K Number
K092081

Validate with FDA (Live)

Device Name
QXLINK
Manufacturer
Date Cleared
2009-07-21

(12 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The QXLink is a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. The software components provide functions for performing operations related to image manipulation, enhancement, compression or quantification. And Images may be acquired from imaging devices such as CR, CT, MR and other devices. This device is not intended for mammographic operations.

Device Description

QXLink is PACS (Picture Archiving and Communication System) software for radiologist and physician doctor. QXLink enables images such as x-ray to be stored electronically and viewed on screens. QXLink conforms to the DICOM standards to allow the sharing of medical information with other digital Imaging system. Also, QXLink is a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. The software components provide functions for performing operations related to image manipulation, enhancement or quantification. And images may be acquired from imaging devices such as CR, CT, MR and other devices. QXLink consists of QXLink Server and QXLink Viewer and the main functions of QXLink are as follows.

AI/ML Overview

This 510(k) submission (K092021) for the device "QXLink" (Picture Archiving and Communication System - PACS) does not contain information regarding a study to prove the device meets specific acceptance criteria.

The document is a 510(k) Summary and the associated FDA clearance letter, which focus on demonstrating substantial equivalence to predicate devices rather than presenting the results of a performance study with defined acceptance criteria.

Therefore, I cannot provide the requested table and information about a study proving acceptance criteria. The 510(k) summary only outlines the device's functions and states its substantial equivalence to other PACS devices.

Here's an explanation of why the requested information is absent from this specific document:

  • Type of Submission: This is a 510(k) notification, which is a premarket submission made to FDA to demonstrate that the device to be marketed is at least as safe and effective as a legally marketed device (predicate device).
  • Focus of 510(k): The primary focus of a 510(k) is to establish "substantial equivalence" to a predicate device. This is often achieved by comparing technical characteristics, indications for use, and safety/performance data (if available and deemed necessary by the FDA for the particular device type).
  • PAC System Nature: For a Picture Archiving and Communication System (PACS) like QXLink, the "performance" typically relates to its ability to correctly accept, transfer, display, store, and process medical images according to industry standards (like DICOM) and its advertised functionalities. These functionalities are usually validated through various software testing methods (e.g., unit testing, integration testing, system testing, user acceptance testing) which are part of a quality management system, but the detailed results of such validation are generally not included in the public 510(k) summary.
  • Lack of Clinical Study: A PACS system, particularly one not intended for mammography, often doesn't require a clinical study with human patients to demonstrate its performance for 510(k) clearance, especially if it's functionally similar to already cleared predicate devices. The "performance" in this context is more about software functionality and adherence to standards than diagnostic accuracy from a clinical perspective.

In summary, the provided document details the description of the device, its intended use, and its comparison to predicate devices for the purpose of 510(k) clearance, but it does not include a specific study or explicit acceptance criteria for its performance in the way a diagnostic AI device would.

{0}------------------------------------------------

K092021

510(k) Summary

JUL 21 2009

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date: May 10, 2009

  1. Company and Correspondent making the submission:

Name - Vieworks Co., Ltd.

Address - #604, Suntechcity 2, 307-2, Sangdaewon-dong, Jungwon-gu, Seongnam-city, Gyeonggi-do, 462-806 South Korea

Telephone - +82-70-7011-6190

Fax - +82-31-737-4954

Contact - Mr. BongGu, Lee

Internet - http://www.vieworks.com

2. Device :

Trade/proprietary name: QXLink
Common Name: Picture archiving and communication system(PACS)
Classification Name: Imaging Processing System, Radiological

3. Predicate Device :

Manufacturer: Mediface Co., Ltd.
Device: MEDIFACE PACS
510(k) Number: K010259(Decision Date - Jan. 29. 2001)
Manufacturer: EMSOMA Co., Ltd.
Device: PetaVision
510(k) Number: K083555(Decision Date - Dec. 16. 2008)
  1. Classifications Names & Cltations :

21CFR 892.2050, LLZ, Imaging Processing System, Radiological, Class2

Vieworks Co., Ltd.

00000017

{1}------------------------------------------------

5. Description :

5.1 General

QXLink is PACS (Picture Archiving and Communication System) software for radiologist and physician doctor. QXLink enables images such as x-ray to be stored electronically and viewed on screens. QXLink conforms to the DICOM standards to allow the sharing of medical information with other digital Imaging system.

Also, QXLink is a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. The software components provide functions for performing operations related to image manipulation, enhancement or quantification. And images may be acquired from imaging devices such as CR, CT, MR and other devices.

QXLink consists of QXLink Server and QXLink Viewer and the main functions of QXLink are as follows.

5.2 Main Function

5.2.1 Image Archive

QXLink Server stores DICOM images that are transmitted in the form of DICOM protocol from each of various modalities. And it saves the information of that image into Database.

5.2.2 Query/Retrieve

QXLink Viewer can search the information of specific DICOM image through DICOM protocol. It queries that information using the several matching attributes (patient name, patient id, sex, study description, study id, accession number, and study date range). It requests QXLink Server to transfer the specific DICOM image:

5.2.3 Image display

QXLink Viewer displays DICOM image on the monitor as if film is read at the film view box. It has several functions to help view images, which are zoom, pan, windowllevel, invert and so on. It can support that a doctor reports and documents the diagnostic conclusion to the Structured Report file (DICOM format). It can also export DICOM images to CD/DVD disk

Vieworks Co., Ltd.

0000018

{2}------------------------------------------------

510(k) Submission - QXLink

and print them to DICOM/Paper printer.

    1. Indication for use :
      The QXLink is a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. The software components may provide functions for performing operations related to image manipulation, enhancement, compression or quantification. And Images may be acquired from imaging devices such as CR, CT, MR and other devices. This device is not intended for mammographic operations.
    1. Comparison with predicate device :
      Vieworks Co., Ltd., believes that the Picture archiving and communication system(QXLink) is substantially equivalent to the MEDIFACE PACS of Mediface Co., Ltd. and the PetaVision of EMSOMA Co., Ltd..
    1. Conclusions :
      In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Vieworks Co., Ltd. concludes that The QXLink is safe and effective and substantially equivalent to predicate devices as described herein.
    1. Vieworks Co., Ltd. will update and include in this summary any other information deemed seasonably necessary by the FDA.
      END

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the caduceus. The logo is black and white.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

L.21 2009

Vieworks, Co., Ltd. % Mr. Marc M. Mouser CAS Manager II / Office Coordinator Underwriters Laboratories, Inc. 2600 NW Lake Road CAMAS WA 98607

Re: K092081

Trade/Device Name: Picture archiving and communication system (PACS)/QXLink Regulation Number: 21 CFR 892.2050

Regulation Name: Picture archiving and communications system Regulatory Class: II

Product Code: LLZ Dated: June 8, 2009 Received: July 9, 2009

Dear Mr. Mouser:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical

{4}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable; the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Lucille M. Morris

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Indications for Use

0920 8 510(k) Number (if known):

Device Name: Picture archiving and communication system (PACS)/ QXLink

Indications for Use:

The QXLink is a device that provides one or more capabilities relating to the acceptance, transfer, display, storage, and digital processing of medical images. The software components provide functions for performing operations related to image manipulation, enhancement, compression or quantification. And Images may be acquired from imaging devices such as CR, CT, MR and other devices. This device is not intended for mammographic operations.

Prescription Use
(Part 21 CFR 801 Subpart D)

Over-The-Counter Use (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

Larry M. Khan

(Division Division of Reproductive. Abdominal and Radiological Devic 510(k) Number

Page 1 of 1

00000016

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).