K Number
K091911

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-07-21

(26 days)

Product Code
Regulation Number
892.5710
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Bolus Compensators are used by radiation therapy professionals for the treatment of cancer patients. They are designed by radiation therapy professionals for a unique patient and are intended to modify the shape of a beam from a radiation therapy source. Each Bolus Compensator must be validated and approved by the radiation therapy professional prior to use on a patient.

Device Description

.decimal's Bolus Compensator manufacturing service manufactures the solid Bolus Compensators for intensity modulation of external beam radiation therapy. The Bolus Compensators are designed by the customer's treatment planning system to block radiation from hitting critical structures and healthy tissue while guiding the radiation to the targeted area.

AI/ML Overview

This document is a 510(k) Premarket Notification summary for a device called "Bolus Compensator". The key information related to acceptance criteria and supporting studies is as follows:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly present a table of acceptance criteria with corresponding performance metrics using an established format. Instead, it states that "Clinically oriented validation test cases were written and executed in house .decimal personnel including Board Certified Medical Physicists where Bolus Compensators was deemed fit for clinical use." This implies that the acceptance criteria were met by these in-house validation tests. The "intended use" section suggests the implicit performance criteria:

  • Manufactures solid Bolus Compensators for intensity modulation of external beam radiation therapy.
  • Designed by the customer's treatment planning system to block radiation from hitting critical structures and healthy tissue.
  • Guides radiation to the targeted area.
  • Must be validated and approved by the radiation therapy professional prior to use on a patient.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for Test Set: The document mentions "Clinically oriented validation test cases were written and executed in house." It does not specify the number of test cases (i.e., sample size).
  • Data Provenance: The tests were "executed in house" by .decimal personnel. This indicates the data is retrospective and generated internally for validation purposes. The country of origin would be the United States, given the company's address is in Sanford, Florida, and the submission is to the FDA.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

  • Number of Experts: The document states "including Board Certified Medical Physicists." It does not specify the exact number of medical physicists involved.
  • Qualifications of Experts: The experts were "Board Certified Medical Physicists." No further details on their experience (e.g., years of experience) are provided.

4. Adjudication Method for the Test Set

The document simply states that the Bolus Compensators "was deemed fit for clinical use" based on the in-house validation. It does not describe a formal adjudication method (like 2+1 or 3+1 consensus). The "Board Certified Medical Physicists" were involved in deeming the product fit, implying their expertise formed the basis of the determination, but not a specific multi-reader adjudication process.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not performed. The document explicitly states: "Clinical testing was not performed as part of the development of this product. Clinical testing is not advantageous in demonstrating substantial equivalence or safety and effectiveness of the device since testing can be performed such that no human subjects are exposed to risk." This implies that no study comparing human readers with and without AI assistance was conducted.

6. If a Standalone Study (Algorithm Only Without Human-in-the-Loop Performance) was Done

The device is a "Bolus Compensator manufacturing service" and the physical bolus compensator itself. It is a physical device, not an AI algorithm. Therefore, the concept of a "standalone (algorithm only) performance" study does not apply here. The validation focused on the physical characteristics and functional performance of the manufactured compensators.

7. The Type of Ground Truth Used

The ground truth was established by "Board Certified Medical Physicists" deeming the manufactured Bolus Compensators "fit for clinical use." This suggests a form of expert judgment/consensus on the functional suitability and accuracy of the manufactured devices based on the "clinically oriented validation test cases." It is not based on pathology, outcomes data, or patient-level clinical results, as no clinical testing with human subjects was performed.

8. The Sample Size for the Training Set

This device is a physical product (.decimal's Bolus Compensator manufacturing service manufactures the solid Bolus Compensators), not an AI algorithm trained on data. Therefore, the concept of a "training set" is not applicable in the context of machine learning.

9. How the Ground Truth for the Training Set Was Established

As stated above, this is not an AI/machine learning device, so there is no training set or ground truth for a training set in the conventional sense. The "ground truth" equivalent would be the engineering specifications and performance expectations for the manufactured bolus compensators, which were assessed by medical physicists during an in-house validation process.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "decimal" in a stylized font, with a registered trademark symbol to the right of the word. Below the word "decimal" is the phrase "the benchmark for custom radiation therapy". The text is in black and the background is white.

K091911/

JUL 21 2009

Section 7 510(k) Summary or 510(k) Statement

Section 807.87 (h) A 510(k) Summary as described in Section 807.92 or a 510(k) statement as described in 807.93

Premarket Notification [510(k)] Summary as required by 21 CFR 807.92

Date summary was prepared:

June 2009

Submitter's Name:

.decimal, Inc. 121 Central Park Pl Sanford, Florida 32771

Contact Person:

Daniel L. Bennett Director of Quality and Regulatory Affairs Phone: 407-330-3300 407-322-7546 Fax: Email:dbennett(@dotdecimal.com

Device Name:

Bolus Compensator

Classification Name:

IXI 21 CFR892.5710 Class II

Predicate Device(s):

.decimal Range Compensator (K071078)

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "decimal" in a stylized font, with a small circle and line above the "l". Below the word "decimal" is the phrase "the benchmark for custom radiation therapy". The text is black and the background is white. The image appears to be a logo or advertisement for a company or product related to radiation therapy.

Intended Use:

.decimal's Bolus Compensator manufacturing service manufactures the solid Bolus Compensators for intensity modulation of external beam radiation therapy. The Bolus Compensators are designed by the customer's treatment planning system to block radiation from hitting critical structures and healthy tissue while guiding the radiation to the targeted area.

Summary of Technological Characteristics:

The device features of Bolus Compensators are similar to the predicate device (.decimal Range Compensators K071078 cleared in 2007). They both are designed by radiation therapy professionals for a unique patient and are intended to modify the shape of a beam from a radiation therapy source. Whereas the Bolus Compensator is made using machinable wax, the Range Compensator is made using acrylic, despite the difference in materials they both share very similar densities. Like Range Compensators Each Bolus Compensator must be validated and approved by the radiation therapy professional prior to use on a patient. The target population is identical and the use parameters are also very similar.

A detailed comparison can be found in section 11 of this submittal.

Summary of Clinical Testing:

Clinical testing was not performed as part of the development of this product. Clinical testing is not advantageous in demonstrating substantial equivalence or safety and effectiveness of the device since testing can be performed such that no human subjects are exposed to risk. Clinically oriented validation test cases were written and executed in house .decimal personnel including Board Certified Medical Physicists where Bolus Compensators was deemed fit for clinical use. A declaration of conformity to this requirement can be found in section 12 of this document.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a circular logo with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol consisting of three curved lines that resemble a person embracing another person. The logo is black and white and appears to be a seal or emblem.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 21 2009

Mr. Daniel L. Bennett Director of Quality and Regulatory Affairs .decimal, Inc. 121 Central Park Pl. SANFORD FL 32771

Re: K091911

Trade/Device Name: Bolus Compensator Regulation Number: 21 CFR 892.5710 · Regulation Name: Radiation therapy beam-shaping block Regulatory Class: II Product Code: IXI Dated: June 24, 2009 Received: June 25, 2009

Dear Mr. Bennett:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

he benchmark for custom radiation therapy

Section 6Indications for Use Statement
510(k) Number (if known)K091911

Device Name: Bolus Compensator

Indication for Use: Bolus Compensators are used by radiation therapy professionals for the treatment of cancer patients. They are designed by radiation therapy professionals for a unique patient and are intended to modify the shape of a beam from a radiation therapy source. Each Bolus Compensator must be validated and approved by the radiation therapy professional prior to use on a patient.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use(Per 21 CFR 801.109)
--------------------------------------------------------------------------------------------------

OR

Over-The-Counter Use________________
----------------------------------------

(Division Sign-Off)
Division of Reproductive, Abdominal and
Radiological Devices

510(k) NumberK091911
------------------------

N/A