K Number
K092830

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-07-21

(310 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Digital X-ray Imaging System, DRC-1000, is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, extremities, and other body parts. Not to be used for fluoroscopy, angiography, or screening mammography. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

Device Description

DRC-1000 is a digital diagnostic x-ray system which consists of Solid State Xray Imager (Flat Panel), tube-collimator assembly mounted on a U-Arm, generator and operation control unit. DRC-1000 is designed for adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, extremities and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

AI/ML Overview

The provided 510(k) summary for the DRC-1000 Digital X-ray Imaging System does not contain information about specific acceptance criteria or a dedicated study proving the device meets particular clinical performance criteria.

This 510(k) submission focuses on demonstrating substantial equivalence to a predicate device (Sedecal X-Plus LP Plus Digital) primarily based on technical specifications and adherence to electrical, mechanical, environmental safety, and EMC standards.

Therefore, many of the requested details cannot be extracted from the provided text.

Here is a summary of what can be inferred or explicitly stated from the document:

1. Table of Acceptance Criteria and Reported Device Performance:

  • No specific clinical acceptance criteria (e.g., sensitivity, specificity, image quality metrics) are provided in the document.
  • The device performance is described in terms of its ability to meet safety and electrical standards rather than clinical diagnostic performance.
Acceptance Criterion (Inferred from document)Reported Device Performance (Inferred from document)
Electrical Safety (IEC 60601-1)All test results were satisfactory.
Mechanical SafetyAll test results were satisfactory.
Environmental SafetyAll test results were satisfactory.
Performance (IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32)All test results were satisfactory.
EMC (IEC 60601-1-2)All test results were satisfactory.
Substantial Equivalence to Predicate DeviceConcluded to be safe, effective, and substantially equivalent to predicate devices.

2. Sample size used for the test set and the data provenance:

  • Not specified. The document does not describe a clinical test set or data involving patient images for performance evaluation. The "test results" mentioned pertain to engineering and safety standards.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No clinical ground truth establishment is described.

4. Adjudication method for the test set:

  • Not applicable. No clinical test set or adjudication process is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This is a traditional X-ray system, not an AI-assisted diagnostic device. Therefore, no MRMC study or AI-related effectiveness is discussed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. This is a hardware device; "standalone algorithm performance" is not relevant here.

7. The type of ground truth used:

  • Not applicable for clinical performance. The "ground truth" for the reported tests would be adherence to the specifications of the mentioned IEC standards (e.g., electrical parameters, radiation output, mechanical stability).

8. The sample size for the training set:

  • Not applicable. This is not an AI/machine learning device that would require a "training set."

9. How the ground truth for the training set was established:

  • Not applicable. As above, no training set is relevant.

In summary: The provided 510(k) summary demonstrates substantial equivalence for the DRC-1000 by

  1. Comparing its general description and indications for use to a legally marketed predicate device.
  2. Stating successful completion of various electrical, mechanical, environmental safety, performance, and EMC standard tests (IEC 60601 series).

It does not include any information regarding clinical performance studies, such as those involving image analysis, diagnostic accuracy, or reader performance, which would typically involve specific acceptance criteria, test sets, expert readers, and ground truth establishment. This is common for traditional imaging hardware where regulatory clearance hinges on safety, basic performance, and equivalence to existing technology, rather than novel clinical diagnostic claims.

{0}------------------------------------------------

K092830

510(k) Submission -- DRC-1000

510(k) Summary

Date: July 27, 2009

    1. Submitter:
      Company Name Company Address

U.S. Agent Address:

Contact Person Telephone Fax

  1. Device :

Reason for 510(k) Manufacturer Model Name: Common Name Classification Name :New Model : Vatech Co., Ltd. : DRC-1000 : Digital X-ray imaging system : SYSTEM, X-RAY, STATIONARY

    1. Predicate Device :
Manufacturer: Sedecal, Inc.
Model Name: Sedecal X-Plus LP Plus Digital
Classification Name: SYSTEM, X-RAY, STATIONARY
510(k). Number: K090238 (Decision Date - Feb. 27. 2009)

: Vatech Co., Ltd.

Secaucus, NJ 07094

: (832)-623-2099

: (713)-464-8880

: Dave Kim

: 23-4, Seogu-Dong, Hwaseong-si,

Gyeonggi-do, 445-170, Korea

333 Meadowlands Parkway, #303

4 . Classifications Names & Citations :

21CFR 892.1680, KPR, System, X-ray, Stationary, Class2

Vatech Co., Ltd.

JUL 2-1 2010

{1}------------------------------------------------

  1. Description :

5.1 General

DRC-1000 is a digital diagnostic x-ray system which consists of Solid State Xray Imager (Flat Panel), tube-collimator assembly mounted on a U-Arm, generator and operation control unit. DRC-1000 is designed for adult and pediatric subjects for taking diagnostic exposures of the skull, spinal column, chest, extremities and other body parts. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

5.2 Product features

a. Condition of Input

1-1. Rated input voltage : AC 220V / AC 110V

1-2. Guaranteed working voltage

l) 110V Mode : 100 ~120V

  1. 220V Mode : 200~240V

1-3. Possible working voltage

  1. 110V Mode : 90~130V

  2. 220V Mode : 180 ~ 250V

1-4. Rated input frequency : 50Hz/60Hz

1-5. Insulation withstanding : below than 1.5KV cap for more than one minute between first test and second test.

b. Capture Mode

Whole body

Shooting Mode : CHEST/RIB/STERNUM/APEX/SELLA ..

  1. Indication for use :

Digital X-ray Imaging System, DRC-1000, is intended for use by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic

{2}------------------------------------------------

510(k) Submission --DRC-1000

exposures of the skull, spinal column, chest, extremities, and other body parts.

Not to be used for fluoroscopy, angiography, or screening mammography. Applications can

be performed with the patient sitting, standing, or lying in the prone or supine position.

  1. Comparison with predicate device :

Vatech Co., Ltd., believes that the DRC-1000 is substantially equivalent to the Sedecal X-Plus LP Plus Digital of Sedecal, Inc.

  1. Safety, EMC and Performance Data :

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28 and IEC 60601-2-32 was performed, and EMC testing was conducted in accordance with standard IEC 60601-1-2. All test results were satisfactory.

9. Conclusion :

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Vatech Co., Ltd. concludes that The DRC-1000 is safe and effective and substantially equivalent to predicate devices as described herein.

  1. Vatech Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.

END

Vatech Co., Ltd.

page 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" arranged in a semi-circular fashion around the symbol. The caduceus is a common symbol associated with healthcare and medicine. The logo is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Vatech Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Consultant Vatech America 333 Meadowlands Parkway, #303 SECAUCUS NJ 07094

JUL 2 1 2010

Re: K092830

Trade/Device Name: Digital X-ray Imaging System / DRC-1000 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: KPR Dated: January 14, 2010 Received: January 19, 2010

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{4}------------------------------------------------

Page 2

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Amold J. Roth

Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

510(k) Number(if known): K092830

Device Name: Digital X-ray Imaging System / DRC-1000

Indications for Üse:

Digital X-ray Imaging System, DRC-1000, is intended for by a qualified/trained doctor or technician on both adult and pediatric subjects for taking diagnostic radiographic exposures of the skull, spinal column, chest, extremities, and other body parts. Not to be used for fluoroscopy, angiography, or screening mammography. Applications can be performed with the patient sitting, standing, or lying in the prone or supine position.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use

(Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Braination Care

ASB

(Division Sign-Off) (Division of Radiological Devices Office of In Vitro Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety .

510K K092803

Page 1 of

N/A