K Number
K093275

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-05-27

(220 days)

Product Code
Regulation Number
888.3660
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.

Device Description

The proposed Equinoxe Reverse Shoulder System 36mm Glenosphere is a modification to the existing Equinoxe Reverse Shoulder System glenospheres previously cleared in K063569. The 36mm Glenosphere mates with previously cleared Equinoxe Reverse Shoulder glenoid baseplate and the glenosphere locking screw (K063569). The rationale for the modification of the devices is to offer a smaller glenosphere for patients in whom the 38mm glenosphere is too large. The only modifications to the predicate devices consist of a proposed dimensional change to decrease the radius of curvature on the articulating surfaces of the glenosphere to 36mm, as well as the removal of the extractor cavities on the side of the glenosphere due to size limitations.

The proposed Equinoxe Reverse Shoulder System 36mm Humeral Liners are a modification to the existing Equinoxe Reverse Shoulder System humeral liners previously cleared in K063569. The 36mm Humeral Liners mate with previously cleared Equinoxe Reverse Shoulder humeral tray (K063569). The rationale for the modification of the devices is to offer a smaller humeral liner for patients in whom the 38mm humeral liner is too large. The only modifications to the predicate devices consist of a proposed dimensional change to decrease the radius of curvature on the articulating surfaces of the humeral liners to 36mm.

AI/ML Overview

This document is a 510(k) Summary of Safety and Effectiveness for the Exactech® Equinoxe® Reverse Shoulder System 36mm Glenosphere and Humeral Liners. It demonstrates substantial equivalence to a previously cleared predicate device rather than presenting a standalone study with acceptance criteria for a new device. Therefore, much of the requested information regarding acceptance criteria, human reader studies, and training set details are not applicable in this context.

However, I can extract information related to the non-clinical performance data and the conclusion of substantial equivalence.

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a substantial equivalence submission for a modification (dimensional change) to an existing device, the "acceptance criteria" are not framed as a specific set of target performance metrics but rather as demonstrating that the modified device performs comparably to the predicate device and meets established engineering standards. The reported device performance is presented as the findings from the non-clinical tests.

Acceptance Criteria (Implied by Substantial Equivalence to Predicate)Reported Device Performance
Stresses on glenosphere are below ASTM yield strength.Finite Element Analysis: Simulated the worst-case loading condition and demonstrated the stresses on the 36mm Equinoxe Glenosphere were below the ASTM yield strength.
Stability of glenosphere and liners are as intended.Dynamic Loading Study: Assessed the stability of the Equinoxe 36mm glenosphere and liners in a polyurethane bone substitute. The study demonstrated that the proposed devices performed as intended at varying degrees of adduction.
Similar relative stability and range of motion to predicate device.Geometric Analysis: Evaluated the jump distance between the proposed Equinoxe 36mm glenospheres and liners and their predicates, demonstrating the devices are expected to provide similar relative stability and range of motion.
Proper articulation of the devices.Polyurethane Bone Substitute Validation: Performed to assess passive range of motion for the proposed devices. Proper articulation was achieved during this validation.
Same intended use, design features, materials, shelf life, packaging, and sterilization as predicate.The document explicitly states these similarities: "The predicate and proposed devices have the same intended use and basic fundamental scientific technology and share the following similarities: The same indications for use, The same design features, Incorporate the same materials, The same shelf life, Are packaged and sterilized using the same materials and processes." (Modification is only dimensional change).

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not explicitly stated in terms of patient data or clinical images. The testing involved:
    • One 36mm Equinoxe Glenosphere for Finite Element Analysis.
    • Unspecified number of 36mm glenospheres and liners for the dynamic loading study.
    • Unspecified number of 36mm glenospheres and liners for geometric analysis.
    • Unspecified number of 36mm glenospheres and liners for polyurethane bone substitute validation.
  • Data Provenance: The data is from non-clinical tests (Finite Element Analysis, dynamic loading study, geometric analysis, polyurethane bone substitute validation) conducted by the manufacturer, Exactech, Inc. It is not clinical data from patients.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. Ground truth, in this context, would typically refer to clinical diagnoses or outcomes. This submission focuses on engineering analysis and mechanical testing, which rely on established engineering principles and measurements, not expert clinical consensus for a "ground truth" derived from patient data.

4. Adjudication Method for the Test Set

Not applicable. There was no clinical test set requiring adjudication of findings by human readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

Not applicable. This device is a shoulder prosthesis, not an AI-powered diagnostic or therapeutic tool. Therefore, MRMC studies involving human readers and AI are not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is a physical medical implant, not an algorithm.

7. The Type of Ground Truth Used

The "ground truth" for the non-clinical tests was based on:

  • Engineering principles and calculations (e.g., ASTM yield strength for FEA).
  • Physical measurements and observations from mechanical and simulation studies (e.g., performance of devices in dynamic loading, proper articulation in bone substitute).
  • Geometric analysis against the known dimensions and performance of the predicate device.

8. The Sample Size for the Training Set

Not applicable. This is a physical device, and the submission does not describe a machine learning algorithm or a "training set" in that context. The design and testing are based on engineering principles and previous device performance.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for a machine learning algorithm involved.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized abstract design of three human profiles facing right, arranged in a stacked formation. The profiles are connected by flowing lines, creating a sense of unity and movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the abstract design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 2, 2016

Exactech, Incorporated Ms. Lindy Knisely Regulatory Affairs Specialist 2320 NW 66th Court Gainesville, Florida 32653

Re: K093275

Trade/Device Name: Exactech Equinoxe® Reverse Shoulder System 36 mm Glenosphere and Humeral Liners Regulation Number: 21 CFR 888.3660 Regulation Name: Shoulder joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: PHX, KWS, KWS, KWT Dated: April 26, 2010 Received: April 29, 2010

Dear Ms. Knisely:

This letter corrects our substantially equivalent letter of May 27, 2010.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

Page 2 - Ms. Lindy Kniselv

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Lori A. Wiggins -S

for

  • Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
    Enclosure

{2}------------------------------------------------

Exactech® Equinoxe® Reverse Shoulder System 36mm Glenosphere and Humeral Liners Special 510(k) - Indications for Use (DRAFT)

Indications for Use Statement

510(k) Number:

Device Name: Exactech® Equinoxe® Reverse Shoulder System 36mm Glenosphere and Humeral Liner

INDICATIONS FOR USE:

The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.

Prescription Use X and/or (Part 21 CFR 801 Subpart D)

K093275

Over-The-Counter Use (21 CFR 807 Subpart C)

Please do not write below this line - use another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Smetu 3 for mxn
(Division Sign)m

vision Sien ivision of Surgical, Orthopedic, and Restorative Devices

510(k) Number K093275

{3}------------------------------------------------

Exactech® Equinoxe® Reverse Shoulder System 36mm Glenosphere and Humeral Liners Special 510(k) - 510(k) Summary of Safety and Effectiveness

MAY 2 7 2010

. . .

510(k) Summary of Safety and Effectiveness

Exactech® Inc. Sponsor: 2320 N.W. 66th Court Gainesville, FL 32653

Phone: (352) 377-1140 Fax: (352) 378-2617

FDA Establishment Number 1038671

  • Lindy Knisely, RN Contact: Regulatory Affairs Specialist
    May 26, 2010 Date:

Trade or Proprietary or Model Name(s):

Exactech® Equinoxe® Reverse Shoulder System 36mm Glenosphere Exactech Equinoxe Reverse Shoulder System 36mm Unconstrained Humeral Liner (+0mm) Exactech Equinoxe Reverse Shoulder System 36mm Unconstrained Humeral Liner (+2.5mm) Exactech Equinoxe Reverse Shoulder System 36mm Constrained Humeral Liner (+0mm) Exactech Equinoxe Reverse Shoulder System 36mm Constrained Humeral Liner (+2.5mm)

Common Name: Reverse Shoulder Prosthesis

Classification Name:

Shoulder joint metal/polymer non-constrained cemented prosthesis (21 CFR 888.3650, Class II, Product Code KWT) Prosthesis, Shoulder, Semi-constrained, metal/polymer cemented (21 CFR 888.3660, Class II, Product Code KWS)

Information on devices to which Substantial equivalence is claimed:

510(k)NumberTrade or Proprietary or Model NameManufacturer
K063569Equinoxe Reverse Shoulder SystemExactech, Inc.
Indications for Use:

{4}------------------------------------------------

Exactech® Equinoxe® Reverse Shoulder System 36mm Glenosphere and Humeral Liners Special 510(k) - 510(k) Summary of Safety and Effectiveness

The Equinoxe Reverse Shoulder System is indicated for use in skeletally mature individuals with degenerative diseases of the glenohumeral joint and a grossly deficient, irreparable rotator cuff. The Equinoxe Reverse Shoulder is also indicated for a failed glenohumeral joint replacement with loss of rotator cuff function resulting in superior migration of the humeral head.

Device Description:

Glenosphere

The proposed Equinoxe Reverse Shoulder System 36mm Glenosphere is a modification to the existing Equinoxe Reverse Shoulder System glenospheres previously cleared in K063569. The 36mm Glenosphere mates with previously cleared Equinoxe Reverse Shoulder glenoid baseplate and the glenosphere locking screw (K063569). The rationale for the modification of the devices is to offer a smaller glenosphere for patients in whom the 38mm glenosphere is too large.

The predicate and proposed devices have the same intended use and basic fundamental scientific technology and share the following similarities:

  • · The same indications for use
  • · The same design features
  • · Incorporate the same materials
  • · The same shelf life
  • · Are packaged and sterilized using the same materials and processes.

The only modifications to the predicate devices consist of a proposed dimensional change to decrease the radius of curvature on the articulating surfaces of the glenosphere to 36mm, as well as the removal of the extractor cavities on the side of the glenosphere due to size limitations.

Humeral Liners

The proposed Equinoxe Reverse Shoulder System 36mm Humeral Liners are a modification to the existing Equinoxe Reverse Shoulder System humeral liners previously cleared in K063569. The 36mm Humeral Liners mate with previously cleared Equinoxe Reverse Shoulder humeral tray (K063569). The rationale for the modification of the devices is to offer a smaller humeral liner for patients in whom the 38mm humeral liner is too large.

The predicate and proposed devices have the same intended use and basic fundamental scientific technology and share the following similarities:

· The same indications for use

{5}------------------------------------------------

Exactech® Equinoxe® Reverse Shoulder System 36mm Glenosphere and Humeral Liners Special 510(k) - 510(k) Summary of Safety and Effectiveness

  • · The same design features
  • · Incorporate the same materials
  • · The same shelf life
  • · Are packaged and sterilized using the same materials and processes.

The only modifications to the predicate devices consist of a proposed dimensional change to decrease the radius of curvature on the articulating surfaces of the humeral liners to 36mm.

Substantial Equivalence of Non-Clinical Performance Data:

Mechanical tests, engineering analyses, and simulated surgical (sawbones) implantations were conducted to demonstrate substantial equivalence to the predicate devices listed above. A summary of these tests and analyses are as follows:

  • . A Finite Element Analysis of 36mm Equinoxe Glenosphere which simulated the worstcase loading condition and demonstrated the stresses on the device were below the ASTM yield strength.
  • . A dynamic loading study in which the stability of the Equinoxe 36mm glenosphere and liners was assessed in a polyurethane bone substitute. The study demonstrated that the proposed devices performed as intended at varying degrees of adduction.
  • A geometric analysis evaluating the jump distance between the proposed Equinoxe . 36mm glenospheres and liners and their predicates demonstrated the devices are expected to provide similar relative stability and range of motion.
  • A polyurethane bone substitute validation was performed to assess passive range of . motion for the proposed devices. Proper articulation was achieved during this validation.

Substantial Equivalence Conclusion:

Results of mechanical test and engineering analysis referenced in this 510(k) submission demonstrate that the Exactech Equinoxe 36mm Glenosphere and Humeral Liners are substantially equivalent to the cleared predicate devices.

N/A