(219 days)
The Epicage Interbody Fusion Device is intended for interbody fusion procedures and is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device is intended to be used with supplemental posterior fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone.
The Epicage Interbody Fusion Device is a PEEK implant that is designed to provide mechanical support to the lumbar spine while biologic fusion takes place. The device is designed with openings for bone growth and fusion. The rib allow for flexure of the device during insertion. The device is available in two lengths to accommodate various patients' anatomy. Each of these sizes is available in 4 heights ranging from 8mm to 14mm in 2mm increments.
The provided text describes pre-clinical performance testing for the Epicage Interbody Fusion Device, which is intended for interbody fusion procedures. This device is a PEEK implant designed to provide mechanical support to the lumbar spine during biologic fusion.
Here's an analysis of the acceptance criteria and study to prove the device meets those criteria:
1. Table of Acceptance Criteria and Reported Device Performance
The text indicates that the device's performance was evaluated against established ASTM standards and a recognized protocol for expulsion testing. The reported performance is a qualitative statement of meeting these standards and being "substantially equivalent" to predicate devices. Specific quantitative acceptance criteria or performance metrics are not explicitly provided in the extract.
| Acceptance Criteria (Implicit from testing standards) | Reported Device Performance |
|---|---|
| Compliance with ASTM F2077-03 for Static Axial Compression | Tested, indicated substantial equivalence to predicate devices and adequate for intended use. |
| Compliance with ASTM F2077-03 for Static Compression Shear | Tested, indicated substantial equivalence to predicate devices and adequate for intended use. |
| Compliance with ASTM F2077-03 for Static Torsion | Tested, indicated substantial equivalence to predicate devices and adequate for intended use. |
| Compliance with ASTM F2077-03 for Dynamic Axial Compression | Tested, indicated substantial equivalence to predicate devices and adequate for intended use. |
| Compliance with ASTM F2077-03 for Dynamic Torsion | Tested, indicated substantial equivalence to predicate devices and adequate for intended use. |
| Compliance with ASTM F2267-04 for Subsidence | Tested, indicated substantial equivalence to predicate devices and adequate for intended use. |
| Resistance to Expulsion (following recognized protocol) | Tested, indicated substantial equivalence to predicate devices and adequate for intended use. |
2. Sample Size and Data Provenance
The provided text focuses on pre-clinical bench testing. It does not describe a clinical study involving human patients. Therefore, there is no "test set" in the context of patient data nor data provenance from specific countries or retrospective/prospective collection methods. The "samples" would refer to the physical Epicage Interbody Fusion Device units used for mechanical testing. The specific number of devices tested for each type of performance test is not mentioned.
3. Number of Experts and Qualifications
This information is not applicable as the study described is pre-clinical bench testing of a mechanical device, not an evaluation requiring expert interpretation of medical images or patient outcomes.
4. Adjudication Method
This information is not applicable for the same reason as point 3. Bench testing results are typically based on objective measurements from laboratory equipment.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The provided text describes pre-clinical bench testing of a mechanical device. It does not involve human readers interpreting images or considering AI assistance.
6. Standalone Performance Study (Algorithm Only)
The device described is a physical medical implant (interbody fusion device), not a software algorithm. Therefore, "standalone" algorithm performance is not applicable. The performance testing focuses on the mechanical integrity and behavior of the implant itself.
7. Type of Ground Truth Used
The ground truth for the performance testing is based on the objective measurements obtained from standardized mechanical tests (ASTM standards and a recognized expulsion protocol). For example, the "ground truth" for static axial compression would be the force/displacement curves and failure points measured according to the ASTM F2077-03 standard. The study aims to demonstrate that the device's mechanical properties meet the expectations for its intended use and are comparable to predicate devices.
8. Sample Size for the Training Set
There is no "training set" in the context of an algorithm or machine learning model, as this is a physical medical device.
9. How Ground Truth for the Training Set Was Established
This information is not applicable as there is no training set for a physical medical device.
{0}------------------------------------------------
4. 510(k) Summary according to 807.92(c)
APR 2 8 2010
| Contact: | Jay Finster |
|---|---|
| R Tree Innovations, LLC | |
| 5956 Timber Ridge Drive | |
| Suite 101 | |
| Prospect, KY 40059 | |
| 502-689-4483 | |
| Trade Name: | Epicage Interbody Fusion Device |
| Product Class: | Class II |
| Classification: | 21 CFR $888.3080 Orthosis, intervertebral fusion device |
| Product Codes: | MAX |
| Panel Code: | 87 |
Indications for Use:
The Epicage Interbody Fusion Device is intended for interbody fusion procedures and is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device is intended to be used with supplemental posterior fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone.
Device Description:
The Epicage Interbody Fusion Device is a PEEK implant that is designed to provide mechanical support to the lumbar spine while biologic fusion takes place. The device is designed with openings for bone growth and fusion. The rib allow for flexure of the device during insertion.
The device is available in two lengths to accommodate various patients' anatomy. Each of these sizes is available in 4 heights ranging from 8mm to 14mm in 2mm increments.
Predicate Device(s):
The Epicage Interbody Fusion Device was shown to be substantially equivalent to previously cleared devices and has the same indications for use, design, function, and materials used. The predicate devices include the BAK Interbody Fusion Device (Spine-Tech, P950002), the Ray Threaded Fusion Cage (P950019), the Brantigan Lumbar I/F Cage (P960025) and the ROI-T device from LDR (K082262)
{1}------------------------------------------------
Performance Testing:
Static axial compression, static compression shear, static torsion, dynamic axial compression and dynamic torsion were completed following ASTM F2077-03. Subsidence was tested following ASTM F2267-04. Expulsion testing was conducted following a recognized protocol to allow comparison evaluation of intervertebral body fusion device assemblies, and characterize their resistance to expulsion. The above pre-clinical testing performed on the Epicage Interbody Fusion Device indicated that the Epicage Interbody Fusion Device is substantially equivalent to the predicate devices and is adequate for the intended use.
Summary:
The Epicage Interbody Fusion Device and predicate devices have the same intended use, to provide mechanical stability in the lumbar disc space to facilitate biologic fusion. The indications for use of the Epicage Interbody Fusion Device contain no new language that is not already included in at least one of the predicate devices. Moreover, the device is very similar in its size to the predicate devices. The materials used are also the same as in some predicate devices. There are no significant differences in technological characteristics compared to the predicates, and the minor differences that do exist do not raise any new types of safety or efficacy issues. Furthermore, bench testing demonstrates that these differences do not adversely impact device performance, as discussed below.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling an emblem. The emblem is a stylized depiction of an eagle, with its wings spread, symbolizing protection and service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002
APR 2 8 2010
R Tree Innovations, LLC % Dr. Richard Jansen Silver Pine Consulting 13540 Guild Avenue Apple Valley, Minnesota 55124
Re: K092901
Trade/Device Name: Epicage Interbody Fusion Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: April 17, 2010 Received: April 20, 2010
Dear Dr. Jansen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{3}------------------------------------------------
Page 2 – Dr. Richard Jansen
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Barbace Buchud
Mark N. Melke Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
3. Statement of Indications for Use
510(k) Number (if known): _K092901
Indications for Use:
The Epicage Interbody Fusion Device is intended for interbody fusion procedures and is indicated for use in skeletally mature patients with Degenerative Disc Disease (DDD) at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. Patients should have received 6 months of non-operative treatment prior to treatment with the device is intended to be used with supplemental posterior fixation. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). It is indicated to be used with autograft bone.
Prescription Use V (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
h
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
K092901 510(k) Number_
Concurrence of CDRH, Office of Device Evaluation (ODE)
N/A