K Number
K093426

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-02-22

(111 days)

Product Code
Regulation Number
868.5260
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is required. It is single patient use, disposable for patients with Tidal Volumes > 150 ml.

May be positioned at the machine end of the expiratory / inspiratory limb of the circuit, or at the patient end of the circuit.

Device Description

The Bact-Trap™ HEPA Midi filter incorporates standard 15 / 22 mm connectors with a gas sampling luer port. It utilizes an electrostatic media for the filtration function. The "HEPA" performance was tested in accordance to DOE 3020-97 and ASTM D2986 - 95a DOP at 30 Ipm.

The filter functions by electrostatic capture.

AI/ML Overview

The acceptance criteria and study proving the device meets these criteria for the Bact-Trap™ HEPA Midi are detailed below.
This device is a bacterial/viral filter, and the studies performed are primarily performance tests against established standards and metrics for such filters, rather than clinical studies involving human patients or ground truth from expert readings in a diagnostic context.

1. Table of Acceptance Criteria and Reported Device Performance

AttributeAcceptance CriteriaReported Device Performance
Resistance to flow< 3.1 cm H2O @ 60 lpm< 3.1 cm H2O @ 60 lpm
Bacterial filtration (BFE)Not explicitly stated as acceptance criteria, but based on predicate device99.9999%
Viral filtration (VFE)Not explicitly stated as acceptance criteria, but based on predicate device99.9999%
HEPA Filtration>99.97% (based on DOE 3020-97 and ASTM D2986 - 95a DOP)>99.97%
Dead Space (ml)Not explicitly stated as acceptance criteria, but based on predicate device46 ml for straight, 48 ml for angled
Standard 15/22 mm connectorsYes (ISO 5356-1)Yes
Gas sampling portYesYes
Housing materialPolystyrenePolystyrene
Filter mediaElectrostaticElectrostatic
Weight (gm)Not explicitly stated as acceptance criteria32 gm
Indications for use (To filter inspired and/or expired gases)YesYes
Intended for single patient useYesYes
PrescriptionYesYes
Intended population (> 150 ml tidal volume)YesYes
Intended Environment of UseHome, Hospital, sub-acute, Emergency servicesHome, Hospital, sub-acute, Emergency services
Placement in various locations in circuitYesYes
Compliance to ISO 5356-1 Conical 15/22YesYes
Compliance to ISO 594-2 Luer FittingsYesYes
Compliance to DOE 3020-97 and ASTM D2986 - 95a DOPYesYes

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify the exact sample sizes (number of devices tested) for each performance criterion. However, the data provenance for the bacterial and viral filtration efficiency (BFE and VFE) is stated as Nelson Lab. This indicates the testing was conducted by an external laboratory specializing in such evaluations. The filtration performance (HEPA, BFE, VFE) and resistance to flow are typical engineering performance tests that would involve a specific number of devices tested under controlled laboratory conditions. The data is retrospective in the sense that it was generated prior to the 510(k) submission to demonstrate compliance. The country of origin for the data is presumably where Nelson Lab is located (typically USA) and potentially Sweden for other internal testing by Pharma Systems AB.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

Not applicable. For this type of medical device (bacterial/viral filter), the "ground truth" is established through physical and engineering performance measurements against established standards (e.g., DOE 3020-97, ASTM D2986 - 95a DOP, ISO standards) and laboratory testing protocols to determine filtration efficiency and resistance. There are no human experts establishing a "ground truth" through interpretation of data in the way radiologists or pathologists would for diagnostic devices. The experts involved would be the laboratory technicians and scientists performing the tests and verifying the methodologies.

4. Adjudication Method for the Test Set

Not applicable. As noted above, the "truth" is determined by direct physical/engineering measurements using standardized testing methods, not by expert consensus or adjudication in this context.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret medical images with or without AI assistance. The Bact-Trap™ HEPA Midi is a filter for breathing circuits, not a diagnostic device.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Yes, in a sense, a "standalone" performance evaluation was conducted. The device's performance was measured intrinsically against established physical and engineering standards (e.g., filtration efficacy, resistance to flow). These measurements are inherent to the device's design and materials and do not involve a "human-in-the-loop" for its function, nor an algorithm in the traditional sense of AI. The device operates mechanically/physically without software control or human intervention for its primary filtration function.

7. Type of Ground Truth Used

The "ground truth" used is based on:

  • Physical and Engineering Standards: Compliance with established international and national standards such as ISO 5356-1, ISO 594-2, DOE 3020-97, and ASTM D2986 - 95a DOP.
  • Laboratory Performance Measurements: Direct measurements of physical properties (e.g., dead space, weight) and functional performance (e.g., resistance to flow, bacterial and viral filtration efficiency) obtained from controlled laboratory tests. These are objective, quantifiable results.

8. Sample Size for the Training Set

Not applicable. The Bact-Trap™ HEPA Midi does not utilize an algorithm or AI that requires a "training set" of data. Its function is based on physical filtration principles and material properties.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as no training set was used.

{0}------------------------------------------------

K093426

FEB 2 2 2010

510(k) Summary Page 1 of 2 11-Feb-10

Pharma Systems ABTel - 011 46 18 34 95 00
Rubanksgatan 9Fax - 011 46 18 34 95 30
S741 71 Knivsta Sweden
Official Contact:Viljar Salumaa, Product Manager
Proprietary or Trade Name:Bact-Trap™ HEPA Midi
Common/Usual Name:Bacterial / Viral Filter
Classification Name:Filter, Bacterial, Breathing Circuit,CAH - 21 CFR 868.5260
Predicate Devices:ARC Medical - FilterFlo™ HEPA – K063125Air Safety - Model 6500/01 - K033008

Device Description

The Bact-Trap™ HEPA Midi filter incorporates standard 15 / 22 mm connectors with a gas sampling luer port. It utilizes an electrostatic media for the filtration function. The "HEPA" performance was tested in accordance to DOE 3020-97 and ASTM D2986 - 95a DOP at 30 Ipm.

The filter functions by electrostatic capture.

Indications for Use

For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is required. It is intended for patients with Tidal Volumes > 150 ml.

May be positioned at the machine end of the expiratory limb of the circuit, or at the patient end of the circuit.

Environment of Use

Home, Hospital, Sub-acute Institutions, Emergency services

: ・
・・・
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・

{1}------------------------------------------------

510(k) Summary Page 2 of 2 11-Feb-10

General Technical Characteristics

AttributeProposed devices
Indications for use - To filter inspired and / orexpired gases.Yes
Intended for single patient useYes
PrescriptionYes
Intended population> 150 ml tidal volume
Intended Environment of UseHome, Hospital, sub-acute, Emergency services
Placement in various locations in circuitYes
Design
Gas sampling portYes
Standard 15/22 mm connectorsYes
Dead Space (ml)46 ml for straight, 48 ml for angled
Resistance to flow< 3.1 cm H2O @ 60 lpm
Bacterial filtration - BFE - Nelson Lab.99.9999%
Viral filtration - VFE - Nelson Lab.99.9999%
HEPA Filtration>99.97%
Weight (gm)32 gm
Materials
Housing polystyreneYes
Filter mediaElectrostatic
Performance Standards
None under Section 514Yes
ISO 5356-1 Conical 15/22Yes
ISO 594-2 Luer FittingsYes
DOE 3020-97 and ASTM D2986 - 95a DOPYes

Differences between Other Legally Marketed Predicate Devices

The data within the submission demonstrates that the proposed devices when compared to the predicate devices are safe and effective and are substantially equivalent to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol on the right side, which is composed of three curved lines. To the left of the caduceus is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter.

FEB 2 2 2010

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002

Pharma Systems AB C/O Mr. Paul E. Dryden President ProMedic, Incorporated 24301 Woodsage Drive Bonita Springs, Florida 34134-2958

Re: K093426

Trade/Device Name: Bact-Trap™ HEPA Midi Regulation Number: 21CFR 868.5260 Regulation Name: Breathing Circuit Bacterial Regulatory Class: II Product Code: CAH Dated: February 11, 2010 Received: February 16, 2010

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Dryden

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Critthon V. Nim

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

{4}------------------------------------------------

Indications for Use Statement

Page 1 of 1

510(k) Number: K093426 (To be assigned)

Device Name: Bact-Trap™ HEPA Midi

Indications for Use:

For use with ventilators, anesthesia machines and open flow systems where filtration of inspired and / or expired gases is required. It is single patient use, disposable for patients with Tidal Volumes > 150 ml.

May be positioned at the machine end of the expiratory / inspiratory limb of the circuit, or at the patient end of the circuit.

Prescription Use XXorOver-the-counter use _
(Part 21 CFR 801 Subpart D)(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Arbuthno

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

11093426 510(k) Number:

§ 868.5260 Breathing circuit bacterial filter.

(a)
Identification. A breathing circuit bacterial filter is a device that is intended to remove microbiological and particulate matter from the gases in the breathing circuit.(b)
Classification. Class II (performance standards).