K Number
K091926

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-12-22

(175 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Choice Spine Anterior Cervical Plate (ACP) System is intended for anterior screw fixation of the cervical spine (levels C2 to C7) and is designed to provide stabilization as an adjunct to spinal fusion at these levels. Indications for the use of this device include: failed previous fusion, pseudoarthrosis, tumor, deformity, spinal stenosis, trauma, spondylolisthesis, or degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies.

Device Description

The Choice Spine Anterior Cervical Plate (ACP) System is an anterior cervical spinal fixation system consisting of plates and screws. The plates are available in multiple lengths to accommodate single or multi-level surgeries and different anatomies.

AI/ML Overview

The provided text is related to a 510(k) premarket notification for a medical device called the "Choice Spine Anterior Cervical Plate (ACP) System." This document primarily focuses on demonstrating substantial equivalence to a predicate device based on intended use, indications, materials, and performance, rather than presenting a study proving a device meets specific performance acceptance criteria with numerical outcomes.

Therefore, many of the requested elements (like sample sizes, expert qualifications, ground truth establishment, MRMC studies, standalone performance, and detailed acceptance criteria tables) are not present in the provided text.

Here's an analysis based on the information available:

1. A table of acceptance criteria and the reported device performance

The document does not specify quantitative acceptance criteria or detailed reported device performance in the context of a clinical or analytical study. Instead, it states that "Documentation was provided that demonstrates the Choice Spine ACP System to be substantially equivalent to a previously cleared device. The substantial equivalence is based upon equivalence in intended use, indications, anatomic sites, primary material of manufacture, and performance." This implies that the device's performance was deemed equivalent to the predicate, likely through design specifications, material testing, and potentially bench testing, but no specific performance metrics or acceptance thresholds are numerically reported.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

Not applicable. The document describes a regulatory submission for substantial equivalence, not a clinical or analytical study with a test set of data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable. No "ground truth" establishment in the context of a diagnostic or AI device is discussed.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. An AI component is not mentioned, and no MRMC study is described.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a spinal implant, not a diagnostic or AI algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

Not applicable.

8. The sample size for the training set

Not applicable.

9. How the ground truth for the training set was established

Not applicable.

In summary, the provided document is a 510(k) summary for a physical medical device (an anterior cervical plate system) seeking market clearance based on substantial equivalence to a predicate device. It does not contain the kind of detailed study information (acceptance criteria, sample sizes, expert involvement, ground truth, MRMC studies) that would be relevant for a diagnostic device or AI algorithm.

{0}------------------------------------------------

K091926 (1/2)

DEC 22 2009

510(k) Summary

Sponsor:

Choice Spine, LP 306 Erin Drive Knoxville, TN 37919 ph: 865.246.3333 fax: 865.588.4045

Contact:

G. Todd Hawkins Director of Regulatory Affairs/Quality Assurance

Trade Name:

Choice Spine Anterior Cervical Plate (ACP) System

Common Name:

Anterior Cervical Plate, Anterior Cervical Spinal Fixation System

Classification Name:

888.3060 - Spinal intervertebral body fixation orthosis

Device Product Code:

KWQ

Device Description:

The Choice Spine Anterior Cervical Plate (ACP) System is an anterior cervical spinal fixation system consisting of plates and screws. The plates are available in multiple lengths to accommodate single or multi-level surgeries and different anatomies.

Intended Use:

The Choice Spine Anterior Cervical Plate (ACP) System is intended for anterior screw fixation of the cervical spine (levels C2 to C7) and is designed to provide stabilization as an adjunct to spinal fusion at these levels. Indications for the use of this device include: failed previous fusion, pseudoarthrosis, tumor, deformity, spinal stenosis, trauma, spondylolisthesis, or degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies.

WARNING: The Choice Spine ACP System is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.

Page 1 of 2

{1}------------------------------------------------

Kc9i926 (²/₂)

510(k) Summary (continued)

Materials:

The Choice Spine Anterior Cervical Plate (ACP) System plates and screws are manufactured from titanium alloy (Ti6Al4V ELI; according to ASTM F136). The plates incorporate a screw-retention mechanism ("clip"), which is manufactured from wrought nickel-titanium alloy (NiTi; according to ASTM F2063).

Substantial Equivalence:

Documentation was provided that demonstrates the Choice Spine ACP System to be substantially equivalent to a previously cleared device. The substantial equivalence is based upon equivalence in intended use, indications, anatomic sites, primary material of manufacture, and performance.

Page 2 of 2.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Choice Spine, LP % G. Todd Hawkins 306 Erin Drive Knoxville, TN 37919

DEC 22 0000

Re: K091926

Trade/Device Name: Choice Spine Anterior Cervical Plate (ACP) System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: December 16, 2009 Received: December 17, 2009

Dear Mr. Hawkins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS)

{3}------------------------------------------------

Page 2 -- Mr. G. Todd Hawkins

regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to . http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/defay4.htm.

Sincerely yours,

erely yours,

for Peter N. D.C
k N. Melkerson

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement

510(k) Number (if known): $691924

Device Name: Choice Spine Anterior Cervical Plate (ACP) System

Indications for Use:

The Choice Spine Anterior Cervical Plate (ACP) System is intended for anterior screw fixation of the cervical spine (levels C2 to C7) and is designed to provide stabilization as an adjunct to spinal fusion at these levels. Indications for the use of this device include: failed previous fusion, pseudoarthrosis, tumor, deformity, spinal stenosis, trauma, spondylolisthesis, or degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies.

WARNING: The Choice Spine Anterior Cervical Plate (ACP) System is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

signature

Page 1 of _1

Over-The-Counter Use

(21 CFR 801 Subpart C)

(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices

Prescription Use X

(Part 21 CFR 801 Subpart D)

510(k) Number K091926
Page 1 of 1

N/A