K Number
K091615

Validate with FDA (Live)

Device Name
MED CONTOUR
Date Cleared
2010-05-06

(337 days)

Product Code
Regulation Number
890.5300
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MC1 is indicated for: a) Therapeutic Massager: 1. Provides temporary relief of minor muscle aches and pains; 2. Relieves muscle spasms, 3. Temporarily improves local blood circulation; 4. Temporarily reduces the appearance of cellulite. b) Ultrasonic Diathermy: 1. Relief of pain; 2. Muscle spasms; 3. Joint contractures; 4. NOT for the treatment of malignancies.

Device Description

The MC1 is a computerized body massager and ultrasound diathermy system. The MC1 is supplied with two handpieces: one ultrasonic handpiece with two 1 MHz transducers mounted on it and one handpiece for computerized body massage.

AI/ML Overview

Here's a summary of the acceptance criteria and study information for the General Project MC1, based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Electrical SafetyConforms with IEC 60601-1
Electromagnetic Compatibility (EMC)Conforms with IEC 60601-1-2
Electrical Safety in Ultrasonic Physiotherapy EquipmentConforms with IEC 60601-2-5
Tissue Heating for Ultrasonic Diathermy (Required for therapeutic effect, specifically for tissue temperature to reach at least 40° C)The MC1 increases tissue temperature as required for ultrasonic diathermy; "The two 1 MHz ultrasound transducers can be considered equivalent to the MED Sculpt in terms of heating tissue temperature to at least 40° C." Results were similar to the predicate device (MED Sculpt).

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a "test set" in the context of patient or image data for evaluating clinical performance. The performance testing described focuses on engineering and physical performance (electrical safety, EMC, tissue heating).

  • Sample Size: Not applicable in the traditional sense of a clinical test set. The evaluation compared the MC1's physical performance to a predicate device, the MED Sculpt.
  • Data Provenance: Not applicable. The "data" refers to measurements from the device itself during controlled testing, not data derived from human subjects or retrospective/prospective studies on patient outcomes.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

  • Number of Experts: Not applicable. Ground truth for clinical performance was not established by experts in this context.
  • Qualifications of Experts: Not applicable.

4. Adjudication Method for the Test Set

  • Adjudication Method: Not applicable. No human readers or adjudicators were involved in evaluating the device's technical performance against a ground truth.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No. The submission describes technical and performance testing of the device itself, not a study evaluating human reader performance with or without AI assistance.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

  • Was a standalone study done? This question is generally relevant for AI/software devices. For this physical medical device (ultrasound and massager), a "standalone" study in the AI sense is not applicable. The performance testing was of the device operating on its own.

7. Type of Ground Truth Used

  • Type of Ground Truth: The ground truth for the device's physical performance was based on established engineering standards (IEC) for safety and electromagnetic compatibility, and a specific temperature threshold (at least 40° C) for tissue heating as a measure of therapeutic efficacy for ultrasonic diathermy. This could be considered measured physical properties and engineering standards. It was not pathology, expert consensus on clinical cases, or outcomes data.

8. Sample Size for the Training Set

  • Sample Size: Not applicable. This device does not use an "AI" or "machine learning" algorithm that requires a training set. The device's function is mechanical and ultrasonic.

9. How the Ground Truth for the Training Set Was Established

  • How Ground Truth Established: Not applicable, as there is no training set for an AI algorithm.

{0}------------------------------------------------

K091615

MAY - 6 2010

510(k) Summary For the General Project MC1

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

1. General Information

Submitter:

General Project, S.r.I. Via della Gora 13/19-15/19 50025 Montespertoli, Florence Italy

Contact Person:

Maureen O'Connell O'Connell Regulatory Consultants, Inc. 5 Timber Lane North Reading, MA 01864 Tel: 978-207-1245 Fax: 978-824-2541

Summary Preparation Date:

  1. Names

Trade Name: Common Name: Classification Name: MC1 Ultrasound and Massager Therapeutic Massager Diathermy, Ultrasonic, For Use in Applying Therapeutic Deep heat. Product Codes: ISA (21 CFR 890.5660), IMI (21 CFR 890.5300)

3. Legally Marketed Predicate Devices

The MC1 is substantially equivalent to the following devices: MED Sculpt, manufactured by General Project (K053041), Dermosonic, manufactured by Sybaritic, Inc. (K024307), Forte CPS 400 Combo (K982830), manufactured by Chattanooga Group, Inc. and Omnisound 3000 (K883893), manufactured by Physio Technology, Inc.

April 26, 2010

{1}------------------------------------------------

4. Device Description

The MC1 is a computerized body massager and ultrasound diathermy system. The MC1 is supplied with two handpieces: one ultrasonic handpiece with two 1 MHz transducers mounted on it and one handpiece for computerized body massage.

5. Intended Use

The MC1 is indicated for:

  • a) Therapeutic Massager:
      1. Provides temporary relief of minor muscle aches and pains
      1. Relieves muscle spasms
      1. Temporarily improves blood circulation
      1. Temporarily reduces the appearance of cellulite
  • b) Ultrasonic Diathermy:
      1. Relief of pain
      1. Muscle spasms
      1. Joint contractures
      1. NOT for the treatment of malignancies

6. Substantial Equivalence

The MC1 is substantially equivalent to the identified predicate devices. The MC1 has the same indications for use as the General Project MED Sculpt and the Sybaritic Dermosonic. Regarding technological characteristics, the MC1 includes both massage and ultrasound diathermy handpieces as do the General Project MED Sculpt and the Sybaritic Dermosonic. The Chattanooga Group, Inc. Forte CPS 400 Combo and the Physio Technology, Inc. Omnisound 3000 are ultrasound based devices only. Differences between the MC1 and the identified predicate devices were evaluated in performance testing and the MC1 was found to be substantially equivalent to the identified predicate devices.

7. Performance Data

The MC1 was tested and found to conform with IEC 60601-1 for electrical safety, IEC 60601-1-2 for electromagnetic compatibility and IEC 60601-2-5 for electrical safety in ultrasonic physiotherapy equipment. The MED Sculpt was found to conform with these same electrical safety standards. Performance data was presented which showed that the MC1 performed similarly to the previously cleared MED Sculpt in terms of tissue heating. Specifically, testing showed that the MC1 increases tissue temperature as required for ultrasonic diathermy. The two 1 MHz ultrasound transducers can be considered equivalent to the MED Sculpt in terms of heating tissue temperature to at least 40° C. These results are similar to those reported for the MED Sculpt.

8. Conclusion

The MC1 is substantially equivalent to the identified predicate devices in terms of indications for use, electrical safety testing and performance testing which indicated that the MC1 and the MED Sculpt were substantially equivalent in terms of tissue heating.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

General Project, S.r.l % O'Connell Regulatory Consultants, Inc. Ms. Maureen O'Connell Regulatory Consultant 5 Timber Lane North Reading, Massachusetts 01864

MAY - 6 2010

Re: K091615

Trade/Device Name: MC1 Regulation Number: 21 CFR 890.5300 Regulation Name: Ultrasonic diathermy Regulatory Class: II Product Code: IMI, ISA Dated: April 26, 2010 Received: April 27, 2010

Dear Ms. O'Connell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Ms. Maureen O'Connell

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedi and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): ---------------------------------------------------------------------------------------------

Device Name:

The MC1 is indicated for:

  • a) Therapeutic Massager:
      1. Provides temporary relief of minor muscle aches and pains;
      1. Relieves muscle spasms,
      1. Temporarily improves local blood circulation;
      1. Temporarily reduces the appearance of cellulite.

b) Ultrasonic Diathermy:

    1. Relief of pain;
    1. Muscle spasms;
    1. Joint contractures;
    1. NOT for the treatment of malignancies.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

  • X
    Prescription Use (Per 21 CFR 801.109) OR

Over The Counter Use (Optional Format 1-2-96)

Page 1 of 1

.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Division of Surgial and Restorative D. Views

Number K091615

510(k) Number

§ 890.5300 Ultrasonic diathermy.

(a)
Ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions —(1)Identification. An ultrasonic diathermy for use in applying therapeutic deep heat for selected medical conditions is a device that applies to specific areas of the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended to generate deep heat within body tissues for the treatment of selected medical conditions such as relief of pain, muscle spasms, and joint contractures, but not for the treatment of malignancies.(2)
Classification. Class II (performance standards).(b)
Ultrasonic diathermy for all other uses —(1)Identification. An ultrasonic diathermy for all other uses except for the treatment of malignancies is a device that applies to the body ultrasonic energy at a frequency beyond 20 kilohertz and that is intended for the treatment of medical conditions by means other than the generation of deep heat within body tissues as described in paragraph (a) of this section.(2)
Classification. Class III (premarket approval).(c)
Date PMA or notice of completion of PDP is required. A PMA or notice of completion of a PDP for a device described in paragraph (b) of this section is required to be filed with the Food and Drug Administration on or before July 13, 1999, for any ultrasonic diathermy described in paragraph (b) of this section that was in commercial distribution before May 28, 1976, or that has, on or before July 13, 1999, been found to be substantially equivalent to an ultrasonic diathermy described in paragraph (b) of this section that was in commercial distribution before May 28, 1976. Any other ultrasonic diathermy described in paragraph (b) of this section shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.