K Number
K093594

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-03-03

(104 days)

Product Code
Regulation Number
878.4800
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MicroFlow Spacer (Frontal) is indicated for use as a postoperative spacer to maintain an opening to the frontal sinuses within the first 14 days following surgery. The MicroFlow Spacer also helps to prevent obstruction.
The MicroFlow Spacer (Ethmoid) is indicated for use as a postoperative spacer to maintain an opening to the ethmoid sinuses within the first 14 days following surgery. The MicroFlow Spacer also helps to prevent obstruction.

Device Description

The MicroFlow Spacer is a device that maintains an opening at the target sinus for up to 14 days postoperatively. There are two models of the MicroFlow Spacer: Ethmoid and Frontal. The modified device can be manually removed during an office follow-up visit at any time.

AI/ML Overview

The provided material describes the Relieva Stratus MicroFlow Spacer, a manual surgical instrument. Because this is a physical medical device and not an AI/ML-driven solution, many of the typical acceptance criteria and study components related to software performance (such as sample sizes for test sets, expert ground truth establishment, MRMC studies, standalone performance, and training set details) are not applicable.

The document indicates that the device's performance was evaluated for substantial equivalence to a predicate device, the Relieva Stratus MicroFlow Spacer (K083574).

Here's a breakdown of the relevant information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document explicitly states: "The MicroFlow Spacer met all performance acceptance criteria." However, it does not provide a detailed table listing specific acceptance criteria and their corresponding reported performance values. It only offers a general statement of compliance. The focus of this 510(k) summary is on establishing substantial equivalence through relevant performance tests, rather than detailing a specific set of clinical performance metrics.

Acceptance CriteriaReported Device Performance
Not explicitly detailed in the provided text. The document states "The MicroFlow Spacer met all performance acceptance criteria."Met all performance acceptance criteria. The device was deemed "substantially equivalent to the predicate device as confirmed through relevant performance tests."

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not specified. The document refers to "relevant performance tests" but does not detail the size or nature of any test sets.
  • Data Provenance: Not specified. As this is a physical device submission focused on substantial equivalence to a predicate, the "data" likely refers to engineering performance testing rather than clinical study data from a specific country or a retrospective/prospective design.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

This is not applicable as the device is a physical surgical instrument, not an AI/ML system requiring expert-adjudicated ground truth for software performance evaluation. The "ground truth" would relate to engineering specifications and clinical outcomes, likely assessed through standard medical device testing and clinical review, not expert consensus on image interpretation.

4. Adjudication Method for the Test Set:

This is not applicable for the reasons stated above.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study:

This is not applicable as the device is a physical surgical instrument, not an AI/ML system intended to assist human readers with interpretation.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study:

This is not applicable as the device is a physical surgical instrument, not a standalone algorithm.

7. Type of Ground Truth Used:

The "ground truth" for this medical device submission would be based on engineering specifications and potentially clinical outcomes reported in existing literature or previous predicate device data, rather than expert consensus, pathology, or outcomes data specifically generated for a new AI model's ground truth. The primary basis for approval is substantial equivalence to a legally marketed predicate device.

8. Sample Size for the Training Set:

This is not applicable as the device is a physical surgical instrument and does not involve an AI/ML training set.

9. How the Ground Truth for the Training Set Was Established:

This is not applicable as the device is a physical surgical instrument and does not involve an AI/ML training set.

{0}------------------------------------------------

Aclarent

MAR - 3 2010

:

Krg3594

APPENDIX A: 510(k) SUMMARY
Sponsor/Submitter:Acclarent, Inc.1525-B O'Brien DriveMenlo Park, California 94025
Contact Person:Courtney KalofRegulatory Affairs CoordinatorPhone: (650) 687-4473Fax: (650) 687-4449
Date of Submission:November 16, 2009
Device Trade Name:Relieva Stratus MicroFlow Spacer
Common Name:MicroFlow Spacer
Device Classification:Class I
Regulation Number:21 CFR 878.4800
Classification Name:Manual surgical instrument for general use
Product Code:KAM
Predicate Device:Relieva Stratus MicroFlow Spacer (K083574)
Device Description:The MicroFlow Spacer is a device that maintains an opening at thetarget sinus for up to 14 days postoperatively. There are two models ofthe MicroFlow Spacer: Ethmoid and Frontal. The modified device canbe manually removed during an office follow-up visit at any time.
Indications for Use:The MicroFlow Spacer (Ethmoid) is indicated for use as a postoperativespacer to maintain an opening to the ethmoid sinuses within the first 14days following surgery. The MicroFlow Spacer also helps to preventobstruction.The MicroFlow Spacer (Frontal) is indicated for use as a postoperativespacer to maintain an opening to the frontal sinuses within the first 14days following surgery. The MicroFlow Spacer also helps to preventobstruction.
TechnologicalCharacteristics:The MicroFlow Spacer is designed to be implanted into the frontal orethmoid sinuses and to maintain its position by a self-retentionmechanism.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "Acclarent" in a bold, sans-serif font. A curved line extends from the top right of the "t", arcing over the word and ending near the right side of the "t". The text is black and the background is white.

Performance Data:

The MicroFlow Spacer met all performance acceptance criteria.

Summary of Substantial Equivalence:

. .

The MicroFlow Spacer is substantially equivalent to the predicate device as confirmed through relevant performance tests.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight. The symbol is composed of three curved lines that suggest the wings and body of the bird.

MAR 3 2010

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Acclarent, Inc. c/o Ms. Courtney Kalof Regulatory Affairs Coordinator 1525-B O'Brien Dr. Meno Park, CA 94025

Re: K093594

Trade/Device Name: Relieva Stratus MicroFlow Spacer Regulation Number: 21 CFR 878.4800 Regulation Name: Manual surgical instrument for general use Regulatory Class: Class I Product Code: KAM Dated: January 26, 2010 Received: January 27, 2010

Dear Ms. Kalof:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Peter Schmidt

Malvina B. Eydelman. M.L Director Division of Ophthalmic. Neurological. and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K093594

Acclarént

APPENDIX B: INDICATIONS FOR USE STATEMENT

510(k) Number (if known):K093594
Trade Name:Relieva Stratus MicroFlow Spacer
Common Name:MicroFlow Spacer
Indications For Use:The MicroFlow Spacer (Frontal) is indicated for use as a postoperative spacer to maintain an opening to the frontal sinuses within the first 14 days following surgery. The MicroFlow Spacer also helps to prevent obstruction.The MicroFlow Spacer (Ethmoid) is indicated for use as a postoperative spacer to maintain an opening to the ethmoid sinuses within the first 14 days following surgery. The MicroFlow Spacer also helps to prevent obstruction.
Prescription Use(Part 21 CFR 801 Subpart D)X
AND/OR
Over-The-Counter Use(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Daniel C. Cupp
(Division Sign-Off)

Page i of /

Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices

:10(k) Number KD93594

§ 878.4800 Manual surgical instrument for general use.

(a)
Identification. A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.