(105 days)
The KRD1™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (Pseudoarthrosis).
In addition, the KRD1™ Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft.
The KRD1™ Pedicle Screw System consists of longitudinal rods, polyaxial screws, transverse connectors, and instrumentation to facilitate installation of this system.
The KRD11M Pedicle Screw is offered in multiple screw diameters, with multiple lengths to accommodate individual patient needs. KRD1™ Pedicle Screw assemblies consist of the screw, a tulip, and a washer that secures the tulip to the screw and also serves as the saddle for the longitudinal rod. The tulip and screw designs allow the screw to have a polyaxial rotation relative to the tulip. A locking cap is placed on top of the pedicle screw assembly to secure the position of the implant and to retain the longitudinal rod.
The KRD1™ Pedicle Screw assembly serves as the central fixation device to which various rods and cross connectors are secured to provide desired fixation.
Longitudinal rods are provided in two configurations, straight and lordotic, have a fixed diameter (5.5mm), and vary by length.
Cross connector assemblies are provided in multiple configurations, varying by length. Cross connectors are used to provide additional fixation support.
The KRD1™ Pedicle Screw System components are fabricated from medical grade titanium alloy conforming to ASTM F-136 specifications.
This document describes the KRD1™ Pedicle Screw System, a spinal fixation device, and its acceptance criteria for regulatory approval.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Substantial equivalence to predicate devices in: | Shown to be substantially equivalent to predicate devices in: |
| - Indications for use | - Indications for use |
| - Function | - Function |
| - Operating principles | - Operating principles |
| - Materials | - Materials |
| Performance (mechanical testing) | Performance (mechanical testing) was conducted and demonstrated substantial equivalence to predicate devices. |
| Clinical data | Not required for this device type and demonstrated substantial equivalence. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not explicitly mention a "test set" in the context of clinical or performance data that would typically have a sample size for evaluation. Instead, the device's performance was established through mechanical testing and comparison to predicate devices, rather than through clinical trials with patient-specific test sets. Therefore, details regarding sample size for a "test set" and data provenance (country of origin, retrospective/prospective) are not applicable in the conventional sense for this type of device submission. The "data" provenance would refer to the mechanical testing results themselves.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
As clinical data or a patient-based "test set" requiring expert ground truth were not required or performed, this information is not applicable. The ground truth for device performance was established through engineering and biomechanical principles in mechanical testing.
4. Adjudication Method for the Test Set
Since no patient-based "test set" was utilized, an adjudication method for such a set is not applicable.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed as this device relies on mechanical testing and comparison to predicates for substantial equivalence, not on human reader performance with or without AI assistance. Therefore, there is no effect size reported for human readers with AI vs. without AI assistance.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
This device is a physical pedicle screw system, not an algorithm or AI-driven software. Therefore, a standalone (algorithm only) performance study is not applicable.
7. Type of Ground Truth Used
The ground truth used for establishing the device's performance and safety was primarily based on mechanical testing results and comparison of these results, as well as design and material specifications, against established performance characteristics of legally marketed predicate spinal fixation devices. This implies that the 'ground truth' is adherence to engineering standards and established safety and efficacy profiles of similar devices.
8. Sample Size for the Training Set
As this is a physical medical device and not an AI/ML algorithm requiring a training set, the concept of a "training set" and its sample size is not applicable.
9. How Ground Truth for the Training Set Was Established
Since there is no "training set" for this physical device, the method for establishing its ground truth is not applicable.
{0}------------------------------------------------
5.0 510(k) Summary
1. Sponsor
SpineFrontier, Inc. 500 Cummings Center Suite 3500 Beverly, MA 01915
NOV 1 9 2009
Primary Contact: John Sullivan 1- 978-232-3990 Telephone:
November 12, 2009 Date Prepared:
2. Device Name and Classification:
KRD1™ Pedicle Screw System " Proprietary Names:
Common/Usual Name: Spine Fixation System
Classification Name: Orthosis, Spondylolisthesis Spinal Fixation Orthosis, Spinal Pedicle Fixation (21 CFR 888.3070(b)(1)), Class II
MNH, MNI Product Code:
3. Predicate Devices
K950099 - Syneray Posterior Spinal System K992739 - Synthes USS Click-X Variable Axis System K052123 - Synthes Pangea System
4. Device Description
The KRD1™ Pedicle Screw System consists of longitudinal rods, polyaxial screws, transverse connectors, and instrumentation to facilitate installation of this system.
The KRD11M Pedicle Screw is offered in multiple screw diameters, with multiple lengths to accommodate individual patient needs. KRD1™ Pedicle Screw assemblies consist of the screw, a tulip, and a washer that secures the tulip to the screw and also serves as the saddle for the longitudinal rod. The
SpineFrontier, Inc. KRD1 Pedicle Screw System - Supplement to 510(k) Premarket Notification K092420 November 12, 2009
Company Confidential
{1}------------------------------------------------
tulip and screw designs allow the screw to have a polyaxial rotation relative to the tulip. A locking cap is placed on top of the pedicle screw assembly to secure the position of the implant and to retain the longitudinal rod.
The KRD1™ Pedicle Screw assembly serves as the central fixation device to which various rods and cross connectors are secured to provide desired fixation.
Longitudinal rods are provided in two configurations, straight and lordotic, have a fixed diameter (5.5mm), and vary by length.
Cross connector assemblies are provided in multiple configurations, varying by length. Cross connectors are used to provide additional fixation support.
The KRD1™ Pedicle Screw System components are fabricated from medical grade titanium alloy conforming to ASTM F-136 specifications.
5. Intended Use
The KRD1™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondvlolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (Pseudoarthrosis).
In addition, the KRD1™ Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft.
6. Technological Characteristics & Substantial Equivalence Determination
The SpineFrontier KRD1™ Pedicle Screw System was shown to be substantially equivalent to predicate devices through comparison of indications for use, function, operating principles, and materials.
The KRD1 The Pedicle Screw System has been shown to be substantially equivalent to predicate devices in terms of performance (mechanical testing). Clinical data was not required for this device.
Company Confidential
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002
NOV 1 9 2009
SpineFrontier, Inc. % Mr. John Sullivan Director, QA and Regulatory Compliance 500 Cummings Center, Suite 3500 Beverly, Massachusetts 01915
Re: K092420
Trade/Device Name: KRD1™ Pedicle Screw System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: October 20, 2009 Received: October 20, 2009
Dear Mr. Sullivan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
Page 2 - Mr. John Sullivan
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Joneta J
Image /page/3/Picture/6 description: The image shows a stylized handwritten letter 'g'. The letter is in a cursive style, with a loop at the top and a curved tail extending downwards. The letter is black against a white background, and the lines are smooth and flowing.
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
4.0 Indications for Use Statement
510(k) Number (if Known): K092420
Indications For Use:
The KRD1™ Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (Pseudoarthrosis).
In addition, the KRD1™ Pedicle Screw System is indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone.
Prescription Use: _X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use: (Part 21 CFR 807 Subpart C)
Smata for mxm
Division Size 80
(Division Sign Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K092420
SpineFrontier, Inc. KRD1 " Pedicle Screw System - Supplement to 510(k) Premarket Notification K092420 November 12, 2009
Company Confidential
gl fi
N/A