K Number
K092953

Validate with FDA (Live)

Device Name
Xpert VanA Assay
Manufacturer
Date Cleared
2009-12-17

(84 days)

Product Code
Regulation Number
866.1640
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cepheid Xpert® vanA Assay performed in the GeneXpert® Dx System is a qualitative in vitro diagnostic test designed for rapid detection of the vanA gene sequence associated with vancomycin resistance in bacteria obtained from rectal swab specimens from patients at risk for intestinal colonization with vancomycin-resistant bacteria. The test utilizes automated real-time polymerase chain reaction (PCR) to detect the vanA gene that is frequently associated with vancomycin-resistant enterococci (VRE). The Xpert vanA Assay is intended to aid in the recognition, prevention, and control of vancomycin-resistant organisms that colonize patients in healthcare settings. The Xpert vanA Assay is not intended to diagnose infections caused by vancomycin-resistant bacteria nor to guide or monitor treatment for vancomycin-resistant bacterial infections. Concomitant cultures are necessary to recover organisms for identification of vancomycin-resistant bacteria, antimicrobial susceptibility testing and for epidemiological typing.

Device Description

The Cepheid Xpert vanA Assay is a rapid, automated in vitro diagnostic test for qualitative detection of the vanA gene sequence associated with vancomycin resistance in bacteria obtained directly from rectal swab specimens. The Xpert vanA Assay system performs real-time multiplex polymerase chain reaction (PCR) for detection of DNA after an initial sample processing step. The assay is performed on the Cepheid GeneXpert® Dx System. The specimen is collected on a double swab, one of which is placed in a tube containing elution reagent. Following brief vortexing, the eluted material and two single-use reagents (Reagent 1 and Reagent 2) that are provided with the assay are transferred to different, uniquely-labeled chambers of the disposable fluidic cartridge (the Xpert vanA cartridge). The user initiates a test from the system user interface and places the cartridge into the GeneXpert® Dx System instrument platform, which performs hands-off realtime, multiplex polymerase chain reaction (PCR) for detection of DNA. In this platform, additional sample preparation, amplification, and real-time detection are all fullyautomated and completely integrated. The GeneXpert® System consists of a GeneXpert instrument, personal computer, and the multi-chambered fluidic cartridges that are designed to complete sample preparation and real-time PCR for detection of the vanA gene that is associated with vancomycin-resistant enterococci (VRE) in less than 45 minutes. Each instrument system has 1 to 16 randomly accessible modules that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection. The Xpert vanA Assay includes reagents for the detection of the vanA resistant gene as well as an internal sample processing control (SPC) to control for adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR assay. The SPC also ensures the PCR conditions (temperature and time) are appropriate for the amplification reaction and that the PCR reagents are functional. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Xpert vanA Assay:

Acceptance Criteria and Device Performance for Xpert vanA Assay

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state pre-defined acceptance criteria in terms of numerical thresholds for sensitivity, specificity, or agreement. Instead, it demonstrates "substantial equivalence" to predicate devices (a PCR assay and a culture method) and the "current standard of care." The reported performance metrics are presented from clinical comparison studies.

For the purpose of this response, I will interpret the achieved performance relative to the reference standard as the demonstrated effectiveness that led to acceptance.

MetricAcceptance Criteria (Implicit from Predicate Comparison)Reported Device Performance (vs. Direct Culture + Sequencing)Reported Device Performance (vs. Enriched Culture + Sequencing)
Percent Positive Agreement (Sensitivity)Substantially equivalent to predicate devices / current standard of care98.4%86.5%
Percent Negative Agreement (Specificity)Substantially equivalent to predicate devices / current standard of care92.4%93.5%
AccuracySubstantially equivalent to predicate devices / current standard of care93.0%92.6%
Positive Predictive Value (PPV)Substantially equivalent to predicate devices / current standard of care60.0%67.1%
Negative Predictive Value (NPV)Substantially equivalent to predicate devices / current standard of care99.8%97.8%
Overall Assay Success RateHigh success rate98.1% (combining first and second attempts)98.1% (combining first and second attempts)

Note: The document emphasizes that the device is "as safe, as effective, and performs as well as the current standard of care," which implies the stated performance metrics meet the (unspecified) acceptance levels for substantial equivalence.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: A total of 1231 specimens were tested in the clinical comparison study.
  • Data Provenance: The study was a multi-site prospective investigation study conducted at three US institutions.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

The document does not specify the "number of experts" used in the typical sense of radiologists or other clinicians reviewing images. Instead, the ground truth was established through a multi-step laboratory process:

  • Reference Culture: Performed by a central laboratory.
  • Bi-directional Sequencing Confirmation: Performed on vancomycin-resistant E. faecalis or E. faecium isolates, sent to a second reference laboratory using alternative vanA specific primers.
  • Identification and Susceptibility Testing: Presumptive VRE was definitively identified using the API20S strip (BioMérieux, France) and VRE isolates were tested for susceptibility to glycopeptides using vancomycin E-test strips and agar dilution for teicoplanin.

While specific "expert qualifications" (e.g., years of experience for individual technicians/scientists) are not detailed, the process involves established laboratory protocols and confirmation by multiple methods and reference laboratories, implying a high level of expertise in molecular biology and microbiology.

4. Adjudication Method for the Test Set

The adjudication method was a multi-step laboratory confirmation process, rather than a typical clinical adjudication by multiple human reviewers. The ground truth was established by:

  • Reference Culture: Starting with bile esculin azide broth with vancomycin, followed by subculture to agar.
  • Presumptive Identification: Gram stain, catalase, and pyr test.
  • Definitive Identification: API20S strip.
  • Susceptibility Testing: Vancomycin E-test strips and agar dilution for teicoplanin.
  • Final Confirmation: Bi-directional sequencing of vanA-positive isolates using independent primers.

This layered approach serves as the "adjudication" to establish the most accurate ground truth for VRE colonization with the vanA gene.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No, a MRMC comparative effectiveness study was not done.

This device is an in vitro diagnostic (IVD) PCR assay, which produces an objective "Positive" or "Negative" result for the presence of the vanA gene. It does not involve human readers interpreting complex data like medical images, so a study comparing human reader performance with and without AI assistance is not applicable.

6. Standalone Performance Study

Yes, a standalone performance study was done.

The clinical comparison study directly evaluated the Xpert vanA Assay's performance (algorithm only, without human-in-the-loop adjustments to the result) against the established reference culture and bi-directional sequencing gold standard. The reported Percent Positive Agreement, Percent Negative Agreement, Accuracy, PPV, and NPV are all measures of the standalone performance of the device.

7. Type of Ground Truth Used

The ground truth used was a combination of expert consensus laboratory methods:

  • Reference culture: Phenotypic detection of vancomycin-resistant enterococci (VRE) growth on selective media.
  • Definitive identification and susceptibility testing: Using established laboratory methods (API20S, E-test strips, agar dilution).
  • Bi-directional sequencing confirmation: Genetic confirmation of the vanA gene using independent primers.

This provides a robust, multi-faceted ground truth derived from both phenotypic and genotypic characteristics.

8. Sample Size for the Training Set

The document does not specify a sample size for a training set. This is common for IVD assays based on established molecular biology principles (PCR), where the assay design is driven by known genetic targets rather than machine learning on large, annotated datasets requiring separate training and test sets.

The "non-clinical studies" (analytical reactivity, sensitivity, linearity, specificity, interfering substances) describe laboratory-based testing of known strains and samples to characterize the assay's fundamental performance. These experiments might involve optimizing assay parameters, but not in the context of a "training set" for a machine learning algorithm.

For example:

  • Analytical Reactivity: Tested 30 vancomycin-resistant enterococci strains and 20 vancomycin-sensitive enterococci strains.
  • Analytical Sensitivity (LoD): Used 4 to 10 replicates for estimation and 20 replicates for confirmation.
  • Linearity: Used replicates of 4 at each of 6 serial dilutions.
  • Analytical Specificity: Tested 42 bacterial and fungal strains.

These are not "training sets" in the AI sense, but rather validation sets for internal development and characterization.

9. How the Ground Truth for the Training Set Was Established

As there is no explicit training set described in the context of a machine learning algorithm, the question of how its ground truth was established is not directly applicable.

The "ground truth" for the analytical studies (e.g., whether a strain is vanA positive or negative, or its CFU/mL concentration) was established by standard microbiological and genetic methods (e.g., strain identification, sequencing, quantitative culturing methods).

{0}------------------------------------------------

K092953

5.0 510(k) Summary

As required by 21 CFR Section 807.92(c).

Submitted by:Cepheid
904 Caribbean Drive
Sunnyvale, CA 90489
Phone number: (408) 400-8230
Fax number: (408) 541-6439
DEC 1 7 2009
----------------
Contact:Russel K. Enns, Ph.D.
Date of Preparation:September 23, 2009
Device:
Trade name:Xpert® vanA
Common names:Xpert® vanA Assay
Type of Test:Qualitative nucleic acid amplification test of the vanA gene directly from rectal swabs.
Classification:II
Classification name:System, test, genotypic detection, resistant marker, Enterococcus species
Regulation number:866.1640
Procode:NIJ
Classification Advisory Committee:Microbiology
Panel:83
Predicate Devices:BD IDI-VanR® Assay [510(k) #K061686]
Remel Esculin Azide Agar w/6 µg/mL vancomycin [510(k) #K972359]

Device Description:

The Cepheid Xpert vanA Assay is a rapid, automated in vitro diagnostic test for qualitative detection of the vanA gene sequence associated with vancomycin resistance in bacteria obtained directly from rectal swab specimens. The Xpert vanA Assay system performs real-time multiplex polymerase chain reaction (PCR) for detection of DNA after an initial sample processing step. The assay is performed on the Cepheid GeneXpert® Dx System.

The specimen is collected on a double swab, one of which is placed in a tube containing elution reagent. Following brief vortexing, the eluted material and two single-use reagents (Reagent 1 and Reagent 2) that are provided with the assay are transferred to

{1}------------------------------------------------

different, uniquely-labeled chambers of the disposable fluidic cartridge (the Xpert vanA cartridge). The user initiates a test from the system user interface and places the cartridge into the GeneXpert® Dx System instrument platform, which performs hands-off realtime, multiplex polymerase chain reaction (PCR) for detection of DNA. In this platform, additional sample preparation, amplification, and real-time detection are all fullyautomated and completely integrated.

The GeneXpert® System consists of a GeneXpert instrument, personal computer, and the multi-chambered fluidic cartridges that are designed to complete sample preparation and real-time PCR for detection of the vanA gene that is associated with vancomycin-resistant enterococci (VRE) in.less than 45 minutes. Each instrument system has 1 to 16 randomly accessible modules that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids, an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

The Xpert vanA Assay includes reagents for the detection of the vanA resistant gene as well as an internal sample processing control (SPC) to control for adequate processing of the target bacteria and to monitor the presence of inhibitor(s) in the PCR assay. The SPC also ensures the PCR conditions (temperature and time) are appropriate for the amplification reaction and that the PCR reagents are functional. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dye stability.

Device Intended Use:

The Cepheid Xpert® vanA Assay performed in the GeneXpert® Dx System is a qualitative in vitro diagnostic test designed for rapid detection of the vanA gene sequence associated with yancomvcin resistance in bacteria obtained from rectal swab specimens from patients at risk for intestinal colonization with vancomycin-resistant bacteria. The test utilizes automated real-time polymerase chain reaction (PCR) to detect the vanA gene that is frequently associated with vancomvcin-resistant enterococci (VRE). The Xpert vanA Assay is intended to aid in the recognition, prevention, and control of vancomycinresistant organisms that colonize patients in healthcare settings. The Xpert vanA Assay is not intended to diagnose infections caused by vancomycin-resistant bacteria nor to guide or monitor treatment for vancomycin-resistant bacterial infections. Concomitant cultures are necessary to recover organisms for identification of vancomvcin-resistant bacteria, antimicrobial susceptibility testing and for epidemiological typing.

Substantial Equivalence:

The Xpert vanA Assay is substantially equivalent to the BD IDI-VanR Assay (510(k) #K061686). Both assays detect vancomycin resistance genes directly from rectal swab specimens from patients at risk for VRE colonization. Both assays use real-time PCR amplification and fluorogenic target-specific hybridization detection.

Table 5.1 shows the similarities and differences between the Xpert vanA Assay and the BD IDI-VanR Assay.

{2}------------------------------------------------

The Xpert vanA is also substantially equivalent to the reference direct culture method. The reference culture method is the Remel Bile Esculin Azide agar with 6 ug/mL vancomycin (BEAV) [510(k) #K972359].

Performance characteristics of the Xpert vanA Assay were determined in a multi-site prospective investigation study at three US institutions by comparing the Xpert vanA Assay to reference culture followed by bi-directional sequencing confirmation on those samples positive for vanA by culture.

Table 5.2 compares the new device with the reference direct culture method. The test results showed the Xpert vanA Assay to be substantially equivalent to the current standard of care, the reference culture method.

{3}------------------------------------------------

:

ı

Similarities
DevicePredicate
ItemXpert vanA AssayIDI-VanR Assay (K061686)
Intended UseThe Cepheid Xpert® vanAAssay performed in theGeneXpert® Dx System is aqualitative in vitro diagnostictest designed for rapiddetection of the vanA genesequence associated withvancomycin resistance inbacteria obtained from rectalswab specimens from patientsat risk for intestinalcolonization withvancomycin-resistant bacteria.The test utilizes automatedreal-time polymerase chainreaction (PCR) to detect thevanA gene that is frequentlyassociated with vancomycin-resistant enterococci (VRE).The Xpert vanA Assay isintended to aid in therecognition, prevention, andcontrol of vancomycin-resistant organisms thatcolonize patients in healthcaresettings. The Xpert vanAAssay is not intended todiagnose infections caused byvancomycin-resistant bacterianor to guide or monitortreatment for vancomycin-resistant bacterial infections.Concomitant cultures arenecessary only to recoverorganisms for epidemiologicaltyping, antimicrobialsusceptibility testing, and forfurther confirmatoryidentification of vancomycin-resistant bacteria.The IDI-VanR® Assay is aqualitative in vitro test for therapid detection ofvancomycin-resistanceenterococci (VRE). The assayis performed on an automatedreal-time PCR instrument withrectal swabs from patients atrisk for VRE colonization.The IDI-VanR® Assay can beused as an aid to identify,prevent and controlvancomycin-resistantcolonization in healthcaresettings. Concomitant culturesare necessary to recoverorganisms for epidemiologicaltyping, susceptibility testingand for further confirmatoryidentification. The IDI-VanR® Assay is not intendedto diagnose VRE infectionsnor to guide or monitortreatment for VRE infections.
Type of testQualitativeSame
Similarities
Technological PrinciplesFully-automated nucleic acid amplification (DNA); real-time PCRSame
Specimen TypeDirect from rectal SwabsSame
Test CartridgeDisposable single-use, multi-chambered fluidic cartridge.Disposable single-use PCR tube
ProbesTaqMan® ProbesMolecular Beacons
ControlsInternal sample processing control (SPC) and probe check control (PCC).External controls available.One internal reagent control and external positive and negative controls required per run
DNA Target SequenceDetects gene sequences for the vanA encoded resistance to vancomycin/ teicoplanin.Detects gene sequences for VanR (vanA and vanB) encoded resistance to vancomycin/teicoplanin.
Rapid test resultsLess than 45 minutes to results.Approximately 120 minutes.
Interpretation of test resultsDiagnostic software of the Cepheid GeneXpert Dx SystemDiagnostic software of the Cepheid SmartCycler Dx System
Differences
DevicePredicate
ItemXpert vanA AssayIDI-VanR Assay (K061686)
Instrument SystemCepheid GeneXpert Dx SystemCepheid SmartCycler
DNA Target SequenceDetects sequences for the vanA gene.Detects sequences for vancomycin resistance [vanR (vanA and vanB)] gene, but does not differentiate vanA from vanB.
Sample Extraction/FluidicsSelf-contained and automated after swab elution and two single-dose reagent additions.Manual
UsersOperators with no clinical lab experience to experienced clinicalCLIA High Complexity Laboratory Users
Similarities
ItemDeviceXpert vanA AssayCulture Method PredicateRemel Bile Esculin Azideagar with 6 µg/mLvancomycin (BEAV)[510(k) #K972359]
Intended UseThe Cepheid Xpert® vanA Assayperformed in the GeneXpert® DxSystem is a qualitative in vitro diagnostictest designed for rapid detection of thevanA gene sequence associated withvancomycin resistance in bacteriaobtained from rectal swab specimensfrom patients at risk for intestinalcolonization with vancomycin-resistantbacteria. The test utilizes automatedreal-time polymerase chain reaction(PCR) to detect the vanA gene that isfrequently associated with vancomycin-resistant enterococci (VRE). The XpertvanA Assay is intended to aid in therecognition, prevention, and control ofvancomycin-resistant organisms thatcolonize patients in healthcare settings.The Xpert vanA Assay is not intended todiagnose infections caused byvancomycin-resistant bacteria nor toguide or monitor treatment forvancomycin-resistant bacterialinfections. Concomitant cultures arenecessary only to recover organisms forepidemiological typing, antimicrobialsusceptibility testing, and for furtherconfirmatory identification ofvancomycin-resistant bacteria.Remel Bile Esculin Azide agarwith 6 µg/mL vancomycin is aplated medium recommendedfor use in qualitativeprocedures as a selective anddifferential medium for theprimary isolation ofvancomycin resistantenterococci from surveillancecultures. This product is notintended for use as a methodof antimicrobial susceptibilitytesting. Confirmation ofvancomycin resistance by anapproved method isrecommended as someorganisms on initial isolationmay overcome the inhibitoryeffects of the medium.
Single useYesSame
Type of testQualitativeSame
Specimen TypeDirect from rectal swabsDirect from rectal swab orstool.

Table 5.1: Similarities and Differences between the Xpert vanA Assay and the IDI-VanR Assay

{4}------------------------------------------------

.

.

{5}------------------------------------------------

CONFIDENTIAL

Table 5.2: Similarities and Differences between the Xpert vanA Assay and the Reference Culture Method Predicate Device

{6}------------------------------------------------

Differences
ItemXpert vanA AssayCulture Method Predicate
TechnologyFully-automated nucleic acid (DNA) preparation and amplification; real-time PCR.Detect sequences specific to vanA genePhenotypic detection of vancomycin-resistant enterococci (VRE) based on culture growth
Mode of DetectionPresence of vanA geneGrowth or no growth on6mg/mL vancomycin agar
Specimen TypeDirect from rectal swabsCulture grown direct fromrectal swab or stool
Internal ControlsSample processing control (SPC) and probe check control (PCC).Not applicable
Interpretation of test resultsDiagnostic software of the GeneXpert Dx SystemVisual interpretation
Remel Bile Esculin Azide agar with 6 µg/mL vancomycin (BEAV) [510(k) #K972359]

Non-Clinical Studies:

Analytical Reactivity (Inclusivity)

Thirty vancomvcin-resistant enterococci strains and 20 vancomycin sensitive enterococci strains, provided by the CDC, were tested using the Xpert vanA Assay. Of the 30 vancomvcin-resistant enterococci strains. 10 were identified as vanA positive. Enterococci strains were selected to broadly represent the genetic diversity found in enterococci. Stock cultures were prepared by suspending the bacterial growth from agar plates in PBS buffer containing 15% glycerol. The concentration of each stock was adjusted to 5.6x10° to 2.1x10'0 CFU/mL. All strains were serially diluted to approximately 360 CFU/swab and tested in triplicate.

Under the conditions of this study, all 20 vancomycin sensitive strains were reported as "vanA NEGATIVE," as expected. Among the 10 vanA positive vancomycin-resistant enterococci strains tested, one strain was reported as "vanA NEGATIVE." When this strain was sequenced the data matched 100% to a reference vanB sequence, confirming that the Xpert vanA Assay accurately reported the strain "vanA NEGATIVE." The remaining 9 vanA positive vancomycin resistant enterococci strains were correctly reported as "vanA POSITIVE." as expected. Among the 20 non-vanA vancomycin resistant enterococci strains, all were reported as "vanA NEGATIVE," as expected.

{7}------------------------------------------------

Analytical Sensitivity

Studies were performed to determine the 95% confidence intervals for the analytical limit of detection (LoD) of Enterococcus facium (vanA) diluted into a fecal matrix of human origin that can be detected by the Xpert vanA Assay. The fecal matrix consisted of autoclaved human liguid feces (vanA negative) diluted 1:10 in Tris buffer. The LoD is defined as the lowest number of colony forming units (CFU) per swab that can be reproducibly distinguished from negative samples with 95% confidence.

The analytical LoD was estimated using 4 to 10 replicates at each dilution. The LoD was confirmed by running a total of 20 replicates at the estimated LoD concentration. Under the conditions of this study, the limit of detection for the Xpert vanA Assay on a simulated rectal swab specimen is 37 CFU.

Linearity

A study was conducted to define the reportable range of the Xpert vanA Assay and demonstrate a linear between target input and assay output. Linearity was evaluated using Enterococcus faecium (vanA) cells serially diluted over 6 logs and processed using the Xpert vanA Assay. The diluted cells resulted in a cell concentration dose range of 50 CFU/test to 5x10' CFU/test. Replicates of four (4) were tested at each concentration.

For enterococci cells, under the conditions of this study, the Xpert vanA Assay responds linearly (r2 = 0.994) with respect to vanA detection as a function of Enterococcus faecium cell input over 6 logs (50 - 5x10' CFU/test). The reportable Ct range is 12.1 to 35.3 (cutoff Ct = 40.0). PCR efficiency for the vanA reaction is 87.7 %.

Analytical Specificity

Forty-two bacterial and fungal strains were collected, quantitated and tested using the Xpert vanA Assay. The strains originated from the American Type Culture Collection (ATCC), Culture Collection University of Göteborg (CCUG), German Collection of Microorganisms and Cell Cultures (DSMZ), and the Centers for Disease Control and Prevention (CDC).

The organisms tested were identified as Gram-positive (22), Gram-negative (18), including antibiotic-resistant strains of Pseudomonas spp. and Acinetobacter spp., and yeast (2). The organisms were further classified as aerobic (24), anaerobic (14) or microaerophillic (2). Of the species tested, 2 vancomycin-sensitive strains of E. faecalis and E. faecium were included.

Each strain was tested in triplicate at concentrations ranging from 8.5x108 to 2.3x1010 CFU/swab. Positive and negative controls were included in the study. Under the conditions of the study, all isolates were reported "vanA NEGATIVE". The analytical specificity was 100%.

{8}------------------------------------------------

Interfering Substances

Sixteen exogenous substances occasionally used or found in stool were tested for interference with the Xpert vanA Assay. The substances tested are listed in Table 5.3. None of the 16 substances tested showed detectable interference for vanA. However, Hydrocortisone cream (1 % Hydrocortisone) and Pepto-Bismol® (1 - 5% Bismuth subsalicylate) may slightly interfere with the Xpert vanA Assay. When tested in the Interference study, Hydrocortisone cream and Pepto-Bismol® resulted in slightly higher Ct values relative to the buffer control.

SubstanceSubstance
Whole BloodVaseline
Karolinska University HospitalUnilever
Mucin (porcine)Dulcolax®
SigmaBoehringer Ingelheim Pharmaceuticals
Kaopectate®Preparation H® Portable Wipes
ChattemWyeth Consumer Healthcare
Imodium®Vancomycin
McNeil-PPCFluka
Fleet®Metronidazole
CB Fleet CompanyActavis
Fecal fatsAnusol® Plus
Karolinska University HospitalTM Warner-Lambert Company
K-Y Jelly/Gelée®E-Z-HDTM High Density Barium Sulfate for suspension
McNeil-PPCE-Z-EM Canada
AHydrocortisone CreamAPepto-Bismol®
Longs DrugsProctor & Gamble
Table 5.3: Substances Tested and Showing No Assay Interference for vanA
---------------------------------------------------------------------------

4 When tested in the Interference study, results showed slightly higher Ct values relative to the buffer r control

Clinical Studies

Clinical Comparison Study

Performance characteristics of the Xpert vanA Assay were determined in a multi-site prospective investigation study at three US institutions by comparing the Xpert vanA Assay to reference culture followed by bi-directional sequencing for confirmation on vancomycin-resistant E. faecalis or E. faecium isolates.

Subjects included individuals whose routine care called for VRE testing. One swab from a double swab set was used for patient management; the other swab was used for the Xpert vanA Assay testing. The leftover swab designated for patient management was sent to a central laboratory for reference culture.

Leftover specimen swabs designated for culture testing were stored at 2-8°C and shipped on ice packs to the central culture laboratory within 48 hours of collection. Reference culture was initiated within 16 hours of receipt or within 5 days of swab collection.

{9}------------------------------------------------

Each swab was subsequently placed into bile esculin azide broth with 8 µg/ml vancomycin. The plates were incubated at 35°C and examined at 48 and 72 hours. The broth was also incubated at 35℃ for 48 hours and subcultured to a bile esculin azide agar with 6 ug/ml of vancomycin.

Small, gray colonies with a black halo were considered suspicious for VRE. Presumptive identification was accomplished by performing a Gram stain, catalase and disc pyr (Lpyrrolidonyl-beta-naphthylamide) test. Presumptive VRE specimens were Gram-positive cocci or coccobacilli and pyr positive. Presumptive VRE was definitively identified using the API20S strip (BioMérieux, France). Finally, VRE isolates were tested for their susceptibility to glycopeptides using vancomycin &-test strips (AB Biodisk, Sweden). Susceptibility to teicoplanin for the isolates was determined by agar dilution. Following reference culture testing. DNA was prepared from vancomycin-resistant E. faecilis or E. faecium isolates, and sent to a second reference laboratory for bi-directional sequencing using alternative vanA specific primers (i.e., different from those used in the Xpert vanA Assay).

Performance of the Xpert vanA Assay was calculated relative to the results of direct culture with bi-directional sequencing, and enriched culture with bi-directional sequencing.

Overall Results

A total of 1231 specimens were tested by Xpert vanA Assay, culture and bi-directional sequencing.

Performance vs. Direct Culture

Relative to direct culture with bi-directional sequencing, the Xpert vanA Assay

demonstrated a percent positive agreement of 98.4% and a percent negative agreement of

92.4% (Table 5.4).

{10}------------------------------------------------

CONFIDENTIAL

Direct Culture + Sequencing
PosNegTotal
Xpert vanAAssayPos12684210
Neg210191021
Total12811031231
% Positive Agreement: 98.4%
% Negative Agreement: 92.4%
Accuracy: 93.0%
PPV: 60.0%
NPV: 99.8%
Prevalence: 10.4%
Table 5.4: Xpert vanA Assay Performance vs.
Direct Culture with Bi-directional Sequencing

Of the Xpert vanA Assays run on eligible specimens, 94.0% (1180/1255) of these specimens were successful on the first attempt. The remaining 75 gave indeterminate results on the first attempt (26 "INVALID", 49 "ERROR" and 0 "NO RESULT"). Sixty two (62) of the75 indeterminates on the first attempt had sufficient sample for retest, 82.3% (51/62) gave a result on the second the attempt. Overall assay success rate (combining the first and second attempts) was 98.1% (1231/1255).

Performance vs. Enriched Culture

Relative to enriched culture with bi-directional sequencing, the Xpert vanA Assay demonstrated a percent positive agreement of 86.5% and a percent negative agreement of 93.5% (Table 5.5).

{11}------------------------------------------------

Enriched Culture + Sequencing
PosNegTotal
Xpert vanAAssayPos141 (137)69 (69)210 (206)
Neg22 (21)999 (953)1021 (974)
Total163 (158)1068 (1022)1231 (1180)
% Positive Agreement:86.5%
% Negative Agreement:93.5%
Accuracy:92.6%
PPV:67.1%
NPV:97.8%
Prevalence:13.2%

Table 5.5: Xpert vanA Assay Performance vs. Enriched Culture with Bi-directional Sequencing

Of the Xpert vanA Assays run on eligible specimens, 94.0% (1180/1255) of these specimens were successful on the first attempt. The remaining 75 gave indeterminate results on the first attempt (26 "INVALID", 49 "ERROR" and 0 "NO RESULT"). Sixty two (62) of the75 indeterminates on the first attempt had sufficient sample for retest, 82.3% (51/62) gave a result on the second the attempt. Overall assay success rate (combining the first and second attempts) was 98.1% (1231/1255)

Antibiotic Usage

Among the 1231 cases included in the main dataset, antibiotic use within the 3 weeks prior to sample collection was reported for 414 and no antibiotic use was confirmed for 483; for 334 cases, antibiotic status was unknown. Antibiotic use did not cause a statistically significant difference in assay performance.

Reproducibility

A panel of four specimens with varying concentrations of vanA was tested on 10 different days by two different operators at each of the three sites (4 specimens x 2 operators/ day x 10 days x 3 sites). One lot of Xpert vanA Assay was used at each of the 3 testing sites. Xpert vanA Assays were performed according to the Xpert vanA Assay procedure. Results are summarized in Tables 5.6.

{12}------------------------------------------------

CONFIDENTIAL

% Agreementa
Specimen IDSite 1Site 2Site 3% TotalAgreement bySample
Neg100%(20/20)90%(18/20)100%(20/20)96.7%(58/60)
vanA High Neg100%(20/20)100%(20/20)95%(19/20)98.3%(59/60)
vanA Low Pos100%(20/20)100%(20/20)100%(20/20)100%(60/60)
vanA Moderate Pos100%(20/20)95%(19/20)100%(20/20)98.3%(59/60)
% Total Agreement by Site100%(80/80)96.3%(77/80)98.8%(79/80)98.3%(236/240)

Table 5.6: Summary of Reproducibility Results (all)4

"For negative and high negative samples, %Agreement = (# negative results/total samples run); for low and moderate positive samples, %Agreement = (# positive results/total samples run).

Conclusions

The results of the nonclinical analytical and clinical performance studies summarized above demonstrate that the Xpert vanA Assay is as safe, as effective, and performs as well as the current standard of care, the reference culture method with confirmation of all results by bi-directional sequence analysis. Therefore, the Xpert vanA Assay is substantially equivalent to the predicate device.

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. Inside the circle is a stylized image of what appears to be an abstract symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-0609 Silver Spring, MD 20993-0002

OEC 1 7 2000

Cepheid® c/o Russel K. Enns. Ph.D. Senior Vice President Regulatory, Clinical & Government Affairs & Quality Systems 904 Carribean Drive Sunnyvale, CA 94089

Re: K092953

Trade/Device Name: Xpert® vanA Assay Regulation Number: 21 CFR §866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: NIJ, OOI Dated: September 23, 2009 Received: September-24,-2009.

Dear Dr. Enns:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other

{14}------------------------------------------------

Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Sayatryns

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the word "Cepheid" in a stylized font, with three curved lines above it. Below the word, there is the text "4.0 Indications for Use Statement". The text is in a smaller font than the word "Cepheid".

510(k) Number (if known): K092953

Device Name: Xpert vanA

Indications for Use:

The Cepheid Xpert® vanA Assay performed in the GeneXpert® Dx System is a qualitative in vitro diagnostic test designed for rapid detection of the vanA gene sequence associated with vancomycin resistance in bacteria obtained from rectal swab specimens from patients at risk for intestinal colonization with vancomycin-resistant bacteria. The test utilizes automated real-time polymerase chain reaction (PCR) to detect the vanA gene that is frequently associated with vancomycin-resistant enterococci (VRE). The Xpert vanA Assay is intended to aid in the recognition, prevention, and control of vancomycinresistant organisms that colonize patients in healthcare settings. The Xpert vanA Assay is not intended to diagnose infections caused by vancomycin-resistant bacteria nor to guide or monitor treatment for vancomycin-resistant bacterial infections. Concomitant cultures are necessary to recover organisms for identification of vancomycin-resistant bacteria, antimicrobial susceptibility testing and for epidemiological typing

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Lueddeke. Cook

ivision Sign-Off

Office of In Vitro Diagnostic Device Evaluation and Safety

510(K) K092953

Page 1

§ 866.1640 Antimicrobial susceptibility test powder.

(a)
Identification. An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.(b)
Classification. Class II (performance standards).