined P+P)100% (5/5)
Sensitivity (Lung Transplant Recipients)30.0-90.3% (95% CI for combined P+P)66.7% (4/6)
Sensitivity (Hematologic Disease Patients)90.8-99.7% (95% CI for combined P+P)98.3% (57/58)
Specificity (Combined SOT & Lung Transplant Recipients, without colonization)94.3-98.2% (95% CI)96.8% (330/341)
Specificity (Combined SOT & Lung Transplant Recipients, with colonization)69.1-88.6% (95% CI)80.6% (50/62)
Specificity (Combined SOT & Lung Transplant Recipients, Total)91.6-96.2% (95% CI)94.3% (380/403)
Specificity (Hematologic Disease Patients)66.0-89.8% (95% CI)80.5% (33/41)

The acceptance criteria appear to be implicitly defined by the sponsor's demonstration of these performance characteristics, particularly with confidence intervals, which indicate variability.

2. Sample Size Used for the Test Set and Data Provenance

  • Serum Samples:

    • Pediatric Patients: 1954 serum samples from 129 immunocompromised pediatric patients (age ≤ 21 years) at high risk for Invasive Aspergillosis (IA), covering those diagnosed with Proven and Probable IA and controls.
      • Controls: 1625 serum samples from 108 pediatric patients.
      • IA Diagnosed: 249 serum samples from 17 pediatric patients.
    • Adult Patients: 1724 serum samples from 172 bone marrow transplant (BMT) and leukemic patients, covering those diagnosed with and without IA.
      • Controls: 1262 serum samples from 143 adult patients.
      • IA Diagnosed: 462 serum samples from 29 adult patients.
    • Provenance: "three testing centers in the United States" for pediatric studies and "three testing centers in North America" for adult studies. The studies are described as clinical studies, implying prospective collection of samples related to patient diagnosis/monitoring.
  • BAL Fluid Samples:

    • Study 1 (SOT & Lung Transplant): 449 BAL samples from 178 Solid Organ Transplant (SOT) and lung transplant recipients.
      • Controls: 403 BAL samples from 167 SOT and lung transplant recipients (United States).
      • IA Diagnosed (SOT): 5 recipients.
      • IA Diagnosed (Lung): 6 recipients.
    • Study 2 (Hematology Patients): 99 evaluable BAL samples from 99 high-risk hematology patients.
      • IA Diagnosed: 58 patients.
      • Controls: 41 patients.
    • Provenance: Two studies from the "United States" (SOT & Lung Transplant) and one "retrospective analysis... from a study outside the United States" (Hematology Patients).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The document does not specify the number or qualifications of experts used to establish the ground truth. It states that diagnoses of Invasive Aspergillosis (Proven or Probable) were defined by EORTC/NIAID definitions. This implies that the ground truth was established based on these standardized clinical and microbiological criteria, likely applied by clinical specialists at the participating sites, rather than an independent panel of experts reviewing cases.

4. Adjudication Method for the Test Set

The document does not describe a specific adjudication method (e.g., 2+1, 3+1) for establishing the ground truth. The reliance on EORTC/NIAID definitions suggests a criteria-based diagnosis rather than a read-by-committee adjudication process.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The studies evaluate the performance of the device itself (standalone algorithm) against a clinical diagnosis (ground truth), not in a human-in-the-loop context or comparing human reader performance with and without AI assistance.

6. Standalone Performance Study

Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The entire performance evaluation section describes the sensitivity and specificity of the Platelia™ Aspergillus EIA (the device/algorithm) when tested against various patient samples. The results presented are exclusively based on the device's output (index values) compared to the patient's clinical diagnosis.

7. Type of Ground Truth Used

The ground truth used was clinical diagnosis based on established criteria, specifically the Invasive Fungal Infection Cooperative Group (IFICG) of the European Organization for Research and Treatment of Cancer (EORTC) and the Mycosis Study Group (MSG) of the National Institute of Allergy and Infectious Diseases (NIAID) definitions for Proven and Probable Invasive Aspergillosis. Controls were defined as patients without signs of Invasive Aspergillosis.

8. Sample Size for the Training Set

The document does not mention a separate training set or its sample size. The provided performance evaluation focuses entirely on the "test set" or clinical validation set. Based on the context, this device is a laboratory assay, not a machine learning algorithm requiring a distinct training phase in the same way an AI imaging algorithm might. The reproducibility studies detail the use of "panel members" which are likely used for internal quality control and assay validation, but not a "training set" in the context of developing an AI model.

9. How the Ground Truth for the Training Set Was Established

As no specific training set for an AI model is described, this question is not applicable. The device is a diagnostic assay, and its development would typically involve assay design, optimization, and analytical validation rather than an AI training process.

{0}------------------------------------------------

.

:

·

510k Summary K093678

SECTION 2

510(k) SUMMARY

{1}------------------------------------------------

510k Summary K093678

510(k) SUMMARY:

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is:
DATE:November 20th, 2009
APPLICANT:Bio-Rad3, Boulevard Raymond Poincar92430 Marnes-la-Coquette, France
OFFICIAL CORRESPONDENT:Dr. Sylvie Confida
TELEPHONE:33-1-47-95-6138
FAX:33-1-47-95-6242
PRODUCT TRADE NAME:Bio-Rad Laboratories Platelia™ Aspergillus EIA
COMMON NAME:Aspergillus Antigen EIA
CLASSIFICATION NAME:Antigen, Galactomannan, Aspergillus spp.
PREDICATE DEVICE:Platelia™ Aspergillus EIA

DEVICE DESCRIPTION

The Platelia™ Aspergillus EIA is a one-stage immunoenzymatic sandwich microplate assay which detects galactomannan in human serum and BAL fluid samples. The assay uses rat EBA-2 monoclonal antibodies, which are directed against Aspergillus galactomannan, and have been characterized in previous studies. The monoclonal antibodies are used, (1) to coat the wells of the microplate and bind the antigen, and (2) to detect the antigen bound to the sensitized microplate reagent: peroxidaselinked monoclonal antibodies).

Serum or BAL fluid samples are heat-treated in the presence of EDTA in order to dissociate immune complexes and to precipitate proteins that could possibly interfere with the test. The treated samples and conjugate are added to the wells coated with monoclonal antibodies, and incubated. A monoclonal antibody - galactomannan monoclonal antibody / peroxidase complex is formed in the presence of galactomannan antigen.

{2}------------------------------------------------

The strips are washed to remove any unbound material. Next, the substrate solution is added, which will react with the complexes bound to the well to form a blue color reaction. The enzyme reaction is stopped by the addition of acid, which changes the blue color to yellow. The absorbance (optical density) of specimens and controls is determined with a spectrophotometer set at 450 and 620/630 nm wavelength.

INTENDED USE

The Platelia™ Aspergillus ElA is an immunoenzymatic sandwich microplate assay for the detection of Aspergillus galactomannan antigen in adult and pediatric serum and Bronchoalveolar Lavage (BAL) fluid samples.

The Platelia™ Aspergillus EIA is a test which, when used in conjunction with other diagnostic procedures such as microbiological culture, histological examination of biopsy samples and radiographic evidence, can be used as an aid in the diagnosis of Invasive Aspergillosis.

INDICATIONS FOR USE

The Platelia™ Aspergillus EIA is an immunoenzymatic sandwich microplate assay for the detection of Aspergillus galactomannan antigen in adult and pediatric serum and Bronchoalveolar Lavage (BAL) fluid samples.

The Platelia™ Aspergillus EIA is a test which, when used in conjunction with other diagnostic procedures such as microbiological culture, histological examination of biopsy samples and radiographic evidence, can be used as an aid in the diagnosis of Invasive Aspergillosis.

TECHNOLOGICAL CHARACTERISTICS

The following tables summarize similarities and differences between the Platelia™ Aspergillus EIA (62793) and the current Platelia™ Aspergillus EIA (K060641):

Table 1(a) Similarities between intended use

Similarities inFunction and UsePlatelia™ Aspergillus EIA,Catalog 62793Platelia™ Aspergillus EIAcurrent (K060641)
Intended UseGalactomannan antigendetection.Galactomannan antigendetection.

{3}------------------------------------------------

.

Differences inFunction and UsePlatelia™ Aspergillus EIA,Catalog 62793Platelia™ Aspergillus EIA(K060641)
Intended UseDetection of Aspergillusgalactomannan antigen inadult and pediatric serum andBronchoalveolar Lavage(BAL) Fluid samples.Detection of Aspergillusgalactomannan antigen in adultand pediatric serum samples.
MatrixSerum and BronchoalveolarLavage (BAL) Fluid samplesSerum
Intended Use/Indications for UseBoth the Intended Use andIndications for Use are same.The Intended Use andIndications for Use aredifferent.

Table 1(b) Differences between intended use

Table 2 Similarities between reagents and materials

Similarities inComponents /MaterialsPlatelia™ Aspergillus EIA,Catalog 62793Platelia™ Aspergillus EIA(K060641)
Microplate96 well microplate -antibody coated microwells96 well microplate - antibodycoated microwells
ReagentsConjugate, Wash Buffer,Substrate, TMB Chromogen,Sample Diluent, PositiveControl, Stop Solution.Conjugate, Wash Buffer,Substrate, TMB Chromogen,Sample Diluent, PositiveControl, Stop Solution.

Table 3 Similarities between assay procedures.

Similarities inAssay ProceduresPlatelia™ Aspergillus EIACatalog 62793Platelia™ Aspergillus EIA(K060641)
Incubationtemperature of themicroplate afteraddition of theconjugate and thetreated sera.Incubation temperature: 37°CIncubation temperature: 37°C
Incubation time ofthe microplate afteraddition of theconjugate and thetreated sera.Incubation time: 90 ± 5minutesIncubation time: 90 ± 5minutes

{4}------------------------------------------------

Differences inLimitations of theProcedure (Section13)Platelia™ Aspergillus EIA,Catalog 62793Platelia™ Aspergillus EIA(K060641)
Limitation: Point 11Addition of Histoplasma andGeotrichum to the list offungi causing cross-reactivity
Limitation: Point 12Addition of a limitation thatcross-reactivity of BAL fluidwith Mycoplasma andanaesthetic drugs used tonumb the neck/throat areahas not been evaluated.
Limitation: Point 13Addition of limitationregarding positive reactionsobserved in patientsreceiving productscontaining sodium gluconate,galactofuranose orgalactomannan.
Limitation: Point 14Addition of limitationregarding positive reactionsobserved in patientsreceiving PLASMA-LYTE™ solution.
Limitation: Point 15Addition of limitation aboutassay results in BAL fluidsamples fromimmunocompromisedpatients to be interpretedwith caution.

Table 4 Differences between Limitations of the Procedure

{5}------------------------------------------------

Limitation: Point 16Results of the Platelia™
Aspergillus EIA in
Bronchoalveolar Lavage
(BAL) fluid samples between
0.5-1.0 index have a lower
predictive value than BAL
sample results > 1.0 index
values, therefore the results
between 0.5-1.0 index values
should be reviewed and
supported by other clinical,
radiological or laboratory
evidence of invasive
aspergillosis8, 17

PERFORMANCE EVALUATION SUMMARY

A. EXPECTED VALUES

I. SERUM

The expected prevalence of Invasive Aspergillosis varies with the patient population: rates from 5-20% have been reported.

The following results have been obtained from clinical studies conducted on pediatric (age < 21 years) patients in the United States and on adult patients in North America.

A.-Pediatrics

A clinical study was conducted on a total of 1954 serum samples from 129 immunocompromised pediatric (Age ≤ 21 years) patients, at high risk for Invasive Aspergillosis (IA) and patients diagnosed with Proven and Probable Invasive Aspergillosis, at three testing centers in the United States to determine the performance characteristics of the Platelia™ Aspergillus EIA. The distribution of index for these populations is shown in the following charts:

Pediatric Patients diagnosed without Invasive Aspergillosis (control population)

A total of 1625* pediatric serum samples obtained from 108 immunocompromised pediatric patients at three testing centers in the United States were tested to determine the performance characteristics of the Platelia™ Aspergillus EIA. The distribution of index values for samples is shown in the following chart:

{6}------------------------------------------------

Image /page/6/Figure/2 description: The image shows the words "Figure I" in a bold, serif font. The word "Figure" is capitalized, and the Roman numeral "I" is also capitalized. The text is black against a white background. The image appears to be a label or title for a figure or diagram.

Image /page/6/Figure/3 description: The image is a bar graph titled "Distribution of the Serum Index Value from the Pediatric Control Population (N=1625)". The x-axis is labeled "Index" and shows values from 0-0.1 to >1.5. The y-axis is labeled "Number of Sera" and ranges from 0 to 900. The bar graph shows that the highest number of sera is in the 0.2-0.3 range, with a value of 805.

*Note: 80 samples, from 4 control patients with positive galactomannan antigen results coinciding with piperacillin/tazobactam (Zosyn®) therapy were excluded.

Pediatric Patients diagnosed with Invasive Aspergillosis

The scatter plot depicts galactomannan assay results for the 249 serum samples from 17 patients in this study diagnosed with proven or probable Invasive Aspergillosis as defined by EORTC/NIAID definitions. Not every serum sample from each patient is expected to be positive. The expected prevalence of Invasive Aspergillosis varies with the patient population; rates from 5-20% have been reported 10.23. The prevalence rate of this study was 13.6%.

Figure 2

Image /page/6/Figure/8 description: The image is a scatter plot titled "Pediatric Proven/ Probable Aspergillosis: Distribution of Index/ Pediatric Patient (N=17)". The x-axis is labeled "Patient Number" and ranges from 0 to 18. The y-axis is labeled "Index" and ranges from 0.0 to >2.0. The scatter plot shows the distribution of index values for each patient, with a horizontal line at 0.5.

{7}------------------------------------------------

B. Adults

A clinical study was conducted on a total of 1724 serum samples from 172 bone marrow transplant (BMT) and leukemic patients diagnosed with and without Invasive Aspergillosis, at three testing centers in North America to determine the performance characteristics of the Platelia™ Aspergillus EIA. The distribution of index values for these populations is represented in the following charts.

Adult Patients diagnosed without Invasive Aspergillosis (control population)

A total of 1262 serum samples obtained from 143 bone marrow transplant (BMT) and leukemic patients at three testing centers in North America were tested with the Platelia™ Aspergillus EIA test. The distribution of index values is shown in the following chart:

Image /page/7/Figure/6 description: The image shows the text "Figure 3" in a bold, serif font. The text is black against a white background. The number 3 is slightly smaller than the word "Figure".

Image /page/7/Figure/7 description: The image is a bar graph titled "Distribution of the Serum Index Value from the Adult Control Population (N=1262)". The x-axis is labeled "Index" and shows index ranges from 0-0.1 to >1.5. The y-axis is labeled "Number of Sera" and ranges from 0 to 800. The bar graph shows the number of sera for each index range, with the highest number of sera in the 0.2-0.3 range (662).

Adult Patients diagnosed with Invasive Aspergillosis

This scatter plot depicts galactomannan assay results for the 462 serum samples from 29 patients in this study diagnosed with proven or probable Invasive Aspergillosis as defined by EORTC/NIAID definitions. Not every serum sample from each patient is expected to be positive. The expected prevalence of Invasive Aspergillosis varies with the patient population; rates from 5-20% have been reported 10,23. The prevalence rate for this study was 16.9%.

{8}------------------------------------------------

Image /page/8/Figure/2 description: The image shows the words "Figure 4" in a serif font. The words are in bold and are black. There is a period after the number 4.

Image /page/8/Figure/3 description: The image is a scatter plot titled "Adult Proven/Probable Aspergillosis: Distribution of Index/Adult Patient (N=29)". The x-axis is labeled "Patient Number" and ranges from 0 to 30. The y-axis is labeled "Index" and ranges from 0.0 to greater than or equal to 2.0. The scatter plot shows the distribution of index values for each patient, with a horizontal line at 0.5.

The following graphs represent examples of a patient without clinical signs or symptoms óf Invasive Aspergillosis (negative for Aspergillus) and a patient with proven or probable Invasive Aspergillosis (positive for Aspergillus) respectively.

Image /page/8/Figure/5 description: This image contains the text "Figure 5 Negative patient". The text is in bold font. The figure number is at the top, and the description of the figure is below it.

Image /page/8/Figure/6 description: The image is a graph titled "CONTROL PATIENT". The x-axis is labeled "Days" and ranges from 0 to 60. The y-axis is labeled "Index" and ranges from 0 to 1.6. The graph shows a line with diamond markers that fluctuates between 0 and 0.3, with a few peaks around 0.2, and a horizontal line at 0.5.

{9}------------------------------------------------

Figure 6 Positive patient

Image /page/9/Figure/3 description: This image is a line graph titled "PROVEN INVASIVE ASPERGILLOSIS PATIENT". The x-axis is labeled "Days" and ranges from 0 to 60. The y-axis is labeled "Index" and ranges from 0 to 1.6. The line on the graph starts at approximately 0.2, rises sharply around day 25, peaks around day 30 at approximately 1.5, and then gradually declines to around 0.4 by day 55.

II. BAL FLUID

Two studies were conducted on a total of 449 BAL samples from 178 Solid Organ transplant (SOT) and lung transplant recipients with and without invasive aspergillosis in the United States to determine the performance characteristics of the Platelia™ Aspergillus EIA kit with Bronchoalveolar Lavage Fluid samples.

Of these, there were 403 BAL samples from 167 solid organ and lung transplant recipients without invasive aspergillosis.

In addition, a retrospective analysis was performed on BAL samples from 99 evaluable high risk haematology patients in a study outside the United States which included 58 patients with proven or probable invasive aspergillosis.

Expected values in BAL samples from the combined SOT and lung transplant recipients without Invasive Aspergillosis are presented in the table below. Results are presented by samples from transplant recipients with and without mold colonization.

Table 1

Expected Values by Sample Combined SOT and Lung Transplant Recipients without Invasive Aspergillosis N =403 BAL Fluids

DiagnosisNPositive (%)Negative (%)
Controls withoutcolonization34111/341(3.2%)330/341(96.8%)
Controls with colonization6212/62(19.4%)50/62 (80.6%)
Control Total40323/403(5.7%)380/403(94.3%)

Expected values in BAL samples from the combined SOT and lung transplant recipients without Invasive Aspergillosis are presented by transplant type in the table below.

{10}------------------------------------------------

Table 2

Expected Values by Sample Combined SOT and Lung Transplant Recipients without Invasive Aspergillosis By Transplant Type N =403 RAL Fluids

N - 403 BAL Fluids
Transplant TypeNPositive (%)Negative (%)
Heart283/28 (10.7%)25/28 (89.3%)
Kidney253/25 (12.0%)22/25 (88.0%)
Liver231/23 (4.3%)22/23 (95.7%)
Lung32716/327 (4.9%)311/327(95.1%)
Control Total40323/403(5.7%)380/403(94.3%)

Expected values were also evaluated in a total of 41 BAL fluid samples from 41 hematological disease patients without Invasive Aspergillosis and are presented in the Table below

Table 3

Expected Values by Sample Hematologic disease patients without Invasive Aspergillosis N =41 BAL Fluids

Alleria Million Art Program And any Art All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All All AlAND AND A B A A Libray on Street StatesDiagnosisALL House of the American"Whileser deberg for working with Address of the Address of the Artistics of Articles of Articles of Articles of Articles of Articles of Articles of Articles of Articles of APositive (%)Negative (%)
Controls8/41 (19.5%)33/41 (80.5%)

B. REPRODUCIBILITY STUDIES

a) Reproducibility Studies In Serum

Inter-assay and Intra-assay variability for the Platelia™ Aspergillus EIA were determined in a study using a panel of 6 pooled patient serum samples (one negative, one low positive, two positive, and two high positive) obtained at three clinical trial sites in North America. Each of the 6 panel members were tested in triplicate (x3) on 3 different days, on one lot, at two sites (total number of replicates at each site = 9). Each of the 6 panel members was tested in duplicate (x2) on 3 different days, on 1 lot, at a third site (total number of replicates at the third site = 6). One (1) operator performed all precision testing at each site. The data were analyzed according to the Clinical Laboratory Standards Institute (CLSI) (formerly National Committee for Clinical Laboratory Standards (NCCLS)). The mean optical density (OD) and mean index value, standard deviation (SD), percent coefficient of variation (%CV), within run precision (intraassay) and within site (inter-assay) precision for each panel member at each site are illustrated below in the following tables.

{11}------------------------------------------------

Table 4

C:
1
Site 1Panel MemberNegLow PosPos #1Pos #2High Pos#1High Pos #2Neg ControlCO ControlPos Control
ODIndexODIndexODIndexODIndexODIndexODIndexODIndexODIndexODIndex
N999999999999336633
Mean0.0520.090.4450.740.7021.170.9311.5631.2272.062.8874.830.0460.080.6061.002.2163.67
Within Run(intra-assay)1 SD0.0020.000.0220.030.0590.090.0440.080.0510.090.0890.17N/AN/A0.020.03N/AN/A
%CVN/AN/A4.8%4.4%8.4%7.6%4.7%5.1%4.2%4.4%3.1%3.6%N/AN/A3.7%3.4%N/AN/A
Total(inter-assay)2 SD0.0360.040.0510.080.0700.140.0440.250.0580.290.1690.58N/AN/A0.1020.030.3170.12
%CVN/AN/A11.5%10.4%10.0%11.6%4.7%15.7%4.7%14.3%5.9%11.9%N/AN/A16.9%2.8%14.3%3.3%
Site 2Panel MemberNegLow PosPos #1Pos #2High Pos#1High Pos #2Neg ControlCO ControlPos Control
ODIndexODIndexODIndexODIndexODIndexODIndexODIndexODIndexODIndex
N999999999999336633
Mean0.0400.100.2800.700.3640.890.6021.490.8012.011.3813.430.0740.180.4151.001.1972.97
Within Run(intra-assay)1 SD0.0060.010.0410.090.0230.070.0450.110.0460.100.0470.11N/AN/A0.000.01N/AN/A
%CVN/AN/A14.5%13.0%6.4%7.6%7.5%7.1%5.7%4.8%3.5%3.2%N/AN/A1.1%1.1%N/AN/A
Total(inter-assay)2 SD0.0060.030.0580.190.0830.180.0570.280.0420.530.0791.00N/AN/A0.0940.010.0680.54
%CVN/AN/A20.8%27.0%22.7%19.8%9.5%18.7%5.3%26.5%5.8%29.2%N/AN/A22.7%0.9%5.7%18.2%
Site 3Panel MemberNegLow PosPos #1Pos #2High Pos#1High Pos #2Neg ControlCO ControlPos Control
ODIndexODIndexODIndexODIndexODIndexODIndexODIndexODIndexODIndex
N666666666666336633
Mean0.0490.100.3880.810.6521.360.8301.731.1582.412.3784.960.0590.120.4801.001.6523.45
Within Run(intra-assay)1 SD0.0030.010.0090.020.0820.170.0680.140.0940.200.1260.25N/AN/A0.0280.06N/AN/A
%CVN/AN/A2.4%2.4%12.5%12.2%8.2%8.2%8.1%8.2%5.3%5.1%N/AN/A5.8%5.8%N/AN/A
Total(inter-assay)2 SD0.0120.030.0780.130.0680.150.1040.250.0820.150.1110.34N/AN/A0.0280.040.0560.23
%CVN/AN/A20.0%15.8%10.5%11.1%12.5%14.3%7.1%6.2%4.7%6.8%N/AN/A5.8%4.1%3.4%6.6%

N/A = not applicable

'NCCLS EP5-A, Vol. 19, No. 2, Page 24, Equation (C2) 2 NCCLS EP5-A, Vol. 19, No. 2, Page 25, Equation (C3) and Equation (C4)

b) Reproducibillity Studies in BAL

Inter-assay and Intra-assay variability for the Platelia™ Aspergillus EIA were determined in a study using a panel of 4 pooled patient BAL samples spiked with purified galactomannan (one negative, one high negative, one low positive and one medium positive) at 3 testing sites (Two US clinical testing sites and one internal site). Each of the 4 panel members and the controls were tested in duplicate (x2) in 2 runs per day on 5 different days on one lot (Total number of replicates at each site = 120). Two (2) operators performed all precision testing at each site. The data was analyzed according to the Clinical Laboratory Standards Institute (CLSI) (formerly National Committee for Clinical Laboratory Standards (NCCLS)). The mean optical density (OD) and mean index value, standard deviation (SD), percent coefficient of variation (%CV), within run precision (intra-assay) and between site, between day, between operator and between run (inter-assay) precision for each panel member are illustrated below in the following table:

{12}------------------------------------------------

Table 5

Combined Sites Summary

SummaryNegativeN= 60High NegativeN=60LowPositiveN=60MediumPositiveN=60PositiveControlN=60NegativeControlN=60
ODIndexODIndexODIndexODIndexODIndexODIndex
Mean0.1210.290.2140.500.3750.880.5751.351.5803.720.0470.11
WithinRunSDN/AN/A0.0370.1030.0350.0780.0290.0670.1110.265N/AN/A
(IntraAssay)%CVN/AN/A17.4%20.5%9.3%8.9%5.0%5.0%7.0%7.1%N/AN/A
TotalSDN/AN/A0.0420.0950.0610.1220.0700.1380.1900.438N/AN/A
(InterAssay)%CVN /AN/A19.6%18.9%16.2%13.9%12.2%10.2%12.0%11.8%N/AN/A

C. PERFORMANCE EVALUATION STUDIES

I. SERUM SAMPLES

Clinical testing to evaluate the sensitivity, specificity, and predictive value of the Platelia™ Aspergillus EIA was conducted on pediatric (age ≤ 21 years) patients at three sites located in the United States and on adult patients at three sites located in North America. The studies were conducted using a total of 1954 serum samples collected from 129 pediatric patients and a total of 1724 serum samples collected from 172 adult patients from the following populations*:

· patients without signs of Invasive Aspergillosis (control patients)

· patients with Probable Invasive Aspergillosis

· patients with Proven Invasive Aspergillosis

  • The Invasive Fungal Infection Cooperative Group (IFICG) of the European Organization for Research and Treatment of Cancer (EORTC) and the Mycosis Study Group (MSG) of the National Institute of Allergy and Infectious Diseases (NIAID) in 2002 have defined criteria for diagnosis of Invasive Aspergillosis (IA) in patients with hematologic malignancy or hematopoetic stem cell transplant. 4

SENSITIVITY

A. Pediatrics

Results from this study have been analyzed in terms of patient sensitivity. Sensitivity testing was conducted using the Platelia™ Aspergillus EIA at three sites on a combined total of 17 immunocompromised pediatric patients diagnosed with Proven or Probable Invasive Aspergillosis.

{13}------------------------------------------------

Table 6

DiagnosisNumber ofpatientsSensitivity95% ConfidenceInterval
Proven Aspergillosis944.4% (4/9)18.9-73.3%
Probable Aspergillosis862.5% (5/8)30.6-86.3%
Combined Proven andProbable Aspergillosis17*52.9% (9/17)31.0-73.8%

*Note: 8 of the 17 patients gave negative Aspergillus galactomannan antigen results. All of the 8 patients with negative Aspergillus galactomannan antigen results received therapy with multiple antifungal agents. The concomitant use of mold-active anti-fungal therapy in some patients with Invasive Aspergillosis may result in reduced sensitivity 1.

B. Adults

Sensitivity testing was conducted using the Platelia™ Aspergillus ElA at three sites on a combined total of 29 Bone Marrow Transplant (BMT) and Leukemia adult patients diagnosed with Proven or Probable Invasive Aspergillosis.

Table 7

DiagnosisNumber ofpatientsSensitivity95% ConfidenceInterval
Proven Aspergillosis1181.8% (9/11)52.3-94.9%
Probable Aspergillosis1877.8% (14/18)54.8-91.0%
Combined Proven andProbable Aspergillosis2979.3% (23/29)61.6-90.2%

SPECIFICITY

A. Pediatrics

Specificity by pediatric patients

Specificity testing was conducted using the Platelia™ Aspergillus EIA at three sites on a combined total of 108* immunocrompomised pediatric patients without signs of Invasive Aspergillosis (control patients).

Table 8

{14}------------------------------------------------

SiteNumber of patientsSpecificity95% ConfidenceInterval
14486.4 % (38/44)73.3-93.6%
25986.4 % (51/59)75.5-93.0%
35100% (5/5)56.6-100%
Combined Sites10887.0% (94/108)79.4-92.1%

*Note: 4 patients with positive galactomannan antigen results coinciding with piperacillin / tazobactam therapy were excluded.

Specificity by pediatric samples

Specificity testing was conducted using the Platelia™ Aspergillus EIA at three sites on a combined total of 1625* samples obtained from 108 immunocrompomised pediatric patients without signs of Invasive Aspergillosis (control patients).

SiteNumber of samplesSpecificity95% ConfidenceInterval
179498.9% (785/794)97.9-99.4%
273197.8% (715/731)96.5-98.6%
3100100% (100/100)96.3-100%
Combined Sites162598.5% (1600/1625)97.7-99.0%

Table 9

*Note: 80 samples from 4 patients with positive galactomannan antigen results coinciding with piperacillin / tazobactam therapy were excluded.

B. Adults

Specificity by adult patients

Specificity testing was conducted using the Platelia™ Aspergillus EIA at three sites on a combined total of 143 Bone Marrow Transplant (BMT) and Leukemia adult patients without signs of Invasive Aspergillosis (control patients).

{15}------------------------------------------------

SiteNumber of patientsSpecificity95% ConfidenceInterval
12878.6% (22/28)60.5-89.8%
27793.4% (71/77)84.0-96.4%
33889.5% (34/38)75.9-95.8%
Combined Sites14388.8% (127/143)82.6-93.0%

Specificity by adult samples

Specificity testing was conducted using the Platelia™ Aspergillus EIA at three sites on a combined total of 1262 samples obtained from 143 Bone Marrow Transplant (BMT) and Leukemia adult patients without signs of Invasive Aspergillosis (control patients).

'able1IT
----------------------
SiteNumber of samplesSpecificity95% ConfidenceInterval
134998.0% (342/349)95.9-99.0%
256098.6% (552/560)97.2-99.3%
335398.9% (349/353)97.1-99.6%
Combined Sites126298.5% (1243/1262)97.7-99.0%

PREDICTIVE VALUE

Positive and negative predictive values have been analyzed for the patient population in this study. Based on the actual average of 13.6% prevalence rate in pediatrics and 16.9% prevalence rate in adults observed in this study, positive and negative predictive values have been calculated as below:

A. Pediatrics

Study Prevalence 13.6%
PPV: 39.1%95% Confidence Interval : 22.2-59.2%
NPV: 92.2%95% Confidence Interval : 85.3-96.0%

B. Adults

Study Prevalence 16.9%
PPV: 59.0%95% Confidence Interval:43.4-72.9%
NPV: 95.5%95% Confidence Interval: 90.5-97.9%

The expected prevalence of Invasive Aspergillosis varies with the patient population; rates from 5-20% have been reported 19,23. For patient populations on the lower end of the

{16}------------------------------------------------

510k Summary K093678

published prevalence, the positive and negative predictive values have been re-calculated using a 5% prevalence rate.

A. Pediatrics

Calculated Prevalence 5%
PPV : 17.6%95% Confidence Interval : 6.5-39.8%
NPV : 97.2%95% Confidence Interval : 92.1-99.1%

B. Adults

Calculated Prevalence 5%
PPV: 27.2%95% Confidence interval: 13.7-46.7%
NPV: 98.8%95% Confidence Interval: 95.4-99.7%

CROSS REACTIVITY

a) Cross Reactivity - Serum

A study to evaluate the effect of potentially interfering medical conditions unrelated to Invasive Aspergillosis was performed with one lot of the Platelia™ Aspergillus EIA kit. The following serum samples were tested for cross-reactivity with the PlateliaTM Aspergillus EIA. A total of 151 sera were tested.

Table 12

. . . . . .

Pathology# Samples Tested# Positives
Rheumatoid Factor100
ANA Positive100
IgG Hypergammaglobulinemia100
IgM Hypergammaglobulinemia100
Cancer*110
Non-Viral Cirrhosis (primary biliary;alcohol induced; drug induced)100
Multiple Transfusions100
Multiparous Females100
HAV100
HCV100
Rubella100
CMV100
Syphilis (RPR+)100
Toxoplasmosis100
Mycoplasma100
  • One each of bladder, breast(2), colon, endometrial, lung, prostate, renal, and squamous(3).

II. BAL FLUID SAMPLES- Performance Characteristics

Sensitivity and specificity of the Platelia™ Aspergillus EIA with BAL fluid samples were evaluated in two studies in the United States on 116 samples from 62 solid organ transplant recipients and 333 samples from 116 lung transplant recipients and one study

{17}------------------------------------------------

outside the United States on 99 samples from 99 high risk hematology patients with and without invasive aspergillosis.

A. SENSITIVITY

Sensitivity was evaluated in Solid Organ Transplant and Lung Transplant recipients diagnosed with invasive aspergillosis as well as hematologic disease patients diagnosed with invasive aspergillosis according to the EORTC/MSG criteria.

I. Solid Organ Transplant recipients with Invasive Aspergillosis

Of the total of 116 samples from 62 Solid Organ Transplant recipients in one study, sensitivity was evaluated in 5 recipients diagnosed with invasive aspergillosis as shown in the table below.

Table 13

Proven or Probable Invasive Aspergillosis in Solid Organ Transplant Recipients

DiagnosisNIndex ≥ 0.5Sensitivity95% ConfidenceInterval
Proven Aspergillosis222/2 (100%)34.2 - 100%
ProbableAspergillosis333/3 (100%)43.8 - 100%
Combined Provenand ProbableAspergillosis555/5 (100%)56.5 - 100%

By Patient

Table 14

Proven or Probable Invasive Aspergillosis in Solid Organ Transplant Recipients By Transplant Type

TransplantTypeNIndex ≥0.5Sensitivity95%ConfidenceInterval
Heart111/1 (100%)20.6 - 100%
Kidney333/3 (100%)43.8 - 100%
Liver111/1 (100%)20.6 - 100%
Total555/5 (100%)56.5 - 100%

{18}------------------------------------------------

II. Lung Transplant recipients with invasive aspergillosis

Of the total of 333 samples from 116 Lung Transplant recipients in another study, sensitivity was evaluated in 6 recipients diagnosed with invasive aspergillosis as shown in the table below.

Table 15

Proven or Probable Invasive Aspergillosis in Lung Transplant Recipients By patient

DiagnosisNIndex ≥ 0.5Sensitivity95%ConfidenceInterval
Proven Aspergillosis211/2 (50.0%)9.4 - 90.6%
Probable Aspergillosis433/4 (75.0%)30.0 - 95.4%
Combined Proven andProbable Aspergillosis644/6 (66.7%)30.0 - 90.3%

III. Hematologic disease patients with invasive aspergillosis

Sensitivity was also evaluated in a third study in 58 samples from 58 hematologic disease patients diagnosed with invasive aspergillosis as shown in the table below. In the study a retrospective analysis was performed on BAL samples from high risk hematology patients using the Platelia™ Aspergillus EIA. The data from this published study below was evaluated to establish the performance characteristics of the Platelia™ Aspergillus EIA on BAL fluid.

Maertens et al. 2009 Bronchoalveolar Lavage Fluid Galactomannan for the Diagnosis of Invasive Pulmonary Aspergillosis in Patients with Hematologic Diseases. Clin. Infec. Diseas.49:1688-93

Table 16

DiagnosisNIndex ≥ 0.5Sensitivity95%ConfidenceInterval
Proven Aspergillosis313131/31 (100%)89.0 - 100%
Probable Aspergillosis272626/27 (96.3%)81.7 - 99.3%
Combined Proven andProbable Aspergillosis585757/58(98.3%)90.8 - 99.7%

Proven or Probable Invasive Aspergillosis in Hematologic Disease Patients

{19}------------------------------------------------

B. SPECIFICITY

Specificity was evaluated in a total of 98 BAL samples from 57 SOT recipients and 305 BAL samples from 110 Lung Transplant recipients without invasive aspergillosis and is shown in the table below. Results are presented by samples from transplant recipients with and without mold colonization.

Table 17

Specificity by Sample Combined SOT and Lung Transplant Recipients without Invasive Aspergillosis N = 403 BAL Fluids

DiagnosisNIndex < 0.5Negative (%)95%ConfidenceInterval
Controls without colonization341330330/341(96.8%)94.3 - 98.2%
Controls with colonization625050/62 (80.6%)69.1 - 88.6%
Control Total403380380/403(94.3%)91.6 - 96.2%

Specificity in BAL samples from the combined SOT and Lung Transplant recipients without invasive aspergillosis is presented by transplant type in the table 16 below.

Table 18

Specificity by Sample Combined SOT and Lung Transplant Recipients without Invasive Aspergillosis By Transplant Type N = 403 BAL Fluids

Transplant TypeNIndex< 0.5Negative (%)95% ConfidenceInterval
Heart25/28 (89.3%)72.8 - 96.3%
Kidney252222/25 (88.0%)70.0 - 95.8%
Liver2322/23 (95.7%)79.0 - 99.2%
Lung327311311/327 (95.1%)92.2 - 97.0%
Control Total403380380/403 (94.3%)91.6 - 96.2%

Specificity was also evaluated in a total of 41 BAL samples from 41 hematologic disease patients without invasive aspergillosis and is shown in the table below.

Table 19

.

Specificity by Sample Hemotalogic Disease Patients without Invasive Aspergillosis N= 41

DiagnosisNIndex <0.5Negative (%)95% Confidence Interval
Control Patients413333/41 (80.5%)66.0 - 89.8%

{20}------------------------------------------------

Image /page/20/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

BIO-RAD Laboratories c/o Priva Bondre Regulatory Affairs Representative 6565 185" Ave. NE Redmond. WA 98052

JAN 13 2011

K093678 Regulation Number: Regulation Name: Regulatory Class: Product Code: Dated: Received:

Trade/Device Name: Platelia "Aspergillus ElA 21CFR §866.3040 Aspergillus spp. Serological reagents. Class I NOM January 7, 2011 January 11, 2011

Dear Ms. Bondre:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section

{21}------------------------------------------------

Page 2 – Priya Bondre

510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Vally Augusto

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{22}------------------------------------------------

Bio-Rad Laboratories Platelia™ Aspergillus EIA-Bronchoalveolar Lavage Fluid

For and the same

Premarket 510(k) Notification

CONFIDENTIAL

Indications for Use

K093678

JAN 1 3 2011

510(k) Number (if known): Not known at this time

Device Name: Platelia™ Aspergillus EIA

Indications For Use:

The Platelia™ Aspergillus EIA is an immunoenzymatic sandwich microplate assay for the detection of Aspergillus galactomannan antigen in adult and pediatric sevurosary Bronchoalveolar lavage (BAL) fluid samples.

The Platelia™ Aspergillus EIA is a test which, when used in conjunction with other diagnostic procedures such as microbiological culture, histological examination aff biopsy samples and radiographic evidence, can be used as an aid in the diagnosis of invasive aspergillosis.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ludolite Poole
Division Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K093678

§ 866.3040

Aspergillus spp. serological reagents.(a)
Identification. Aspergillus spp. serological reagents are devices that consist of antigens and antisera used in various serological tests to identify antibodies toAspergillus spp. in serum. The identification aids in the diagnosis of aspergillosis caused by fungi belonging to the genusAspergillus. Aspergillosis is a disease marked by inflammatory granulomatous (tumor-like) lessions in the skin, ear, eyeball cavity, nasal sinuses, lungs, and occasionally the bones.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.