K Number
K092786

Validate with FDA (Live)

Date Cleared
2010-03-10

(181 days)

Product Code
Regulation Number
870.2770
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

For use only in healthy subjects for Measurement of: Estimated: Skeletal muscle mass, Extra-Cellular Water (ECW), Intra-Cellular Water (ICW), Total Body Water, (TBW), Body Fat, Body Lean + Dry Lean, Metabolic Rates, Segmental Lean Mass, Segmental Body Fat mass, % Segmental Body Fat, and Energy expenditure of activity. Actual : Weight, Body Mass Index (BMI),and Impedance Values

Device Description

Models InBody 270, InBody R20, and InBody R20B are impedance plethysmograph body composition analyzers. These devices determine body composition parameters based on bioelectrical impedance analysis (BIA). BIA relies on the differing behavior of biological tissues in response to an applied electrical current. Lean tissue is generally highly conductive because it contains large amounts of bound water and electrolytes, while fat tissue and bone are relatively poor conductors. By analyzing the response to electrical signals, BIA thereby permits differentiation of lean tissue, fat, and water and, in some instances, derivation of related body composition parameters. The total impedance resulting from BIA incorporates both resistance and capacitance components.

AI/ML Overview

The provided FDA 510(k) summary for the Biospace Body Composition Analyzers (Models InBody 270, InBody R20, and InBody R20B) focuses on demonstrating substantial equivalence to a predicate device (Biospace Body Composition Analyzer Model InBody 230, K062603) rather than presenting a detailed study with specific acceptance criteria and performance metrics for the new devices being cleared.

The core of the submission relies on the assertion that "The results of bench and software testing indicates that the new devices are as safe and effective as the predicate device." and "After analyzing bench testing data, software validation, and the risk analysis, it is the conclusion of Biospace that the Models InBody 270, InBody R20, and InBody R20B, BODY COMPOSITION ANALYZERS are as safe and effective as the predicate devices, and have few technological differences, thus rendering them substantially equivalent to the predicate device."

This means the submission does not contain the level of detail requested for a typical study that proves acceptance criteria based on direct performance measurements. It's a comparison study, arguing the new device is sufficiently similar to an already cleared device. Therefore, a direct table of acceptance criteria and reported device performance for the new devices is not explicitly provided in the document.

However, we can infer the "acceptance criteria" through the lens of substantial equivalence: the new devices must perform similarly and as safely and effectively as the predicate device for all shared indications. Any new indications (Segmental Body Fat mass, % Segmental Body Fat-Energy expenditure of activity) would implicitly have an acceptance criterion of accurate measurement, but no specific performance targets are given.

Here's an attempt to address your questions based on the provided text, acknowledging the limitations inherent in a 510(k) summary focusing on substantial equivalence:


1. Table of Acceptance Criteria and Reported Device Performance

As noted, the 510(k) summary does not provide a table of explicit numerical acceptance criteria for the new devices and their direct performance. The acceptance criterion is primarily substantial equivalence to the predicate device (Biospace Body Composition Analyzer Model InBody 230, K062603) in terms of safety and effectiveness for shared indications, and the ability to accurately measure the new indications.

The 'reported device performance' is therefore a general statement of equivalency.

Acceptance Criterion (Inferred from Substantial Equivalence)Reported Device Performance (as stated in the 510(k))
For all shared indications (e.g., Estimated: Skeletal muscle mass, ECW, ICW, TBW, Body Fat, Body Lean + Dry Lean, Metabolic Rates, Segmental Lean Mass; Actual: Weight, BMI, Impedance Values): The new devices (InBody 270, R20, R20B) must be as safe and effective as the predicate device (InBody 230)."The results of bench and software testing indicates that the new devices are as safe and effective as the predicate device."
For new indications (Segmental Body Fat mass, % Segmental Body Fat, Energy expenditure of activity): The devices must accurately measure these parameters for use in healthy subjects.Implied by the statement of "as safe and effective" and the clearance for these indications. No specific performance metrics (e.g., accuracy, precision relative to a gold standard) are provided in this summary.
Technical Specifications: Operate within specified parameters (e.g., Bioelectrical Impedance analysis method, 20/100kHz frequency, 8 tactile electrodes placed on thumbs, palms, heels, fore-feet, measuring at Right Arm, Left Arm, Trunk, Right Leg, Left Leg, Upright patient position).These parameters are identical to the predicate device, indicating conformity with established specifications.

No information is provided regarding:

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective) The document does not specify a separate "test set" in the context of clinical performance data for the new devices, as the primary assessment is substantial equivalence through bench and software testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience) Not applicable, as detailed clinical performance data using expert-established ground truth is not presented for the new devices.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance Not applicable, as this device is a body composition analyzer, not an AI-assisted diagnostic tool requiring human reader studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The device itself is a standalone body composition analyzer. The "bench and software testing" mentioned would apply to the algorithm's performance, but no details of such a standalone performance study (e.g., accuracy against a gold standard like DEXA) are provided in this summary. The substantial equivalence argument often relies on the predicate device already having demonstrated acceptable standalone performance.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not specified for the new devices. For the predicate device, ground truth for body composition analysis typically involves methods like DEXA (Dual-energy X-ray absorptiometry) or hydrostatic weighing, but these details are not in this summary.

8. The sample size for the training set Not applicable, as this is not an AI/machine learning device that involves a "training set" in the typical sense. Any internal algorithm development for the predicate or the new devices would have used internal data, but details are not provided.

9. How the ground truth for the training set was established Not applicable.


In summary, this 510(k) submission successfully argued for substantial equivalence based on technical similarities and general statements of safety and effectiveness derived from bench and software testing, without providing granular clinical study data with specific acceptance criteria and performance metrics for the new devices.

{0}------------------------------------------------

K092786
PAGE 1 OF 2

510(k) Summary K09

Biospace Corporation Limited Donghyun Building, 518-10 Dogok 2 - Dong Gangnam-Gu, Seoul, KOREA 135-854 Tel : +82-2-501-3939, Fax : +82-2-501-3978 Homepage : http:// www.biospace.co.kr

MAR 1 0 2010

August 27, 2009 Contact: Kichul Cha, CEO

  • Identification of the Device: 1. Proprietary-Trade Name: Biospace Body Composition Analyzers, Models InBody 270, InBody R20, and InBody R20B Classification Names: 74 MNW ANALYZER, BODY COMPOSITION Common/Usual Name: Body fat meter
    1. Equivalent legally marketed devices: Biospace Body Composition Analyzer Model InBody 230 K062603
    1. Indications for Use (intended use) For use only in healthy subjects for Measurement of: Estimated: Skeletal muscle mass, Extra-Cellular Water (ECW), Intra-Cellular Water (ICW), Total Body Water, (TBW), Body Fat, Body Lean + Dry Lean, Metabolic Rates, Segmental Lean Mass, Segmental Body Fat mass, % Segmental Body Fat, and Energy expenditure of activity. Actual : Weight, Body Mass Index (BMI),and Impedance Values
    1. Description of the Device: Models InBody 270, InBody R20, and InBody R20B are impedance plethysmograph body composition analyzers. These devices determine body composition parameters based on bioelectrical impedance analysis (BIA). BIA relies on the differing behavior of biological tissues in response to an applied electrical current. Lean tissue is generally highly conductive because it contains large amounts of bound water and electrolytes, while fat tissue and bone are relatively poor conductors. By analyzing the response to electrical signals, BIA thereby permits differentiation of lean tissue, fat, and water and, in some instances, derivation of related body composition parameters. The total impedance resulting from BIA incorporates both resistance and capacitance components.
    1. Safety and Effectiveness, comparison to predicate device. The results of bench and software testing indicates that the new devices are as safe and effective as the predicate device.

{1}------------------------------------------------

K092786
PAGE 2 OF 2

Biospace Body Composition AnalyzerModels InBody 270, InBody R20, and
Model InBody 230InBody R20B
510(k) numberK062603(NEW)
Indications for UseMeasurement of EstimatedExtra-Cellular Water(ECW),Intra-Cellular Water(ICW),Total Body Water,Skeletal Muscle MassBody Fat,Body Lean + Dry Lean,Basal Metabolic Rates,Segmental Lean MassMeasurement of EstimatedExtra-Cellular Water(ECW),Intra-Cellular Water(ICW),Total Body Water,Skeletal Muscle MassBody Fat,Body Lean + Dry Lean,Basal Metabolic Rates,Segmental Lean MassNew:Segmental Body Fat mass% Segmental Body Fat- Energyexpenditure of activity
Measurement of Actual :WeightBody Mass Index (BMI)Impedance ValuesMeasurement of Actual :WeightBody Mass Index (BMI)Impedance Values
Analysis MethodBioelectrical ImpedanceBioelectrical Impedance
Operating parametersFrequency :20, 100kHzFrequency :20, 100kHz
Electrode TypeTactileTactile .
Number /8 electrodes8 electrodes
Placement ofplaced on thumbs, palms, heels, andplaced on thumbs, palms, heels, and
Electrodesfore-feetfore-feet
Impedance MeasuringRight Arm, Left Arm, Trunk,Right Arm, Left Arm, Trunk,
SiteRight Leg, Left LegRight Leg, Left Leg
Patient PositionUprightUpright
Patient populationHealthy individualsSAME
Power SourceAC Line .SAME, Except Model R20 operatesfrom 4-AA alkaline batteries.
  1. Substantial Equivalence Chart, Model InBody 270, InBody R20, and InBody R20B .

Differences between models InBody 270, InBody R20, and InBody R20B: InBody 270 has a pedestal stand for the hand contacts, and InBody R20/R20B use a coiled cord for the hand contacts. Model R20B has Bluetooth, whereas the other two models do not.

7. Conclusion

After analyzing bench testing data, software validation, and the risk analysis, it is the conclusion of Biospace that the Models InBody 270, InBody R20, and InBody R20B, BODY COMPOSITION ANALYZERS are as safe and effective as the predicate devices, and have few technological differences, thus rendering them substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its wings and body. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" around the perimeter.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G61 Silver Spring, MD 20993-0002

MAR ] 0 2010

Biospace Corporation Ltd. c/o Mr. Daniel Kamm Kamm & Associates 8726 Ferrara Court NAPLES FL 34114

Re: K092786

Trade/Device Name: Biospace Body Composition Analyzer, Models InBody 270, InBody R20, and InBody R20B

Regulation Number: 21 CFR 8870.2770 Regulation Name: Impedance plethysmograph Regulatory Class: II Product Code: MNW · Dated: January 26, 2010 Received: February 2, 2010

Dear Mr. Kamm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{3}------------------------------------------------

Page 2 –

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K09 2786

Device Name: Biospace Body, Composition Analyzer, Models InBody 270, InBody R20, and InBody R20B.

Indications For Use:

For use only in healthy subjects for Measurement Of:

Estimated:

Skeletal muscle mass, Extra-Cellular Water (ECW), Intra-Cellular Water (ICW), Total Body Water, (TBW), Body Fat, Body Lean + Dry Lean, Metabolic Rates, Segmental Lean Mass, Segmental Body Fat mass, % Segmental Body Fat, and Energy expenditure of activity.

Actual : Weight, Body Mass Index (BMI),and Impedance Values

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices

510(k) Number_

Page 1 of 1

§ 870.2770 Impedance plethysmograph.

(a)
Identification. An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.(b)
Classification. Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.