(77 days)
GALILEOS Implant is intended for use as planning and simulation software to aid qualified dental professionals in the placement of dental implants and the planning of surgical implant treatments. GALILEOS Implant is based on medical imaging information presented by the Sirona GALAXIS 3D viewer and produced by Sirona GALILEOS medical cone beam scanner. The dental professionals' input information may be exported from GALILEOS Implant and used as input data for CAD (Computer-Aided Design) or Rapid Prototyping , Systems.
GALILEOS Implant V1.7 is a pure software device. GALILEOS Implant V 1.7 is a planning and simulation software to aid qualified dental professionals in the placement of dental implants and the planning of surgical implant treatments. GALILEOS Implant is based on medical imaging information presented by the Sirona GALAXIS 3D viewer and produced by Sirona GALILEOS medical cone beam scanner. GALILEOS Implant V1.7 allows to name, position, move, rotate, resize and visualize dental implants and other planning objects (i.e. nerve canals) within the visualized 3D volume. Thus, dental professionals like implantologists are enabled to precisely plan the positions, orientations, types and sizes of Implants to be placed in the patient's mandible/maxilla together with the related surgical procedures. The dental professionals' planning data may be exported from GALILEOS Implant V1.7 and used as input data for CAD (Computer-Aided Design) or Rapid Prototyping Systems.
The provided document is a 510(k) summary for the GALILEOS Implant V 1.7 software. It describes the device's intended use and claims substantial equivalence to predicate devices. However, it does not contain information about specific acceptance criteria, a study that proves the device meets those criteria, or the detailed performance metrics of the device itself.
The submission focuses on establishing substantial equivalence based on intended use, features, and technical characteristics, and mentions "Performance testing to validate the safety and effectiveness of the GALLEOS Implant V 1.7 system included validation testing and bench tests of the software functions." However, the results of this performance testing are not provided in this summary.
Therefore, I cannot provide the requested information. The document explicitly states:
- There is no table of acceptance criteria and reported device performance.
- There is no sample size used for a test set, nor data provenance.
- There is no information on the number or qualifications of experts used for ground truth.
- There is no adjudication method mentioned.
- There is no multi-reader multi-case (MRMC) comparative effectiveness study mentioned, nor any effect size.
- There is no standalone (algorithm only) performance study detailed.
- The type of ground truth used is not specified.
- There is no sample size for the training set.
- There is no information on how ground truth for the training set was established.
The document is a regulatory filing focused on establishing substantial equivalence, not a detailed technical report of performance studies.
{0}------------------------------------------------
510(k) Summary
for
GALILEOS Implant V 1.7
DEC 1 7 2009
I Company Name and Address
- 1.1 Sponsor
SICAT GmbH & Co. KG Brunnenallee 6 D-53177 Bonn Germany
Manufacturer
SICAT GmbH & Co. KG Brunnenallee 6 D-53177 Bonn Germany Registration Number: 3006098230 Operations: Manufacturer Status: Active
1.2 Contact
SICAT GmbH & Co. KG Brunnenallee 6 D-53177 Bonn Germany
Telephone: +49-228/854 697 84 Facsimile: +49-228/854 697 99
Primary Contact: Mr. Markus Pfister . Secondary Contact: Mr. Dr. Manfred Breuer
2 Device Name and Classification
Proprietary Name: GALILEOS Implant Common/Usual Name: Preoperative Dental Implant Planning Software Classification Name: System, Image Processing, Radiological Regulation Description: Picture archiving and communications system
Product Code: Regulation Number; Classification Class:
LLZ 892.2050 Class II Product
{1}------------------------------------------------
3 Predicate Device
GALILEOS Implant V 1.7 is claimed to be substantially equivalent in material, design and function to the SimPlant 12 product which was cleared by FDA under 510(k) K033849 (May 25, 2004), K053592 (Feb 8, 2006), K081402 (JUL 18, 2008) and the GALILEOS Implant V1,0 product which was cleared by FDA under 510(k) K061472 on June 9, 2006,
4 Device Description
GALILEOS Implant V1.7 is a pure software device.
GALILEOS Implant V 1.7 is a planning and simulation software to aid qualified dental professionals in the placement of dental implants and the planning of surgical implant treatments. GALILEOS Implant is based on medical imaging information presented by the Sirona GALAXIS 3D viewer and produced by Sirona GALILEOS medical cone beam scanner.
GALILEOS Implant V1.7 allows to name, position, move, rotate, resize and visualize dental implants and other planning objects (i.e. nerve canals) within the visualized 3D volume. Thus, dental professionals like implantologists are enabled to precisely plan the positions, orientations, types and sizes of Implants to be placed in the patient's mandible/maxilla together with the related surgical procedures.
The dental professionals' planning data may be exported from GALILEOS Implant V1.7 and used as input data for CAD (Computer-Aided Design) or Rapid Prototyping Systems.
5 Intended Use
GALILEOS Implant V1.7 is intended for use as planning and simulation software to aid qualified dental professionals in the placement of dental implants and the planning of surgical implant treatments. GALLEOS Implant is based on medical imaging information presented by the Sirona GALAXIS 3D viewer and produced by Sirona GALILEOS medical cone beam scanner. The dental professionals' input information may be exported from GALILEOS Implant and used as input data for CAD (Computer-Aided Design) or Rapid Prototyping Systems.
ବ Substantial Equivalence
The GALILEOS Implant V 1.7 system is substantially equivalent to the SimPlant 12 System (K033849, K053592, K081402) and the GALILEOS Implant V1.0 System (K061472) based on the equivalence of the intended use, similar features and technical characteristics. Performance festing to validate the safety and effectiveness of the GALLEOS Implant V 1.7 system included validation testing and bench tests of the software functions.
7 Conclusion
GALLEOS Implant V1.7 is considered to be substantially equivalent in design, material and function to the SimPlant 12 System and the GALILEOS Implant V1.0 System.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
SICAT GmbH & Co. KG % Mr. Stefan Preiss Responsible Third Party Official TÜV SÜD America 1775 Old Hwy 8 NW, Ste 104 NEW BRIGHTON MN 25112-1891
Re: K093090
Trade/Device Name: GALILEOS Implant V1.7 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 1, 2009 Received: December 4, 2009
Dear Mr. Preiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
DEC 1 7 2009
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 -
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Singerely yours
Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
l INDICATIONS FOR USE STATEMENT
for GALILEOS Implant V1.7
510(k) Number (if known):
KO93090
Device Name: GALILEO$ Implant V1.7
Indications for Use:
GALILEOS Implant is intended for use as planning and simulation software to aid qualified dental professionals in the placement of dental implants and the planning of surgical implant treatments. GALILEOS Implant is based on medical imaging information presented by the Sirona GALAXIS 3D viewer and produced by Sirona GALILEOS medical cone beam scanner. The dental professionals' input information may be exported from GALILEOS Implant and used as input data for CAD (Computer-Aided Design) or Rapid Prototyping , Systems.
Prescription Use X (Part 21 CFR 801 Subpart D) ાર
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K093090
N/A