Intended Use

A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Device Description

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

AI/ML Overview

The provided document describes the acceptance criteria and the study performed for a medical device: Better Care Plastic Products Co., Ltd. Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color.

Here's an analysis based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria / Test PerformedReported Device Performance
ASTM D-5250-06e1 (Physical and Dimensions Testing)All testing meets requirements for physical and dimensions testing.
FDA 1000 ml. Water Fill TestMeets requirements (AQL 2.5, inspection level I).
Primary Skin IrritationNo primary skin irritant reactions.
Skin Sensitization (Allergic Contact Dermatitis)No sensitization reactions.
Residual Powder Test (ASTM D-6124-06)Meets "powder-free" claims (contains no more than 2 mg powder per glove).
Pinhole FDA requirementsMeets pinhole FDA requirements.
Biocompatibility requirementsMeets biocompatibility requirements.
Labeling claimsMeets labeling claims.

2. Sample Size Used for the Test Set and Data Provenance

  • ASTM D-5250-06e1 Testing: The document states "Inspection level S-2, AQL 4.0" for physical and dimensions testing. AQL (Acceptable Quality Limit) and inspection level are metrics for sampling, but the precise sample size in numbers is not explicitly stated.

  • FDA 1000 ml. Water Fill Test: "Samplings of AQL 2.5, inspection level I" were conducted. Again, a precise numerical sample size is not provided, but these metrics define the sampling plan.

  • Primary Skin Irritation and Skin Sensitization: The sample size for these tests is not specified in the document.

  • Residual Powder Test: The sample size for this test is not specified.

  • Data Provenance: The studies were performed by Better Care Plastic Products Co., Ltd. for their product. The location of the company is Fuqian Xi Road, West district of Shenze Industrial Base, Shenze County, Hebei Province 050000 (China). The data is retrospective as it was collected before the 510(k) submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • This information is not applicable as the tests performed are primarily laboratory-based performance and safety tests for examination gloves, not diagnostic studies requiring expert review of medical images or data. Ground truth here refers to the objective results of the specified physical, chemical, and biological tests.

4. Adjudication Method for the Test Set

  • This is not applicable. The tests are objective and performed according to established ASTM standards and FDA protocols (e.g., water fill test). There is no "adjudication" in the sense of reconciling differing expert opinions, as the results are determined by the test procedures themselves.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

  • This is not applicable. The device is an examination glove, not an AI-powered diagnostic tool. Therefore, MRMC studies are not relevant.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

  • This is not applicable. The device is an examination glove and does not involve any algorithms or AI for standalone performance evaluation.

7. The Type of Ground Truth Used

  • The ground truth for this device is based on objective measurements and laboratory test results adhering to recognized standards (ASTM D-5250-06e1, ASTM D-6124-06) and regulatory requirements (FDA 1000 ml. Water Fill Test, biocompatibility assessments). It is not based on expert consensus, pathology, or outcomes data in the clinical diagnostic sense. It's the inherent physical and chemical properties of the glove.

8. The Sample Size for the Training Set

  • This is not applicable. The device is a physical medical product (examination gloves) that undergoes testing based on established standards, not a machine learning model that requires a "training set."

9. How the Ground Truth for the Training Set Was Established

  • This is not applicable for the same reason as point 8.

{0}------------------------------------------------

MAY - 5 2010

EXHIBIT #1 Page 1 of 2

510(K) SUMMARY

This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(K) number is: K092302

l . Submitter's Identification:

Mr. Guixi Liu Better Care Plastic Products Co., Ltd. Fuqian Xi Road, West district of Shenze Industrial Base, Shenze County, Hebei Province 050000

Date Summary Prepared: June 30, 2009

2. Name of the Device:

Better Care Plastic Products Co., Ltd. Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color

3. Predicate Device Information:

Grand Work Plastic Products Co., Ltd Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color (K042830)

4. Device Description:

Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250, Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color, 80 LYZ, and meets all requirements of ASTM standard D-5250-06e1.

5. Intended Use:

A patient examination glove is disposable device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

{1}------------------------------------------------

K092302

EXHIBIT #1 Page 2 of 2

6. Comparison to Predicate Devices:

Better Care Plastic Products Co., Ltd.'s Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color is substantially equivalent in safety and effectiveness to the Grand Work Plastic Products Co., Ltd. Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color.

7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:

The standards used for Better Care Plastic Products Co., Ltd.'s glove production are based on ASTM-D-5250-06e1. All testing meets requirements for physical and dimensions testing conducted on gloves. Inspection level S-2, AQL 4.0.

The FDA 1000 ml. Water Fill Test was also conducted with samplings of AQL 2.5, inspection level I, meeting these requirements.

Primary Skin Irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.

A Residual Powder Test that based on ASTM D-6124-06 for powder at finished inspection is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).

8. Labeling:

There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.

9. Discussion of Clinical Tests Performed:

Not Applicable - There is no hypoallergenic Claim.

10. Conclusions:

Better Care Plastic Products Co., Ltd.'s Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color conform fully to ASTM D-5250-06e1 standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three branches of government. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The logo is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Bettercare Plastic Products Company, Limited C/O Mr. Jie Liu Surprotect, Incorporated 3973 Schaefer Avenue Chino, California 91710

Re: K092302

Trade/Device Name: Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera. Green Color Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: April 14, 2010 Received: April 15, 2010

MAY - 5 2010

Dear Mr. Liu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2- Mr. Liu

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), ' please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/ CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Suom Runner

-Anthony Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Attachment A

INDICATION FOR USE

510 (k) NUMBER (if KNOW): K 092 302 APPLICANT: DEVICE NAME:

Better Care Plastic Products Co., Ltd. Powder Free Vinyl Synthetic Examination Gloves with Aloe Vera, Green Color

INDICATIONS FOR USE:

A patient examination glove is disposable device intended for medical a purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter-Use _ √ (Optional Format 1-2-96)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

A

Concurrent of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K092302

§ 880.6250 Non-powdered patient examination glove.

(a)
Identification. A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.(b)
Classification. Class I (general controls). The device, when it is a finger cot, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 880.9.