K Number
K093333

Validate with FDA (Live)

Date Cleared
2010-04-02

(158 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The PRIMUS PSS8 Steam Sterilizer Series are designed for use in the Hospital operating suites, central sterile supply and clinical laboratories. The PRIMUS PSS8 Steam Sterilizer Series provide efficient steam sterliization of non-porous and porous, heat and moisture stabile materials. The intended use for the sterilizers models listed above is to provide efficient steam sterilization of non-porous and porous, heat and moisture stabile materials, wrapped and unwrapped surgical instruments, hard goods, and linens.

Device Description

The PRIMUS PSS8 Steam Sterilizers (or Autoclaves) are Class II, Product Code FLE Medical Devices as defined by CFRS880.6880 and defined for use in Hospital operating suites, central sterile supply and clinical laboratories. The PRIMUS PSS8 Steam Sterilize Series provide efficient steam sterilization of non-porous and porous, heat and moisture stable materials. The models contained within this submission request utilize the same technology, materials and updated software as predicate devices cleared under K991575 and K082817. The proposed PRIMUS Steam Sterilizer chambers offered within this submission is equipped with the same options offered under the predicate device, in design and construction except for ASME approved (optional) carbon steel reinforced doors, the vessel size and (optional 304 stainless) vessel jackets configured with 316L internal surface stainless steel chambers: offered in either horizontal or vertical sliding door applications. The PSS8-J, K and M units may be pit-mounted, allowing the optional floor carts to roll directly in to the chamber. All sterilizer doors are designed to be efficient, reliable and inherently safe. Pass through (double-door) models are also available. PSS sterilizer units offered under this request for clearance are available in the following configurations: PSS8-AA-M, PSS8-A-M, PSS8-B-M, PSS8-C-M, PSS8-D-M, PSS8-E-M, PSS8-F-M, PSS8-G-M, PSS8-G.1-M, PSS8-J-M, PSS8-K-M, PSS8-M-M. The Multifunction units also offer vacuum, gravity, liquids and test (VAC) configuration.

AI/ML Overview

This document describes the validation of the PRIMUS PSS8 Steam Sterilizers, a Class II medical device, against established safety and effectiveness standards.

1. Acceptance Criteria and Reported Device Performance

The acceptance criteria for the PRIMUS PSS8 Steam Sterilizers are derived from FDA-recognized consensus standards for electrical safety and steam sterilization effectiveness. The reported device performance indicates conformance to these standards.

Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance
Electrical SafetyConformance to: - AAMI / ANSI / IEC 60601-1-2, (Second Edition, 2001), Medical Electrical Equipment -- Part 1-2: Collateral Standard - General Requirements for Safety; Electromagnetic Compatibility - Requirements and Tests. - UL 61010A-1, IEC 61010-1 Amendment 2, and the Part 2, Particular Requirements for Autoclaves Using Steam for the Treatment of Medical Materials & for Laboratory Process, IEC 61010-2-041, UL 61010A-2-041. - CAN/CSA-C22.2 No. 1010 and CAN/CSA-C22.2 No. 1010.2-041-96 (Canadian Standard for Laboratory Equipment).PRIMUS declares conformance to all listed electrical safety standards and applicable industry and electrical codes. Software-designed fail-safe controls provide appropriate warnings and signals. Safety warnings and signals are challenged and verified during a 100% Factory Acceptance Test, with results documented in the Device History Record (DHR).
Sterilization EffectivenessValidation against ANSI/AAMI ST8:2001, Hospital steam sterilizers, ensuring that the sterilizer achieves a Sterility Assurance Level (SAL) of $10^{-6}$ for each load cycle type. The standard establishes minimum construction and performance requirements for hospital sterilizers using saturated steam for volumes greater than 2 cubic feet (56 liters) with no exclusions.PRIMUS has validated sterilizer performance for each load cycle type. Internal validation protocols comply with ANSI/AAMI ST8:2001. Product performance data was independently validated with Biological Indicators (BIs) exposed during validation for each cycle type. Data output from the validation report met a Sterility Assurance Level (SAL) of $10^{-6}$ when the sterilizer is used and maintained as directed.
Pressure Vessel SafetyDesign, manufacture, and testing in accordance with American Society of Mechanical Engineers (ASME), Section VIII, Division 1 Unfired Pressure Vessels.PRIMUS PSS8 Pressure Vessels are designed, manufactured, and tested in accordance with ASME, Section VIII, Division 1 Unfired Pressure Vessels.
User InformationProvision of information in the User's Manual to ensure safe and effective use. Recommendation for periodic monitoring of sterilizer performance in accordance with ANSI/AAMI ST79:2006.PRIMUS provides information in the User's Manual. PRIMUS recommends that a suitable chemical indicator or biological challenge test be used periodically according to health care facilities' documented plan for monitoring SAL $10^{-6}$ performance. PRIMUS also recommends using validated factory preset cycles as per ANSI/AAMI ST79:2006.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a "test set" in the context of a dataset for an AI/algorithm, as the device is a steam sterilizer. Instead, the validation involved testing the sterilizer's performance with various load types and configurations for all models.

  • Sample Size: The sample size for effectiveness testing is implied by the range of sterilizer models and load configurations tested. The table details testing for 12 different sterilizer models (e.g., PSS8-AA-M, PSS8-K-M) across 6 different cycle types (Vacuum, Gravity, Liquids, Test (VAC)) and varying load configurations (e.g., "Unwrapped Nonporous Single Instrument," "Double wrapped instrument trays 16 pounds each tray," "Textile packs 9x9x6, 12 lbs.," "Vented borosilicate glass containers"). Each cycle type and load configuration combination for each sterilizer model represents an individual test scenario. While the exact number of individual runs for each scenario is not provided, the validation covered a comprehensive range of operating conditions.
  • Data Provenance: The study was conducted internally by PRIMUS Sterilizer Company, LLC. The data is prospective, generated through systematic testing and validation protocols to ensure compliance with recognized standards. The country of origin of the data is not explicitly stated but is implicitly the United States, given the company's location (Omaha, NE), the regulatory body (FDA), and the standards referenced (AAMI/ANSI, UL, ASME).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not applicable. The device is a steam sterilizer, not an AI/algorithm that requires interpretation by human experts to establish ground truth for a test set. The efficacy of sterilization is objectively measured using biological indicators (BIs) to demonstrate a Sterility Assurance Level (SAL), which serves as the "ground truth" for sterilization effectiveness.

4. Adjudication Method for the Test Set

This information is not applicable as the device is a steam sterilizer. The "test set" in this context refers to the physical sterilization tests conducted, and the results are objectively measured (e.g., viability of biological indicators), not subject to human adjudication for discrepancies.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

This information is not applicable. The PRIMUS PSS8 Steam Sterilizers are physical medical devices designed for sterilization, not AI-powered diagnostic or assistive tools that would involve human readers or MRMC studies.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This information is not applicable as the device is a steam sterilizer. While the sterilizer does contain "validated software" for fail-safe controls, this software operates as an integral part of the mechanical and thermal system, not as a standalone AI algorithm whose performance is evaluated independently in comparison to human performance.

7. The Type of Ground Truth Used

The ground truth used for effectiveness validation was the Sterility Assurance Level (SAL) of $10^{-6}$, as measured by biological indicators (BIs).

  • Biological Indicators (BIs): These contain a known number of highly resistant bacterial spores. After a sterilization cycle, the BIs are incubated, and their viability (or lack thereof) indicates whether the sterilization process was successful in killing all microorganisms, achieving the specified SAL.
  • Conformance to ANSI/AAMI ST8:2001: This standard defines the performance criteria for hospital steam sterilizers, including the acceptable SAL.

8. The Sample Size for the Training Set

This information is not applicable in the context of typical AI/ML training sets. While there is a "validated software" designed into the sterilizers, it is not described as an AI or machine learning model that undergoes a "training" phase with a dataset in the conventional sense. The software's reliability and functionality are likely established through rigorous software validation and verification processes according to industry standards, rather than statistical training on a large dataset.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the same reasons mentioned in point 8. The "ground truth" for the software refers to its ability to correctly execute control functions, provide accurate readings, and trigger appropriate warnings/signals in various operational and malfunction scenarios. This is established through software testing, simulation, and hardware-in-the-loop testing, not through a "training set" of data with pre-established ground truth labels as would be used for AI/ML models.

{0}------------------------------------------------

1093333

APR - 2 2010

Section 5

A 510(k) SUMMARY PERTAINING TO THE SAFETY AND EFFECTIVENESS OF PRIMUS PSS8 STEAM STERILIZERS

Manufacturer:PRIMUS Sterilizer Company, LLC.117 South 25th StreetOmaha NE 68131Phone: 402-344-4200Fax: 402-344-4242
Contact:Connie Mansfield, Mgr., Marketing Communications & Regulatory ComplianceOmaha NE 68131Phone: 402-344-4200 Ext 1229
Date Prepared:March 12, 2009

Introduction:

The PRIMUS PSS8 Steam Sterilizers (or Autoclaves) are Class II, Product Code FLE Medical Devices as defined by CFRS880.6880 and defined for use in Hospital operating suites, central sterile supply and clinical laboratories. The PRIMUS PSS8 Steam Sterilize Series provide efficient steam sterilization of non-porous and porous, heat and moisture stable materials. The models contained within this submission request utilize the same technology, materials and updated software as predicate devices cleared under K991575 and K082817.

The proposed PRIMUS Steam Sterilizer chambers offered within this submission is equipped with the same options offered under the predicate device, in design and construction except for ASME approved (optional) carbon steel reinforced doors, the vessel size and (optional 304 stainless) vessel jackets configured with 316L internal surface stainless steel chambers: offered in either horizontal or vertical sliding door applications. The PSS8-J, K and M units may be pit-mounted, allowing the optional floor carts to roll directly in to the chamber. All sterilizer doors are designed to be efficient, reliable and inherently safe. Pass through (double-door) models are also available.

PSS sterilizer units offered under this request for clearance are available in the following configurations. The Multifunction units also offer vacuum, gravity, liquids and test (VAC) configuration.

PSS8-AA-M**(16" x 16" x 26" Multi-Functional Sterilizer)Single or Double Drains
PSS8-A-M**(20" x 20" x 38" Multi-Functional Sterilizer)Single or Double Drains
PSS8-B-M**(26" x 26" x 39" Multi-Functional Sterilizer)Single or Double Drains
PSS8-C-M**(26" x 26" x 49" Multi-Functional Sterilizer)Single or Double Drains
PSS8-D-M**(26" x 26" x 67" Multi-Functional Sterilizer)Single or Double Drains
PSS8-E-M**(26" x 36" x 39" Multi-Functional Sterilizer)Single or Double Drains
PSS8-F-M**(26" x 36" x 48" Multi-Functional Sterilizer)Single or Double Drains
PSS8-G-M**(26" x 36" x 60" Multi-Functional Sterilizer)Single or Double Drains
PSS8-G.1-M***(32" x 36" x 48" Multi-Functional Sterilizer)Single or Double Drains
PSS8-J-M**(26" x 63" x 48" Multi-Functional Sterilizer)Single or

or Door OL Door OT Door DI Door or Door or Door Dr Double Door

Door

Door

Door

Section 5 Page 1 of 4

{1}------------------------------------------------

PSS8-K-M**(26" x 63" x 76" Multi-Functional Sterilizer)Single or
Double Door
PSS8-M-M***(35" x 57" x 60" Multi-Function Sterilizer)Single or
Double Door

PRIMUS Steam Sterilizers models defined above are as described in Section 4 "Indications for Use Statement" and contained within this submission

The PRIMUS PSS8 series sterilizers are offered with factory pre-set sterilization cycles described within the Indication for Use Table (See attached Table og 3).

Safety

The PRIMUS PSS8 Steam Sterilizer Series have been validated against FDA recognized consensus standards for electrical safety:

  • AAMI / ANSI / IEC 60601-1-2, (Second Edition, 2001), Medical Electrical Equipment -- Part 1-2: . Collateral Standard - General Requirements for Safety; Electromagnetic Compatibility - Requirements and Tests.
    Additionally PRIMUS declares conformance to applicable industry and electrical codes as follows

  • . UL 61010A-1, IEC 61010-1 Amendment 2, and the Part 2, Particular Requirements for Autoclaves Using Steam for the Treatment of Medical Materials & for Laboratory Process, IEC 61010-2-041, UL 61010A-2-041.

  • CNL indicates the product was evaluated to the Canadian Standard for Laboratory Equipment, CAN/CSA-● C22.2 No. 1010 and the Part 2, Particular Requirements for Autoclaves Using Steam for the Treatment of Medical Materials and Laboratory Process, CAN/CSA-C22.2 No. 1010.2-041-96.

  • PRIMUS PSS8 Pressure Vessels are designed, manufactured and tested in accordance with American . Society of Mechanical Engineers (ASME), Section VIII, Division 1 Unfired Pressure Vessels.

Validated software designed into the PRIMUS PSS8 series sterilizer provides for fail-safe controls that give appropriate warnings and signals when required conditions have not been met or if unit malfunctions. The technology designed into the PRIMUS PSS Steam Sterlizer Series provides for fail safe controls that provide end users with appropriate warnings and signals when required conditions have not been met or malfunctions occurs. Safety warnings and signals are challenged and verified on each unit, as a function of routine process control testing, during a 100% Factory Acceptance Test conducted at final inspection. Results are documented and maintained in the Device History Record (DHR).

Effectiveness:

PRIMUS has validated sterilizer performance for each load cycle type to ensure that the exposure time, provided in the directions for use statements, have been proven to assure the safety and effectiveness of the PRIMUS PSS8 series sterilizer. Internal validation protocols are documented and deemed to be comply to the ANSI/AAMI ST8:2001, Hospital steam sterilizers - (Sterility); a harmonized standard recognized FDA as applicable to the validation of steam sterilizers intended for use in hospitals and other health care facilities and that have a volume greater than 2 cubic feet (ft') (56 liter {L] ) with no exclusions taken. The recognized standard established the minimum construction requirements and performance requirements for hospital sterilizers, using saturated steam as the sterilizing agent.

Declarations of Conformance made by PRIMUS to recognized standards are based on a verification of product performance data, that was independently validated with BI indicators exposed during validation for each cycle type published within the Indications for Use statement. Data output from the validation report was reviewed and found to meet Sterility Assurance Level (SAL) of 10th when the sterilizer is used and maintained as directed.

Operator Information

PRIMUS provides information in the User's Manual that is intended to ensure safe and effective use of the sterilizer. Additional information concerning recommended practices for end users in monitoring sterilizer performance can be found in the ANSI/AAMI ST79:2006 Comprehensive guide to steam sterility assurance in the health

{2}------------------------------------------------

care facilities or other applicable standards to assure safe and effective use of the steam sterilization processes for application.

Based on the testing and comparison to the consensus standard, PRIMUS concludes that each of the chamber sizes included under this submission performs as intended and raises no new safety or effectiveness issues when used as directed.

Connie L. Mangold

Mgr., Marketing Communications and Regulation Compliance PRIMUS Sterilizer Company, LLC

{3}------------------------------------------------

Cycle #Cycle TypeSterilize TempSterilize Time (min)Dry Time (min)No. of Pre-vacs16x16x2620x20x3826x26x3926x26x4926x26x6726x36x3926x36x4826x36x6032x36x4826x63x4826x63x7635x57x60
2Vacuum132°C(270°F)413N/A111N/A11N/A1N/AN/AN/APorous Single
3Vacuum132°C(270°F)4303N/A11111111111Instrumenteach tray
4Vacuum132°C(270°F)"4303N/A246N/A612N/A15N/AN/AN/A9x6, 12 lbs.
6Gravity121.1°C(250°F)30300N/A224612161616306042maximum size:pounds
7Liquids121.1°C(250°F)3015*057111149189N/A207255N/A284N/AN/AN/APlate glassml or smaller.
8Test (VAC)132°C(270°F)3-1/223N/A136181228313252309388390543860615
1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack1 TestPack
ShelvingLoading Eq.ShelvingLoading Eq.ShelvingLoading Eq.ShelvingLoading Eq.ShelvingLoading Eq.ShelvingLoading Eq.
Maximum Load Configurations

** Dry time in LIQUIDS cycle is liquid cool time.

WARNING

THE INTENDED USE OF A LIQUID CYCLE IS NOT FOR STERILIZING MATERIALS FOR DIRECT PATIENT CONTACT

WARNING

PRIMUS RECOMMENDS THE VALIDATED FACTORY PRESET CYCLES. CHANGES TO THE CYCLE PARAMETERS,
PER ANSI/AAM ST79:2006, ARE NOT RECOMMENDED. CONTACT PRIMUS FOR FURTHER INFORMATION.

NOTE

STERILIZE TEMP PARAMETER IS PROGRAMMED TO ENSURE STERILIZE TEMP MEETS ANSI/AAM ST8
REQUIREMENTS FOR EACH FACTORY VALIDATED CYCLE.

{4}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features an abstract image of an eagle with its wings spread, symbolizing protection and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle, indicating the department's name and national affiliation.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR - 2 2010

Ms. Connie Mansfield Manager, Marketing Communications Primus Sterilizer Company, LLC 117 South 250 Street Omaha, Nebraska 68131

Re: K093333

Trade/Device Name: PRIMUS Steam Sterilizers Regulation Number: 21CFR 880.6880 Regulation Name: Steam Sterilization Regulatory Class: II Product Code: FLE Dated: March 26, 2010 Received: March 26, 2010

Dear Ms. Mansfield:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Ms. Mansfield

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

for

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Section 4

Indications for Use Statement

Applicant: PRIMUS Sterilizer Company, LLC

510(k) Number: K0933333

Device Name: PRIMUS Steam Sterilizers

Indication for Use:

The PRIMUS PSS8 Steam Sterilizer Series are designed for use in the Hospital operating suites, central sterile supply and clinical laboratories. The PRIMUS PSS8 Steam Sterilizer Series provide efficient steam sterliization of non-porous and porous, heat and moisture stabile materials. The PRIMUS PSSS Steam Sterilizer Series are available in the following configurations:

PSS8-AA-M**(16" x 16" x 26" Multi-Functional Sterilizer)Single orDouble Door
PSS8-A-M**(20" x 20" x 38" Multi-Functional Sterilizer)Single orDouble Door
PSS8-B-M**(26" x 26" x 39" Multi-Functional Sterilizer)Single orDouble Door
PSS8-C-M**(26" x 26" x 49" Multi-Functional Sterilizer)Single orDouble Door
PSS8-D-M**(26" x 26" x 67" Multi-Functional Sterilizer)Single orDouble Door
PSS8-E-M**(26" x 36" x 39" Multi-Functional Sterilizer)Single orDouble Door
PSS8-F-M**(26" x 36" x 48" Multi-Functional Sterilizer)Single orDouble Door
PSS8-G-M**(26" x 36" x 60" Multi-Functional Sterilizer)Single orDouble Door
PSS8-G.1-M***(32" x 36" x 48" Multi-Functional Sterilizer)Single orDouble Door
PSS8-J-M**(26" x 63" x 48" Multi-Functional Sterilizer)Single orDouble Door
PSS8-K-M**(26" x 63" x 76" Multi-Functional Sterilizer)Single orDouble Door
PSS8-M-M***(35" x 57" x 60" Multi-Function Sterilizer)Single orDouble Door

Sterilizers are available in multi-functional models which feature vacuum, gravity, liquids and test (VAC) cycles.

· PRIMUS Sterilizer Company, LLC recommends that a suitable chemical indicator or biological challenge test be used on a periodic basis to test effectiveness of the sterilizer in accordance with health care facilities docurented plan for monitoring SAL 10d performance.

The intended use for the sterilizers models listed above is to provide efficient steam sterilization of non-porous and porous, heat and moisture stabile materials, wrapped and unwrapped surgical instruments, hard goods, and linens. The cycles to be cleared for each of the models listed above are found in the table below:

{7}------------------------------------------------

Load TypeCycle #Cycle TypeSterilize TempSterilize Time (min)Dry Time (min)No. of Pre-vacs16x16x2620x20x3826x26x3926x26x4926x26x6726x36x3926x36x4826x36x6032x36x4826x63x4826x63x7635x57x60
Unwrapped Nonporous Single Instrument2Vacuum132°C(270°F)413N/A111N/A11N/A1N/AN/AN/A
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AShelving
Loading Eq.
Double wrapped instrument trays 16 pounds each tray3Vacuum132°C(270°F)4303N/A24686121415306042
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AShelving
Loading Eq.
Textile packs 9x9x6, 12 lbs.4Vacuum132°C(270°F)4303N/A4612121412161816306042
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AShelving
Loading Eq.
Fabric packs maximum size:12x12x20, 12 pounds6Gravity121.1°C(250°F)30300N/A22469121416205040
N/AN/AN/AN/AN/AN/AN/AN/AN/AN/AN/AShelving
Loading Eq.
Vented borosilicate glass containers, 1000 ml or smaller, 4"x7"7Liquids121.1°C(250°F)3015*0N/A57111149189207255N/A284N/AN/AN/A
N/A136181228313252309388390543860615
Loading Eq.
Bowie-Dick Test8Test (VAC)132°C(270°F)3-1/22
1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack1 Test Pack
Shelving
Loading Eq.

.

.

:

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).