(301 days)
The EMS9UA Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
- For the measurement of cerebral artery blood velocities to determine the presence of hemodynamically significant deviations from normal values
- To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.
The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal. The EMS9UA Transcranial Doppler is intended for use during:
a) Diagnostic exams
b) Surgical interventions
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
Shenzhen Delicate believes that the Model EMS-9UA is substantially equivalent to its EMS OU Transcranial Doppler which was cleared on May 5, 2006 510kH K060112. The EMS-9UA has the same device description except that the head frame used for longer term monitoring has the ability to track the Doppler signal and therefore not lose the signal with patient movement and time. The tracking is accomplished by adding to the EMS9U an additional circuit which detects the ultrasound Doppler return and positions the face of the probe in the headband to maximize the detected ultrasound return. The headband electronics does not change or interfere with the transmitted ultrasound. Except for the servo motor controller added to the circuitry of the EMS9U range and the software added to control it, and the modifications to the INNSO Tange and the servo motor controlled probe, the EMS9U range and the EMS- 9UA are identical internally and functionally. The probes are identical to those cleared in K060112.
Here's a breakdown of the acceptance criteria and study information for the EMS9UA Transcranial Doppler with Robotic Probe Headband, based on the provided document:
Acceptance Criteria and Device Performance
The document does not explicitly state quantitative performance acceptance criteria (e.g., specific accuracy thresholds) for the EMS9UA device. Instead, the "acceptance criteria" are implied by compliance with safety and industry standards, and the demonstration of safety and effectiveness through performance testing and a clinical trial, showing substantial equivalence to the predicate device.
The reported device performance emphasizes its capabilities as a diagnostic ultrasound fluid flow analysis system for measuring cerebral artery blood velocities and detecting micro embolic signals, with the added functionality of the robotic headband to track the Doppler signal.
Table of Acceptance Criteria and Reported Device Performance:
| Acceptance Criteria Category | Specific Criteria (Implicit/Explicit) | Reported Device Performance/Evidence |
|---|---|---|
| Safety | Compliance with relevant safety standards (UL 2601-1, IEC60601-1-2, IEC60601-2-37) | "Extensive safety... testing... before release. Safety tests have been performed to ensure the device complies with applicable industry and safety standards." |
| Incorporation of safeguards based on literature review. | "A review of the literature pertaining to the safety of the EMS9UA Transcranial Doppler has been conducted and appropriate safeguards have been incorporated in the design..." | |
| No known contra-indications. | "No known at this time." (Contra-indications section) | |
| Effectiveness/Functionality | Meet all functional specifications. | "Final testing of the EMS9UA included various performance tests designed to ensure that the device met all of its functional specifications." |
| Ability to measure cerebral artery blood velocities to determine hemodynamically significant deviations. | "Intended for use as a diagnostic ultrasound fluid flow analysis system... for the measurement of cerebral artery blood velocities to determine the presence of hemodynamically significant deviations from normal values." (Indications for Use) | |
| Ability to assess arterial cerebral blood flow for micro embolic signals. | "Intended for use... to assess arterial cerebral blood flow for the occurrence of micro embolic signals." (Indications for Use) | |
| Robotic Headband's ability to track the Doppler signal (unique feature compared to predicate). | "The EMS-9UA has the same device description except that the head frame used for longer term monitoring has the ability to track the Doppler signal and therefore not lose the signal with patient movement and time." "The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal." | |
| Substantial Equivalence | Demonstrated substantial equivalence to the predicate device (EMS9U Transcranial Doppler, K060112 and Spencer Technologies Marc 600). | "The conclusion drawn from these tests is that the EMS9UA Transcranial Vascular Doppler with Robotic Probe Headband and it's transducers is substantially equivalent in safety and efficacy to the predicate devices listed in the comparison table above." "A clinical trial involving 100 patients was conducted comparing the EMS9UA Transcranial Doppler with Robotic Headband with the Spencer Technologies Marc 600 predicate headband and was found to be safe and effective." |
| Labeling | Include instructions for safe and effective use, warnings, cautions, and guidance for use. | "The Model EMS9UA Transcranial Doppler device labeling includes instructions for safe and effective use, warnings, cautions and guidance for use." |
Study Information
The document describes a clinical trial, though details are limited.
-
Sample Size Used for the Test Set and Data Provenance:
- Sample Size: 100 patients.
- Data Provenance: Not explicitly stated (e.g., country of origin). The study is referred to as a "clinical trial," which typically implies prospective data collection, but this is not explicitly confirmed.
-
Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
- This information is not provided in the document. The study compared the EMS9UA with a predicate device, but it doesn't describe who evaluated the outputs or established a ground truth for the comparison.
-
Adjudication Method for the Test Set:
- This information is not provided in the document.
-
If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:
- Yes, a comparative study was done. A "clinical trial involving 100 patients was conducted comparing the EMS9UA Transcranial Doppler with Robotic Headband with the Spencer Technologies Marc 600 predicate headband."
- Effect Size: The document states that the device "was found to be safe and effective" in comparison to the predicate, implying non-inferiority or similar performance. However, no specific quantitative effect size or details on how human readers improved with AI vs. without AI assistance are provided. The robotic headband's function is to track the Doppler signal and not lose the signal with patient movement and time, which is a practical improvement in monitoring, rather than an AI-driven interpretive enhancement for human readers. This device's primary improvement is in signal stability and monitoring duration, not necessarily in diagnostic interpretation per se.
-
If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
- The document implies that the device is a diagnostic ultrasound system intended for use by "trained medical personnel." It is not described as an AI algorithm providing standalone diagnoses. The robotic probe head is an automated feature for signal tracking, which enhances the device's performance, but it's not suggested that it operates entirely without human oversight or interpretation of the ultrasound data. Therefore, a standalone algorithm-only performance study as typically understood for AI diagnostics is not applicable/not described.
-
The Type of Ground Truth Used:
- This is not explicitly stated. Given that it's a comparative study against a predicate device, the "ground truth" for the comparison would likely be the diagnostic output or clinical utility of the predicate device. However, details on how this was established or if an independent clinical ground truth (e.g., pathology, follow-up outcomes) was used are not provided.
-
The Sample Size for the Training Set:
- This is not applicable/not provided. The document describes a medical device (Transcranial Doppler with a robotic headband for signal tracking), not a typical AI model that undergoes a separate training phase with a distinct training dataset. The "training" for such a device would involve engineering design, calibration, and internal verification.
-
How the Ground Truth for the Training Set Was Established:
- This is not applicable/not provided for the reasons stated above.
{0}------------------------------------------------
Shenzhen Delicate Electronics Co. Ltd
K09 2164
p. 1 of 5
SUMMARY
MAY 1 3 2010
This summary of 510k safety and effectiveness information is being submitted in accordance with 21CFR part 807.92
Submitters name, address, phone number, contact person and 1. preparation date:
Name:
Shenzhen Delicate Electronics Co. Ltd. 6C, Block 8, Tian-an Ind. Area 518054 Nanshan District Shenzhen People's Republic of China Phone: 086 0755-26412665 Fax: 086 0755-26492090 Responsible person: Frank Qui
Official Correspondent:
William Stern Multigon Industries, Inc. 1 Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 X27 Fax: . 914 376 6111
Date of Preparation:
April 7, 2009
{1}------------------------------------------------
K09264
P. 2 of 5
- Device:
Proprietary Name: EMS9UA Transcranial Doppler with Robotic Probe Headband
Common Name: Transcranial and Vascular Doppler Diagnostic Ultrasound Transducer Classification Name: 21 CFR892.1550 System,Imaging,Pulsed Doppler, Ultrasonic 21 CFR892.1570 Diagnostic Ultrasound Transducer
Classification Number: 90IYN 90ITX
Manufactured By:
Shenzhen Delicate Electronics Co. Ltd. 6C, Block 8, Tian-an Ind. Area 518054 Nanshan District Shenzhen People's Republic of China Phone: 086 0755-26412665 Fax: 086 0755-26492090
- Substantially Equivalency
Device Description:
Shenzhen Delicate believes that the Model EMS-9UA is substantially equivalent to its EMS OU Transcranial Doppler which was cleared on May 5, 2006 510kH K060112. The EMS-9UA has the same device description except that the head frame used for longer term monitoring has the ability to track the Doppler signal and therefore not lose the signal with patient movement and time. The tracking is accomplished by adding to the EMS9U an additional circuit which detects the ultrasound Doppler return and positions the face of the probe in the headband to maximize the detected ultrasound return. The headband electronics does not change or interfere with the transmitted ultrasound. Except for the servo motor controller added to the circuitry of the EMS9U range and the software added to control it, and the modifications to the INNSO Tange and the servo motor controlled probe, the EMS9U range and the EMS- 9UA are identical internally and functionally. The probes are identical to those cleared in K060112.
{2}------------------------------------------------
K092164
P. 3 of 5
... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
.
Hence the EMS-9UA is substantially equivalent to the EMS9U range of Trans Cranial Dopplers.
Product Comparison Chart
.
:
.
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
| Parameters | EMS9U range | EMS-9UA | SPENCER TCD 100M MARC 600 |
|---|---|---|---|
| 510k Number | K060112 | K092164 | K002533 |
| Transducer Frequency | 2,4, and 8 mHz | 2,4, and 8 mHz | 2 mHz |
| Frequency Spectrum | FFT256/512 dots | FFT256/512 dots | FFT256/512 dots |
| Frequency Ranges | 1 to 16 mHz | 1 to 16 mHz | 2 mHz |
| Depth Measurement | 5 to 136 mm | 5 to 136 mm | 5 to 140 mm |
| Gain | 0 to 40 dB | 0 to 40 dB | 0 to 40 dB |
| Clinical ApplicationFor 2 mHz | |||
| PWD | Opthalmic, Adult Cephalic& Peipheral Vascular | Opthalmic, Adult Cephalic& Peipheral Vascular | Opthalmic, Adult Cephalic |
| CWD | Opthalmic & PeripheralVascular | Opthalmic & PeripheralVascular | |
| Clinical ApplicationFor 4 mHz and 8 mHzCWD | Peripheral Vascular | Peripheral Vascular | |
| Head Frame | Bilateral Probes AdjustedManually | Bilateral Probes AdjustedManually and/ormatically | Bilateral Probes AdjustedManually |
{3}------------------------------------------------
4. PERFORMANCE STANDARDS
No performance standards have been established for the EMS9UA Transcranial Doppler under section 514 of the Federal Food and Drug Act. However the EMS9UA Transcranial Doppler has been designed to meet the following standards:
UL 2601-1 Safety Requirements for Medical Equipment AIUM/NEMA UD 2 Standard for Real Time Display of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment AIUM/NEMA UD 3 Standard for Real Time Display of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment IEC 1157 Declaration of Acoustic Power IEC60601-1-2 IEC60601-2-37
5. INDICATIONS FOR USE
Indications for Use:
The EMS9UA Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
I ) For the measurement of cerebral artery blood velocities to determine the presence of hemodynamically significant deviations from normal values 2) To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the
vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.
The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal.
The EMS9UA Transcranial Doppler is intended for use during:
a) Diagnostic exams
b) Surgical interventions
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
6. CONTRA-INDICATIONS
None known at this time.
7. COMPARISON TO PREDICATE DEVICES
The EMS9UA Transcranial Doppler has the same device characteristics as the approved predicate device listed above with the commonality of ultrasound transducers, priprively of operation, and display of blood flow waveforms.
{4}------------------------------------------------
8. TEST DATA
The EMS9UA Transcranial Doppler with Robotic Probe Headband has been subjected to extensive safety, performance testing, and validation before release. Final testing of the EMS9UA included various performance tests designed to ensure that the device met all of its functional specifications. Safety tests have been performed to ensure the device complies with applicable industry and safety standards.
A clinical trial involving 100 patients was conducted comparing the EMS9UA Transcranial Doppler with Robotic Headband with the Spencer Technologies Marc 600 predicate headband and was found to be safe and effective.
The Model EMS9UA Transcranial Doppler device labeling includes instructions for safe and effective use, warnings, cautions and guidance for use. It has therefore shown to be safe and effective.
9. LITERATURE REVIEW
A review of the literature pertaining to the safety of the EMS9UA Transcranial Doppler has been conducted and appropriate safeguards have been incorporated in the design of the EMS9UA Transcranial Doppler.
10. CONCLUSIONS
The conclusion drawn from these tests is that the EMS9UA Transcranial Vascular Doppler with Robotic Probe Headband and it's transducers is substantially equivalent in safety and efficacy to the predicate devices listed in the comparison table above
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the eagle. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY 1 3 2010
Shenzhen Delicate Electronics Co., Ltd. % Mr. William Stern Official Correspondent Multigon Industries, Inc. 1 Odell Plaza YONKERS NY 10701
Re: K092164
Trade/Device Name: EMS9UA Transcranial Doppler Robotic Probe Headband Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: OQQ, IYN and ITX Dated: April 19, 2010 Received: April 21, 2010
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the EMS9UA Transcranial Doppler Robotic Probe Headband, as described in your premarket notification:
Transducer Model Number
| 2 MHz |
|---|
| 4 MHz |
| 8 MHz |
{6}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
If you have any questions regarding the content of this letter, please contact Brendan O'Leary at (301) 796-6898.
Sincerely yours
Donald Trump
Donald St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
Indications for Use
510(k) Number (if known): K092164
Device Name: EMS9UA Transcranial Doppler with Robotic Probe Headband
Indications for Use:
The EMS9UA Transcranial Doppler Ultrasound System is intended for use as a diagnostic ultrasound fluid flow analysis system:
-
For the measurement of cerebral artery blood velocities to determine the presence of hemodynamically significant deviations from normal values
-
To assess arterial cerebral blood flow for the occurrence of micro embolic signals. Vessels intended for observation include, but are not limited to the middle, anterior and posterior cerebral arteries, via the temporal windows, the vertebral mid basilar arteries via the foramen magnum and the ophthalmic artery and intracranial internal carotid artery via the eye.
The Roboprobe Headband facilitates monitoring use by its ability to track the Doppler signal, The EMS9UA Transcranial Doppler is intended for use during:
a) Diagnostic exams
b) Surgical interventions
The device is not intended to replace other means of evaluating vital patient physiological processes, is not intended to be used in fetal applications, and is not intended to be used inside the sterile field.
It is to be used by trained medical personnel in hospitals, clinics and physicians offices by prescription or doctor's orders. --
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .
Concurrence of CDRH, Office of Device Evaluation (ODE)
Robattzeckey
K092164
(División Sign-om)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety
{8}------------------------------------------------
APPENDIX G: Diagnostic Ultrasound Indications for use for whole
SYSTEM: EMS9UA TRANSCRANIAL DOPPLER K092164
INTENDED USE: diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| TRANSDUCER | CLINICAL APPLICATION | MODE OF OPERATION | PREVIOUSLY |
|---|---|---|---|
| FREQUENCY | CLEARED | ||
| 2 MHZ | OPTHALMIC | PWD | K060112 |
| 2 MHZ | ADULT CEPHALIC | PWD | K060112 |
| 2 MHZ | PERIPHERAL VESSEL | PWD | K060112 |
| 4 MHZ | PERIPHERAL VESSEL | CWD | K060112 |
| 8 MHZ | PERIPHERAL VESSEL | CWD | K060112 |
PWD= PULSED WAVE DOPPLER
CWD=CONTINUOUS WAVE DOPPLER
(Division Sign-Off Division of Radiological Devices ice Evaluation and Safety Office of In Vitro
510K Kag2164
K092164
P. 2 of 5
{9}------------------------------------------------
Contains Nonbinding Recommendations Appendix G
Appendix G: Example Diagnostic Ultrasound Indications For Use Format
System: EMS9UA_Transcranial_Doppler
Transducer: 2 mHz
Previously cleared under K060112 May 5 2006
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track 1 Only) | Specific(Tracks 1 & 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | Other*(Specify) |
| Ophthalmic | Ophthalmic | P | ||||||
| Fetal | ||||||||
| Abdominal | ||||||||
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro) | ||||||||
| Laparoscopic | ||||||||
| FetalImaging& Other | Pediatric | |||||||
| Small Organ (Specify) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | P | |||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skeletal(Conventional) | ||||||||
| Musculo-skeletal(Superficial) | ||||||||
| Intravascular | ||||||||
| Other (Specify) | ||||||||
| Cardiac Adult | ||||||||
| Cardiac | Cardiac Pediatric | |||||||
| Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | ||||||||
| Intra-cardiac | ||||||||
| Other (Specify) | ||||||||
| PeripheralVessel | Peripheral vessel | P | ||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix
- Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging Haymonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
(Division Sign-Off)
510K
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation an
K09 2/64
P. 3 of 5
ನಿ ಒ
{10}------------------------------------------------
Contains Nonbinding Recommendations Appendix G
Appendix G: Example Diagnostic Ultrasound Indications For Use Format
System: Transducer:
EMSQUA 4-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Intended blagnostics intrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General(Track 1 Only) | Specific(Tracks 1 & 3) | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | Other*(Specify) |
| Ophthalmic | Ophthalmic | |||||||
| FetalImaging& Other | Fetal | |||||||
| Abdominal | ||||||||
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Small Organ (Specify) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skeletal(Conventional) | ||||||||
| Musculo-skeletal(Superficial) | ||||||||
| Intravascular | ||||||||
| Other (Specify) | ||||||||
| Cardiac | Cardiac Adult | |||||||
| Cardiac Pediatric | ||||||||
| Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | ||||||||
| Intra-cardiac | ||||||||
| Other (Specify) | ||||||||
| PeripheralVessel | Peripheral vessel | |||||||
| Other (Specify) | P |
N = new indication; P = previously cleared by FDA; E = added under this appendix * Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging
510K
(Division Sign-Off) Division of Radiological Devic Office of In Vitro Diagnostic Device Evaluation
K092164
P. 4 of 5
{11}------------------------------------------------
Contains Nonbinding Recommendations Appendix G
Appendix G: Example Diagnostic Ultrasound Indications For Use Format
EMS9UA System:
Transducer:
-8_mHz
Previously_Cleared under K060112 May 5 2006
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | |||||||
|---|---|---|---|---|---|---|---|---|
| General | Specific | B | M | PWD | CWD | ColorDoppler | Combined(Specify) | Other*(Specify) |
| (Track 1 Only) | (Tracks 1 & 3) | |||||||
| Ophthalmic | Ophthalmic | |||||||
| FetalImaging& Other | Fetal | |||||||
| Abdominal | ||||||||
| Intra-operative (Specify) | ||||||||
| Intra-operative (Neuro) | ||||||||
| Laparoscopic | ||||||||
| Pediatric | ||||||||
| Small Organ (Specify) | ||||||||
| Neonatal Cephalic | ||||||||
| Adult Cephalic | ||||||||
| Trans-rectal | ||||||||
| Trans-vaginal | ||||||||
| Trans-urethral | ||||||||
| Trans-esoph. (non-Card.) | ||||||||
| Musculo-skeletal(Conventional) | ||||||||
| Musculo-skeletal(Superficial) | ||||||||
| Intravascular | ||||||||
| Other (Specify) | ||||||||
| Cardiac | Cardiac Adult | |||||||
| Cardiac Pediatric | ||||||||
| Intravascular (Cardiac) | ||||||||
| Trans-esoph. (Cardiac) | ||||||||
| Intra-cardiac | ||||||||
| Other (Specify) | ||||||||
| PeripheralVessel | Peripheral vessel | P | ||||||
| Other (Specify) |
N = new indication; P = previously cleared by FDA; E = added under this appendix
- Examples of other modes of operation may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, and Color Velocity Imaging,
(Division Sign-Off)
510K
Division of Radiological Dev Office of In Vitro Division of Hadlollogical Devices
Office of In Vitro Diagnostic Device Evaluation
K092164
P. 5 of 5
N/A