K Number
K093300

Validate with FDA (Live)

Date Cleared
2009-10-30

(9 days)

Product Code
Regulation Number
878.4460
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

These powder-free surgeon's gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination in the environments within hospitals and other healthcare facilities.

Device Description

Esteem SMT Polyisoprene Powder-Free Surgical gloves are formulated using synthetic rubber latex. They are offered powder-free and sterile.

AI/ML Overview

The Esteem SMT Polyisoprene Powder-Free Surgical Gloves were subject to several performance tests to demonstrate substantial equivalence. The acceptance criteria and reported device performance are summarized below:

1. Table of Acceptance Criteria and Reported Device Performance

Test:Acceptance Criteria:Reported Device Performance:
Primary Skin IrritationGloves must be non-irritating.Gloves are non-irritating.
Guinea Pig MaximizationGloves must not display any potential for sensitization.Gloves do not display any potential for sensitization.
DimensionsMust meet requirements of ASTM D3577.Gloves meet requirements of ASTM D3577.
Physical CharacteristicsMust meet requirements for rubber surgical gloves per ASTM D3577.Gloves meet requirements for rubber surgical gloves per ASTM D3577.
Freedom from HolesMust meet requirements of 21 CFR 800.20 and ASTM D3577.Gloves meet requirements of 21 CFR 800.20 and ASTM D3577.
Powder ResidualMust meet powder level requirements for "Powder-Free" designation per ASTM D3577, tested using ASTM standard D6124.Results generated values below 2mg of residual powder per glove, meeting the "Powder-Free" designation.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify the exact sample sizes used for each test. The tests were performed by Cardinal Healthcare 222 LTD. in Rayong, Thailand, which is the country of origin. The study appears to be prospective testing of the manufactured gloves.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This product is a medical glove, not an AI or imaging device that requires interpretation by medical experts. Therefore, the concept of "experts establishing ground truth for the test set" and their qualifications as typically understood in AI/imaging studies (e.g., radiologists) is not applicable. The ground truth for glove performance is established by standardized material and physiological testing methods.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable as this is not an AI or imaging device requiring human adjudication of results from a diagnostic test. Performance is determined by established physical and biological testing standards.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable as this is a physical medical device (surgical gloves), not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable as this is a physical medical device. The "standalone" performance here refers to the glove's intrinsic properties as measured by the specified tests, without human intervention in the measurement of those properties (although humans would be performing the tests).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for this device is based on established industry standards and regulatory requirements for medical gloves. This includes:

  • ASTM D3577: Standard Specification for Rubber Surgical Gloves.
  • 21 CFR 800.20: Regulations regarding freedom from holes in surgeon's gloves.
  • ASTM D6124: Standard Test Method for Residual Powder on Medical Gloves.
  • Biological safety standards: For skin irritation and sensitization (e.g., Primary Skin Irritation and Guinea Pig Maximization tests).

These standards define the objective criteria and methods against which the glove's performance is measured.

8. The sample size for the training set

Not applicable. This is a physical device, not a machine learning model, hence no "training set" in the context of AI.

9. How the ground truth for the training set was established

Not applicable for a physical device.

{0}------------------------------------------------

K 093300

Page 1 of 2

Image /page/0/Picture/2 description: The image shows the logo for Cardinal Health. The logo consists of a stylized bird-like graphic above the company name. The text "CardinalHealth" is written in a bold, sans-serif font, with the first letter of each word capitalized.

OCT 8 0 2009

510(K) SUMMARY OF SAFETY AND EFFECTIVENESS

ESTEEM SMT POLYISOPRENE POWDER-FREE SURGICAL GLOVES

Manufacturer:Cardinal Healthcare 222 LTD.
7/111 Moo 4, Highway 331 Mabyangporn Pluakdaeng,
Rayong, 21140, Thailand
Regulatory Affairs Contact:Tatyana Bogdan, RAC
Cardinal Health, Inc.
1430 Waukegan Road
McGaw Park, IL 60085
Telephone:847-578-2325
Date Summary Prepared:October 14, 2009
Product Trade Name:Esteem SMT Polyisoprene Powder-Free Surgical Gloves
Common Name:Surgeon's Gloves
Classification Name:Surgeon's Gloves
Device Description:Esteem SMT Polyisoprene Powder-Free Surgical gloves are formulated using synthetic rubber latex. They are offered powder-free and sterile.
Intended Use:Esteem SMT Polyisoprene Powder-Free Surgical Gloves are a disposable device that is intended to be worn by operating room personnel to protect a surgical wound from contamination in the environments within hospitals and other healthcare facilities.
Predicate Devices:Esteem Polyisoprene Powder-Free Surgical Gloves previously cleared under 510(k) K011721 (product code KGO);
Substantial Equivalence:Esteem SMT Polyisoprene Powder-Free Surgical gloves are substantially equivalent to Esteem Polyisoprene Powder-Free

·

5

{1}------------------------------------------------

K093300

page 2 of 2

Surgical gloves. Substantial equivalence can be established in regard to intended use, physical characteristics, design and product features. Both gloves are made of synthetic rubber latex using similar manufacturing process and packaging.

Performance Testing:

Test:Result:
Primary Skin IrritationGloves are non-irritating.
Guinea Pig MaximizationGloves do not display any potential for sensitization.
DimensionsGloves meet requirements of ASTM D3577.
Physical CharacteristicsGloves meet requirements for rubber surgical gloves per ASTMD3577.
Freedom from HolesGloves meet requirements of 21 CFR 800.20 and ASTM D3577
Powder ResidualGloves meet powder level requirements for "Powder-Free"designation per ASTM D3577 tested using ASTM standardD6124, Standard test method for residual powder on medicalgloves. Results generated values below 2mg of residual powderper glove.
Clinical Data:No clinical data was required.
Conclusion:The Esteem SMT Polyisoprene Powder-Free Surgical Gloves meetthe technological characteristics of ASTM D3577 and aresubstantially equivalent to predicate devices identified in this510(k) summary.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is centered in the image. The font is a serif font, and the text is black. The background is white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Cardinal Health, Incorporated C/O Mr. Ned Devine Responsible Third Party Official Underwriters Laboratories Incorporated 333 Pfingsten Road Northbrook, Illinois 60062-2096

OCT 3 0 2009

Re: K093300

Trade/Device Name: Esteem SMT Polyisoprene Powder-Free Surgical Sterile Gloves Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class: I Product Code: KGO Dated: October 19, 2009 Received: October 21, 2009

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Devine

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthem Orenton

Susan Runner, D.D.S., M.A. Acting Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

FDA CDRH DMC

OCT 21 2009

Received

Indications for Use

510(k) Number (if known): _ | 093300

Device Name:Esteem SMT Polyisoprene Powder-Free Surgical Sterile Gloves
---------------------------------------------------------------------------

Indications for Use:

These powder-free surgeon's gloves are a disposable device made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination in the environments within hospitals and other healthcare facilities.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)
----------------------------------------------------------

Shippy to the paris Wills

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K O 93300

Cardinal Health, Inc. Premarket Notification Submission - Traditional 510(k)

§ 878.4460 Non-powdered surgeon's glove.

(a)
Identification. A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls).