(267 days)
The AEGIS and AEGIS II Spinal Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
When used as an anterior screw fixation system, The AEGIS I and AEGIS II Spinal Systems are indicated for patients with deqenerative disc disease which is defined as back pain of the discogenic origin with degeneration of the disc confirmed by history and radiographic studies, Spondylolisthesis, fracture, spinal stenosis, spinal deformities such as scoliosis, lordosis, turnor, pseudoarthrosis, or revision of failed fusion attempts.
The AEGIS® Pedicle Screw System is comprised of pedicle screw with diameters from 4.5mm to 8.5mm with increments of 1mm and length ranging from 20mm to 55 mm with increments of 5mm, a sleeve of standard diameter of 14 mm, and a set screw of M10XP1.0 & rods with standard diameter of 6 mm. All the components are manufactured from medical grade titanium alloy (Ti6Al4V-Eli).
The mono axial pedicle screw is used as an adjunct to spinal fusion surgery, provides a means of gripping a spinal segment. The screws themselves do not fixate the spinal segment, but act as firm anchor points that can then be connected with a rod. The screws are placed at two or three consecutive spine segments (e.g. lumbar segment 4 and 5) and then a short rod is used to connect the screws. This construct prevents motion at the segments that are being fused.
The Corentec® AEGIS II® Spinal system consists of various Pedicle Screws (mono / poly) with standard and guided type designs, Rod and Rod Link with a set screw, the assembly of which is intended to provide temporary stabilization following surgery to fuse the spine. This system is designed on the basis of long standing spinal technology which is already in the market for more than few decades.
The provided text describes mechanical performance testing for the AEGIS & AEGIS II Spinal Systems. This is not a study involving AI, human readers, or image analysis, but rather physical stress tests on the device itself. Therefore, many of the requested categories are not applicable.
Here's the information that can be extracted from the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria (Test Method) | Reported Device Performance |
|---|---|
| Static test (ASTM F1717-04): Tension | Demonstrated equivalence to legally marketed predicate devices. |
| Static test (ASTM F1717-04): Compression | Demonstrated equivalence to legally marketed predicate devices. |
| Static test (ASTM F1717-04): Torsion | Demonstrated equivalence to legally marketed predicate devices. |
| Dynamic test (ASTM F1717-04): Fatigue | Demonstrated equivalence to legally marketed predicate devices. |
| Axial pull-out strength test (ASTM 543-07) | Demonstrated equivalence to legally marketed predicate devices. |
| Axial Gripping Capacity test (ASTM F1798-97) | Demonstrated equivalence to legally marketed predicate devices. |
2. Sample size used for the test set and the data provenance
The document does not specify the exact sample size for each mechanical test. The "data provenance" in this context refers to the manufacturing and testing of the physical device components. The tests were performed on the AEGIS and AEGIS II Spinal Systems themselves. The country of origin for the submitter is Korea. The tests are prospective in the sense that they are conducted on newly manufactured devices to ensure they meet standards.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable. This was mechanical testing of a medical device, not a human reader study or clinical evaluation requiring expert interpretation of data. The ground truth refers to the physical properties and performance of the device under stress, as measured by standard engineering tests.
4. Adjudication method for the test set
Not applicable. This was mechanical testing against established ASTM standards, not a diagnostic or interpretive study requiring adjudication of human readings.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. This was a mechanical performance study of a spinal implant system, not a study involving human readers or AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. This was mechanical testing of a physical medical device, not an algorithm.
7. The type of ground truth used
The ground truth used was the performance of the device under specific mechanical stress tests, compared against the performance of "legally marketed predicate devices" as defined by established ASTM (American Society for Testing and Materials) standards (F1717-04, 543-07, F1798-97).
8. The sample size for the training set
Not applicable. This was not a machine learning study with a training set.
9. How the ground truth for the training set was established
Not applicable. There was no training set.
{0}------------------------------------------------
510(k) PREMARKET NOTIFICATION CORENTEC CO. LTD. AEGIS & AEGIS II SPINAL SYSTEMS
510(k) Summary
APR - 2 2010
Sponsor
Submitter: Corentec Co., Ltd. 247 Giro-ri, Ipjang-myeon, cheonan-si, Chungchongnam-do, 330-822 Korea Telephone: 82-41-585-7114 Fax: 82-41-585-7113
| Contact Person: | Robert Schiff |
|---|---|
| Telephone: | 973-227-1830 |
Date Prepared: June 15, 2009
Device Name
| Proprietary Name: | AEGIS & AEGIS II Spinal Systems |
|---|---|
| Common/Usual Name: | Pedicle Screw Spinal System |
| Classification Name: | Pedicle Screw Spinal System per 21 CFR §888.3070 |
| Spondylolisthesis Spinal Fixation Device System (MNH) per 21 CFR §888.3060 | |
| Spinal Intervertebral Body Fixation Orthosis (KWQ) per 21 CFR §888.3060Class II | |
| Classification Code: | MNI, MNH, KWQ |
| Predicate Devices |
| Predicate Device #1: | XIA Spinal System and XIA 4.5 Spinal System (K060361) |
|---|---|
| Predicate Device #2: | XIA Titanium and 4.5 Spinal System (K060979) |
| Predicate Device #3: | GGS Pedicle Screw System (K053573) |
| Predicate Device #4: | U & I, Optima, Spinal System (K031585) |
Pg 1 of 3
{1}------------------------------------------------
510(k) PREMARKET NOTIFICATIO CORENTEC CO. I TO AEGIS & AEGIS II SPINAL SYSTEMS
Device Description
The AEGIS® Pedicle Screw System is comprised of pedicle screw with diameters from 4.5mm to 8.5mm with increments of 1mm and length ranging from 20mm to 55 mm with increments of 5mm, a sleeve of standard diameter of 14 mm, and a set screw of M10XP1.0 & rods with standard diameter of 6 mm. All the components are manufactured from medical grade titanium alloy (Ti6Al4V-Eli).
The mono axial pedicle screw is used as an adjunct to spinal fusion surgery, provides a means of gripping a spinal segment. The screws themselves do not fixate the spinal segment, but act as firm anchor points that can then be connected with a rod. The screws are placed at two or three consecutive spine segments (e.g. lumbar segment 4 and 5) and then a short rod is used to connect the screws. This construct prevents motion at the segments that are being fused.
The Corentec® AEGIS II® Spinal system consists of various Pedicle Screws (mono / poly) with standard and guided type designs, Rod and Rod Link with a set screw, the assembly of which is intended to provide temporary stabilization following surgery to fuse the spine. This system is designed on the basis of long standing spinal technology which is already in the market for more than few decades.
Device Intended Use
The AEGIS and AEGIS II Spinal Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondvlolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
When used as an anterior screw fixation system, The AEGIS I and AEGIS II Spinal Systems are indicated for patients with deqenerative disc disease which is defined as back pain of the discogenic origin with degeneration of the disc confirmed by history and radiographic studies, Spondylolisthesis, fracture, spinal stenosis, spinal deformities such as scoliosis, lordosis, turnor, pseudoarthrosis, or revision of failed fusion attempts.
Materials
All products are made of titanium (Ti-6Al-4V ELI, ASTM F136-98) approved for medical use.
{2}------------------------------------------------
510(k) PREMARKET NOTIFICATIO CORENTEC CO. LTD AEGIS & AEGIS II SPINAL SYSTEMS
Performance Testing
Mechanical testing performed with the AEGIS and AEGIS II Spinal Systems demonstrated equivalence of the device to legally marketed predicate devices.
Mechanical test reports were completed for the following test methods:
- Static test: Tension, compression and torsion test report (ASTM F1717-04)
- Dynamic test: Fatigue test report (ASTM F1717-04)
- Axial pull-out strength test (ASTM 543-07)
- Axial Gripping Capacity test (ASTM F1798-97)
Substantial Equivalence
Corentec Co Ltd. has submitted information to demonstrate that, for the purposes of FDA's regulation of medical devices, the AEGIS and AEGIS II Spinal Systems are substantially equivalent in indications and design principles to predicate devices that have been determined by FDA to be substantially equivalent to preamendment devices.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is positioned above the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA", which are arranged in a circular pattern around the eagle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Corentec Co., Ltd % Schiff & Company, Inc. Mr. Robert Schiff 1129 Bloomfield Avenue West Caldwell, New Jersey 07006
APR - 2 2010
Re: K092076
Trade/Device Name: Aegis and Aegis II Spinal Systems Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II -Product Code: MNI, MNH, KWO Dated: March 31, 2010 Received: April 1, 2010
Dear Mr. Schiff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542, of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Robert Schiff
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHQffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark H. Milliman
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) PREMARKET NOTIFICATION CORENTEC CO. LTD. AEGIS & AEGIS II SPINAL SYSTEMS
INDICATIONS FOR USE STATEMENT
510(k) Number (if known): Device Name:
Not assigned yet AEGIS and AEGIS II Spinal Systems
Indications for Use:
The AEGIS and AEGIS II Spinal Systems are intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis).
When used as an anterior screw fixation system, The AEGIS and AEGIS II Spinal Systems are indicated for patients with degenerative disc disease which is defined as back pain of the discogenic origin with degeneration of the disc confirmed by history and radiographic studies, Spondylolisthesis, fracture, spinal deformities such as scoliosis, kyphosis, lordosis, tumor, pseudoarthrosis, or revision of failed fusion attempts.
Prescription Use
(Per 21 CFR 801 Subpart D)
Over-The-Counter Use_
(21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
(Division Sign-Off)
Division of Surgical, Orthopedic,
and Restorative Devices
SUBMITTED BY SCHIFF & COMPANY, WEST CALDWELL, NJ
510(k) Number K092076 14
Pg 1 of 1