K Number
K092500

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-08-20

(6 days)

Product Code
Regulation Number
866.3980
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ProFlu+™ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C.

Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative RSV results be confirmed by culture.

Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Device Description

The ProFlu+ Assay enables detection and differentiation of Influenza A Virus, Influenza B Virus, Respiratory Syncytial Virus (RSV) (Types A and B), and Internal Control. Nasopharyngeal swab specimens are collected from symptomatic patients using a polyester, rayon or nylon tipped swab and place into viral transport medium.

An Internal Control (IC) is added to each sample prior to nucleic acid isolation to monitor for inhibitors present in the specimens. The isolation and purification of the nucleic acids is performed using either a MagNA Pure LC Instrument (Roche) and the MagNA Pure Total Nucleic Acid Isolation Kit (Roche) or a NucliSENS® easyMAGTM System (bioMérieux) and the Automated Magnetic Extraction Reagents (bioMérieux).

The purified nucleic acids are added to ProFlu+ Supermix along with enzymes included in the ProFlu+ Detection Kit. The ProFlu+ Supermix contains oligonucleotide primers and target-specific oligonucleotide probes. The primers are complementary to highly conserved regions of genetic sequences for these respiratory viruses. The probes are dual-labeled with a reporter dye attached to the 5'-end and a quencher dye attached to the 3'-end.

Reverse transcription of the RNA in the sample into complementary DNA (cDNA) and subsequent amplification of DNA is performed in a Cepheid SmartCycler® II instrument. In this process, the probe anneals specifically to the template followed by primer extension and amplification. The ProFlu+ Assay is based on Tagman chemistry, which utilizes the 5' - 3' exonuclease activity of the Taq polymerase to cleave the probe thus separating the reporter dye from the quencher. This generates an increase in fluorescent signal upon excitation from a light source. With each cycle, additional reporter dye molecules are cleaved from their respective probes, further increasing fluorescent signal. The amount of fluorescence at any given cycle is dependent on the amount of amplification products present at that time. Fluorescent intensity is monitored during each PCR cycle by the SmartCycler instrument.

AI/ML Overview

This document describes the ProFlu+ Assay, a multiplex Real Time RT-PCR in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids. However, the provided text does not contain the information requested in your prompt regarding acceptance criteria, specific study details, sample sizes, ground truth establishment, expert qualifications, or comparative effectiveness study results.

The document primarily focuses on:

  • Identification of the device: ProFlu+ Assay
  • Contact information: Prodesse, Inc.
  • Predicate device: K081030 - ProFlu+ Assay, Prodesse, Inc. and K091667 - ID-Tag Respiratory Virus Panel, Luminex Molecular Diagnostics, Inc.
  • Intended Use: Aid in differential diagnosis of Influenza A, Influenza B, and RSV viral infections from nasopharyngeal swab specimens from symptomatic patients. Explicitly states it's not for Influenza C, and negative results for RSV should be confirmed by culture.
  • Product Description: Details the methodology (multiplex Real Time RT-PCR, use of internal control, MagNA Pure LC Instrument or NucliSENS® easyMAGTM System for nucleic acid isolation, Cepheid SmartCycler® II for RT-PCR, Taqman chemistry).
  • Substantial Equivalence: Notes that the intended use remains the same, and the device is reactive to the 2009 H1N1 Influenza Virus.
  • FDA Clearance: Provides formal communication from the FDA regarding the 510(k) clearance (K092500).

Therefore, I cannot fulfill your request for the specific details outlined in your prompt based on the provided text. The document does not include a table of acceptance criteria, reported device performance metrics against such criteria, sample sizes for test/training sets, ground truth methods, expert details, or information on MRMC studies.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "Prodesse" in a serif font, with the words "Real Time Solutions" underneath in a smaller, italicized font. To the left of the word "Prodesse" is a geometric shape that looks like a stylized hexagon. Underneath the geometric shape is a smaller, more abstract shape.

KO92500

ATTACHMENT 9

ProFlu+ Special 510(k) Submission

510(k) SUMMARY

August 13, 2009

CONTACT

Kristine Schraufnagel Prodesse, Inc. W229 N1870 Westwood Dr. Waukesha, WI 53186 Phone: 262-446-0700 ext. 109 Fax: 262-446-0600 Email: kschraufnagel@prodesse.com

AUG 2 D 2009

NAME OF DEVICE

Trade Name: ProFlu+ Assay Regulation Number: 21 CFR 866.3980 Classification Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay

PREDICATE DEVICE

K081030 - ProFlu+ Assay, Prodesse, Inc. K091667 - ID-Tag Respiratory Virus Panel, Luminex Molecular Diagnostics, Inc.

INTENDED USE

The ProFlu+™ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isotated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C.

Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative RSV results be confirmed by culture.

Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

CONFIDENTIAL TO PRODESSE

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Prodesse. The logo consists of a geometric shape on the left and the word "Prodesse" in a serif font on the right. Below the word "Prodesse" are the words "Real Time Solutions" in a smaller font.

ATTACHMENT 9

ProFlu+ Special 510(k) Submission

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

PRODUCT DESCRIPTION

The ProFlu+ Assay enables detection and differentiation of Influenza A Virus, Influenza B Virus, Respiratory Syncytial Virus (RSV) (Types A and B), and Internal Control. Nasopharyngeal swab specimens are collected from symptomatic patients using a polyester, rayon or nylon tipped swab and place into viral transport medium.

An Internal Control (IC) is added to each sample prior to nucleic acid isolation to monitor for inhibitors present in the specimens. The isolation and purification of the nucleic acids is performed using either a MagNA Pure LC Instrument (Roche) and the MagNA Pure Total Nucleic Acid Isolation Kit (Roche) or a NucliSENS® easyMAGTM System (bioMérieux) and the Automated Magnetic Extraction Reagents (bioMérieux).

The purified nucleic acids are added to ProFlu+ Supermix along with enzymes included in the ProFlu+ Detection Kit. The ProFlu+ Supermix contuins oligonucleotide primers and target-specific oligonucleotide probes. The primers are complementary to highly conserved regions of genetic sequences for these respiratory viruses. The probes are dual-labeled with a reporter dye attached to the 5'-end and a quencher dye attached to the 3'-end.

Reverse transcription of the RNA in the sample into complementary DNA (cDNA) and subsequent amplification of DNA is performed in a Cepheid SmartCycler® II instrument. In this process, the probe anneals specifically to the template followed by primer extension and amplification. The ProFlu+ Assay is based on Tagman chemistry, which utilizes the 5' - 3' exonuclease activity of the Taq polymerase to cleave the probe thus separating the reporter dye from the quencher. This generates an increase in fluorescent signal upon excitation from a light source. With each cycle, additional reporter dye molecules are cleaved from their respective probes, further increasing fluorescent signal. The amount of fluorescence at any given cycle is dependent on the amount of amplification products present at that time. Fluorescent intensity is monitored during each PCR cycle by the SmartCycler instrument.

CONFIDENTIAL TO PRODESSE

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image contains the word "Prodesse" in a bold, serif font. Below the word "Prodesse" is the phrase "Real Time Solutions" in a smaller, italicized font. To the left of the word "Prodesse" is a geometric shape that appears to be a stylized logo.

ATTACHMENT 9

SUBSTANTIAL EQUIVALENCE

The intended use remains the same.

Nasopharyngeal swab samples can be collected in Remel M4, M5, and M6, Copan Universal Transport Medium and Becton Dickinson Universal Viral Transport and tested using the ProFlu+ Assay.

ProFlu+ Assay is reactive to the 2009 H1N1 Influenza Virus ("Swine Flu") at a concentration near the ProFlu+ Assays' limit of detection for Influenza A for all the five culture isolates tested. In addition, the ProFlu+ Assay detected S confirmed clinical positive samples.

CONFIDENTIAL TO PRODESSE

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human figures connected at the arms, representing health, services, and the human aspect of the department's mission.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

MAR 22 2010

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center-WO66-G609 Silver Spring, MD 20993-0002

Kristen Schraufnagel QA Officer Prodesse. Inc. W229 N1870 Westwood Dr. Waukesha, WI 53186

Re: K092500

Trade/Device Name: Proflu+™ Assay Regulation Number: 21 CFR 866.3980 Regulation Name: Respiratory Viral Panel Multiplex Nucleic Acid Assay Regulatory Class: Class II Product Code: OCC Dated: August 13, 2009 Received: August 14, 2009

Dear Ms. Schraufnagel:

This letter corrects our substantially equivalent letter of August 20, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR). Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{4}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific information about the application of our labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 594-3084. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Salla atgr

Sally A. Hojvat. M.Sc., Ph Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Prodesse, Inc. ProFlu+ Assay Special 510(k) Submission

Attachment 5 Indication for Use

510(k) Number (if known):

Device Name: ProFlu+ Assay

Indication For Use:

The ProFlu+™ Assay is a multiplex Real Time RT-PCR in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C.

Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. It is recommended that negative RSV results be confirmed by culture.

Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation. When other Influenza A viruses are emerging, performance characteristics may vary.

If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.

Prescription Use X (21 CFR Part 801 Subpart D)

And/Or

Over the Counter Use _ (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

for U. Scher

ee

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K092500

§ 866.3980 Respiratory viral panel multiplex nucleic acid assay.

(a)
Identification. A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B;
(2) Influenza A subtype H1 and Influenza A subtype H3;
(3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B;
(4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus;
(5) Human Metapneumovirus;
(6) Rhinovirus; and
(7) Adenovirus.
(b)
Classification. Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.