K Number
K092421

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-02-12

(189 days)

Product Code
Regulation Number
884.1690
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

AlphaScope Hysteroscope is used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. AlphaScope Hysteroscope is used with Hysteroscope Sheaths to provide access to the uterine cavity during diagnostic and operative hysteroscopic procedures.

Device Description

AlphaScope Hysteroscope is used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. AlphaScope Hysteroscope is used with Hysteroscope Sheaths to provide access to the uterine cavity during diagnostic and operative hysteroscopic procedures. The use of AlphaScope Hysteroscope is restricted for the use of endoscopic surgeons and qualified assistants. AlphaScope Hysteroscopes are re-useable, they can be sterilized by 134 C Degrees Steam Autoclave.

AI/ML Overview

The provided text describes the regulatory clearance for the AlphaScope Hysteroscope, a medical device. However, it does not include information about specific acceptance criteria, a study proving the device meets those criteria, or any of the detailed performance metrics asked for in the prompt. This document is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than presenting detailed new clinical or non-clinical performance studies with acceptance criteria.

The key statement regarding performance is: "Based on the equivalence in design and materials to predicate devices, performance testing was not warranted. The device meets the same criteria and is as effective and safe as SE devices." This indicates that no new performance studies were conducted to define specific acceptance criteria for this particular device.

Therefore, I cannot populate the table or answer the specific questions about sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, or ground truth for a performance study, as this information is not present in the provided text.

Based on the provided document, here's what can be inferred and what cannot:

1. A table of acceptance criteria and the reported device performance

  • Cannot provide. The document states that performance testing was not warranted due to equivalence with predicate devices. Therefore, specific acceptance criteria and detailed performance metrics for this new device are not reported. The device is deemed to meet the "same criteria" and be "as effective and safe" as existing, substantially equivalent (SE) devices.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Cannot provide. No test set for performance evaluation is described.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Cannot provide. No ground truth establishment for a test set is described.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Cannot provide. No test set or adjudication process is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Cannot provide. No MRMC study is mentioned. This is a purely mechanical device, not an AI-driven one.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Cannot provide. This is a purely mechanical device, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Cannot provide. No ground truth for performance evaluation is described.

8. The sample size for the training set

  • Cannot provide. There is no mention of a training set, as this is not an AI/algorithm-based device.

9. How the ground truth for the training set was established

  • Cannot provide. There is no mention of a training set or its ground truth.

Summary based on the provided document:

The AlphaScope Hysteroscope received 510(k) clearance based on its substantial equivalence to legally marketed predicate devices, specifically the MicroSpan Gold Hysteroscope (K972426) and Alpha Hysteroscope, Rod Lenses (K012869).

The basis for this substantial equivalence claim relies on:

  • Identical intended use.
  • Similar materials (primarily surgical grade stainless steel, with verification of biocompatibility).
  • Similar basic design, with slight modifications (design, diameter, material) that were determined not to "adversely affect the safety and effectiveness."
  • Conformity with European Norms and directives (CE-marked) and relevant international standards (ISO, EN, ASTM, AAMI, ANSI, IEC).
  • Validation of reprocessing and sterilization protocols by Nelson Lap.
  • Market observation of a rigid AlphaScope Hysteroscope (2.7 mm diameter) which "has shown that the used materials are not effect the safety of the patients."

Crucially, the document explicitly states: "Based on the equivalence in design and materials to predicate devices, performance testing was not warranted. The device meets the same criteria and is as effective and safe as SE devices." Therefore, no new, specific acceptance criteria or performance studies for this device are detailed in this 510(k) summary.

{0}------------------------------------------------

92421

510(k) Summary

FEB 1 2 2010

Responsible Persons:

Bernd Maier, Director of OEM Business (responsible for Design evaluation) Reinhold Blazejewski, Fa.MediTech, (responisible for production) Klaus Moser, Head of Service (responsible for Technical Requests) Ulrich Henzler/Gabriela Trompler, QAM (Responsible for Requlations and Documentation

Date Summary prepared: June 22, 2009

Device Name: AlphaScope Hysteroscope, fiber optic

Device Description (807.92):

AlphaScope Hysteroscope is used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. AlphaScope Hysteroscope is used with Hysteroscope Sheaths to provide access to the uterine cavity during diagnostic and operative hysteroscopic procedures.

The use of AlphaScope Hysteroscope is restricted for the use of endoscopic surgeons and qualified assistants.

AlphaScope Hysteroscopes are re-useable, they can be sterilized by 134 C Degrees Steam Autoclave

Classification accord. MDD 93/42 in Risk Class IIa Accord. CFR in Risk Class II

Common Name: Hysteroscopes and accessories

Classification namesProduct codeCFR Regulation #
Hysteroscope, fiberoptic85 HIH884.1690

{1}------------------------------------------------

Indications for Use (807.92)

AlphaScope Hysteroscope is used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.

AlphaScope Hysteroscope is used with Hysteroscope Sheaths to provide access to the uterine cavity during diagnostic and operative hysteroscopic procedures.

Industry Standards/Performance Data (808.92)

We certify the conformity with all European Norms and directives. AlphaScope Hysteroscopes and their accessories are CE-marked.

Also they are in conformity with relevant ISO/EN/ASTM/AAMI/ANSI/IEC and other device-related standards that apply to the manufacture, packaging, labelling, reprocessing, traceability including the validation of these processes. Also we are monitoring these devices using statistics and market observation.

Summary of Testing

All materials used in the composition of AlphaScope Hyseroscopes and their accessories were subjected to performance and physical tests to evaluate safety, effectiveness, and reliability of the devices.

The market observation of AlphaScope Hysteroscope, rigid, with diameter of 2,7 mm has shown that the used materials are not effect the safety of the patients.

All results were in conformance with the cited harmonized device standards and filed.

Information Bearing on the Safety and Effectiveness (807.92)

AlphaScope Hysteroscop has the same intended use as the predicate devices. It is made of the same materials as one of the predicate device (Alpha Hysteroscope, K012869) and is produced to the same international and FDA-recognized standards. Slight modifications in design, diameter and material do not adversely affect the safety and effectiveness of these devices.

In summary, the

  • . Intended use
  • Performance attributes .
  • . Materials and
  • Basic design .

are identical and substantially equivalent to SE devices and to K012869 application.

{2}------------------------------------------------

Labeling

Package Label: All devices are packed in special designed boxes. The Packing was validated (Drop Test).

Affixed to each Box is a label that identifies the enclosed product. Please see Labels in Appendix V

Every product is carrying the product item number, CE-Mark and Serial Number of the device (Please see description, Appendix VII)

Operating Instructions are delivered with the AlphaScopeHysteroscope. Please see sample in Appendix I .

Reprocessing and Sterilization

AlpahaScope Hysteroscopes are delivered non-sterile and they were tested for effective Cleaning Possibility and for effective Sterilisation (see Validation report of Nelson Lap, Appendix IV).

All validated Reprocessing and Sterilization instructions are given in the Operating Manual.

Material

Most of the components are Surgical grade stainless steel and in conformance with FDA consensus standards. SE devices are using the same materials for their scopes.

Biocompatibility was tested by MDT (Rigid AlphaHysteroscope)

Standards

DIN 58105 Medical Endoscopes ISO 8600-1 Endoscopes and Phototonic; General Requirements ISO 8600-2 Endoscopes and Phototonic; Special Requirements for Bronchoscopes ISO 8600-3 Endoscopes and Phototonic; Definition of Viewing Field and Viewing Angle ISO 8600-4 Endoscopes and Phototonic; Definition of maximal size of Insertion part ISO 8600-5 Endoscopes and Phototonic; Definition of the resolution ISO 8600-6 Endoscopes and Phototonic; Definitions DIN EN ISO 9001; Quality Management System DIN EN 13485 Quality Management System for Medical devices DIN EN 14971 Requirements for Risk Management Medical devices DIN EN 60601-2-18; Special Requirements for the Safety including of the basic performance of Endoscopic Equipment DIN EN ISO 10088-1; Stainless Steel DIN EN ISO 7153-1; Surgical Instruments; Part 1 Stainless Steel

{3}------------------------------------------------

DIN EN 17664; Sterilization of Medical Devices DIN EN 1041; Requirements for the Information provided by the Manufacturer DIN EN 980; Symbols for Identification of Medical Devices DIN EN 22 248 (ISO 2248) Drop Test USFDA (21 CFR Part 58) Sterilization Validation AAMI TIR12:2004 AAMI TIR.2003 ANSI/AAMI ST81:2004 ANSI/AAMI ST89:2006 ASTM E 1837 (1996) FDA: CRD 183 FDA; CRD 255 FDA; CRD 256 HTM; CRD 259

Non-Clinical Test Results

Based on the equivalence in design and materials to predicate devices, performance testing was not warranted. The device meets the same criteria and is as effective and safe as SE devices.

Tuttlingen, June 22, 2009

Thilo Henzler President Bernd Maier Director

Klaus Moser Head of Service Ulrich Henzler QAM

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of an eagle or bird-like figure, represented by three curved lines that suggest the shape of a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

GIMMI GmbH c/o Mr. Ken Blake V.P. and General Manager Scanlan International One Scanlan Plaza SAINT PAUL MN 55107

FFB 1 2 2010

Re: K092421

Trade Name: AlphaScope Hysteroscope Regulation Number: 21 CFR §884.1690 Regulation Name: Hysteroscope and accessories Regulatory Class: II Product Code: HIH Dated: January 20, 2010 Received: January 25, 2010

. Dear Mr. Blake:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device-Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{5}------------------------------------------------

Page 2 -

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Janine M. Morris

Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Abbreviated 510(k)

510(k) Number: Hysteroscope, fiberoptic, Common Name: AlphaScope Hysteroscope, fiberoptic, operative and diagnostic Device Name: Predicate Device The AlphaScope Hysteroscope is substantially equivalent to the following Name: predicate devices: MicroSpan Gold Hysteroscope (K972426) Alpha Hysteroscope, Rod Lenses (K012869) 21 CFR § 884.1690 Product code 85 HIH Classification:

Indications for Use

Class II

AlphaScope Hysteroscope is used to permit viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures.

AlphaScope Hysteroscope is used with Hysteroscope Sheaths to provide access to the uterine cavity during diagnostic and operative hysteroscopic procedures.

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) (Division Sign-Un)
Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

OT

Over-The-Counter Use

.

Prescription Use

§ 884.1690 Hysteroscope and accessories.

(a)
Identification. A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.(b)
Classification. (1) Class II (performance standards).(2) Class I for hysteroscope accessories that are not part of a specialized instrument or device delivery system; do not have adapters, connectors, channels, or do not have portals for electrosurgical, laser, or other power sources. Such hysteroscope accessory instruments include: lens cleaning brush, cannula (without trocar or valves), clamp/hemostat/grasper, curette, instrument guide, forceps, dissector, mechanical (noninflatable), and scissors. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.