The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle fixation. Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications:
- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
- spondylolisthesis,
- trauma (i.e., fracture or dislocation),
- spinal stenosis.
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- tumor,
- pseudoarthrosis, and
- failed previous fusion
The Firebird Spinal Fixation Screw System components are used with certain components of the Blackstone SFS system, including rods, rod connectors and cross-connectors.
The Firebird Spinal Fixation System is a temporary, titanium alloy, multiple component system comprised of a variety of non-sterile, single use components that allow the surgeon to build a spinal implant construct. The system is attached to the vertebral body by means of screws to the non-cervical spine. The spinal construct is completed by connecting the screws with titanium alloy or cobalt chrome rods.
The document describes the K092624 premarket notification for the Blackstone Medical, Inc. Firebird Spinal Fixation System with the addition of a Cobalt Chrome Rod. This is a medical device submission, and as such, the concept of "acceptance criteria" and "device performance" as it might apply to software or AI algorithms is not directly applicable in the same way. Instead, the "acceptance criteria" for a medical device like this is typically demonstrating substantial equivalence to a predicate device, primarily through mechanical testing to ensure safety and effectiveness.
Here's a breakdown based on the provided text, adapted to the context of a medical hardware device:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Demonstration of Substantial Equivalence) | Reported Device Performance (Study Results) |
|---|---|
| Same intended use as predicate device | The modified system has the same intended use as the predicate system. |
| Similar indications for use as predicate device | The modified system has similar indications for use as the predicate system. |
| Similar technological characteristics to predicate device | The modified system has similar technological characteristics and principles of operation as the predicate system. |
| Meets established mechanical testing standards | Mechanical testing was conducted in accord with ASTM standards. |
| Mechanical performance substantially equivalent to predicate | Mechanical testing demonstrates that the system with cobalt chrome rods is substantially equivalent to the predicate system when using titanium alloy rods. |
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: The document does not specify the exact sample size for the mechanical testing. For medical devices, "sample size" in this context typically refers to the number of devices or components tested. ASTM standards for medical devices usually specify the required number of samples for each test type to achieve statistical significance.
- Data Provenance: The document does not explicitly state the country of origin or whether the data was retrospective or prospective. Given that the submission is to the FDA (USA), it is reasonable to infer the testing was likely conducted in the US or by a lab adhering to US standards. Mechanical testing on medical devices is inherently prospective, as it involves newly produced components subjected to defined test protocols.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)
This question is not applicable. For mechanical testing of a medical device, "ground truth" is established by the specifications of the ASTM standards themselves and the objective measurements obtained from standardized testing equipment, not by a panel of human experts. Expert clinical opinion would be relevant in the initial design and indication definition, but not in validating the mechanical performance.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
This question is not applicable. Mechanical testing of physical medical devices does not involve human adjudication in the way clinical studies or image analysis might. The results are based on objective physical measurements and adherence to pre-defined test protocols.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
This question is not applicable. The device is a spinal fixation system (hardware), not an AI or imaging diagnostic tool. Therefore, MRMC studies or human reader improvement with AI assistance are irrelevant to this submission.
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
This question is not applicable. The device is a spinal fixation system, which is a physical implant, not an algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The "ground truth" for the mechanical testing of this device is established by objective physical measurements (e.g., force, displacement, cycles to failure) in accordance with recognized industry standards (ASTM standards). These standards define the criteria for acceptable performance (e.g., fatigue life, strength) for spinal fixation systems.
8. The sample size for the training set
This question is not applicable. There is no "training set" in the context of mechanical testing for a medical device like a spinal fixation system. The device's performance is evaluated against engineering standards and comparison to a predicate device.
9. How the ground truth for the training set was established
This question is not applicable, as there is no training set for this type of device submission.
{0}------------------------------------------------
K09262y
SEP 2 5 2009
Premarket Notification 510(k) Blackstone Medical, Inc. Firebird Spinal Fixation System Modification Confidential
510(k) SUMMARY
Spinal Fixation System Modification - Cobalt Chrome Rod
Blackstone Medical, Inc. Sponsor: 1211 Hamburg Turnpike Suite 300 Wayne, NJ 07470 Registration Number: 3004606875 Whitney G. Torning, Senior Director of Regulatory Affairs Contact Person: & Quality Assurance Telephone Number: 973.406.2838 Fax Number: 973.406.2938 Email: wtorning@blackstonemedical.com Martin G. Sprunck Submitter: Regulatory Affairs Manager Manufacturer: Blackstone Medical, Inc. 1211 Hamburg Turnpike, Ste. 300 Wayne, NJ 07470 Registration Number: 3004606875 Precision Forms, Inc. Contract Manufacturer:
97 Decker Rd. Butler, NJ 07405
System Name:
Trade Name (Component): Firebird Cobalt Chrome Rod
Common Name (System): Posterior Thoracolumbar System
NKB - Orthosis, Spinal Pedicle Fixation, for Degenerative Product Code: Disc Disease
Firebird Spinal Fixation System
MNI~ Orthosis, Spinal Pedicle Fixation Subsequent Product Codes: MNH - Orthosis, Spondylolisthesis Spinal Fixation
{1}------------------------------------------------
Premarket Notification 510(k) Blackstone Medical, Inc. Firebird Spinal Fixation System Modification Confidential . . . .
| Regulatory Classifications: | Class III Preamendment Device, 888.3070 - Pedicle ScrewSpinal System - *Class III Summary and CertificationRequiredClass II – 888.3070 – Pedicle Screw Spinal System |
|---|---|
| Review Panel: | Orthopedic Device Panel |
| Predicate Devices: | Blackstone Pedicle Screw System (K081684 SE 9/15/08)Blackstone Pedicle Screw System, 4.0 mm Screws(K082797 SE 10/17/08)K2M, Inc. CoCr Rod (K080792 SE 4-8-08)Applied Spine Technologies, Inc. Bar Pedicle Screw SpinalFixation System (K061162 SE 7-26-06) |
Intended Use / Indications for Use
The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle fixation. Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft or allograft. The device is indicated for all of the following indications:
-
- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
-
- spondylolisthesis,
-
- trauma (i.e., fracture or dislocation),
-
- spinal stenosis.
-
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
-
- tumor,
-
- pseudoarthrosis, and
-
- failed previous fusion
The Firebird Spinal Fixation Screw System components are used with certain components of the Blackstone SFS system, including rods, rod connectors and cross-connectors.
Technological Characteristics
The Firebird Spinal Fixation System is a temporary, titanium alloy, multiple component system comprised of a variety of non-sterile, single use components that allow the surgeon to build a spinal implant construct. The system is attached to the vertebral body by means of screws to the non-cervical spine. The spinal construct is completed by connecting the screws with titanium alloy or cobalt chrome rods.
{2}------------------------------------------------
Performance Data
Mechanical testing of the Firebird Spinal Fixation system with the cobalt chrome rods was conducted in accord with ASTM standards, and demonstrates that the system is substantially equivalent to the predicate system when using titanium alloy rods. The modified system has the same intended use, similar indications, technological characteristics and principles of operation as the predicate system.
Basis of Substantial Equivalence
Mechanical testing was conducted to demonstrate that the Firebird Spinal Fixation System with the addition of cobalt chrome alloy rods is substantially equivalent to the current Firebird Spinal Fixation System, (K081684 SE 9/15/08 and K082797 SE 10/17/08), which has been cleared by FDA for the purpose of building a spinal implant construct in the noncervical spine.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle in flight, with three curved lines representing the wings and tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002
SEP 2 5 2009
Blackstone Medical, Inc. % Ms. Whitney G. Törning Senior Director of Regulatory Affairs & Quality Assurance 1211 Hamburg Turnpike, Suite 300 Wayne, New Jersey 07470
Re: K092624
Trade/Device Name: Firebird Spinal Fixation System, Firebird Cobalt Chrome Rods Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: III Product Code: NKB, MNI, MNH Dated: August 24, 2009 Received: August 26, 2009
Dear Ms. Törning:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Ms. Whitney G. Törning
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Haibau Buelup
Mark N. Melkers Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Premarket Notification 510(k) Blackstone Medical, Inc. Firebird Spinal Fixation System Modification Confidential
Indications for Use Statement
510(k) Number (if known):
System Name: Firebird Spinal Fixation System Device Name: Firebird Cobalt Chrome Rods
Indications for Use:
The Firebird Spinal Fixation System is intended for posterior, non-cervical pedicle fixation. Pedicle screw fixation is limited to skeletally mature patients and is intended to be used as an adjunct to fusion using autograft. The device is indicated for all of the following indications:
- degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies)
-
- spondylolisthesis.
-
- trauma (i.e., fracture or dislocation),
-
- spinal stenosis.
-
- deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
-
- tumor,
-
- pseudoarthrosis, and
-
- failed previous fusion
The Firebird Spinal Fixation System components are used with certain components of the Blackstone SFS system, including rods, rod connectors and cross-connectors.
| Prescription Use X | AND/OR | Over-The-Counter Use_ |
|---|---|---|
| (Part 21 C.F.R. 801 Subpart D) | (21 C.F.R. 807 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Kareem S. Burney for MXM
(Division Sign-Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K092624
N/A