K Number
K091783

Validate with FDA (Live)

Date Cleared
2009-07-07

(20 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Avantis Medical Systems, Inc. Third Eye is indicated for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

Device Description

The Third Eye System consists of both disposable portions and facility equipment. The disposable portions include the Third Eye Retroscope and Caps. The facility equipment portions of the device include the Third Eye Video Processor and accessories. The Third Eye is designed as an auxiliary device for use with a colonoscope during a colonoscopy procedure. After a colonoscope has been advanced to the cecum, the Third Eye Retroscope is inserted through the instrument channel of the colonoscope. As the Third Eye Retroscope emerges from the distal tip of the instrument channel of the colonoscope, it automatically bends 180 degrees to form a "J" shape. The Third Eye then provides a continuous retrograde image of the colon throughout the process of withdrawal of the colonoscope. The Third Eye Video Processor is a piece of electrical equipment that acts as an interface for the Third Eye Retroscope, the video output signals, and user input controls. The output video signals from the Video Processor are displayed on a monitor for viewing by the physician. Application software that runs on the Video Processor allows the user to adjust the light intensity and various picture settings of the Third Eye Retroscope image.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Avantis Medical Systems, Inc. Third Eye® Retroscope® Auxiliary Endoscopy System, structured to answer your specific questions.

Based on the provided document, this is a Special 510(k) submission, which means the device is a modified version of an already cleared device (K083180). Special 510(k)s typically focus on demonstrating that modifications do not adversely affect safety and effectiveness, rather than proving initial efficacy. Therefore, a comprehensive de novo clinical study demonstrating efficacy against specific acceptance criteria is not expected or presented here.


1. Table of Acceptance Criteria and Reported Device Performance

No specific, quantified acceptance criteria or detailed device performance metrics (e.g., sensitivity, specificity, accuracy for diagnostic purposes) are provided in this Special 510(k) summary. The document states:

Acceptance CriteriaReported Device Performance
Design inputs and intended performance characteristics for the modified device (change in hydrophilic coating on camera polarizer)"Results demonstrate that the modified Third Eye Retroscope Auxiliary Endoscopy System performs as intended." (Paragraph 510(k) Summary h))

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not applicable. No clinical or diagnostic performance test set (e.g., images for diagnostic accuracy) is described.
  • Data Provenance: Not applicable. The "test dataset" described is limited to "Bench verification testing."

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. No expert-based ground truth establishment is described, as the testing was limited to bench verification.

4. Adjudication Method for the Test Set

Not applicable. No test set requiring adjudication is described.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

No. An MRMC comparative effectiveness study was not done or described in this document. The submission is a Special 510(k) for a device modification, focusing on engineering verification.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

Yes, in a conceptual sense. The "testing" mentioned is "Bench verification testing." This means the device itself (the modified Retroscope and its components) was tested to ensure it met design inputs and performed as intended without human intervention in the diagnostic interpretation process. However, this is for the device's functionality (e.g., image quality, mechanics, material compatibility) and not for its diagnostic accuracy in identifying pathologies.

7. The Type of Ground Truth Used

For the bench verification testing, the "ground truth" would be the engineering specifications and design requirements for the device components. The testing aimed to verify that the modified components (specifically the hydrophilic coating) still met these pre-defined standards.

8. The Sample Size for the Training Set

Not applicable. This device is an endoscope system, not an AI/ML algorithm that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This device is an endoscope system, not an AI/ML algorithm that requires a training set.


Summary of the Study Presented (Special 510(k) Context):

The submission is a Special 510(k) for a minor modification to an already cleared device (K083180). The modification involved changing the hydrophilic coating on the camera polarizer from one manufacturer to another.

The study presented is "Bench verification testing." The purpose of this testing was to confirm that the modified device still met its original "design inputs and intended performance characteristics." The results, as stated, "demonstrate that the modified Third Eye Retroscope Auxiliary Endoscopy System performs as intended." This implies that physical and functional tests were conducted to ensure the new coating did not negatively impact the device's operation, sterility, or mechanical properties, thus confirming it remained substantially equivalent to the predicate device. No clinical efficacy study or diagnostic performance study was conducted or required for this type of submission.

{0}------------------------------------------------

K091783
py 1 of 3

JUL - 7 2009

510(k) SUMMARY

n medical systems

ਹੈ

ಗು V

Applicant Information: a)

Date Summary Prepared15 June 2009
Sponsor/SubmitterAvantis Medical Systems, Inc.263 Santa Ana CourtSunnyvale, CA 94085-4511
Correspondent Contact InformationAmrita SethiDirector of Regulatory and QualityAffairsPhone: 408-733-1901 x227Fax: 408-733-1847E-mail: asethi@avantismedical.com

b) Device Information:

Device Common NameEndoscope
Device Trade & Proprietary NameThird Eye® Retroscope® AuxiliaryEndoscopy System (Third Eye)
Device Classification NameColonoscope (per 21CFR 876.1500)
Device Classification Regulation21CFR 876.1500
Device ClassificationClass II (special controls)
Device Classification & Product CodeFDF, FDS, KNW, KOG

Identification of Predicate Device: c)

The Avantis Third Eye is substantially equivalent in operation and fundamental scientific technology to its previous version cleared under K083180.

d) Device Description Summary:

The Third Eye System consists of both disposable portions and facility equipment. The disposable portions include the Third Eye Retroscope and Caps. The facility equipment portions of the device include the Third Eye Video Processor and accessories.

{1}------------------------------------------------

K 091783
pg 2 of 3

15 June 2009 Special 510(k)

The Third Eye is designed as an auxiliary device for use with a colonoscope during a colonoscopy procedure. After a colonoscope has been advanced to the cecum, the Third Eye Retroscope is inserted through the instrument channel of the colonoscope. As the Third Eye Retroscope emerges from the distal tip of the instrument channel of the colonoscope, it automatically bends 180 degrees to form a "J" shape. The Third Eye then provides a continuous retrograde image of the colon throughout the process of withdrawal of the colonoscope.

The Third Eye Video Processor is a piece of electrical equipment that acts as an interface for the Third Eye Retroscope, the video output signals, and user input controls. The output video signals from the Video Processor are displayed on a monitor for viewing by the physician. Application software that runs on the Video Processor allows the user to adjust the light intensity and various picture settings of the Third Eye Retroscope image.

Intended Use: e)

The Third Eye Retroscope is intended for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

Discussion of Substantial Equivalence to Predicate Device: f)

The Third Eye has been modified to change the hydrophilic coating on the camera polarizer from a Surmodics-manufactured hydrogel to a Hydromer-manufactured hydrogel. No changes have been made to the facility equipment portion of the Third Eye system. The modified device is substantially equivalent to the cleared predicate device in fundamental design, materials, processes, sterile barrier packaging, and sterilization. All changes have been subjected to risk analysis and have been verified to meet current design specifications.

Differences between the modified Third Eye System and the predicate device have no impact on safety and effectiveness.

Scientific Technology g)

The Third Eye® Retroscope® Auxiliary Endoscopy System includes the Retroscope, a Video Processor, Cap and accessories. Modifications made to the device and accessories did not change the fundamental scientific technology of the device. The testing described demonstrates that the differences in the device components and accessories do not raise any new or unresolved issues for safety and efficacy.

263 Santa Ana Court, Sunnyvale, CA 94085 TEL: (408) 733-1901, FAX: (408) 733-1847

{2}------------------------------------------------

K 091783

Pg 3 of 3
15 June 2009
Special 510(k)

Summary of Supporting Non-Clinical Performance Data h)

Bench verification testing was conducted to verify that the modified device meets the design inputs and intended performance characteristics. Results demonstrate that the modified Third Eye Retroscope Auxiliary Endoscopy System performs as intended.

Any statement regarding "substantial equivalence" made in this 510(k) submission and summary only relates to whether the product addressed in this submission may be lawfully marketed without premarket approval or reclassification, and is not intended to be interpreted as an admission or any other type of evidence in any patent proceeding, including patent infringement, litigation or proceeding before any Patent Office. The present submission and statements therein therefore should not be construed as affecting or relating to the scope of any patent or patent application, or to whether the product addressed in this submission, or its use, may be considered indistinct, from a patentability perspective, from any of the other devices referenced in this filing.

263 Santa Ana Court, Sunnyvale, CA 94085 TEL: (408) 733-1901, FAX: (408) 733-1847

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with a more abstract and modern design.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL - 7 2009

Ms. Amrita Sethi Director of Regulatory Affairs and Quality Affairs Avantis Medical Systems, Inc. 263 Santa Anna Court SUNNYVALE CA 94085-4511

Re: K091783

Trade/Device Name: Third Eye® Retroscope® Auxiliary Endoscopy System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: June 15, 2009 Received: June 17, 2009

Dear Ms. Sethi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{4}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/indr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours

Sincerely yours,

Janine M. Morris

Janine M. Morris Acting Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

INDICATIONS FOR USE STATEMENT

510(k) Number (if known): _ K 09 / 7 83

Device Name: Third Eye® Retroscope® Auxiliary Endoscopy System (Third Eye)

Indications for Use: The Avantis Medical Systems, Inc. Third Eye is indicated for use in the instrument channel of a conventional colonoscope to provide retrograde illumination and visualization of the colon for diagnostic purposes.

Prescription Use Over-The-Counter Use X - AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hula Cluss

Page I of 1

(Division Sign-Off) (Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.