K Number
K091638

Validate with FDA (Live)

Date Cleared
2009-09-14

(102 days)

Product Code
Regulation Number
888.3080
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DORADO™ Wide Intervertebral Body Cage is a spinal intervertebral body cage intended for a posterior approach and uses autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolithesis at the involved level(s). The SpineFrontier DORADO™ Wide Intervertebral Body Cage is intended to be used with supplemental spinal fixation system(s) (Example: Posterior Pedicle Screw). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Device Description

The DORADO™ Wide Intervertebral Body Cage is a spinal intervertebral body cage intended for a posterior approach and uses autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. The system is comprised of devices of various fixed heights and footprints to fit the anatomical needs of a wide variety of patients. The device has raised contours on the superior and inferior surfaces that will resist implant expulsion.

AI/ML Overview

This submission (K091638) is for a spinal intervertebral body cage, a medical device, not an AI/ML powered device. Therefore, many of the typical acceptance criteria and study components related to AI performance (e.g., sample size for test set, number of experts, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable.

The "acceptance criteria" for this device are based on its substantial equivalence to previously cleared predicate devices and its compliance with relevant guidance documents and mechanical testing. The submission does not involve clinical data or AI performance metrics.

Here's a breakdown of the relevant information from the provided document, adapted to the context of a traditional medical device submission:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance / Basis for Equivalence
Indications for UseSubstantially equivalent to predicate devices L2-S1, DDD, Grade I spondylolisthesis/retrolisthesis, posterior approach, autogenous bone graft, supplemental fixation, skeletally mature, 6 months non-operative treatment.The Dorado™ Wide IBC has been shown to be substantially equivalent to predicate devices (K072289, P960025) in terms of indications for use. The stated intended use aligns with the predicate devices.
Technological CharacteristicsSubstantially equivalent to predicate devices in terms of function, operating principles, and materials.Comparison of function, operating principles, and materials to K072289 and P960025 demonstrated substantial equivalence.
MaterialsMaterials are equivalent to those used in predicate devices.Stated as equivalent to predicate devices. (Specific materials not detailed in summary, but assumed to be comparable to cleared devices).
Performance (Mechanical Testing)Meets criteria defined in "Class II Special Controls, Guidance Document: Intervertebral Fusion Device, June 12, 2007."Demonstrated to meet criteria defined in the guidance document for Intervertebral Fusion Devices.
Regulatory ComplianceCompliance with 21 CFR 888.3080 (Intervertebral body fusion device) and Class II Special Controls.The device was evaluated in accordance with the FDA Guidance Document and classified as Class II, product code MAX. Received 510(k) clearance, indicating compliance.

Regarding the AI-specific questions, these are NOT APPLICABLE to this traditional medical device submission:

  • 2. Sample size used for the test set and the data provenance: Not applicable. No test set for AI performance.
  • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No expert-derived ground truth for AI.
  • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No test set for AI performance.
  • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
  • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical implant, not an algorithm.
  • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for this device's acceptance is primarily regulatory compliance, comparison to predicate devices, and mechanical testing results, rather than clinical outcomes or diagnostic accuracy.
  • 8. The sample size for the training set: Not applicable. No training set for an AI algorithm.
  • 9. How the ground truth for the training set was established: Not applicable. No training set for an AI algorithm.

Summary of the Study that Proves Acceptance Criteria are Met:

The "study" in this context is the 510(k) Pre-Market Notification process itself, which focused on demonstrating substantial equivalence to legally marketed predicate devices and compliance with relevant mechanical testing standards and guidance.

  • Basis for Substantial Equivalence: The primary "proof" for the Dorado™ Wide Intervertebral Body Cage was the comparison of its indications for use, technological characteristics (function, operating principles, materials), and performance (mechanical testing) against established predicate devices:
    • K072289: SpineFrontier Inc., Dorado Intervertebral Body Cage
    • P960025: Saber Lumbar I/F Cage and Jaguar Lumbar I/F Cage
  • Guidance Document Compliance: The device was evaluated in accordance with the "FDA Document, Class II Special Controls, Guidance Document: Intervertebral Fusion Device, June 12, 2007." Compliance with the criteria defined in this document, particularly regarding mechanical testing, served as a key component of demonstrating safety and effectiveness.
  • Clinical Data: The submission explicitly states: "Clinical data was not required for this device." This reinforces that the acceptance was based on non-clinical evidence of substantial equivalence and adherence to recognized standards.

In essence, the device met its acceptance criteria by demonstrating that it is as safe and effective as existing legally marketed devices through a thorough comparison of its design, materials, and performance with those predicates, and by conforming to relevant FDA guidance.

{0}------------------------------------------------

SEP 1 4 2009

510(k) Summarv

1. Sponsor

SpineFrontier, Inc. 500 Cummings Center Suite 3500

Beverly, MA 01915

Primary Contact: John Sullivan Telephone: 1- 978-232-3990

Date Prepared: May 29, 2009

2. Device Name and Classification:

Dorado ™ Wide Intervertebral Body Cage, Proprietary Name: Dorado™ Wide IBC, Dorado™ Wide IBF Common/Usual Name: Intervertebral Fusion Device With Bone Graft. Lumbar Classification Name: Intervertebral Fusion Device With Bone Graft, Lumbar, (21 CFR 888.3080), Class II Product Code: MAX

3. Predicate Devices

K072289 - SpineFrontier Inc., Dorado Intervertebral Body Cage P960025 - Saber Lumbar I/F Cage and Jaguar Lumbar I/F Cage

4. Device Description

The DORADO™ Wide Intervertebral Body Cage is a spinal intervertebral body cage intended for a posterior approach and uses autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. The system is comprised of devices of various fixed heights and footprints to fit the anatomical needs of a wide variety of patients. The device has raised contours on the superior and inferior surfaces that will resist implant expulsion.

SpineFrontier, Inc. Dorado " Wide Intervertebral Body Cage - Supplement to Abbreviated 510(k) K091638 September 1, 2009

Company Confidential

5.0

{1}------------------------------------------------

K09638 page 2/2

5. Intended Use

The DORADO™ Wide Intervertebral Body Cage is a spinal intervertebral body cage intended for a posterior approach and uses autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolithesis at the involved level(s). The SpinaFrontier DORADO™ Wide Intervertebral Body Cage is intended to be used with supplemental spinal fixation system(s) (Example: Posterior Pedicle Screw). Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six months of nonoperative. treatment.

6. Technological Characteristics

The SpineFrontier Dorado™ Wide Intervertebral Body Cage was shown to be substantially equivalent to predicate devices through comparison of indications for use, function, operating principles, and materials,

7. Basis for Substantial Equivalence

The Dorado™ Wide Intervertebral Body Cage was evaluated in accordance with FDA Document, Class II Special Controls, Guidance Document: Intervertebral Fusion Device, June 12, 2007, and has been found to meet criteria defined in the guidance document; and has been demonstrated to be substantially equivalent to predicate devices in terms of indications for use. function, materials, and performance (mechanical testing). Clinical data was not required for this device.

Company Confidential

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle. In the center of the circle is a stylized image of an eagle or bird-like figure with three curved lines representing its wings or feathers.

SEP 1 4 2009

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center - WO66-G609 Silver Spring, MD 20993-0002

Spinefrontier, Inc. % Mr. John Sullivan Director of OA and Regulatory Compliance 500 Cummings Center, Suite 3500 Beverly, Massachusetts 01915

Re: K091638

Trade/Device Name: Dorado™ Wide Intervertebral Body Cage Regulation Number: 21CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: II Product Code: MAX Dated: September 1, 2009 Received: September 3, 2009

Dear Mr. Sullivan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. John Sullivan

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Daniel Keane

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use Statement 4:0

510(k) Number (if Known):

Indications For Use:

The DORADO™ Wide Intervertebral Body Cage is a spinal intervertebral body cade

intended for a posterior approach and uses autogenous bone graft in patients with degenerative disc disease (DDD) at one or two contiguous spinal levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolithesis at the involved level(s). The SpineFrontier DORADO™ Wide Intervertebral Body Cage is intended to be used with supplemental spinal fixation system(s) (Example: Posterior Pedicle Screw).

Degenerative disc disease is defined as discogenic back pain with degeneration of the disc confirmed by history or radiographic studies. These patients should be skeletally mature and have had six months of non-operative treatment.

Prescription Use: X (Part 21 CFR 801 Subpart D)

ANDIOR

Over-The-Counter Use: (Part 21 CFR 807 Subpart C)

Daniel Krone for Nixon

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091638

Company Confidential

§ 888.3080 Intervertebral body fusion device.

(a)
Identification. An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.(b)
Classification. (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
Date premarket approval application (PMA) or notice of product development protocol (PDP) is required. Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.