(114 days)
The PowerPump is a dual arthroscopic pump system intended to provide fluid distension and irrigation of the knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid suction during diagnostic and operative arthroscopic procedures.
The PowerPump provides liquid irrigation and aspiration for arthroscopic procedures in one unit through two individual roller pumps. Both roller pumps are software controlled and automatically manage fluid and joint pressure based on procedure settings chosen by the user. If needed, both flow and pressure settings can be individually adjusted. By controlling both inflow and outflow, the PowerPump accurately regulates pressure and flow in the joint. The PowerPump also provides suction when used in conjunction with a shaver. The PowerPump consists of the following articles:
- PowerPump Unit
- Non-sterile Foot Control
- Non-sterile Power Cord
- Disposable, sterile Patient Cassette
- Disposable, sterile Day Cassette
- Disposable, sterile PowerPump Irrigation Tubing
The provided document is a 510(k) summary for the PowerPump® Arthroscopy System. It focuses on demonstrating substantial equivalence to predicate devices for its intended use as an arthroscopic pump. The document does not contain the detailed information typically found in a study demonstrating how a device meets specific acceptance criteria in terms of performance metrics like sensitivity, specificity, accuracy, or similar quantitative assessments.
Therefore, most of the requested information cannot be extracted from this document, as it is primarily a regulatory filing for substantial equivalence, not a performance study report.
Here's a breakdown of what can and cannot be answered based on the provided text:
1. A table of acceptance criteria and the reported device performance
- Cannot be provided directly. The document states, "Summary of Testing: The PowerPump has been tested for function, performance and safety. Test results show that all specifications have been met." However, it does not list the specific acceptance criteria (e.g., pressure regulation accuracy within X%, flow rate within Y%) or the reported device performance values against those criteria. It only provides a high-level statement that specifications were met.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Cannot be provided. The document does not describe a clinical or performance study with a 'test set' in the context of typical AI/diagnostic device evaluation. The assessment seems to be based on engineering and safety testing.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
- Not applicable / Cannot be provided. This information is relevant for studies where expert consensus or interpretation is used to establish ground truth, typically in image analysis or diagnostic device assessments. This document describes an arthroscopy pump, a mechanical and fluid management device, where "ground truth" for performance is established through direct measurement against engineering specifications rather than expert interpretation of data.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
- Not applicable / Cannot be provided. This refers to consensus methods for expert decisions in diagnostic studies. It doesn't apply to the engineering and safety testing of a fluid management system.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. This type of study (MRMC, human-AI interaction) is specific to AI-powered diagnostic or assistive devices, particularly in image interpretation. The PowerPump is an arthroscopy pump and does not involve human readers or AI assistance in this context.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This question is also specific to AI algorithms. The PowerPump's performance is inherently its standalone mechanical/fluid management function; there isn't a separate "algorithm only" performance to evaluate in this context.
7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)
- Cannot be explicitly determined, but inferred. For a device like an arthroscopy pump, "ground truth" for performance would likely be established through:
- Engineering specifications and standards: e.g., pressure gauges demonstrating actual intra-articular pressure, flow meters measuring fluid delivery/aspiration rates.
- Bench testing: In vitro models simulating joint cavities.
- Safety standards compliance: Electrical safety, biocompatibility, etc.
The document doesn't detail these, but these are the typical methods for such a device.
8. The sample size for the training set
- Not applicable / Cannot be provided. This refers to machine learning models. The PowerPump is not described as an AI/ML device requiring a training set.
9. How the ground truth for the training set was established
- Not applicable / Cannot be provided. Same reason as above.
In summary, the provided 510(k) document is a regulatory filing focused on establishing substantial equivalence for an arthroscopy pump. It does not provide the details of a performance study in the way one would for an AI or diagnostic device, and therefore, most of the detailed questions cannot be answered from this text.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a handwritten sequence of characters. The sequence starts with the letter 'K', followed by the numbers '093409'. The handwriting is cursive and the numbers are written in a connected manner.
510(k) Summary
Submitter: Medical Vision Nacka, Sweden
FEB 2 4 2010
Contact Information: C. G. Bundy Associates, Inc. 435 Rice Creek Terrace Fridley, MN 55432
Submission Date: October 29, 2009
Device Name and Classification: Power Pump® Arthroscopy System, Class II Product Code: HRX
Equivalent Device Identification:
Device: Future Medical Systems FMS DUO, K954465 Smart Vision® Feature: Gambro Dialysis Machine, K771227 and Baxter Altratouch system 1000 dialysis machine. K954987
Device Description:
The PowerPump provides liquid irrigation and aspiration for arthroscopic procedures in one unit through two individual roller pumps. Both roller pumps are software controlled and automatically manage fluid and ioint pressure based on procedure settings chosen by the user. If needed, both flow and pressure settings can be individually adjusted. By controlling both inflow and outflow, the PowerPump accurately regulates pressure and flow in the joint. The PowerPump also provides suction when used in conjunction with a shaver. The PowerPump consists of the following articles:
-
- PowerPump Unit
-
- Non-sterile Foot Control
-
- Non-sterile Power Cord
-
- Disposable, sterile Patient Cassette
-
- Disposable, sterile Day Cassette
-
- Disposable, sterile PowerPump Irrigation Tubing
Intended Use:
The PowerPump is a dual arthroscopic pump system intended to provide fluid distension and irrigation of the knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluid suction during diagnostic and operative arthroscopic procedures.
Summary of Testing: The PowerPump has been tested for function, performance and safety. Test results show that all specifications have been met.
{1}------------------------------------------------
Conclusion:
The PowerPump arthroscopy pump and the predicate device FMS DUO are substantially equivalent. Both pumps are intended to be used for providing distention pressure of body joints by pressurizing irrigation liquid, and to aspirate the liquid from the joint. Differences that exist between the devices are minor and do not affect the safety and effectiveness of the PowerPump.
The optical detection (SmartVision) feature is equivalent to blood and debris detection features used with dialysis machines. The addition of this well-known technology to the PowerPump does not affect the safe and effective intended use of the device as an Arthroscopy pump.
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of a bird, possibly an eagle, with its wings spread, symbolizing protection and care.
Public Health Service
FEB 2 4 2010
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
Medical Vision AB % C. G. Bundy Associates, Inc. Constance G. Bundy 435 Rice Creek Terrace Fridley, Minnesota 55432
Re: K093409
Trade/Device Name: PowerPump® Arthroscopy System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: February 17, 2010 Received: February 22, 2010
Dear Constance G. Bundy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 - Constance G. Bundy
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice.for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
yours,
Melkerson
Mark N. Melkerson Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Indications for Use
510(k) Number: K093409
Device Name: PowerPump® Arthroscopy System
Indications For Use:
The PowerPump is a dual arthroscopic pump system intended to provide fluid distension and irrigation of the knee, shoulder, hip, elbow, ankle and wrist joint cavities and fluirds motion during diagnostic and operative arthroscopic procedures.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|---|
| -------------------------------------------------------- |
| (Division Sign-Off |
|---|
| Division of Surgica Orthopedic, |
| and Restorative Devices |
| 510(k) Number | K093409 |
|---|---|
| --------------- | --------- |
N/A