K Number
K091616

Validate with FDA (Live)

Date Cleared
2010-03-29

(299 days)

Product Code
Regulation Number
866.3510
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The BioPlex™ 2200 MMRV IgG kit is a multiplex flow immunoassay intended for the qualitative detection of IgG antibodies to Measles, Mumps, Rubella, and Varicella-zoster virus (VZV) in human serum and EDTA or heparinized plasma. The BioPlex 2200 MMRV IgG kit is intended for use with the Bio-Rad BioPlex 2200 System.

This kit is intended as an aid in the determination of serological status to Measles, Mumps, Rubella, and VZV. This kit is not intended for use in screening blood or plasma donors.

The performance of this assay has not been established for use in neonates, pediatrics and immunocompromised patients, or for use at point of care facilities.

Device Description

The BioPlex 2200 MMRV IgG kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube. Four (4) different populations of dyed beads are coated with antigens to identify the presence of IgG class antibodies associated with Measles, Mumps, Rubella and Varicella-zoster. The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead set reagent into a reaction vessel. The mixture is incubated at 37°C. After a wash cycle, anti-human IgG antibody, conjugated to phycoerythrin (PE), is added to the dyed beads and this mixture is incubated at 37°C. The excess conjugate is removed in another wash cycle, and the beads are resuspended in wash buffer. The bead mixture then passes through the detector.

The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity (RFI). Three additional dyed beads, an Internal Standard Bead (ISB), a Serum Verification Bead (SVB) and a Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum to the reaction vessel and the absence of significant non-specific binding in serum.

The instrument is calibrated using a set of three (3) distinct calibrator vials, supplied separately by Bio-Rad Laboratories.

AI/ML Overview

The BioPlex 2200 MMRV IgG kit is a multiplex flow immunoassay intended for the qualitative detection of IgG antibodies to Measles, Mumps, Rubella, and Varicella-zoster virus (VZV) in human serum and EDTA or heparinized plasma. It is intended for use with the Bio-Rad BioPlex 2200 System as an aid in determining serological status to these viruses. It is not intended for screening blood/plasma donors, and performance has not been established for neonates, pediatrics, immunocompromised patients, or point-of-care facilities.

Here's an analysis of the acceptance criteria and the studies performed:

1. Table of Acceptance Criteria and Reported Device Performance

The submission does not explicitly state pre-defined acceptance criteria (e.g., minimum sensitivity/specificity percentages). Instead, the performance is demonstrated through comparative testing against legally marketed predicate devices and reproducibility studies. The tables below summarize the reported device performance.

Comparative Testing: BioPlex 2200 MMRV IgG vs. Commercially Available EIAs

Target AnalyteStudy PopulationAgreement Type (EIA vs. BioPlex)Reported Performance (95% CI)
Measles IgGPregnant Women (N=396)Positive % Agreement93.3% (90.4 - 95.4%)
Negative % Agreement100% (67.5 - 100%)
Test Ordered (N=1183)Positive % Agreement94.7% (93.2 - 95.9%)
Negative % Agreement80.2% (70.3 - 87.5%)
Mumps IgGPregnant Women (N=396)Positive % Agreement94.3% (91.5 - 96.3%)
Negative % Agreement96.2% (81.1 - 99.3%)
Test Ordered (N=1183)Positive % Agreement90.4% (88.5 - 92.0%)
Negative % Agreement91.0% (81.8 - 95.8%)
Rubella IgGPregnant Women (N=396)Positive % Agreement97.9% (95.9 - 98.9%)
Negative % Agreement73.7% (51.2 - 88.2%)
Test Ordered (N=1183)Positive % Agreement94.8% (93.3 - 96.0%)
Negative % Agreement86.2% (75.7 - 92.5%)
VZV IgGPregnant Women (N=396)Positive % Agreement95.1% (92.4 - 96.9%)
Negative % Agreement100% (88.6 - 100%)
Test Ordered (N=1183)Positive % Agreement94.0% (92.4 - 95.2%)
Negative % Agreement98.4% (91.3 - 99.7%)

Retrospective Negative Samples: BioPlex 2200 MMRV IgG vs. EIA

Target AnalyteSample Count (N)BioPlex Pos (+)BioPlex EqvBioPlex Neg (-)Negative % Agreement (95% CI)
Measles IgG930093100% (96.0 - 100%)
Mumps IgG966783100% (95.6 - 100%)
Rubella IgG2683825795.9% (92.8 - 97.7%)
VZV IgG14300143100% (97.4 - 100%)

Reproducibility (Total %CV for Serum, across sites, days, runs)

AnalyteHigh Positive Range (Total %CV)Low Positive Range (Total %CV)Near Cutoff Range (Total %CV)High Negative Range (Total %CV)Positive Control (Total %CV)
Measles IgG10.2 - 13.913.5 - 16.010.4 - 14.89.4 - 11.512.3
Mumps IgG7.2 - 10.25.7 - 6.97.4 - 7.65.3 - 8.47.5
Rubella IgG10.1 - 15.28.0 - 12.19.0 - 11.010.9 - 11.510.7
VZV IgG9.0 - 10.06.8 - 9.38.7 - 9.86.1 - 13.67.9

2. Sample Sizes Used for the Test Set and Data Provenance

  • Prospective Comparative Testing:
    • Pregnant Women: N = 396 serum samples. Data provenance: Not explicitly stated, but the study was conducted at 3 U.S. clinical trial sites. Retrospective or prospective nature is not specified for collection.
    • Measles, Mumps, Rubella, or VZV Test Ordered: N = 1183 serum samples. This population included (N = 790) samples submitted for routine testing and (N = 393) samples for pre-employment evaluation. Data provenance: Not explicitly stated, but the study was conducted at 3 U.S. clinical trial sites.
  • Retrospective Comparative Testing (Negative Samples):
    • Measles IgG: N = 93 serum samples.
    • Mumps IgG: N = 96 serum samples.
    • Rubella IgG: N = 268 serum samples.
    • VZV IgG: N = 143 serum samples.
    • Data provenance: Not explicitly stated, but implies these were pre-existing samples.
  • CDC Rubella Evaluation Serum Panel: N = 100 serum samples (82 positive, 18 negative). Data provenance: Provided by the CDC.
  • Reproducibility Studies: A "reproducibility panel" was prepared, consisting of positive panel members in serum, EDTA plasma, and sodium heparin plasma, along with a positive and negative control. The exact number of panel members is not explicitly stated beyond "varying levels of antibodies" and a "positive control." Each panel member and controls were tested in quadruplicate on 1 run per day over 5 days at each of 3 sites (4 replicates x 1 run x 5 days x number of panel members = total Sample N for reproducibility analysis, which for each measured point was N=60).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

The ground truth for the comparative testing was established using commercially available EIA (Enzyme Immunoassay) predicate devices, which are conventional laboratory tests. No human experts establishing a ground truth in the manner of, for example, radiologists interpreting images, are mentioned for these studies.

For the CDC Rubella Evaluation Serum Panel, the samples were "masked, characterized" by the CDC. This implies expert characterization by the CDC, but the specific number and qualifications of experts are not described.

4. Adjudication Method for the Test Set

  • Prospective Comparative Testing:
    • For Mumps and Measles, equivocal results on the commercially available EIAs were "further tested on two additional commercially available EIAs for consensus." This implies a 2 out of 3 consensus method for adjudicating equivocal results from the primary predicate EIA, using the two additional EIAs. The exact consensus rule (e.g., majority or all agree) is not explicitly stated but "consensus" suggests agreement among at least two.
    • For Rubella, equivocal results on the commercially available EIA were "re-tested on the commercially available EIA, following the manufacturer's recommendations." This is a re-testing procedure, not a multi-reader adjudication.
    • For VZV, no specific adjudication method for equivocal results is mentioned in the tables provided.
  • Retrospective Comparative Testing:
    • For Measles and Mumps negative samples, they were "selected using a consensus of two out of three commercially available EIAs." This is a pre-selection method for the test samples based on consensus, rather than adjudication of results from the BioPlex device.
    • For Rubella and VZV negative samples, they were "selected using the respective commercially available EIAs used for the comparative analysis." No consensus method is mentioned here for selection.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Its Effect Size

No MRMC comparative effectiveness study was described. The study focuses on comparing the BioPlex device's performance to existing immunoassays, rather than assessing human reader performance with and without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the studies described are standalone performance evaluations of the BioPlex 2200 MMRV IgG kit against predicate devices. The device is an automated immunoassay system, and its performance metrics (agreement, reproducibility) reflect its standalone operational capabilities. There is no human-in-the-loop component mentioned in its operational or evaluation methodology that would constitute an "AI assistance" scenario.

7. The Type of Ground Truth Used

The primary ground truth for evaluating the BioPlex 2200 MMRV IgG kit's performance was established using results from legally marketed, commercially available EIA (Enzyme Immunoassay) predicate devices.

  • For Measles and Mumps, this involved consensus results from two out of three commercially available EIAs for equivocal samples.
  • For Rubella, re-testing with the predicate EIA was used for equivocal results.
  • For the CDC Rubella panel, the ground truth was "masked, characterized" CDC reference sera.

8. The Sample Size for the Training Set

No training set is explicitly mentioned, as this is a traditional immunoassay device, not a machine learning or AI-driven algorithm in the context of typical training/validation/test sets for pattern recognition. The device is calibrated using a "set of three (3) distinct calibrator vials, supplied separately by Bio-Rad Laboratories" and controls are used for quality assurance.

9. How the Ground Truth for the Training Set Was Established

Since no traditional "training set" in the context of AI/ML was used, this question is not applicable. The device's operational parameters are based on established immunoassay principles, calibration, and quality control using defined calibrator and control materials.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Bio-Rad Laboratories. The logo is in black and white, with the words "BIO RAD" in white letters on a black oval background. The letters are bold and sans-serif.

BIOPLEX 2200 MMRV IgG 510(k) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807 92.

510(k) Number510(k) Summary Report Date
K091616March 25, 2010

MANUFACTURER INFORMATION

Manufacturer
Manufacturer AddressBio-Rad Laboratories, Inc.Clinical Systems Division4000 Alfred Nobel DriveHercules, CA 94547
Telephone(510) 724-7000
Establishment Registration No.2915274
Owner / OperatorBio-Rad Laboratories, Inc.4000 Alfred Nobel DriveHercules, CA 94547
Owner / Operator No.9929003
Official Correspondent for the BioPlex 2200 MMRV IgG
Official Correspondent AddressBio-Rad Laboratories6565 185th Ave NERedmond, WA 98052
Telephone425-881-8300
Establishment Registration No.3022521
Owner / OperatorBio-Rad Laboratories6565 185th Ave NERedmond, WA 98052
Official CorrespondentMr. Christopher Bentsen
Telephone(425) 498-1709
Fax(425) 498-1651

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Bio-Rad Laboratories. The logo is white text on a black background. The text is in a bold, sans-serif font. The logo is simple and recognizable.

CLASSIFICATION INFORMATION

Classification NameMultiplex immunoassay for measles virus, mumps virus,rubella and varicella zoster virus
Common Name:Multi-Analyte Detection System - MMRV IgG
Product Trade NameBioPlex 2200 MMRV IgG on the BioPlex 2200 Multi-AnalyteDetection System
Device ClassClass II
Classification PanelMicrobiology
Regulation Number866.3520
Product CodeOPL

LEGALLY MARKETED EQUIVALENT (SE) DEVICES

Comparative FDA Cleared510(k) NumberDecision
PREDICATE DEVICEDate
BioMerieux VIDAS Measles IgG (MSG)510(k) ExemptN/A
BioMerieux VIDAS Mumps IgG (MPG)510(k) ExemptN/A
Bio-Rad Rubella IgG EIAK9610539/14/1996
BioMerieux VIDAS Varicella-Zoster IgG (VZV)K92312210/26/1992

DEVICE DESCRIPTION

The BioPlex 2200 MMRV IgG kit uses multiplex flow immunoassay, a methodology that greatly resembles traditional EIA, but permits simultaneous detection and identification of many antibodies in a single tube. Four (4) different populations of dyed beads are coated with antigens to identify the presence of IgG class antibodies associated with Measles, Mumps, Rubella and Varicella-zoster. The BioPlex 2200 System combines an aliquot of patient sample, sample diluent, and bead set reagent into a reaction vessel. The mixture is incubated at 37°C. After a wash cycle, anti-human IgG antibody, conjugated to phycoerythrin (PE), is added to the dyed beads and this mixture is incubated at 37°C. The excess conjugate is removed in another wash cycle, and the beads are resuspended in wash buffer. The bead mixture then passes through the detector.

The identity of the dyed beads is determined by the fluorescence of the dyes, and the amount of antibody captured by the antigen is determined by the fluorescence of the attached PE. Raw data is calculated in relative fluorescence intensity (RFI). Three additional dyed beads, an Internal Standard Bead (ISB), a Serum Verification Bead (SVB) and a Reagent Blank Bead (RBB) are present in each reaction mixture to verify detector response, the addition of serum to the reaction vessel and the absence of significant non-specific binding in serum.

The instrument is calibrated using a set of three (3) distinct calibrator vials, supplied separately by Bio-Rad Laboratories.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black background. The text is bold and sans-serif.

KIT COMPONENTS

BioPlex 2200 MMRV IgG Reagent Pack ( 665-2450). The reagent pack contains supplies sufficient for 100 tests.

VialDescription
Bead SetOne (1) 10 mL vial, containing dyed beads coated with Measles (Rubeola),Mumps, Rubella, and VZV antigens plus an Internal Standard bead (ISB), aSerum Verification bead (SVB), and a Reagent Blank bead (RBB) in bufferwith Glycerol and protein stabilizers (bovine). ProClin™ 300 (0.3%), sodiumbenzoate (0.1%) and sodium azide (< 0.1%) as preservatives.
ConjugateOne (1) 5 mL vial, containing phycoerythrin conjugated murine monoclonalanti-human IgG antibody and phycoerythrin conjugated murine monoclonalanti-human FXIII antibody, in buffer with protein stabilizers (bovine).ProClin™ 300 (0.3%), sodium benzoate (0.1%) and sodium azide (< 0.1%)as preservatives.
Sample DiluentOne (1) 10 mL vial, containing buffer with protein stabilizers (bovine andmurine). ProClin™ 300 (0.3%), sodium benzoate (0.1%) and sodium azide(<0.1%) as preservatives.

ADDITIONAL REQUIRED ITEMS, AVAILABLE FROM BIO-RAD

Catalog #Description
663-2400BioPlex 2200 MMRV IgG Calibrator Set: Three (3) 0.5 mL vials, eachcontaining human IgG antibodies to Measles, Mumps, Rubella, and VZV, ina human serum matrix made from defibrinated plasma. All antibodies arederived from human disease state plasma. All calibrators contain ProClin™300 (0.3%), sodium benzoate (0.1%) and sodium azide (<0.1%) aspreservatives
663-2430BioPlex 2200 MMRV IgG Control Set: Two (2) 1.5 mL Positive Controlserum vials, each containing human IgG antibodies to Measles, Mumps,Rubella, and VZV, in a human serum matrix made from defibrinatedplasma; and two (2) 1.5 mL Negative Control serum vials, in a humanserum matrix made from defibrinated plasma. All antibodies are derivedfrom human disease state plasma. All controls contain ProClin™ 300(0.3%), sodium benzoate (0.1%) and sodium azide (<0.1%) aspreservatives.
660-0817BioPlex 2200 System Sheath Fluid: Two (2) 4 L bottles containingPhosphate Buffered Saline (PBS). ProClin® 300 (0.3%) and Sodium azide(0.1%) as preservatives.
660-0818BioPlex 2200 System Wash Solution: One (1) 10 L bottle containingPhosphate Buffered Saline (PBS) and Tween 20. ProClin® 300 (0.3%) andSodium azide (0.1%) as preservatives.
660-0000BioPlex 2200 Instrument and Software.

INTENDED USE / INDICATIONS FOR USE

The BioPlex™ 2200 MMRV IgG kit is a multiplex flow immunoassay intended for the qualitative detection of IgG antibodies to Measles, Mumps, Rubella, and Varicella-zoster virus (VZV) in human serum and EDTA or heparinized plasma. The BioPlex 2200 MMRV IgG kit is intended for use with the Bio-Rad BioPlex 2200 System.

This kit is intended as an aid in the determination of serological status to Measles, Mumps, Rubella, and VZV. This kit is not intended for use in screening blood or plasma donors.

The performance of this assay has not been established for use in neonates, pediatrics and immunocompromised patients, or for use at point of care facilities.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is white text on a black background. The text is in a bold, sans-serif font.

TECHNOLOGICAL CHARACTERISTICS

The following tables summarize similarities and differences between the BioPlex 2200 MMRV IgG and the predicate Measles IgG, Mumps IgG, Rubella IgG, and VZV IgG devices used in comparative studies with the BioPlex 2200 MMRV IgG Kit.

A. Measles IgG

Table: Similarities between reagents and materials - Measles

Similarities betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate Measles IgG ELFA
ReagentsSample Diluent, Wash BufferSample Diluent, Wash.
ControlsNegative Control and Multi-Analyte Positive Control.Negative Control and PositiveControl.

Table: Similarities between reagents with regard to function and use - Measles

Similarities betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate Measles IgG ELFA
Analyte DetectionQualitative detection of IgGantibodies to Measles.Qualitative detection of IgGantibodies to Measles.

Table: Differences between reagents and materials - Measles

Differences betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate Measles IgG ELFA
Solid PhaseBead reagent - dyed antigencoated beads.60 - antigen coated Solid PhaseReceptacle
ReagentsConjugate: murine anti-humanIgG / Phycoerythrin.Conjugate: murine anti-humanIgG conjugated to alkalinephosphatase,fluorescent substrate, andblock/pre-wash
Sheath FluidSheath Fluid is used to suspendthe bead reagent and introduce itinto the detector.Not similar; not utilized by thisassay.
Calibrator(s)Calibrators.Standard.

Table: Differences between reagents with regard to function and use - Measles

Differences betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate Measles IgG ELFA
Analyte DetectionQualitative multi-analyte detectionof human IgG antibodies toMeasles, Mumps, Rubella, andVaricella Zoster Virus.Qualitative detection of humanIgG antibodies to Measles.
MatricesSerum, EDTA, or HeparinizedPlasma.Serum.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Bio-Rad Laboratories. The logo consists of the text "BIO-RAD" in white letters against a black background. The text is bold and sans-serif, and the logo has rounded corners.

B. Mumps IgG

Table: Similarities between reagents and materials - Mumps

Similarities betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate Mumps IgG ELFA
ReagentsSample Diluent, Wash Buffer.Sample Diluent, Wash.
ControlsNegative Control and Multi-Analyte Positive Control.Negative Control and PositiveControl.

Table: Similarities between reagents with regard to function and use - Mumps

Similarities betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate Mumps IgG ELFA
Analyte DetectionQualitative detection of IgGantibodies to Mumps.Qualitative detection of IgGantibodies to Mumps.

Table: Differences between reagents and materials - Mumps

Differences betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate Mumps IgG ELFA
Solid PhaseBead reagent - dyed antigencoated beads.60 - antigen coated Solid PhaseReceptacle
ReagentsConjugate: murine anti-humanlgG / Phycoerythrin.Conjugate: murine anti-humanIgG conjugated to alkalinephosphatase,fluorescent substrate, andblock/pre-wash
Sheath FluidSheath Fluid is used to suspendthe bead reagent and introduce itinto the detector.Not similar; not utilized by thisassay.
Calibrator(s)Calibrators.Standard.

Table: Differences between reagents with regard to function and use - Mumps

Differences betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate Mumps IgG ELFA
Analyte DetectionQualitative multi-analyte detectionof human IgG antibodies toMeasles, Mumps, Rubella, andVaricella Zoster Virus.Qualitative detection of humanlgG antibodies to Mumps.
MatricesSerum, EDTA, or HeparinizedPlasma.Serum.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black background. The text is bold and sans-serif.

C. Rubella IgG

Table: Similarities between reagents and materials - Rubella

Similarities betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate Rubella IgG EIA
ReagentsSample Diluent, Wash Buffer.Sample Diluent, WashConcentrate.
ControlsNegative Control and multi-analyte Positive Control.Negative Control, Low PositiveControl, High Positive Control.
Calibrator(s)Calibrators.Calibrators.

Table: Similarities between reagents with regard to function and use - Rubella

Similarities betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate Rubella IgG EIA
Analyte DetectionQualitative detection of IgGantibodies to Rubella.Qualitative detection of IgGantibodies to Rubella.

Table: Differences between reagents and materials - Rubella

Differences betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate Rubella IgG EIA
Solid PhaseBead reagent - dyed antigencoated beads.96 well microplate - antigencoated microwells.
ReagentsConjugate: murine anti-humanIgG / Phycoerythrin.Conjugate: goat anti-human IgG,Substrate, Stop Reagent.
Sheath FluidSheath Fluid is used to suspendthe bead reagent and introduce itinto the detector.Not similar; not utilized in EIA's.

Table: Differences between reagents with regard to function and use - Rubella

Differences betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate Rubella IgG EIA
Analyte DetectionQualitative, multi-analytedetection of human IgGantibodies to Measles, Mumps,Rubella, and Varicella ZosterVirus.Qualitative, semi-quantitative,and quantitative detection of IgGantibodies to Rubella.
MatricesSerum, EDTA, or HeparinizedPlasma.Serum.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Bio-Rad. The logo is in white text on a black, rounded rectangle. The text is bold and sans-serif.

D. VZV IgG

Table: Similarities between reagents and materials - VZV

Similarities betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate VZV IgG ELFA
ReagentsSample Diluent, Wash Buffer.Sample Diluent, Wash.
ControlsNegative Control and Multi-Analyte Positive Control.Negative Control and PositiveControl.

Table: Similarities between reagents with regard to function and use - VZV

Similarities betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate VZV IgG ELFA
Analyte DetectionQualitative detection of IgGantibodies to VZV.Qualitative detection of IgGantibodies to VZV.

Table: Differences between reagents and materials - VZV

Differences betweenComponents /MaterialsBioPlex 2200 MMRV IgG KitPredicate VZV IgG ELFA
Solid PhaseBead reagent - dyed antigencoated beads.60 - antigen coated Solid PhaseReceptacle
ReagentsConjugate: murine anti-humanlgG / Phycoerythrin.Conjugate: murine anti-humanIgG conjugated to alkalinephosphatase,fluorescent substrate, andblock/pre-wash
Sheath FluidSheath Fluid is used to suspendthe bead reagent and introduce itinto the detector.Not similar; not utilized by thisassay.
Calibrator(s)Calibrators.Standard.

Table: Differences between reagents with regard to function and use - VZV

Differences betweenFunction and UseBioPlex 2200 MMRV IgG KitPredicate VZV IgG ELFA
Analyte DetectionQualitative multi-analyte detectionof human IgG antibodies toMeasles, Mumps, Rubella, andVaricella Zoster Virus.Qualitative detection of humanlgG antibodies to VZV.
MatricesSerum, EDTA, or HeparinizedPlasma.Serum.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in black and white, with the words "BIO-RAD" in a bold, sans-serif font. The logo is simple and recognizable.

PERFORMANCE SUMMARY

A. Expected Values

The prevalence and expected values for each of the MMRV IgG assays were determined in a prospective study using serum samples from pregnant women (N = 396); and serum samples with a Measles, Mumps, Rubella, or VZV IgG test ordered for routine testing. The routine testing population included samples from pre-employment screening (N=393) and samples from subjects with a Measles test ordered (N=197), Mumps test ordered (N=199), Rubella test ordered (N=199), or VZV Test ordered (N=195). Results are shown in the tables below.

ResultNAgePos (%)Eqv (%)Neg (%)
Measles IgG5914 - 2056 (94.9%)1 (1.7%)2 (3.4%)
33621 - 47305 (90.8%)8 (2.4%)23 (6.8%)
396Total*362 (91.4%)9 (2.3%)25 (6.3%)
Mumps IgG5914 - 2054 (91.5%)1 (1.7%)4 (6.8%)
33621 - 47295 (87.8%)10 (3.0%)31 (9.2%)
396Total*349 (88.1%)11 (2.8%)36 (9.1%)
Rubella IgG5914 - 2057 (96.6%)0 (0.0%)2 (3.4%)
33621 - 47314 (93.5%)8 (2.4%)14 (4.2%)
396Total*372 (93.9%)8 (2.0%)16 (4.0%)
VZV IgG5914 - 2051 (86.4%)1 (1.7%)7 (11.9%)
33621 - 47296 (88.1%)7 (2.1%)33 (9.8%)
396Total*348 (87.9%)8 (2.0%)40 (10.1%)

Table: Pregnant Women

*The total includes one sample from a subject of unknown age.

Note: Due to rounding, numbers across columns may not total 100%.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in white text on a black background. The text is bold and sans-serif.

ResultGenderNAgePos (%)Eqv (%)Neg (%)Total N
Measles IgGF168-2014 (87.5%)1 (6.3%)1 (6.3%)590
M148-2011 (78.6%)0 (0.0%)3 (21.4%)
F23221-87209 (90.1%)3 (1.3%)20 (8.6%)
M32821-87297 (90.5%)12 (3.7%)19 (5.8%)
Total531 (90.0%)16 (2.7%)43 (7.3%)
Mumps IgGF198-2017 (89.5%)2 (10.5%)0 (0.0%)592
M188-2016 (88.9%)0 (0.0%)2 (11.1%)
F27021-87239 (88.5%)9 (3.3%)22 (8.1%)
M28521-87252 (88.4%)11 (3.9%)22 (7.7%)
Total524 (88.5%)22 (3.7%)46 (7.8%)
Rubella IgGF278-2022 (81.5%)0 (0.0%)5 (18.5%)592
M118-208 (72.7%)2 (18.2%)1 (9.1%)
F31121-87278 (89.4%)8 (2.6%)25 (8.0%)
M24321-87210 (86.4%)10 (4.1%)23 (9.5%)
Total518 (87.5%)20 (3.4%)54 (9.1%)
VZV IgGF188-2014 (77.8%)0 (0.0%)4 (22.2%)588
M198-2013 (68.4%)0 (0.0%)6 (31.6%)
F25421-87229 (90.2%)5 (2.0%)20 (7.9%)
M29721-87270 (90.9%)7 (2.4%)20 (6.7%)
Total526 (89.5%)12 (2.0%)50 (8.5%)

Table: Measles, Mumps, Rubella, or VZV Test Ordered

Note: Due to rounding numbers across columns may not total 100%.

B. Reproducibility Studies

To assess reproducibility of each of the assays in the BioPlex 2200 MMRV IgG kit, a reproducibility panel was prepared at Bio-Rad Laboratories. The positive panel members were prepared by combining one or more antibody positive patient samples in 3 different matrices (serum, EDTA plasma, and sodium heparin plasma) with one or more of the 4 analytes contained in the BioPlex 2200 MMRV IgG (Measles, Mumps, Rubella, and VZV). The panel contained members with varying levels of antibodies to the analytes in the BioPlex 2200 MMRV IgG kit, and a positive control (antibody positive for all analytes). Reproducibility testing was performed at 3 U.S. clinical trial sites. Three lots of BioPlex 2200 MMRV IgG Reagent Packs, 3 lots of BioPlex 2200 MMRV IgG Calibrator Set and 3 lots of BioPlex 2200 MMRV IgG Control Set were used to evaluate reproducibility. Each site evaluated 1 lot of the BioPlex 2200 MMRV IgG kit. Each of the panel members and a positive and negative control was tested in quadruplicate on 1 run per day over 5 days at each of 3 sites (4 replicates x 1 run x 5 days = 20 replicates per panel member per site = 60 total replicates for 3 sites). The data were analyzed for intra-assay and inter-assay reproducibility according to the Clinical and Laboratory Standards Institute (CLSI) guidance EP15-A2 (Vol. 25, No. 17). The mean Antibody Index (Al), standard deviation (SD), and percent coefficient of variation (%CV) for each panel member were calculated. Serum results are shown in the tables below.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Bio-Rad Laboratories. The logo is white text on a black background. The text is in a bold, sans-serif font. The logo is surrounded by a rounded rectangle.

Table: Reproducibility: BioPlex 2200 Measles IgG Serum

Measles IgGPanel MembersSampleNGrandMean(AI)Within-RunBetween-DayBetween-Site*Total
SD%CVSD%CVSD%CVSD%CV
High Positive 1602.50.1124.40.1385.50.1857.30.25610.2
High Positive 2602.10.0833.90.1266.00.25011.90.29213.9
Low Positive 1601.50.0775.30.0573.90.21214.60.23316.0
Low Positive 2601.80.0774.40.1327.40.18610.40.24113.5
Near Cutoff 1600.90.0879.80.0637.20.0758.50.13114.8
Near Cutoff 2600.90.0667.30.0667.40.0000.00.09310.4
High Negative 1600.70.0425.70.0354.80.0658.80.08511.5
High Negative 2600.50.0316.00.0173.40.0336.40.0489.4
Positive Control603.00.0913.10.1194.00.33211.20.36412.3
  • Between site includes between lot variance.

Table: Reproducibility: BioPlex 2200 Mumps IgG Serum

Mumps IgGPanel MembersSampleNGrandMean(AI)Within-RunBetween-DayBetween-Site*Total
SD%CVSD%CVSD%CVSD%CV
High Positive 1602.70.0833.10.0782.90.1605.90.1967.2
High Positive 2603.10.1083.40.1043.30.2839.00.32110.2
Low Positive 1601.70.0603.60.0442.60.0583.50.0945.7
Low Positive 2601.70.0623.60.0693.90.0754.30.1196.9
Near Cutoff 1600.80.0475.60.0354.20.0242.80.0647.6
Near Cutoff 2600.90.0445.00.0222.50.0424.80.0657.4
High Negative 1600.50.0244.70.0112.10.0051.00.0265.3
High Negative 2600.70.0365.30.0365.30.0263.80.0578.4
Positive Control602.60.0612.40.0692.70.1696.60.1937.5
  • Between site includes between lot variance.

Table: Reproducibility: BioPlex 2200 Rubella IgG Serum

Rubella IgGPanel MembersSampleNGrandMean(AI)Within-RunBetween-DayBetween-Site*Total
SD%CVSD%CVSD%CVSD%CV
High Positive 1601.90.0834.40.0512.70.1638.70.19010.1
High Positive 2602.80.0923.30.0833.00.40314.50.42115.2
Low Positive 1601.40.0685.00.0735.30.0453.30.1108.0
Low Positive 2601.80.0834.70.0995.60.1679.50.21112.1
Near Cutoff 1600.80.0395.00.0435.50.0395.10.0709.0
Near Cutoff 2600.80.0516.20.0334.10.0668.20.09011.0
High Negative 1600.50.0295.40.0163.00.0489.00.05810.9
High Negative 2600.50.0295.80.0377.40.0336.60.05711.5
Positive Control602.10.0683.30.0964.60.1919.10.22510.7

*Between site variance includes between lot variance.

. ﭘ

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Bio-Rad. The text "BIO-RAD" is written in white, bold, sans-serif font. The text is set against a black, rounded-corner rectangular background.

VZV IgGPanel MembersSampleNGrandMean(AI)Within-RunBetween-DayBetween-Site*Total
SD%CVSD%CVSD%CVSD%CV
High Positive 1602.80.0702.40.0551.90.2388.40.2549.0
High Positive 2603.50.0832.40.1033.00.3239.30.34910.0
Low Positive 1601.10.0474.30.0696.30.0585.30.1019.3
Low Positive 2601.20.0554.50.0645.20.0000.00.0846.8
Near Cutoff 1600.80.0394.90.0678.50.0000.00.0779.8
Near Cutoff 2601.00.0494.70.0757.30.0000.00.0908.7
High Negative 1600.70.0243.40.0304.20.0212.90.0436.1
High Negative 2600.50.0357.30.05110.60.0224.70.06613.6
Positive Control602.40.0582.40.0622.50.1747.10.1947.9

Table: Reproducibility: BioPlex 2200 VZV IgG Serum

  • Between site includes between lot variance.

C. Comparative Testing

Comparative Testing: Prospective

Performance of the MMRV IgG kit was evaluated against corresponding commercially available Measles, Mumps, Rubella, and VZV immunoassays. Serum samples from pregnant women (N = 396) and serum samples with a Measles, Mumps, Rubella, or VZV IgG test ordered (N = 1183) were tested at 3 U.S. clinical trial sites. Serum samples with a Measles, Mumps, Rubella, or VZV IgG test ordered included serum samples submitted for routine testing (N = 790) and serum samples for pre-employment evaluation (N = 393). Results for all populations are shown in the tables below. Equivocal results obtained for Mumps and Measles on the commercially available ElAs were further tested on two additional commercially available EIAs for consensus. The consensus results are presented in the tables for Mumps and Measles, below. Equivocal results obtained for Rubella on the commercially available EIA were re-tested on the commercially available EIA, following the manufacturer's recommendations. Re-test results are presented in the tables for Rubella, below.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Bio-Rad. The text "BIO-RAD" is written in white, bold letters. The background is black and shaped like a rounded rectangle.

Pregnant WomenBioPlex 2200 MMRV IgG
Pos (+)EqvNeg (-)TotalPos (+)% AgreementNeg (-)% Agreement
Commercially Available EIAMeasles IgGPos (+)3629938093.3%(362/388)100%(8/8)
Eqv*0088
Neg (-)0088
Total36292539695% CI90.4 - 95.4%95% CI67.5 - 100%
Mumps IgGPos (+)349101137094.3%(349/370)96.2%(25/26)
Eqv*0000
Neg (-)01252695% CI
Total349113639695% CI91.5 - 96.3%95% CI81.1 - 99.3%
Rubella IgGPos (+)3696237797.9%(369/377)73.7%(14/19)
Eqv**0000
Neg (-)32141995% CI
Total37281639695% CI95.9 - 98.9%95% CI51.2 - 88.2%
VZV IgGPos (+)3488436095.1%(348/366)100%(30/30)
Eqv0066
Neg (-)00303095% CI
Total34884039695% CI92.4 - 96.9%95% CI88.6 - 100%

Table: Pregnant Women: BioPlex 2200 MMRV IgG vs. EIA (N = 396)

  • Results obtained by a consensus of two out of three commercially available EIAs.

** Result after re-testing with the commercially available EIA following the manufacturer's recommendations.

BioPlex 2200 MMRV IgG
Measles, Mumps, Rubella, or VZV Test OrderedPos (+)EqvNeg (-)TotalPos (+)% AgreementNeg (-)% Agreement
Measles IgGPos (+)10422526109394.7%(1042/1100)80.2%(65/81)
Eqv*22711
Neg (-)4106579
Total10483798118395% CI93.2 - 95.9%95% CI70.3 - 87.5%
Mumps IgGPos (+)10065149110690.4%(1006/1113)91.0%(61/67)
Eqv*03710
Neg (-)156167
Total100759117118395% CI88.5 - 92.0%95% CI81.8 - 95.8%
Rubella IgGPos (+)10603027111794.8%(1060/1118)86.2%(56/65)
Eqv**0011
Neg (-)455665
Total10643584118395% CI93.3 - 96.0%95% CI75.7 - 92.5%
VZV IgGPos (+)10471636109994.0%(1047/1114)98.4%(60/61)
Eqv181524
Neg (-)006060
Total104824111118395% CI92.4 - 95.2%95% CI91.3 - 99.7%

Table: Test Ordered Samples: BioPlex 2200 MMRV IgG vs. EIA (N = 1183)

  • Results obtained by a consensus of two out of three commercially available EIAs.

** Result after re-testing with the commercially available EIA following the manufacturer's recommendations.

BioPlex 2200 MMRV IgG 510(k) Summary (3-25-10)

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in white text on a black background. The text is bold and sans-serif.

Comparative Testing: Retrospective

Performance of the MMRV IgG kit was evaluated against corresponding commercially available Measles, Mumps, Rubella, and VZV immunoassays using retrospective serum samples negative for antibodies to Measles (N = 93), Mumps (N = 96), Rubella (N = 268), or VZV (N = 143). The negative samples for Measles and Mumps were selected using a consensus of two out of three commercially available EIAs. Negative samples for Rubella and VZV were selected using the respective commercially available ElAs used for the comparative analysis. Results are shown in the table below.

RetrospectiveMeasles, Mumps,Rubella, or VZVNegative SamplesBioPlex 2200 MMRV IgGNeg (-)% Agreement
Measles IgG(N = 93)Pos (+)0000100%(93/93)
Eqv0000
Neg (-)00939395% CI
Total00939396.0 - 100%
Mumps IgG(N = 96)Pos (+)0066100%(83/83)
Eqv0077
Neg (-)00838395% CI
Total00969695.6 - 100%
Rubella IgG(N = 268)Pos (+)000095.9%(257/268)
Eqv0000
Neg (-)3825726895% CI
Total3825726892.8 - 97.7%
VZV IgG(N = 143)Pos (+)0000100%(143/143)
Eqv0000
Neg (-)0014314395% CI
Total0014314397.4 - 100%

Table: Retrospective Negative Samples: BioPlex 2200 MMRV IgG vs. EIA

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for Bio-Rad Laboratories, a company that manufactures and distributes products for the life science research and clinical diagnostics markets. The logo is in white text on a black background. The text is in a bold, sans-serif font. The logo is simple and recognizable.

Correlation with CDC Rubella Evaluation Serum Panel

The performance of the BioPlex 2200 MMRV IgG kit was assessed using a masked, characterized Rubella IgG serum panel from the CDC. The panel consisted of 82 positive samples and 18 negative samples. The results are presented as a means to convey further information on the performance of the test kit and do not imply endorsement of the assay by the CDC. Results are shown in the table below.

Table: Characteristics of CDC Rubella IgG Reference Sera (N = 100)

CDC SamplesReference ResultBioPlex 2200 Pos (+)BioPlex 2200 Neg (-)
Rubella IgGPos (+)811*
Neg (-)018
  • Note: The panel consisted of 100 specimens (50 pairs). There are 9 negative sera, resulting in 18 negative sera resulting in 82 posive samples. One of the positive samples tested negative for rubella with the BioPlex 2200 MMRV IgG kit while the other member of this pair tested positive.

Correlation of the BioPlex 2200 MMRV IgG Rubella IgG Assay with the WHO Standard

Dilutions of the WHO anti-Rubella immunoglobulin, 1st International Standard, RUBI-1-94, in the range of 0 - 20 IU/ml, were analyzed with the BioPlex 2200 Rubella IgG assay. Results can be found in the table below.

Expected IU/mLBioPlex 2200
Measured Mean Al (1.0 Al= 10 IU/ml)Mean Percent of Expected
0.00.0--
5.00.6121
7.50.8109
10.01.099
12.51.191
15.01.388
17.51.482
20.01.680

Table: Rubella World Health Organization RUBI-1-94 Sample Evaluation

CDC Rubella Low Positive Control Testing

A lyophilized CDC low positive control serum was prepared by and tested at one clinical trial site. The control was tested neat and dilution was run in duplicate. Results are shown in the table below.

Table: Mean Low-Titer Anti-Rubella Human Reference Serum CDC Biological Standard Results (N = 2)

DilutionMean(AI)
Neat - Prepared per package Insert2.3
1/2 Dilution1.4

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image contains the words BIO RAD in white font on a black rounded rectangle background. The words are written in all capital letters. The font is sans-serif and bold.

D. Cross-Reactivity

A cross-reactivity study was performed to determine if samples from various disease states and other potentially cross-reacting agents interfere with test results when tested with the BioPlex 2200 MMRV IgG kit. Samples known to be positive for one of the thirteen potential crossreactants listed in the table below were evaluated with the BioPlex 2200 MMRV IgG assays. Due to the high prevalence of IgG antibodies to Measles, Mumps, Rubella, and VZV in the population, and in order to properly evaluate the potential cross-reactivity, all samples were pre-tested by commercially available Measles, Mumps, Rubella and VZV assays and only those that tested negative by the commercially available assay were further tested by the BioPlex 2200 MMRV IgG kit. The table below summarizes negative agreement between the BioPlex 2200 MMRV IgG assays and the corresponding commercially available Measles, Mumps, Rubella and VZV assays within each of the thirteen cross-reactant panels. The results demonstrate that the various disease state samples evaluated do not cross-react with the 4 antigens in the BioPlex 2200 MMRV IgG kit. Please note that for HCV Ab and ANA IgG, potential cross-reactivity was not well assessed due to the limitation of the sample size tested.

Number of Negative BioPlex 2200 Results /Number of Negative Commercially Available Assay Results
PotentialCross-ReactantMeaslesMumpsRubellaVZV
NNegativeAgreementNNegativeAgreementNNegativeAgreementNNegativeAgreement
ANA IgG55/555/52*2/255/5
CMV IgG1010/101010/101010/101010/10
EBV VCA IgG1010/101010/101010/101010/10
HBsAb IgG1010/101010/101010/101010/10
HCV Ab3*3/33*3/31*1/11*1/1
HSV-1 IgG1010/101010/101010/101010/10
HSV-2 IgG1010/101010/101010/101010/10
Measles IgG-N/A1010/101010/101010/10
Mumps IgG1010/10-N/A1010/101010/10
Parvovirus B19 IgG1010/101010/101010/101010/10
Rubella IgG1010/101010/10-N/A1010/10
Toxoplasma IgG1010/1066/688/81010/10
VZV IgG1010/101010/101010/10-N/A

Table: Cross-Reactivity

  • Potential cross-reactivity was not well assessed due to the limitation of the sample size tested.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Mail Center-WO66-G609 Silver Spring, MD 20993-0002

Bio-Rad Laboratories, Inc. c/o Mr. David Bhend Regulatory Affairs Representative Bio-Rad Laboratories - Redmond Operations 6565 185th Ave NE Redmond, WA 98052

MAR 2 9 2010

Re: K091616

Trade/Device Name: BioPlex 2200 MMRV IgG on the BioPlex 2200 Multi-Analyte Detection System Regulation Number: 21 CFR& 866.3510 Regulation Name: Rubella Virus Serological Reagents Regulatory Class: Class II Product Code: OPL, LJB, LJB, LJY, LFY Dated: March 25, 2010 Received: March 26, 2010.

Dear Mr. Bhend:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{16}------------------------------------------------

Page 2 - David Bhend

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Sary astoyma

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{17}------------------------------------------------

Indication for Use

510(k) Number (if known): K091616

Device Name: BioPlex 2200 MMRV IgG kit on the BioPlex 2200 Multi-Analyte Detection System

Indication For Use:

The BioPlex™ 2200 MMRV IgG kit is a multiplex flow immunoassay intended for the qualitative detection of IgG antibodies to Measles, Mumps, Rubella, and Varicella-zoster virus (VZV) in human serum and EDTA or heparinized plasma. The BioPlex 2200 MMRV IgG kit is intended for use with the Bio-Rad BioPlex 2200 System.

This kit is intended as an aid in the determination of serological status to Measles, Mumps, Rubella, and VZV. This kit is not intended for use in screening blood or plasma donors.

The performance of this assay has not been established for use in neonates, pediatrics and immunocompromised patients, or for use at point of care facilities.

Prescription Use X (21 CFR Part 801 Subpart D) And/Or

Over the Counter Use (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OVD)

Uwe Schirf

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) K091616

§ 866.3510 Rubella virus serological reagents.

(a)
Identification. Rubella virus serological reagents are devices that consist of antigens and antisera used in serological tests to identify antibodies to rubella virus in serum. The identification aids in the diagnosis of rubella (German measles) or confirmation of a person's immune status from past infections or immunizations and provides epidemiological information on German measles. Newborns infected in the uterus with rubella virus may be born with multiple congenital defects (rubella syndrome).(b)
Classification. Class II. The special controls for this device are:(1) National Committee for Clinical Laboratory Standards':
(i) 1/LA6 “Detection and Quantitation of Rubella IgG Antibody: Evaluation and Performance Criteria for Multiple Component Test Products, Speciment Handling, and Use of the Test Products in the Clinical Laboratory, October 1997,”
(ii) 1/LA18 “Specifications for Immunological Testing for Infectious Diseases, December 1994,”
(iii) D13 “Agglutination Characteristics, Methodology, Limitations, and Clinical Validation, October 1993,”
(iv) EP5 “Evaluation of Precision Performance of Clinical Chemistry Devices, February 1999,” and
(v) EP10 “Preliminary Evaluation of the Linearity of Quantitive Clinical Laboratory Methods, May 1998,”
(2) Centers for Disease Control's:
(i) Low Titer Rubella Standard,
(ii) Reference Panel of Well Characterized Rubella Sera, and
(3) World Health Organization's International Rubella Standard.