K Number
K092528

Validate with FDA (Live)

Date Cleared
2010-05-21

(276 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Biomet Forerunner™ Plating System is used for adult or pediatric patients as indicated for pelvic, small and long bone fracture fixation of bones that have been surgically prepared (osteotomy) for correction of deformity or arthrodesis. Indications for use include internal fixation of the tibia, fibula, femur, metacarpals, metatarsals, humerus, ulna, radius, middle hand, and middle foot bones. This 510(k) expands the indications for use to the forefoot, and adds specific indications for midfoot/forefoot procedures:

  1. Treatment of fractures and fracture-dislocations of the midfoot/forefoot
  2. Malunions
  3. Non-unions
  4. Joint fusions/arthrodesis
  5. Corrective osteotomies for deformities
    These procedures in the midfoot/forefoot may be indicated as a result of trauma, deformity, osteoarthritis, and rheumatoid arthritis.
Device Description

The Biomet Forerunner" Plating System is an internal fixation system consisting of various sized plates and screws. The plates and screws are fabricated from Ti-6Al-4V.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Biomet Forerunner Plating System, which is a metallic bone fixation device. This specific submission expands the indications for use of the system to include the forefoot and adds specific midfoot/forefoot procedures.

However, the document is a regulatory submission describing a device and its intended use, rather than a study evaluating its performance against specific acceptance criteria. The FDA's review concludes that the device is "substantially equivalent" to predicate devices, meaning it does not raise new issues of safety or effectiveness compared to existing, legally marketed devices.

Therefore, the requested information regarding acceptance criteria, reported device performance, study details (sample sizes, data provenance, ground truth, experts, adjudication, MRMC, standalone performance), and training set information is not present in this document.

The "Summary of Technologies" section states: "The technological characteristics of the Forerunner" Plating System are the same as, or similar to, the predicate devices." This implies that the device's performance would be considered similar to the predicate devices, which have already been deemed safe and effective.

In summary, this document does not contain the information required to populate the table or answer the specific questions about acceptance criteria and a study proving device performance, as it is a substantial equivalence claim for a physical medical device, not an AI/software performance study.

{0}------------------------------------------------

Biomet Trauma Traditional 510(k) Premarket Notification

MAY 2 1 2010

K092528 (pg 1/2)

510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR § 807.93

Submitter:

Biomet Trauma 100 Interpace Parkway Parsippany, NJ 07054

2242816

Establishment Registration Number:

Contact:

Margaret F. Crowe Regulatory Affairs Project Manager Biomet Trauma 100 Interpace Parkway Parsippany, NJ 070654 Tel .: 973-299-9300, ext. 2260 Fax: 973-257-0232 E-mail: margaret.crowe@ebimed.com

May 21, 2010 Date Prepared:

Biomet Forerunner " Plating System Trade/Proprietary Name:

Common/Usual Name: Plates/Screws

Classification Name: Single/multiple metallic bone fixation appliances and accessories (21 CFR 888.3030)

Device Panel/Product Code: Orthopedics HRS/HWC

Device Description:

The Biomet Forerunner" Plating System is an internal fixation system consisting of various sized plates and screws. The plates and screws are fabricated from Ti-6Al-4V.

Indications for Use:

The indications for use for this system have been modified. The Biomet Forerunner™ Plating System is currently indicated for use in adult or pediatric patients as indicated for pelvic, small and long bone fracture fixation and fixation of bones that have been surgically

{1}------------------------------------------------

Biomet Trauma Traditional 510(k) Premarket Notification

prepared (osteotomy) for correction of deformity or arthrodesis. Indications for use include internal fixation of the tibia, fibula, femur, metacarpals, humerus, ulna, radius, middle hand and middle foot bones.

This premarket notification expands the indications for use to the forefoot, and adds the following specific indications for midfoot/forefoot procedures:

  • Treatment of fractures and fracture-dislocations of the midfoot/forefoot 1.
    1. Malunions
    1. Non-unions
    1. Joint fusions/arthrodesis
    1. Corrective osteotomies for deformities

These procedures in the midfoot/forefoot may be indicated as a result of trauma, deformity, osteoarthritis, and rheumatoid arthritis.

Summary of Technologies:

The technological characteristics of the Forerunner" Plating System are the same as, or similar to, the predicate devices.

Substantial Equivalence:

The Forerunner™ Plating System is substantially equivalent to its predicate devices with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness. Examples of predicate devices include:

  • DARCO Locked Universal Plate marketed by Wright Medical Technology, Inc. -. cleared under premarket notification K061808,
  • Charlotte Claw Compression Plate marketed by Wright Medical Technology, Inc .-. cleared under premarket notifications K080295 and K051908

{2}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAY 2 1 2010

Biomet Trauma (aka EBI, LP) % Ms. Margaret Crowe Regulatory Affairs Project Manager 100 Interpace Parkway Parsippany, NJ 07054

Re: K092528

Trade/Device Name: Biomet Forerunner Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: May 19, 2010 Received: May 20, 2010

Dear Ms. Crowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely vours.

Mark N. M kerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Biomet Trauma Traditional 510(k) Premarket Notification

Indications for Use

510(k) Number (if known): _ KO92852( pcg

Device Name: Biomet Forerunner " Plating System

The Biomet Forerunner™ Plating System is used for adult or pediatric patients as indicated for pelvic, small and long bone fracture fixation of bones that have been surgically prepared (osteotomy) for correction of deformity or arthrodesis. Indications for use include internal fixation of the tibia, fibula, femur, metacarpals, metatarsals, humerus, ulna, radius, middle hand, and middle foot bones. This 510(k) expands the indications for use to the forefoot, and adds specific indications for midfoot/forefoot procedures:

    1. Treatment of fractures and fracture-dislocations of the midfoot/forefoot
    1. Malunions
    1. Non-unions
    1. Joint fusions/arthrodesis
    1. Corrective osteotomies for deformities

These procedures in the midfoot/forefoot may be indicated as a result of trauma, deformity, osteoarthritis, and rheumatoid arthritis.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Temperature

Amuta for mxn
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K092852

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.