K Number
K093627

Validate with FDA (Live)

Date Cleared
2011-01-18

(421 days)

Product Code
Regulation Number
878.4810
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

iPulseLight IPL System (HS 300C, HS 650) are identical with regard to indications for use including recommended filters to be used with Fitzpatrick skin type. Both models are intended for medical use in the treatment of the following dermatologic conditions:

· Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen.

· Treatment of:

  • Moderate inflammatory acne vulgaris .
  • Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, . melasma, ephelides (freckles).
  • . Cutaneous lesions including scars
  • Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, leg veins,

spider anglomas and venous malformations.

The integrated thermal cooling is indicated for use in cooling the treatment site prior to, during, and after light treatment in general aesthetic dermatologic and plastic surgery procedures.

  • Reduce pain during light treatment (via partial anesthesia from cooling) ..
  • Reduce discomfort during and/or associated with light treatment .

Minimize thermal injury, including thermal necrosis, to non-target skin and skin i s tructures during and/or associated with light treatment, thus reducing possible complications such as scabbing, scarring, hyper - and/or hypo pigmentation

Allow the use of higher light or laser fluencies for light treatments (such as for hair removal and the treatment of . vascular or pigmented lesions)

Reduce potential side effects of light treatments (such as for hair removal and the treatment of vascular or pigmented i lesions)

Device Description

iPulseLight IPL System is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. With these special properties, the IPL has a wide application in non-ablative therapies based on theory of human skin tissue's selective absorption and photothermolysis of light sources. Meanwhile, IPL treatment is more effective, with no downtime and can make the patients get recovered more quickly than conventional therapies.

AI/ML Overview

This document is a 510(k) Pre-Market Notification for the iPulseLight IPL System (HS 300C, HS 650). It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a performance study with acceptance criteria in the typical sense of a clinical trial for diagnostic AI.

Here's an analysis based on your request, explaining why certain sections are not applicable to this type of submission:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategoryAcceptance Criteria (Not explicitly stated as such, but inferred from comparison)Reported Device Performance (from the iPulseLight IPL System)
Intended UseMust be identical or similar to the predicate device for various dermatologic conditions and cooling functions.Identical to the Accelawave System, covering: permanent hair reduction, moderate inflammatory acne vulgaris, benign pigmented epidermal lesions (dyschromia, hyperpigmentation, melasma, ephelides), cutaneous lesions (scars), benign cutaneous vascular lesions (port wine stains, hemangiomas, telangiectasias, rosacea, spider anglomas, venous malformations, leg veins). Also identical for integrated thermal cooling functions (pain/discomfort reduction, thermal injury minimization, allowing higher fluences, reducing side effects).
TechnologyIntense Pulsed Light (IPL)/broad spectrum light/touch screen operation.Intense Pulsed Light (IPL)/broad spectrum light/touch screen operation.
Wavelength Range420-1200 nm420-1200 nm
Energy Output/Setting Range10-50 J/cm²10-50 J/cm²
Pulse Duration5-50 ms5-50 ms
Output ModePulse methodPulse method
Pulse Width2-15 ms (from predicate)2-20 ms (slightly extended range compared to predicate's stated 2-15 ms)
AccessoriesFoot switch, protection glasses, protection goggles.Foot switch, protection glasses, protection goggles.
Deliver MaterialsSannhireSannhire
Max. Power Consumption1200W1200W
Applicator/Hand-piece Spot Size15x45mm (from predicate)12x35mm and 15x50mm (different sizes, but within typical variations for IPL systems)
Charge Time/Repeat Rate1.5-2.0 Sec.1.5-2.0 Sec.
Cooling MethodCooling handpiece by TE cooler and circulating water & air.Cooling handpiece by TE cooler and circulating water & air.
Device ClassificationClass II; 21 CFR 878.4810, GEXClass II; 21 CFR 878.4810, GEX

Explanation of "Acceptance Criteria" in this context: For a 510(k) submission, "acceptance criteria" are generally not defined as quantitative performance targets from a clinical study for a novel device. Instead, the primary "acceptance criterion" is demonstrating substantial equivalence to a previously legally marketed predicate device. This means the new device must have the same intended use, and the same technological characteristics, or if there are differences in technological characteristics, those differences must not raise new questions of safety or effectiveness and be demonstrated to be as safe and effective as the predicate. The table above reflects this comparison.

2. Sample size used for the test set and the data provenance

This document does not describe a specific study with a test set of data. The submission is a 510(k) premarket notification for substantial equivalence, which primarily relies on comparing the new device's technical specifications and intended use to a predicate device. It does not present clinical performance data from a "test set" in the way an AI/diagnostic device would. Therefore, sample size and data provenance are not applicable here.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. There is no mention of a test set, ground truth, or experts establishing it in this document.

4. Adjudication method for the test set

Not applicable. There is no mention of a test set or adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device, nor does the document describe an MRMC study.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a medical device (IPL system), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

Not applicable. The document does not describe a study involving ground truth determination.

8. The sample size for the training set

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

Not applicable. As above, this is not an AI/machine learning device.

Summary of the Study (as described in the document):

The "study" or evaluation presented in this 510(k) submission is a comparison to a predicate device (Accelawave System K082484) to claim substantial equivalence. The document doesn't detail a traditional clinical trial or performance study with acceptance criteria in the sense of quantitative metrics for a novel technology. Instead, it systematically compares the iPulseLight IPL System (HS 300C, HS 650) to the Accelawave System across various specifications, including:

  • Intended Use: The intended uses are listed as identical, covering a wide range of dermatologic conditions and thermal cooling benefits.
  • Technological Characteristics: Parameters like wavelength range, energy output, pulse duration, output mode, pulse width, accessories, power consumption, applicator spot size, charge time, and cooling method are compared. Most are identical or very similar. The pulse width for the iPulseLight is 2-20 ms compared to the predicate's 2-15 ms, and the spot sizes are different, but these are implied to be within acceptable variation and not to raise new safety or efficacy questions.
  • Performance (implied): The document states "The device complies with the European Medical Directive Standards: 93/42/EEC concerning medical devices," indicating adherence to European safety and performance standards relevant for medical devices. However, no specific clinical performance data from a study is presented.

The conclusion is that based on this comparison, the iPulseLight IPL System is substantially equivalent to the predicate device and does not raise new questions of safety or efficacy.

{0}------------------------------------------------

K043627 p. 1 of 3

510(K) SUMMARY

[as required by 807.92(c)]

JAN 1 8 2011 .

A.510k Number:

B. Applicant: C ompany name: PATS CORP Address: 49 Candlewood Way, Buena Park, CA 90621, USA Phone: 714-523-1592 FAX: 714-523-1592

C. Proprietary and Established Names: Shanghai APOLO Medical Technology Co., Ltd Address: 3/F, Building A. No. 388, Yindu Road, Xuhui District,Shanghai, China Tel: +86-21-3462 2842 Fax: +86-21-3462 2840

D. Regulatory Information

Common Name: Intense Pulsed Light (IPL) system.

Classification name: Laser surgical instrument for use in general and plastic surgery and

in dermatology (21 CFR Part 878.4810).

Device classification: Class II. Product code: GEX

E. Intended use

iPulseLight IPL System (HS 300C, HS 650) are identical with regard to indications for use including recommended filters to be used with Fitzpatrick skin type. Both models are intended for medical use in the treatment of the following dermatologic conditions:

· Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen.

· Treatment of:

  • Moderate inflammatory acne vulgaris .
  • Benign pigmented epidermal lesions including dyschromia, hyperpigmentation, . melasma, ephelides (freckles).
  • . Cutaneous lesions including scars
  • Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, leg veins,

spider anglomas and venous malformations.

The integrated thermal cooling is indicated for use in cooling the treatment site prior to, during, and after light treatment in general aesthetic dermatologic and plastic surgery procedures.

  • Reduce pain during light treatment (via partial anesthesia from cooling) ..
  • Reduce discomfort during and/or associated with light treatment .

Minimize thermal injury, including thermal necrosis, to non-target skin and skin i s tructures during and/or associated with light treatment, thus reducing possible complications such as scabbing, scarring, hyper - and/or hypo pigmentation

Allow the use of higher light or laser fluencies for light treatments (such as for hair removal and the treatment of . vascular or pigmented lesions)

Reduce potential side effects of light treatments (such as for hair removal and the treatment of vascular or pigmented i lesions)

F. Description

iPulseLight IPL System is a type of intensive, broadband, coherent light source which has a wavelength spectrum of 420 nm -1200 nm. With these special properties, the IPL has a wide application in non-ablative therapies based on theory of human skin tissue's selective absorption and photothermolysis of light sources. Meanwhile, IPL treatment is more effective, with no downtime and can make the patients get recovered more quickly than conventional therapies.

{1}------------------------------------------------

K.093627

20 P.

G. Substantial Equivalence Information

Classification name: Laser surgical instrument for use in general and plastic surgery and in dermatology (per 21 CFR Part 878.4810).

Device classification: C lass II. Product code: GEX Predicate devices: Accelawave System (K082484)

Comparison of Ellipse I PL to predicate devices:

iPulse Li ght IP L ( HS- 650 & H S-300C)Accelawave S ystem
510(k) referenceCurrent submissionK082484
Technology/Intense Pulsed LightIntense Pulsed Light
Operation/(IPL)/broad spectrum(IPL)/broad spectrum
Devicelight/touch screenlight/touch screen
descriptionoperation.operation.
Intended Use ( waiPulseLight IPL System (HS 300C, HS 650) are identicalAccelawave System is identical with regard to indications for
velength/ E nergywith regard to indications for use including recommendeduse including recommended filters to be used
see at tachment t abfilters to be used with Fitzpatrick skin type. Both modelsare intended for medical use in the treatment of thewith Fitzpatrick skin type. Both models are intended formedical use in the treatment of the following dermatologic
le)following dermatologic conditions:conditions:
· Permanent hair reduction- long-term stable reduction in· Permanent hair reduction- long-term stable reduction in
number of hairs re-growing after anumber of hairs re-growing after a
treatment regimen.Treatment of:treatment regimen.· Treatment of:
Moderate inflammatory acne vulgaris-- Moderate inflammatory acne vulgaris
Benign pigmented epidermal lesions including- Benign pigmented epidermal lesions including dyschromia,
dyschromia, hyperpigmentation,melasma,hyperpigmentation, melasma, ephelides
ephelides (freckles).Cutaneous lesions including scars(freckles).- Cutaneous lesions including scars
Benign cutaneous vascular lesions including port- Benign cutaneous vascular lesions including port wine stains,
wine stains, hemangiomas, facial, truncal and leghemangiomas, facial, truncal and leg
telangiectasias, erythema of rosacea, leg veins.telangiectasias, erythema of rosacea, leg veins and venous
spider anglomas and venous malformations.malformations.
The integrated thermal cooling is indicated for use inThe integrated thermal cooling is indicated for use in coolingthe epidermis at the treatment site prior to, during,
cooling the epidermis at the treatment site prior to, during,and after light treatment in general aesthetic dermatologic and
and after light treatment in general aesthetic dermatologicplastic surgery procedures.
and plastic surgery procedures.Reduce pain during light treatment (via partial-- Reduce pain during light treatment (via partial anesthesiafrom cooling)
anesthesia from cooling)- Reduce discomfort during and/or associated with light
Reduce discomfort during and/or associated with lighttreatment
treatment- Minimize thermal injury, including thermal necrosis, to
Minimize thermal injury, including thermal necrosis,(to non-target skin and skinstructures duringnon-target skin and skin s tructures during and/orassociated with light treatment, thus reducing possible
and/or associated with light treatment, thus reducingcomplications such as scabbing, scarring, hyper -
possible complications such as scabbing, scarring, hyper -and/or hypo pigmentation
and/or hypo pigmentation- Allow the use of higher light or laser fluencies for light
Allow the use of higher light or laser fluencies for light,treatments (such as for hair removal and the treatment oftreatments (such as for hair removal and the treatmentof vascular or pigmented lesions)
vascular or pigmented lesions)- Reduce potential side effects of light treatments (such as for
Reduce potential side effects of light treatments (suchhair removal and the treatment of vascular or
as for hair removal and the treatment of vascular orpigmented lesions)
pigmented lesions)
Wavelength Range420-1200 nm420-1200 nm
Energy ouput/10-50 J/cm 210-50 J/cm 2
Setting Range
Pulse duration5-50 ms5-50 ms
Output ModePulse methodPulse method
Pulse width2-20 ms2-15 ms
AccessoriesFoot switchFoot switch
Protection glassesProtection glasses
Protection gogglesProtection goggles
Deliver Materialssannhiresannhire

{2}------------------------------------------------

K093627 p. 385

1200W1200W
Max.powerconsumption
Applicator/hand-piece spot size12x35mm and 15x 50mm15x45mm
Charge time/Repeat rate1.5-2.0 Sec.1.5-2.0 Sec.
Cooling methodCooling handpiece by TE co oler an dCirculating water & AirCooling handpiece by TE co oler an dCirculating water & Air
DeviceclassificationII; 21 CFR 878.4810, GEXII; 21 CFR 878.4810, GEX

use for Treatment region and dose rate.
The UDI - Sustem(HS-650,HS-300C)

Intended use for iPulse Light IPL System(HS-650,HS-300C)SKIN TYPES
CONDITIONSIIIIIIIVVVI
Hair(course)610~1200610~1200610~1200640~1200690~1200N/A
Hair(fine)420-1200420-1200420~1200510~1200560-1200N/A
Acen VulgarisN/A
Pigmented EpidermalLesions510-1200510~1200510-1200560-1200560-1200N/A
a) Dyschromia510-1200510-1200510-1200560-1200560-1200N/A
b)Hyperpigmentation560-1200560-1200560-1200560-1200560-1200N/A
c)Melasma560-1200560-1200560-1200560-1200560-1200N/A
d) Ephelides560-1200560-1200560-1200560-1200560-1200N/A
Cutaneous Lesions
Scars
Cutaneous VascularLesions510-1200510-1200510-1200560-1200560-1200N/A
a) Port Wine Stain (Child)510-1200510-1200510-1200560-1200560-1200N/A
b) Port Wine Stain (Adult)560-1200560-1200560-1200560-1200560-1200N/A
c) Hemangiomas510-1200510-1200510-1200560-1200560-1200N/A
d) Telaugiectasias560-1200560-1200560-1200560-1200560-1200N/A
e) Rosacea560-1200560-1200560-1200560-1200560-1200N/A
f) Spider Anglomas560-1200560-1200560-1200560-1200560-1200N/A
g) Venous Malformations560-1200560-1200560-1200560-1200560-1200N/A
Leg Veins510-1200510-1200510-1200560-1200560-1200N/A
a) Small560-1200560-1200560-1200560-1200560-1200N/A
b) Medium560-1200560-1200560-1200560-1200560-1200N/A

stem (HS-650, HS-300C)"

teristics of the iPulse Light II, LED's, that the pertinent characteristi

195-60. of Present in:
Concludios of the interiminal characteristics of the Press. Internent of this concerners of the President on the President of the President of the Primer of Pr Conclusion:
A companison of the indical characteristics of the Propention In Line Collection C
A comprison of the Accelanes System, led onte oncludes Conclusion of the indival characeae.
A computer died devel be Acelana o System. Ind to the concuision .

the iPulse Light IPL system is substantially equivalent to device approved.
IPL System is substantially equivalent of safety or efficacy.
performance or raise new questions

IPL System to

  • a chance or raise new

perform.

H. Performance Characteristics (If/when applicable)
H. Performance wolies with the European Medical Director I Performance Characteristics (It/when appean Medical Prestive
The device complies with the European Medical devices, and the county of the complies 4. Performance Characterisatis (1) - 1.
The device complies with the European Medical Directive
Standards: 93/42/EEC conceming medical devices, and will comply with
enting ar

. Performance Start - 1 - 1 - European Medical Direct - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1

:

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle or bird symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002

JAN 1 8 2011

Shanghali Apolo Medical Technologies Co., Ltd. % Mr. Brandon Choi Authorized Agent, General Manager 49 Candlewood Way Buena Park, CA 90621

Re: K093627

Trade/Device Name: Ipulselight IPL System, Models HS 300C and HS 650 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONF Dated: January 06, 2011 Received: January 11, 2011

Dear Mr. Choi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act).that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

and the country of the county of the

{4}------------------------------------------------

Page 2 - Mr. Brandon Choi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections. 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

for
Mark N. Melkerson

Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

-K093627

e. l.öfz

Indications for Use

510(k) Number (if known):

Device Name: iPulseLight IPL System (HS 300C, HS 650)

Indications For Use:

iPulseLight IPL System (HS 300C, HS 650) are identical with regard to indications for use including recommended filters to be used with Fitzpatrick skin type. Both models are intended for medical use in the treatment of the following dermatologic conditions:

· Permanent hair reduction- long-term stable reduction in number of hairs re-growing after a treatment regimen.

· Treatment of:

  • Moderate inflammatory acne vulgaris ।
  • Benian pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, ephelides (freckles).
  • Cutaneous lesions including scars -
  • Benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, leg veins, spider anglomas and venous malformations.

The integrated thermal cooling is indicated for use in cooling the epidermis at the treatment site prior to, during, and after light treatment in general aesthetic dermatologic and plastic surgery procedures.

  • Reduce pain during light treatment (via partial anesthesia from cooling) เ
  • Reduce discomfort during and/or associated with light treatment ı
  • Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during and/or associated with light treatment, thus reducing possible complications such as scabbing, scarring, hyper - and/or hypo pigmentation
  • Allow the use of higher light or laser fluencies for light treatments (such as for hair i removal and the treatment of vascular or pigmented lesions)
  • Reduce potential side effects of light treatments (such as for hair removal and the treatment of vascular or pigmented lesions)

{6}------------------------------------------------

K093627

p. 20 A 2

JAN 1 8 2011

SKIN TYPES
CONDITIONS1हाાિVVI
Hair(course)610~1200610~1200610~1200640~1200690~1200N/A
Hair(fine)610~1200610~1200610-1200640~ 1200690-1200NIA
Acne Vulgaris420~1200420-1200420~1200510~1200560~1200NIA
Pigmented Epidermal
Lesions
a)Dyschromiaורדה510~1200510-1200510~1200560~1200560~1200N/A
Hyperpigmentation510-1200510~1200510~1200560~1200560~1200N/A
c)Melasma560~1200560~1200560~1200560~1200560~1200N/A
d)Ephelides560~1200560~1200560~1200560~1200560~1200NA
Cutaneous Lesions
Scars560~1200560~1200560~1200560~1200560~1200
Cutaneous Vascular
Lesions
a) Port Wine Stain (Child)AAR HTONETEVAN DUE SOMITTES510~1200510~1200510~1200560~1200560~1200ALC
b) Port Wine Stain (Adult)510~1200510~1200510~1200560~1200560~1200NA
c) Hemangiomas560~1200560~1200560~1200550~1200560~1200N/A
d) Telaugiectasias510~1200510~1200510~1200560~1200560~1200N/A
e) Rosacea560~1200560~1200560~1200560~1200560~1200N/A
f) Spider Anglomas560~1200560~1200580-1200560~1200560~1200NA
g) Venous Malformations560~1200560~1200560~1200560~1200560~1200NA
Leg Veins
a) Small510-1200510~1200510-1200560~1200560~1200N/A
b) Medium560~1200560~1200560~1200560~1200560~1200NA
c) Large560~1200560~1200560~1200560~1200NAN/A

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

штановнымичный иштемии инининин иматичинии и иматимии и имаан имальным интеменным и

Concurrence of CDRH, Office of Device Evaluation (ODE)

Ohm

(Division Sign-Off) ( ( Crision Sign-Off)
Division of Surgical, Orthopedic.
and Restorative Devices and Resion of Surgical, Ort
and Restorative Devices

510(k) Number L093627

§ 878.4810 Laser surgical instrument for use in general and plastic surgery and in dermatology.

(a)
Identification. (1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
(b)
Classification. (1) Class II.(2) Class I for special laser gas mixtures used as a lasing medium for this class of lasers. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.