K Number
K091833

Validate with FDA (Live)

Date Cleared
2009-10-16

(116 days)

Product Code
Regulation Number
890.5850
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The GEMORE GEM-TONE Abdominal Training System/model GEM-TONE 310PE/320PE/340PE/340PE/350PE are the electrical muscle stimulator intended for the following indication for use:

  • Strengthening, toning and firming of the abdomen region.
    The GEMORE GEM-TONE Abdominal Training System and its stimulation programs are not designed for injured or ailing muscles, and use on such muscles is contraindication.
Device Description

The GEM-TONE Abdominal Training System model GEM-TONE 310PE/ 320PE /330PE /340PE / 350PE is battery-operated programmable muscle stimulator intended to improve or facilitate muscle performance by applying an electrical current to electrodes, which is attached on abdomen region.
GEM-TONE Abdominal Training System model GEM-TONE 310PE / 320PE /330PE /340PE /350PE, consists mainly of three parts: the stimulator , electrodes, and support belts fitting for special parts of body. The stimulator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the specified region, so as to transmit this stimulus current to the region of abdomen for the following intended purposes:

  • Strengthening, toning and firming of the abdomen region.
    The GEMORE GEM-TONE Abdominal Training System and its stimulation programs are not designed for injured or ailing muscles, and use on such muscles is contraindication.
    To adequately locate the stimulation on the intended treatment area, the following support belt is provided together with the stimulator:
  • · The abdomen belt, which is capable of connecting to the both two output channels of stimulator.
    The stimulation mode for GEM-TONE Abdominal Training System includes several different operation modes as mentioned on the comparison table. These operation modes are generated from the software control by using the microprocessor as its main control unit.
AI/ML Overview

This submission describes a premarket notification (510(k)) for the GEM-TONE Abdominal Training System, a powered muscle stimulator. The purpose of the 510(k) is to demonstrate substantial equivalence to a predicate device, the Slendertone Flex (K010335).

Here's an analysis of the provided information regarding acceptance criteria and the supporting study:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state "acceptance criteria" in terms of performance metrics like sensitivity, specificity, or accuracy, as would be typical for diagnostic or AI-driven devices. Instead, the "acceptance criteria" for this muscle stimulator appear to be centered around compliance with recognized electrical safety and regulatory standards, and demonstrating equivalent functionality to the predicate device.

Acceptance Criteria (Implied)Reported Device Performance/Evidence
Electrical Safety and Performance StandardsCompliance to applicable voluntary standards including ANSI/AAMI. NS4-1985 (likely AAMI/ANSI NS4-1985 for transcutaneous electrical nerve stimulators, though not explicitly stated for this exact device type, it's a common standard for similar devices), EN 60601-1 (medical electrical equipment - general requirements for basic safety and essential performance), EN 60601-1-1 (collateral standard for safety requirements for medical electrical systems), and EN 60601-1-2 (collateral standard for electromagnetic compatibility).
Software VerificationSoftware verification has been carried out according to FDA software guidance.
Substantial Equivalence to Predicate Device (K010335)The device "has the same intended use and technological characteristics as the cleared device of Slendertone Flex (K010335)." "Verification tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted model could maintain the same safety and effectiveness as that of cleared device." This implies functional equivalence in muscle stimulation.
Intended Use"Strengthening, toning and firming of the abdomen region." This is identical to the implied intended use of the predicate device.

2. Sample Size for the Test Set and Data Provenance

The provided document does not specify any clinical studies or test sets with human participants that would typically involve a "sample size" for performance evaluation in the way an AI/diagnostic device would. The "verification tests" mentioned are likely technical and engineering tests to ensure compliance with electrical and software standards, rather than a clinical trial.

Therefore, information on:

  • Sample size used for the test set: Not applicable/not provided for clinical performance testing.
  • Data provenance (e.g., country of origin of the data, retrospective or prospective): Not applicable, as no clinical data studies are described.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not applicable as the submission focuses on demonstrating electrical safety, software compliance, and functional equivalence to a predicate device through non-clinical testing. There is no mention of a ground truth established by experts for a test set in the context of this 510(k) summary.

4. Adjudication Method for the Test Set

This information is not applicable as there is no described test set of clinical data with expert ground truth to be adjudicated.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study being performed, nor any effect size for human readers improving with or without AI assistance. This type of study is typically relevant for interpretative diagnostic devices, not for a muscle stimulator.

6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study

This concept is not applicable to the GEM-TONE Abdominal Training System described. This device is a powered muscle stimulator, and its functionality does not involve an "algorithm only" performance or human-in-the-loop interpretation in the way a diagnostic AI would. Its performance is directly tied to its electrical output and user interaction (attaching electrodes).

7. Type of Ground Truth Used

There is no mention of a specific "ground truth" (e.g., expert consensus, pathology, outcomes data) being established for clinical performance as no such clinical study is described. The "ground truth" for this submission refers to the established safety and effectiveness standards of the predicate device and the relevant electrical safety standards. The device's performance is measured against these technical and regulatory benchmarks.

8. Sample Size for the Training Set

This information is not applicable. The device is a muscle stimulator, not an AI/machine learning model that requires a "training set" of data. The software verification mentioned refers to adherence to good software engineering practices and FDA guidance, not to training an AI model.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable as there is no training set for an AI model.

{0}------------------------------------------------

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K091833

    1. Submitter's Identifications: Gemore Technology Co., Ltd. 11FL., NO. 29-5, Sec. 2. Chung Cheng E. RD., Tan Shui, Taipei Hsien, Taiwan
      OCT 1 6 2009

Contact: Mr.Boden S.P. Lai President & Official Correspondent

Date of Summary Preparation: Auqust 30,2009.

    1. Name of the Device:
Trade name:GEN-TONE Abdominal Training system; model GEM-TONE310PE/320PE/330PE/340PE/350PE
Common name:Powered Muscle Stimulator
Classification name:Stimulator, Muscle, Powered
Product Code:NGX

3. Information of the 510(k) Cleared Device (Predicate Device):

For this 510(k) submission, we compared our models to the following FDA cleared OTC devices:

  • K010335: Slendertone Flex, made by Bio-Medical Research Ltd.

4. Device Description:

The GEM-TONE Abdominal Training System model GEM-TONE 310PE/ 320PE /330PE /340PE / 350PE is battery-operated programmable muscle stimulator intended to improve or facilitate muscle performance by applying an electrical current to electrodes, which is attached on abdomen region.

GEM-TONE Abdominal Training System model GEM-TONE 310PE / 320PE /330PE /340PE /350PE, consists mainly of three parts: the stimulator , electrodes, and support belts fitting for special parts of body. The stimulator generates the output current specified as the input of controller. The output port transmits the output current to the electrode, which is attached to the specified region, so as to transmit this stimulus current to the region of abdomen for the following intended purposes:

  • Strengthening, toning and firming of the abdomen region.

The GEMORE GEM-TONE Abdominal Training System and its stimulation programs are not designed for injured or ailing muscles, and use on such muscles is contraindication.

Page 1 of 82

{1}------------------------------------------------

Gemore Technology Co., Ltd.

To adequately locate the stimulation on the intended treatment area, the following support belt is provided together with the stimulator:

  • · The abdomen belt, which is capable of connecting to the both two output channels of stimulator.
    The stimulation mode for GEM-TONE Abdominal Training System includes several different operation modes as mentioned on the comparison table. These operation modes are generated from the software control by using the microprocessor as its main control unit.

5. Intended Use:

Basically the indication for use is defined clearly as the description of the following statement:

The GEMORE GEM-TONE Abdominal Training System/model GEM-TONE 310PE/320PE/330PE/340PE/350PE are the electrical muscle stimulator intended for the following indication for use:

  • . Strengthening, toning and firming of the abdomen region.
    The GEMORE GEM-TONE Abdominal Training System and its stimulation programs are not designed for injured or ailing muscles, and use on such muscles is contraindication.

  • Discussion of Non-Clinical Tests Performed Determination of Substantial Equivalence are as 6. follows:
    Compliance to applicable voluntary standards includes ANSI/AAMI. NS4-1985, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement.

In addition to the compliance of voluntary standards, the software verification has been carried out according to the FDA software guidance.

7. Conclusions

The GEM-TONE Abdominal Training System, model GEM-TONE 310PE / 320PE /330PE /340PE /350PE, has the same intended use and technological characteristics as the cleared device of Slendertone Flex (K010335) . Moreover, verification tests contained in this submission demonstrate that the difference in the submitted demonstrate that the difference in the submitted model could maintain the same safety and effectiveness as that of cleared device.

In the other words, those engineering difference do not: (1) affect the intended use or (2) alter the fundamental scientific technology of the device. Therefore, the GEM-TONE Abdominal Training System, model GEM-TONE 310PE / 320PE /340PE /350PE, is substantial equivalent to the chosen predicate model.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized eagle with three lines forming its body and wings, symbolizing service to the nation. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Gemore Technology Co., Ltd. % Mr. Boden Lai 11FL. No. 29-5. Sec. 2. Chung Cheng E. Road, Tan Shui, Taipei Hsien China (Taiwan) 251

OCT 1 6 2009

Re: K091833

Trade/Device Name: GEM-TONE Abdominal Training System, Models 310PE, 320PE, 330PE, 340PE, and 350PE

Regulation Name: 21 CFR 890.5850

Regulation Name: Powered muscle stimulator Regulatory Class: II Product Code: NGX Dated: September 5, 2009 Received: September 8, 2009

Dear Mr. Lai:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Boden Lai

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/McdicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Mark N. Melkerson Director

Division of Surgical, Orthopedic. and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications For Use

510(k) Number (if known):

Device Name: GEM-TONE Abdominal Training System / Model: GEM-TONE 310PE/320PE/330PE/340PE/350PE.

Indications For Use:

The GEMORE GEM-TONE Abdominal Training System/model GEM-TONE 310PE/320PE/340PE/340PE/350PE are the electrical muscle stimulator intended for the following indication for use:

  • Strengthening, toning and firming of the abdomen region. t
    The GEMORE GEM-TONE Abdominal Training System and its stimulation programs are not designed for injured or ailing muscles, and use on such muscles is contraindication.

Prescription Use (Part 21 CFR 801 Subpart D)

OR

Over-The-Counter Use V (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Yousuf M. Al-Otaibi FOR M. MELKERSON

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

K0918 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Page 1 of 1

§ 890.5850 Powered muscle stimulator.

(a)
Identification. A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.(b)
Classification. Class II (performance standards).