K Number
K092811

Validate with FDA (Live)

Date Cleared
2010-01-13

(121 days)

Product Code
Regulation Number
870.1130
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

KD-7035B, KD-7921 and KD-7903 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm- 25cm.

Device Description

KD-7035B, KD-7921 and KD-7903 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm. It is designed and manufactured according to ANSI/AAMI SP10-manual. electronic or automated sphygmanometers. The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result of KD-7035B can also be classified and displayed by the function of blood pressure classification indicator, the memory capability of KD-7035B, KD-7921 and KD-7903 are 2 × 60 times, 60 times and 3 × 40 times. If any irregular heartbeat is detected, it can be shown on the LCD of KD-7035B and KD-7921.

AI/ML Overview

The provided text describes three models of Fully Automatic Electronic Blood Pressure Monitors (KD-7035B, KD-7921, and KD-7903) and their substantial equivalence to a predicate device (KD-738). The summary states that the devices conform to several standards, including AAMI SP10 which outlines performance requirements for sphygmomanometers. However, the document does not present a detailed study with specific acceptance criteria and reported device performance in table format. It also lacks explicit information regarding sample sizes, data provenance, ground truth establishment for test and training sets, the number and qualifications of experts, adjudication methods, or MRMC comparative effectiveness studies.

Based on the information provided in the 510(k) summary, here's what can be extracted and what is missing:


1. Table of Acceptance Criteria and Reported Device Performance

The document states that the devices conform to "AAMI SP10:2002, Manual, electronic or automated sphygmomanometers" and its amendments. AAMI SP10 typically sets accuracy requirements for blood pressure devices. While the exact acceptance criteria from AAMI SP10 are not explicitly listed in the submission, the standard generally requires that:

  • Mean difference: The mean difference between the device reading and the reference standard should be within ±5 mmHg.
  • Standard deviation: The standard deviation of the differences should be 8 mmHg or less.

However, the document does not provide a table with these specific acceptance criteria and the reported device performance results (e.g., actual mean difference and standard deviation achieved by KD-7035B, KD-7921, and KD-7903 during testing). It only states that "the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness," implying that the devices met the relevant standards.


2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: Not specified in the provided text.
  • Data Provenance: Not specified, though the submitter is based in P.R. China, which might suggest data could originate from that region. The study is implicitly retrospective, as it's a premarket submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • Number of Experts: Not specified.
  • Qualifications of Experts: Not specified. For blood pressure device validation, experts are typically trained clinicians who manually measure blood pressure using auscultation with a reference sphygmomanometer.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not specified. In AAMI SP10 validation, typically two or more observers are used for reference measurements, and their readings are averaged or adjudicated if they differ significantly.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • The device is a blood pressure monitor, not an AI-powered diagnostic imaging system. Therefore, an MRMC comparative effectiveness study involving human readers and AI assistance is not applicable and was not performed.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • The devices are standalone electronic blood pressure monitors designed to provide direct measurements. The performance implicitly refers to the standalone performance of the algorithm within the device compared to a reference standard (manual auscultation in a clinical trial). The document does not describe "human-in-the-loop" performance as it's not relevant for this type of device.

7. The Type of Ground Truth Used

  • Type of Ground Truth: For blood pressure monitors conforming to AAMI SP10, the ground truth is typically established by simultaneous, direct auscultatory measurements performed by trained observers using a mercury sphygmomanometer or another validated reference device. This method provides the clinically accepted "true" blood pressure values for comparison.

8. The Sample Size for the Training Set

  • Sample Size for Training Set: Not specified. For this type of device, a "training set" in the context of deep learning or complex AI algorithms is not typically used for basic oscillometric blood pressure monitors. The "training" or calibration would have occurred during the device's engineering and design phases, likely using internal data and algorithms based on known physiological principles.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth for Training Set Was Established: Not specified. As noted above, the concept of a distinct "training set" with established ground truth, as defined in AI studies, is not directly applicable here. The device's algorithms are developed and refined based on engineering principles and potentially internal test data.

{0}------------------------------------------------

Kn92811 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirement of SMDA and 21 CFR 807.92.

1.0 submitter's information

JAN 1 3 2010

Name:Andon Health Co., Ltd.
Address:No 31, Changjiang Road, Nankai District, Tianjin,P.R. China
Phone number:86-22-6052 6161
Fax number:86-22-6052 6162
Contact:Liu Yi
Date of Application:09/04/2009

2.0 Device information

Trade name:Fully Automatic Electronic Blood Pressure Monitor
Common name:Noninvasive blood pressure measurement system
Classification name:Noninvasive blood pressure measurement system

3.0 Classification

Production code: DXN- Noninvasive blood pressure measurement system. Regulation number: 870.1130 Classification: II Cardiovascular Panel:

4.0 Predict device information

Andon Health Co., Ltd. Manufacturer: KD-738 Fully Automatic Electronic Blood Pressure Monitor Device: K092045 510(k) number:

5.0 Device description

KD-7035B, KD-7921 and KD-7903 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm-25cm.

{1}------------------------------------------------

It is designed and manufactured according to ANSI/AAMI SP10-manual. electronic or automated sphygmanometers.

The operational principle is based on oscillometric and silicon integrate pressure sensor technology, the result will be shown on a LCD with an electronic interface module, the result of KD-7035B can also be classified and displayed by the function of blood pressure classification indicator, the memory capability of KD-7035B, KD-7921 and KD-7903 are 2 × 60 times, 60 times and 3 × 40 times. If any irregular heartbeat is detected, it can be shown on the LCD of KD-7035B and KD-7921.

6.0 Intended use

KD-7035B, KD-7921 and KD-7903 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm to 25cm.

The intended use and the indication for use of KD-7035B, KD-7921 and KD-7903, as described in its labeling are the same as the predict device KD-738.

Technological CharacteristicsComparison result
Design principleIdentical
AppearanceSimilar
Patients contact MaterialsIdentical
PerformanceSimilar
BiocompatibilityIdentical
Mechanical safetyIdentical
Energy sourceIdentical
Standards metIdentical
Electrical safetyIdentical
EMCIdentical
FunctionSimilar

7.0 Summary comparing technological characteristics with predicate device

{2}------------------------------------------------

8.0 Performance summary

KD-7035B, KD-7921 and KD-7903 Fully Automatic Electronic Blood Pressure Monitor conforms to the following standards:

  • · IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 1988; Amendment 1, 1991-11, Amendment 2. 1995.
  • · IEC 60601-1-2, Medical Electrical Equipment Part 1-2: General Requirements for Safety - Collateral standard: Electromagnetic Compatibility - Requirements and Tests (Edition 2:2001 with Amendment 1:2004; Edition 2.1 (Edition 2:2001 consolidated with Amendment 1:2004)).
  • · AAMI SP10:2002, Manual, electronic or automated sphygmomanometers.
  • · AAMI / ANSI SP10:2002/A1:2003 -, Amendment 1 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.
  • · AAMI / ANSI SP10:2002/A2:2006 --, Amendment 2 to ANSI/AAMI SP10:2002 Manual, electronic, or automated sphygmomanometers.

9.0 Comparison to the predict device and the conclusion

Our device KD-7035B, KD-7921 and KD-7903 Fully Automatic Electronic J Blood Pressure Monitor is substantially equivalent to the Fully Automatic Electronic Blood Pressure Monitor KD-738 whose 510(k) number is K092045.

KD-7035B and KD-738 are very similar in the design principle, the material, the energy source and the applicable standards. Their cuff circumference are different, their appearance are different. As regard to functions, KD-7035B has 2 × 60 times memory, the averaging of the last 3 times memory and LCD backlight. The MCU of KD-7035B is different from the predicted device KD-738.

KD-7921 and KD-738 are very similar in the design principle, the material, the energy source, the functions and the applicable standards. Their cuff circumference are different, their appearance are different. The MCU of KD-7921 is different from the predicted device KD-738.

KD-7903 and KD-738 are very similar in the design principle, the material, the energy source, the functions and the applicable standards. Their cuff circumference are different, their appearance are different. The performance parameter of cuff pressure range and overpressure limit are different from the predicted device KD-738. KD-7903 and KD-738 also has different MCU. KD-7903 has a memory of 3 × 40 times. KD-7903 didn't have a blood pressure

{3}------------------------------------------------

4/4

....

classification function and a IHB detection function.

However, the test in this submission provides demonstration that these small differences do not raise any new questions of safety and effectiveness.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized, curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Mr. Liu Yi President Andon Health Co., Ltd. No. 31 Changjiang Road, Nankai District Tianjin, 300190 CHINA

JAN 1 3 2010

Re: K092811

Trade/Device Name: Fully Automatic Electronic Blood Pressure Monitors, Models KD-7035B, KD-7921 and KD-7903. Regulatory Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (Two) Product Code: DXN Dated: Undated Received: December 15 2009

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Liu Yi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1.15809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Statement of Indications for Use

SEP 1 4 2009

510(k) Number :K092811Received
Device name:KD-7035B, KD-7921 and KD-7903 Fully AutomaticElectronic Blood Pressure Monitor

Indications for use:

KD-7035B, KD-7921 and KD-7903 Fully Automatic Electronic Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic and systolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. The cuff circumference is limited to 14cm- 25cm.

Prescription use Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use YES (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-COUNTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Rambler

Page 1 of 1

(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number |69281|

§ 870.1130 Noninvasive blood pressure measurement system.

(a)
Identification. A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.(b)
Classification. Class II (performance standards).