K Number
K093499

Validate with FDA (Live)

Date Cleared
2009-11-23

(11 days)

Product Code
Regulation Number
890.5500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The HairMax LaserComb is indicated to promote hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood Hamilton Classifications of IIa to V patterns of hair loss and Fitzpatrick Skin Types I to IV.

Device Description

Similar to the HairMax LaserComb K060305, the modified HairMax LaserComb consists of a hand-held low level laser device that emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp. The predicate unit operates from solely on a class 2, UL and CUL rated wall transformer. The modified unit adds a Lithium ion battery power supply option. In addition, cosmetic changes are being made to the device color and exterior plastic housing. Changes do not alter form fit or function as compared to the predicate device.

AI/ML Overview

Here's an analysis of the provided 510(k) summary, specifically addressing the acceptance criteria and study information:

The document (K093499) describes a modified HairMax LaserComb (version 5.2) and seeks substantial equivalence to a previously cleared device (K060305). Crucially, this 510(k) pertains to modifications (adding a Li-ion battery, cosmetic changes) to an already cleared device, not a new device requiring a full de novo clinical study to prove efficacy. The submission explicitly states that the changes "do not alter form fit or function as compared to the predicate device" and "do not affect the safety or performance of the device and do not change the intended use of the HairMax LaserComb."

Therefore, the document does not contain information about a clinical study proving the modified device's performance against acceptance criteria in terms of hair growth promotion. Instead, it relies on the predicate device's existing clearance and demonstrates that the modifications do not negatively impact safety or effectiveness.

Based on the provided text, the following information can be extracted:


1. Table of Acceptance Criteria and Reported Device Performance:

Since this 510(k) is for a modified device and not a new device requiring efficacy trials for hair growth, there are no specific performance acceptance criteria for hair growth or reported device performance for this particular submission. The focus is on demonstrating that the modifications do not affect safety or performance compared to the predicate.

The document mentions: "Based on the Risk Analysis, the verification and validation tests that were performed and the acceptance criteria applied for each are listed in Section 10." However, Section 10 is not included in the provided text. These tests would typically be related to electrical safety, mechanical integrity, battery performance, and electromagnetic compatibility to ensure the modified device is as safe as the predicate, rather than clinical efficacy.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not applicable for a hair growth efficacy study in this submission. The document refers to "verification and validation tests" in Section 10, which would have their own sample sizes for engineering/safety tests, but these are not specified.
  • Data Provenance: Not applicable for a hair growth efficacy study in this submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable as no clinical efficacy test set is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

  • Not applicable as no clinical efficacy test set is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not applicable. This device is a low-level laser therapy device for hair growth, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable. This device is a physical therapeutic device, not a standalone algorithm.

7. The type of ground truth used:

  • Not applicable for hair growth efficacy in this submission. For the safety and performance of the modified device, the "ground truth" would be compliance with internal design specifications and relevant industry standards (e.g., electrical safety standards).

8. The sample size for the training set:

  • Not applicable. No AI/machine learning model is being trained here.

9. How the ground truth for the training set was established:

  • Not applicable. No AI/machine learning model is being trained here.

Summary of what the document does provide regarding acceptance/equivalence:

  • Predicate Device: HairMax LaserComb K060305. The current device (K093499) is considered substantially equivalent to this predicate.
  • Basis for Equivalence:
    • Same indicated use: "to promote hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood Hamilton Classifications IIa -V patterns of hair loss and Fitzpatrick Skin Types I to IV."
    • Same operating principle.
    • Same basic device design and physical properties.
    • Same materials.
    • Same technological characteristics (laser power, wavelength, laser delivery method, comb component, instructions for use, audible timer).
  • Modifications: Adding a Lithium ion battery power supply option and cosmetic changes (device color and exterior plastic housing dimensions).
  • Verification of Modifications: The submission states these modifications "do not alter form fit or function as compared to the predicate device" and "do not affect the safety or performance of the device and do not change the intended use." Performance testing was conducted "to confirm compliance to design specifications; all functions were verified to operate as designed." (Details of these tests and their acceptance criteria are in a section not provided).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Lexington International, LLC. The word "Lexington" is in a bold, sans-serif font, with a curved line above it. Below "Lexington" is the text "International, LLC" in a smaller font. To the right of the text is a star.

K093499

Lexington International, LLC 777 Yamato Road, Suite 105 Boca Raton, FL. 33431

6. 510(k) Summary

Lexington International, LLC

HairMax LaserComb ver. 5.2

NOV 2 3 2009

Submitter's Contact Information

Name:David Michaels, Managing Director
Address:Lexington International, LLC
777 Yamato Rd. Suite 105
Boca Raton, FL 33431
Telephone:(561) 314-2430
Facsimile:(561) 892-0747
E-mail:DM@HairMax.com
Manufacturer:Lexington International, LLC

Establishment Registration Number: 3006182775

Official Correspondent

Name:Cherita James, Regulatory Consultant
Address:M Squared Associates, Inc.
901 King Street, Suite 200
Alexandria, VA 22314
Telephone:(703)562-9800
Facsimile:(703) 562-9797
E-mail:CJames@MSquaredAssociates.com

Trade Name: HairMax LaserComb Common or Usual Name: Lamp, nonheating, for promotion of hair growth. Product Code: OAP

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows a logo for Lexington International, LLC. The word "Lexington" is in a bold, sans-serif font, with a horizontal line running through the middle of the word. Above the word "Lexington" is a curved line. Below the word "Lexington" is the text "International, LLC" in a smaller, sans-serif font. To the right of the word "Lexington" is a star.

ton International. LLC Yamato Road, Suite 105 Boca Raton, FL. 33431

Classification Name: 21 CFR 890.5500 Infrared lamp Predicate Device: HairMax LaserComb K060305

November 9, 2009

Device Description:

Similar to the HairMax LaserComb K060305, the modified HairMax LaserComb consists of a hand-held low level laser device that emits laser light with the intention to promote hair growth. The device provides distributed laser light to the scalp while the comb teeth simultaneously part the user's hair to ensure the laser light reaches the user's scalp.

The predicate unit operates from solely on a class 2, UL and CUL rated wall The modified unit adds a Lithium ion battery power supply option. In transformer. addition, cosmetic changes are being made to the device color and exterior plastic housing. Changes do not alter form fit or function as compared to the predicate device.

Intended Use / Indications for Use

The HairMax LaserComb is indicated to promote hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood Hamilton Classification IIa -V patterns of hair loss and Fitzpatrick Skin Types I to IV.

Technological Characteristics

The modifications to the HairMax LaserComb since its previous clearance in K060305 include changes to available power supply, and consequently the device casing dimensions and color. These minor differences do not affect the safety or performance of the device and do not change the intended use of the HairMax LaserComb.

Nonclinical Testing

Based on the Risk Analysis, the verification and validation tests that were performed and the acceptance criteria applied for each are listed in Section 10. The modified HairMax LaserComb was subject to the same preclinical requirements as the

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Lexington International, LLC. The word "Lexington" is in a bold, sans-serif font, with a line underneath it. Below that is the word "International, LLC" in a smaller, sans-serif font. There is a curved line above the word "Lexington" and a star to the right of the word.

aton International. LLC 777 Yamato Road, Suite 105 Boca Raton, FL. 33431

predicate device. Performance testing was conducted to confirm compliance to design specifications; all functions were verified to operate as designed.

Substantial Equivalence

The modified HairMax LaserComb is as safe and effective as the predicate device. The modified device has the same intended use of affecting hair growth as the predicate device. The modified HairMax LaserComb has the same indications, i.e., treating androgenetic alopecia, and the same specific indication of promoting hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood-Hamilton Classifications of I to V patterns of hair loss and Fitzpatrick Skin types of I-IV as the predicate device.

The modified HairMax LaserComb is identical in technological characteristics as the device cleared in K060305, including its laser power, wavelength, laser delivery method, its comb component, its instructions for use and its audible timer. The cosmetic change to device color and dimensions of the exterior plastic housing and addition of a battery does not affect the HairMax LaserComb's fundamental scientific technology.

Conclusion:

The modified HairMax LaserComb has the following similarities to the HairMax LaserComb:

  • . has the same indicated use,
  • uses the same operating principle, .
  • incorporates the same basic device design and physical properties, .
  • . incorporates the same materials.

Therefore the modification to the modified HairMax LaserComb can be found substantially equivalent to the HairMax LaserComb cleared in K060305.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Lexington International, LLC % M Squared Associates, Inc. Ms. Cherita James Regulatory Consultant 901 King Street, Suite 200 Alexandria, Virginia 22314

NOV 2 3 2009

Re: K093499

Trade/Device Name: HairMax LaserComb Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP, GEX Dated: November 10, 2009 Received: November 12, 2009

Dear Ms. James:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it . may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Ms. Cherita James

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark M Millerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Lexington International, LLC. The word "Lexington" is in a larger, bold font, with a line underneath it. Below that, the words "International, LLC" are in a smaller font. There is a curved line above the word "Lexington" and a star to the right of the logo.

Lexington International, LLC 777 Yamato Road, Suite 105 Boca Raton, FL. 33431

5. Indications for Use

510(k) Number (if known): Ko 9 3 499

Device Name: HairMax LaserComb

Indications for Use:

The HairMax LaserComb is indicated to promote hair growth in males, 30-60 years of age with androgenetic alopecia who have Norwood Hamilton Classifications of IIa to V patterns of hair loss and Fitzpatrick Skin Types I to IV.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark N. Mulhausen

(Division Sign-Oft) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K093499

§ 890.5500 Infrared lamp.

(a)
Identification. An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.(b)
Classification. Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.