1ML UNITRACT INSULIN SYRINGE 27G X 1/2 AND 29G X 1/2

K093400 · Unilife Medical Solutions Limited · FMF · Apr 1, 2010 · General Hospital

Device Facts

Record IDK093400
Device Name1ML UNITRACT INSULIN SYRINGE 27G X 1/2 AND 29G X 1/2
ApplicantUnilife Medical Solutions Limited
Product CodeFMF · General Hospital
Decision DateApr 1, 2010
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 880.5860
Device ClassClass 2
AttributesTherapeutic

Intended Use

The syringe, as supplied, is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention.

Device Story

1mL Unitract Insulin Syringe is a sterile, single-use, fixed-needle syringe for subcutaneous insulin delivery. Device features include automatic needle retraction and auto-disable mechanism to prevent reuse and sharps injuries. Operation involves manual filling and injection; upon completion of the injection stroke, the plunger engages a spring-based mechanism at the proximal end, triggering automatic needle retraction into the barrel. Simultaneously, the plunger locks into the outer collar, preventing further movement. Used by patients or clinicians for insulin administration. Benefits include reduced risk of needle-stick injuries and prevention of syringe reuse.

Clinical Evidence

Bench testing only. Performance testing verified design specifications, including peak activation forces for retraction, spring reaction forces, and auto-disable override forces. Chemical testing included extractable trace metals, non-volatile residue, and pH. Biocompatibility testing included cytotoxicity and pyrogens, with fluid path materials meeting USP Class VI requirements. Sterilization validated per DIN EN 550:1994.

Technological Characteristics

Fixed-needle piston syringe; 27G x 1/2" or 29G x 1/2" needle sizes. Materials: biocompatible plastics and stainless steel. Energy source: mechanical spring-based retraction. Sterilization: ethylene oxide (EO) per DIN EN 550:1994. Connectivity: none (standalone).

Indications for Use

Indicated for subcutaneous administration of U-100 insulin in patients requiring insulin therapy.

Regulatory Classification

Identification

A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Unilife medical solutions. The word "UNILIFE" is in bold, sans-serif font, with the words "medical solutions" in a smaller font below. Above the logo is the number "2093400" written in cursive. To the right of the word "UNILIFE" is a square logo with a black and white design. ## SECTION 6 ## Special 510(k) Summary APR - 1 2010 This 510(k) summary has been prepared in compliance with 21 CFR 807.92. | Applicant: | Unilife Medical Solutions Limited<br>633 Lowther Rd<br>Lewisberry, PA 17339<br>Tel: 717-938-9323<br>Fax: 717-938-9364 | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Melissa L. DeHass, Senior Manager, Regulatory Affairs<br>Email: Melissa.dehass@unilife.com | | Date Prepared: | 30 October 2009 | | Trade Name: | 1mL Unitract® Insulin Syringe | | Common Name: | Piston Syringe | | Classification: | Syringe, Piston; 21 CFR 880.5860; FMF; Class II | | Predicate Devices: | Unitract Insulin Syringe (K080877) | | Device Description: | The 1mL Unitract® Insulin Syringe is a sterile, fixed needle, single<br>use, syringe, consisting of a barrel and plunger assembly. The<br>syringe is supplied in two needle sizes, being 27G x ½" and 29G x<br>½".<br><br>The devices function in a manner similar to standard syringes for<br>filling and mixing insulin and subcutaneous injection, but also<br>incorporates a sharps injury prevention (automatic needle retraction)<br>and reuse prevention (auto disable) feature.<br><br>The syringe, as supplied, is a device with a small barrel, plunger and<br>fixed needle, calibrated in units of insulin (U-100), to be used to<br>administer (infuse) insulin subcutaneously. It incorporates features<br>that include reuse prevention and sharps injury prevention.<br><br>The sharps injury prevention feature allows for automatic and full<br>retraction of the needle into the syringe barrel immediately after the<br>full dose has been expelled; where the rate of retraction may be<br>controlled by the user. The reuse prevention feature also engages<br>immediately after retraction, which prevents further movement of<br>the plunger in either direction. | | Indications for Use: | The syringe, as supplied, is a device with a small barrel, plunger and<br>fixed needle, calibrated in units of insulin (U-100), to be used to<br>administer (infuse) insulin subcutaneously. It incorporates features<br>that include reuse prevention and sharps injury prevention. | {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for UNILIFE medical solutions. The word "UNILIFE" is written in a stylized font, with the letters connected to each other. Below the word "UNILIFE" is the phrase "medical solutions" in a smaller font. To the right of the word "UNILIFE" is a graphic that looks like a backwards "L" shape. Substantial Equivalence: The 1mL Unitract® Insulin Syringe has identical indications for use to the cited predicate # K080877 which is to infuse drugs; whilst incorporating a sharps injury prevention feature. The device design and technology is identical to K080877. The label changes to the 1mL Unitract® Insulin Syringe do not raise new questions of safety and effectiveness based on its similarities to the cited predicates in terms of design and intended use. The ImL Unitract® Insulin Syringe has identical needle retraction mechanisms to the cited predicate # K080877. The syringe is filled; and at the end of the injection stroke once the full dose has been expelled, the plunger engages the needle to allow retraction into the syringe barrel. The auto disable feature of the 1mL Unitract® Insulin Syringe is integrated within the plunger design, where after full retraction, the plunger engages with the collar outer to prevent movement of the plunger in either direction. This feature requires no secondary action by the user to disable the syringe. The characteristics of the 1mL Unitract® Insulin Syringe spring based mechanism are identical to the technology of the cited predicate # K080877. The 1mL Unitract® Insulin Syringe spring mechanism is located on the proximal end of the syringe to allow retraction into the barrel. The design of the 1mL Unitract® Insulin Syringe integrates the technological design characteristics of the cited predicate, and having identical operating procedures for aspiration and injection, has been demonstrated to have no significant differences. The syringe components are manufactured from biocompatible materials and stainless steel. Mechanical performance tests were conducted to verify the device meets design specifications based on the requirements of the FDA Consensus Standard and the voluntary standards. Additional performance testing was conducted on the safety features relating to peak activation forces of the retraction mechanism, spring reaction forces and forces to override the auto disable feature. Chemical testing included Reducing (Oxidisable) Matter, Extractable Trace Metals, Non-Volatile Residue, pH and chemical residues (EO Sterilisation). Biocompatibility testing of the finished devices included Cytotoxicity and Pyrogens; with all materials in the fluid path meeting USP Class VI requirements. Sterilisation processes are validated to European Standard DIN EN 550:1994, Sterilisation of medical devices - Validation and routine control of ethylene oxide sterilisation to a sterility assurance level (SAL) of 10°. Manufacturing process control requires functional, chemical and biological tests be performed on each production lot number. The 1mL Unitract® Insulin Syringe with 27G x 1/2" or 29G x 1/2" needle submitted in this 510(k) is determined as substantially equivalent to the cited predicate devices based on performance testing, intended use and design principles of operation and technology, used for the subcutaneous infusion of insulin. Any Characteristics: Technological Non Clinical Performance Data: Conclusion: {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Unilife medical solutions. The word "UNILIFE" is in bold, sans-serif font. Below the word "UNILIFE" is the phrase "medical solutions" in a smaller font. To the right of the word "UNILIFE" is a graphic of a curved line. differences cited do not raise any new questions of safety and effectiveness. [INTENTIONALLY LEFT BLANK BELOW LINE] {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its wing and body. The eagle faces to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Ms. Melissa L. DeHass Senior Manager, Regulatory Affairs Unilife Medical Solutions Limited 633 Lowther Road Lewisberry, Pennsylvania 17339 APR - 1 2010 Re: K093400 Trade/Device Name: 1mL Unitract® Insulin Syringe Regulation Number: 21CFR 880.8560 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: March 3, 2010 Received: March 4, 2010 Dear Ms. DeHass: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address > http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Th for Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for "UNILIFE medical solutions". The logo is in black and white and features a stylized font. To the right of the text is a graphic that resembles a leaf. ## SECTION 5 ## Indications for Use 510(k) Number (if known): K093400 1mL Unitract® Insulin Syringe Device Name: Indications for Use: The syringe, as supplied, is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use ______________________________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Anesthesiology, General Hospital infection Control, Dental Devices 510(k) Number. K093400
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%