1ML UNITRACT INSULIN SYRINGE 27G*1/2 , 1ML UNITRACT SYRINGE 29G*1/2
K080877 · Unilife Medical Solutions Limited · MEG · Oct 21, 2008 · General Hospital
Device Facts
| Record ID | K080877 |
| Device Name | 1ML UNITRACT INSULIN SYRINGE 27G*1/2 , 1ML UNITRACT SYRINGE 29G*1/2 |
| Applicant | Unilife Medical Solutions Limited |
| Product Code | MEG · General Hospital |
| Decision Date | Oct 21, 2008 |
| Decision | SESE |
| Submission Type | Abbreviated |
| Regulation | 21 CFR 880.5860 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The syringe as supplied is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention.
Device Story
1mL Unitract® Insulin Syringe is a sterile, single-use, fixed-needle syringe for subcutaneous insulin (U-100) delivery. Device features automatic needle retraction and auto-disable mechanisms to prevent sharps injuries and reuse. Operation involves standard filling, mixing, and injection steps; upon completion of the injection stroke, the plunger engages a spring-based mechanism to retract the needle into the barrel. Once retracted, the plunger locks with the outer collar to prevent further movement. Used in clinical or home settings by patients or healthcare providers. Benefits include reduced risk of accidental needlestick injuries and prevention of syringe reuse.
Clinical Evidence
Bench testing only. Mechanical performance verified against ISO 8537:1991 and AS 1077:1992. Testing included peak activation forces for retraction, spring reaction forces, and auto-disable override forces. Chemical testing included extractables, residues, and pH. Biocompatibility testing (cytotoxicity, pyrogens) confirmed USP Class VI compliance.
Technological Characteristics
Materials: USP Class VI polymers and stainless steel. Mechanism: Spring-based needle retraction and integrated plunger-collar auto-disable lock. Dimensions: 1mL capacity, 27G x 1/2" or 29G x 1/2" fixed needles. Sterilization: Ethylene oxide per DIN EN 550:1994. Connectivity: None (mechanical device).
Indications for Use
Indicated for subcutaneous administration of U-100 insulin in patients requiring insulin therapy.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
- 1cc Insulin SafePro® Safety Syringe (K050134)
- BD Integra 1mL Syringe (K023752)
Related Devices
- K093400 — 1ML UNITRACT INSULIN SYRINGE 27G X 1/2 AND 29G X 1/2 · Unilife Medical Solutions Limited · Apr 1, 2010
- K062318 — BIOTOP HOOKSAFE U-100 INSULIN SYRINGE, MODELS 0.5CC/ML AND 1CC/ML · Biotop Technology Co., Ltd. · Jan 3, 2007
- K040036 — INVIRO SNAP SAFETY INSULIN SYRINGE, 1 CC, MODEL 102905 · Inviro Medical · Mar 18, 2004
- K100998 — 1ML UNITRACT 27G X 1/2 TUBERCULIN SYRINGE, MODEL 132BY2-1 · Unilife Medical Solutions Limited · Aug 30, 2010
- K152808 — U&U Insulin Syringe with/without Safety Retractable Device · U&U Medical Technology Co, Ltd. · Mar 25, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K0800877" at the top. Below the text is the logo for "UNILIFE medical solutions". The logo is black and white and features a stylized "U" and a curved arrow.
# OCT 21 2008 SECTION 5
## Abbreviated 510(k) Summary
This 510(k) summary has been prepared in compliance with 21 CFR 807.92.
| Applicant: | Unilife Medical Solutions Limited<br>35 Clarence Street<br>Sydney NSW 2000<br>Australia<br>Tel: +61 2 8346 6500<br>Fax: +61 2 8346 6511 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Michelle Gow<br>Email: michelle.gow@unilife.com.au |
| Date Prepared: | March, 2008 |
| Trade Name: | 1mL Unitract® Insulin Syringe |
| Common Name: | Insulin Syringe |
| Classification: | Syringe, Piston; 21 CFR 880.5860; FMF; Class II |
| Predicate Devices: | 1cc Insulin SafePro® Safety Syringe (K050134)<br>BD Integra 1mL Syringe (K023752) |
| Device Description: | The 1mL Unitract® Insulin Syringe is a sterile, fixed needle, single use, insulin (U-100) syringe, consisting of a barrel and plunger assembly. The syringe is supplied in two needle sizes, being 27G x ½" and 29G x ½".<br>The devices function in a manner similar to standard syringes for filling and mixing insulin and subcutaneous injection, but also incorporates a sharps injury prevention (automatic needle retraction) and reuse prevention (auto disable) feature.<br>The sharps injury prevention feature allows for automatic and full retraction of the needle into the syringe barrel immediately after the full dose has been expelled; where the rate of retraction may be controlled by the user. The reuse prevention feature also engages immediately after retraction, which prevents further movement of the plunger in either direction. |
| Indications for Use: | The syringe as supplied is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention. |
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the handwritten text "KO8087" at the top. Below that is the logo for "UNILIFE" in a stylized font. Underneath the logo, in smaller letters, is the text "medical solut on:", which appears to be part of the company's tagline or description.
| Substantial<br>Equivalence: | The 1mL Unitract® Insulin Syringe has identical indications for use and similar features to the cited predicate # K050134 which is to infuse insulin subcutaneously; whilst incorporating a sharps injury prevention feature. In addition, the spring based mechanism of the cited predicate # K023752; and how this mechanism is integrated within the design of the 1mL Unitract® Insulin Syringe do not raise new questions of safety and effectiveness. |
|-----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>Characteristics: | The 1mL Unitract® Insulin Syringe has similar needle retraction mechanisms to the cited predicate # K050134. The syringe is filled; and at the end of the injection stroke once the full dose has been expelled, the plunger engages the needle to allow retraction into the syringe barrel. The cited predicate device relies on manual retraction and plunger fracture, where the 1mL Unitract® Insulin Syringe uses a spring based mechanism to retract the needle. The auto disable feature of the 1mL Unitract® Insulin Syringe is integrated within the plunger design, where after full retraction, the plunger engages with the collar outer to prevent movement of the plunger in either direction. This feature requires no secondary action by the user to disable the syringe. The characteristics of the 1mL Unitract® Insulin Syringe spring based mechanism are similar to the technology of the cited predicate # K023752. The spring mechanism of this predicate device is located at the distal end of the syringe which facilitates the retraction of the needle into the plunger. The 1mL Unitract® Insulin Syringe spring mechanism is located on the proximal end of the syringe to allow retraction into the barrel. The design of the 1mL Unitract® Insulin Syringe integrates the technological design characteristics of the cited predicates, and having similar operating procedures for filling and mixing insulin, aspiration and injection, has been demonstrated to have no significant differences. |
| | The syringe components are manufactured from USP Class VI materials and stainless steel. |
| Non Clinical<br>Performance Data: | Mechanical performance tests were conducted to verify the device meets design specifications based on the requirements of the FDA Consensus Standard ISO 8537:1991 Sterile single-use syringes, with or without needle, for insulin and the voluntary standard AS 1077:1992, Single-use syringes (sterile) for the injection of 100 units per millilitre insulin (U- 100). Additional performance testing was conducted on the safety features relating to peak activation forces of the retraction mechanism, spring reaction forces and forces to override the auto disable feature. Chemical testing included Reducing (Oxidisable) Matter, Extractable Trace Metals, Non-Volatile Residue, pH and chemical residues (EO Sterilisation). Biocompatibility testing of the finished devices included Cytotoxicity and Pyrogens; with all materials in the fluid path meeting USP Class VI requirements. |
.
.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Unilife medical solutions. The logo is in black and white and features the word "Unilife" in a stylized font. Below the word "Unilife" is the phrase "medical solutions" in a smaller font. To the right of the word "Unilife" is a graphic element that appears to be a stylized letter "C".
.... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Sterilisation processes are validated to European Standard DIN EN 550:1994, Sterilisation of medical devices - Validation and routine control of ethylene oxide sterilisation to a sterility assurance level (SAL) of 108.
Manufacturing process control requires functional, chemical and biological tests be performed on each production lot number.
==============================================================================================================================================================================
Conclusion: The 1mL Unitract® Insulin Syringe with 27G x ½" and 29G x ½" needles submitted in this 510(k) is determined as substantially equivalent to the cited predicate devices based on performance testing, intended use and design principles of operation and technology, used for the subcutaneous infusion of insulin. Any differences cited do not raise any new questions of safety and effectiveness.
[INTENTIONALLY LEFT BLANK BELOW LINE]
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
#### OCT 2 1 2008
Ms. Michelle Gow Quality and Regulatory Affairs Director Unilife Medical Solutions Limited Level 5, 35 Clarence Street Sydney, NSW AUSTRALIA 2000
Re: K080877
Trade/Device Name: 1mL Unitract® Insulin Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: September 4, 2008 Received: September 8, 2008
Dear Ms. Gow:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Gow
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Unilife medical solutions. The logo is in black and white and features the word "UNILIFE" in a stylized font, with the words "medical solutions" underneath. Above the logo are some handwritten characters.
## SECTION 4
### Indications for Use
510(k) Number (if known):
Unknown
Device Name:
1mL Unitract® Insulin Syringe
Indications for Use:
The syringe as supplied is a device with a small barrel, plunger and fixed needle, calibrated in units of insulin (U-100), intended to be used to administer (infuse) insulin subcutaneously. It incorporates features that include reuse prevention and sharps injury prevention
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
C
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K080877