K Number
K091731

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-09-17

(98 days)

Product Code
Regulation Number
880.6880
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Amsco Evolution Medium Steam Sterilizer Models HC-800 and HC-1000 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

  • Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
  • Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
Device Description

The Amsco Evolution Medium Steam Sterilizer models HC-800 and HC-1000 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

  • Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and . DART (Bowie-Dick) cycles.
  • Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, ● Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.
    The chamber sizes of the proposed Amsco Evolution Medium Steam Sterilizer Models are as follows:
  • 26" x 26" x 49" (Model HC 800) .
  • 26" x 26" x 61" (Model HC 1000) .
AI/ML Overview

The provided text describes a steam sterilizer and its specifications rather than an AI/ML powered device. Therefore, typical acceptance criteria, study methodologies, and performance metrics associated with AI/ML devices (e.g., sensitivity, specificity, AUC, human reader improvement with AI assistance, ground truth establishment for training/test sets) are not applicable.

The document mainly focuses on the factory-programmed sterilization cycles, cycle values, and recommended load capacities for the STERIS Amsco Evolution Medium Steam Sterilizer Models HC-800 and HC-1000. It's a 510(k) summary for a conventional medical device seeking substantial equivalence to a predicate device.

However, I can extract the acceptance criteria (defined by the device's operational parameters) and the "study" that proves it meets these criteria based on the provided information, interpreting "performance" as the ability to achieve the specified sterilization cycles.

Here's the information re-framed for a non-AI/ML device:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are its ability to perform the specified sterilization cycles at defined temperatures, times, and dry times, with recommended load capacities. The reported device performance is that it is equipped with and capable of executing these cycles.

Cycle TypeSterilize TemperatureSterilize TimeDry TimeRecommended Load Performance (Examples)
Prevac (270°F)270°F (132°C)4 minutes30 minutesAble to sterilize Double wrapped instrument trays (max 25 lbs (11.3 kg) each) and Fabric Packs; up to 12 wrapped instrument trays / 16 fabric packs (HC-800), 15/20 (HC-1000).
Prevac (270°F)270°F (132°C)4 minutes5 minutesAble to sterilize Single Fabric Pack.
Gravity (250°F)250°F (121°C)30 minutes15 minutesAble to sterilize Fabric Packs; up to 16 fabric packs (HC-800), 20 (HC-1000).
Liquid (250°F)250°F (121°C)45 minutesN/AAble to sterilize 3 containers with 1000 ml liquid each. (Non-patient contact use only).
DART270°F (132°C)3 ½ minutes1 minuteAble to perform Bowie-Dick Test Pack, DART Test Pack cycle.
SFPP (270°F)270°F (132°C)4 minutes30 minutesAble to sterilize Double wrapped instrument trays (max 25 lbs (11.3 kg) each) and Fabric Packs; up to 12 wrapped instrument trays / 16 fabric packs (HC-800), 15/20 (HC-1000).
Other cycles listed for Prevacuum and SFPP models are similarly specified.TBDTBDTBDTBD

Study to Prove Device Meets Acceptance Criteria:

The document is a 510(k) summary, which typically references underlying validation studies but does not provide details of the studies themselves. For a sterilizer, the "study" would be sterilization validation testing in accordance with recognized standards (e.g., AAMI, ISO) to demonstrate that the device consistently achieves sterility assurance levels (SAL) under its specified cycles and loading conditions. The summary states:

  • "A summary of the technological characteristics of the device subject of this premarket notification in comparison to those of the predicate devices is included in Table 5-5." (Table 5-5 is not provided in the excerpt).
  • The FDA's letter (Page 5) states: "We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices..."

This indicates that STERIS performed the necessary validation testing (e.g., biological indicator testing, physical parameter monitoring, microbial challenge testing) to demonstrate that the Amsco Evolution Steam Sterilizer could reliably achieve the specified sterilization parameters and sterilize the indicated loads. The results of these tests (summarized but not detailed here) were submitted to the FDA to support the claim of substantial equivalence to the predicate device.

2. Sample size used for the test set and the data provenance
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
4. Adjudication method for the test set
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
7. The type of ground truth used
8. The sample size for the training set
9. How the ground truth for the training set was established

These questions are not applicable to this submission. The document describes a steam sterilizer, a non-AI/ML medical device. Therefore, concepts like "test set," "data provenance," "experts," "adjudication," "MRMC studies," "AI assistance," "standalone algorithm performance," "ground truth," and "training set" in the context of AI/ML are not relevant.

For a steam sterilizer, the "ground truth" would be the successful killing of microbial spores (demonstrating sterility) confirmed through laboratory methods (e.g., biological indicator results, microbial challenge tests), and the "test set" would be the series of sterilization cycles performed under various worst-case loading and operating conditions during validation. The number of 'samples' would refer to the number of sterilization cycles performed and the number of biological and physical indicators evaluated. However, those details are not present in this 510(k) summary.

{0}------------------------------------------------

K091731 STERIS®

510(k) Summary For Amsco Evolution Medium Steam Sterilizer Models HC-800 and HC-1000

STERIS Corporation 5960 Heisley Road Mentor, OH 44060 Phone: (440) 354-2600 Fax No: (440) 639-4459

Contact:

Robert F. Sullivan Senior Director FDA Regulatory Affairs Telephone:(440) 392-7695 Fax No: (440) 357-9198

Summary Date:

June 10, 2009

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

{1}------------------------------------------------

1. Device Name

Trade Name:Amsco Evolution Medium Steam Sterilizer
Models:HC-800 and HC-1200
Common/Usual Name:Steam Sterilizer
Classification Name:Steam Sterilizer (21 CFR 880.6880)Product Code 80 FLE

2. Predicate Device

K082435, Amsco Evolution Medium Steam Sterilizer, product code [FLE] cleared December 30, 2008.

Description of Device 3.

The Amsco Evolution Medium Steam Sterilizer models HC-800 and HC-1000 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

  • Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and . DART (Bowie-Dick) cycles.
  • Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, ● Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

The chamber sizes of the proposed Amsco Evolution Medium Steam Sterilizer Models are as follows:

  • 26" x 26" x 49" (Model HC 800) .
  • 26" x 26" x 61" (Model HC 1000) .

Intended Use 4.

The Amsco Evolution Medium Steam Sterilizer models HC-800 and HC-1000 are designed for sterilization of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

  • Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and . DART (Bowie-Dick) cycles.
  • Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, . Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

{2}------------------------------------------------

STERIS ABBREVIATED 510(k) PREMARKET NOTIFICATION STERIO ABDREVIEW Steam Sterilizer (Models HC-800 and HC-1000)

The Amsco Evolution Medium Prevacuum Steam Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-1):

CyclesSterilize TemperatureSterilize TimeDry TimeRecommended Load
Prevac270°F (132°C)4 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities.
Prevac270°F (132°C)4 minutes5 minutesSingle Fabric Pack
Gravity250°F (121°C)30 minutes15 minutesFabric Packs. Refer to Table 4-3 for recommended quantities.
Liquid250°F (121°C)45 minutesN/ARefer to Table 4-4 for recommended quantities.
Prevac275°F (135°C)3 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities.
Gravity270°F (132°C)15 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities.
Gravity250°F (121°C)30 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities.
Gravity270°F (132°C)25 minutes15 minutesFabric Packs. Refer to Table 4-3 for recommended quantities.
DART Warm-Up270°F (132°C)3 minutes1 minuteN/A
DART270°F (132°C)3 ½ minutes1 minuteBowie-Dick Test Pack, DART Test Pack
Leak TestN/AN/AN/AN/A
CyclesSterilize TemperatureSterilize TimeDry TimeRecommended Load
SFPP270°F (132°C)4 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities.
SFPP275°F (135°C)3 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities.
Prevac270°F (132°C)4 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities.
Gravity250°F (121°C)30 minutes15 minutesFabric Packs. Refer to Table 4-3 for recommended quantities.
SFPP270°F (132°C)4 minutes5 minutesSingle Fabric Pack
Prevac275°F (135°C)3 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities.
Gravity270°F (132°C)15 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities.

Table 4-1. Amsco Evolution Medium Prevacuum Steam Sterilizer factory-programmed sterilization cycles and cycle values

*The liquid cycle is for non-patient contact use only.

{3}------------------------------------------------

The Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) Sterilizer is The Amseo Lyonation Necatall Brown and sterilization cycles and cycle values (Table 4-2):

Table 4-2. Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP)
Sterilizer factory-programmed sterilization cvcles and cvcle values

{4}------------------------------------------------

CyclesSterilizeTemperatureSterilizeTimeDry TimeRecommended Load
Liquid250°F (121°C)45 minutesN/ARefer to Table 4-4 forrecommended quantities.
DARTWarm-Up270°F (132°C)3 minutes1 minuteN/A
DART270°F (132°C)3 ½ minutes1 minuteBowie-Dick Test Pack,DART Test Pack
Leak TestN/AN/AN/AN/A

*The liquid cycle is for non-patient contact use only.

The following table lists STERIS's recommended loads by sterilizer size:

Sterilizer recommended loads per sterilizer size
Sterilizer SizeWrapped InstrumentTraysFabric Packs
26" x 26" x 49"1216
26" x 26" x 61"1520

Table 4-3. Amsco Evolution Medium Steam

The following table is a guideline for liquid cycle processing:

Table 4-4. Amsco Evolution Medium Steam Sterilizer Liquid Cycle Guideline

management of the comments of the comments of the many of the many of the many of the many of the many of the many of the many of the many of the many of the many of the many· · R. R. L. Trenspearer seconomics .------------------------------------------------------------------------------------------------------------------------------------------------------------------------------. CON an assa so " . " ..Can REEBER I
----------American Article Art10-00-00

The Amsco Evolution Medium Steam Sterilizer is offered in the following chamber sizes:

  • · 26" x 26" x 49" (Model HC 800)
  • 26" x 26" x 61" (Model HC 1000) .

Description of Safety and Substantial Equivalence 5.

A summary of the technological characteristics of the device subject of this premarket notification in comparison to those of the predicate devices is included in Table 5-5.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The words are arranged on a single line and are centered horizontally.

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized graphic of an abstract human figure with three arms or branches extending upwards. To the left of the graphic, the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the graphic.

SEP 17 2009

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Mr. Robert F. Sullivan Senior Director of FDA Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060-1834

Re: K091731

Trade/Device Name: Amsco Evolution Medium Steam Sterilizer Regulation Number: 21 CFR 880. 6880 Regulation Name: Steam Sterilizer Regulatory Class: II Product Code: FLE Dated: August 13, 2009 Received: August 14, 2009

Dear Mr. Sullivan;

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{6}------------------------------------------------

Page 2 - Mr. Sullivan

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA. has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Anthony O. Winston ber.

Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

Indications for Use

510(k) Number (if known):

Device Name:

Amsco Evolution Medium Steam Sterilizer (Models HC-800 and HC-1000)

Indications For Use:

The Amsco Evolution Medium Steam Sterilizer Models HC-800 and HC-1000 are r no rimboo of heat and moisture-stable materials used in healthcare facilities and are available in two configurations:

  • Prevacuum is equipped with Prevacuum, Gravity, Liquid, Leak Test and DART . (Bowie-Dick) cycles.
  • Steam Flush Pressure-Pulse (SFPP) is equipped with SFPP, Prevacuum, . Gravity, Liquid, Leak Test and DART (Bowie-Dick) cycles.

The Amsco Evolution Medium Prevacuum Steam Sterilizer is equipped with the Fillowing factory-programmed sterilization cycles and cycle values (Table 4-1):

CyclesSterilizeTemperatureSterilize TimeDry TimeRecommended Load
Prevac270°F (132°C)4 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each andFabric Packs. Refer toTable 4-3 for recommendedquantities.
Prevac270°F (132°C)4 minutes5 minutesSingle Fabric Pack
Gravity250°F (121°C)30 minutes15 minutesFabric Packs. Refer toTable 4-3 for recommendedquantities.
Liquid250°F (121°C)45 minutesN/ARefer to Table 4-4 forrecommended quantities.
Prevac275°F (135°C)3 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 4-3 for recommendedquantities.

Table 4-1. Amsco Evolution Medium Prevacuum Steam Sterilizer factory-programmed sterilization cycles and cycle values

{8}------------------------------------------------

CyclesSterilizeTemperatureSterilizeTimeDry TimeRecommended Load
Gravity270°F (132°C)15 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 4-3 for recommendedquantities.
Gravity250°F (121°C)30 minutes30 minutesDouble wrapped instrumenttrays, maximum weight 25lbs (11.3 kg) each. Refer toTable 4-3 for recommendedquantities.
Gravity270°F (132°C)25 minutes15 minutesFabric Packs. Refer toTable 4-3 for recommendedquantities.
DARTWarm-Up270°F (132°C)3 minutes1 minuteN/A
DART270°F (132°C)3 ½ minutes1 minuteBowie-Dick Test Pack,DART Test Pack
Leak TestN/AN/AN/AN/A

*The liquid cycle is for non-patient contact use only.

The Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) Sterilizer is equipped with the following factory-programmed sterilization cycles and cycle values (Table 4-2): '

CyclesSterilize TemperatureSterilize TimeDry TimeRecommended Load
SFPP270°F (132°C)4 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each and Fabric Packs. Refer to Table 4-3 for recommended quantities.
SFPP275°F (135°C)3 minutes30 minutesDouble wrapped instrument trays, maximum weight 25 lbs (11.3 kg) each. Refer to Table 4-3 for recommended quantities.

Table 4-2. Amsco Evolution Medium Steam Flush Pressure-Pulse (SFPP) Sterilizer factory-programmed sterilization cycles and cycle values

{9}------------------------------------------------

CyclesSterilizeTemperatureSterilizeTimeDry TimeRecommended Load
Prevac270°F (132°C)4 minutes30 minutesDouble wrappedinstrument trays,maximum weight 25 lbs(11.3 kg) each and FabricPacks. Refer to Table 4-3for recommendedquantities.
Gravity250°F (121°C)30 minutes15 minutesFabric Packs. Refer toTable 4-3 forrecommended quantities.
SFPP270°F (132°C)4 minutes5 minutesSingle Fabric Pack
Prevac275°F (135°C)3 minutes30 minutesDouble wrappedinstrument trays,maximum weight 25 lbs(11.3 kg) each. Refer toTable 4-3 forrecommended quantities.
Gravity270°F (132°C)15 minutes30 minutesDouble wrappedinstrument trays,maximum weight 25 lbs(11.3 kg) each. Refer toTable 4-3 forrecommended quantities.
Liquid250°F (121°C)45 minutesN/ARefer to Table 4-4 forrecommended quantities.
DARTWarm-Up270°F (132°C)3 minutes1 minuteN/A
DART270°F (132°C)3 ½ minutes1 minuteBowie-Dick Test Pack,DART Test Pack
Leak TestN/AN/AN/AN/A

*The liquid cycle is for non-patient contact use only.

The following table lists STERIS's recommended loads by sterilizer size:

Sterilizer recommended loads per sterilizer size
Sterilizer SizeWrapped Instrument TraysFabric Packs
26" x 26" x 49"1216
26" x 26" x 61"1520

Table 4-3. Amsco Evolution Medium Steam

{10}------------------------------------------------

The following table is a guideline for liquid cycle processing:

Table 4-4. Amsco Evolution Medium Steam

Number ofContainersVolume of Liquid In OneContainerMinimum RecommendedSterilize Time at 250°F (121°C)
31000 ml45 minutes

Sterilizer Liquid Cycle Guideline

The Amsco Evolution Medium Steam Sterilizer is offered in the following chamber sizes:

  • 26" x 26" x 49" (Model HC 800) .
  • 26" x 26" x 61" (Model HC 1000)

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

K 09173 i Shuli M. Murphy, MD

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K091731

§ 880.6880 Steam sterilizer.

(a)
Identification. A steam sterilizer (autoclave) is a device that is intended for use by a health care provider to sterilize medical products by means of pressurized steam.(b)
Classification. Class II (performance standards).