K Number
K091436

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-05-29

(15 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DIGITAL RADIOGRAPHY CXDI-55C provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures. This device is not intended for mammography applications.

Device Description

The DIGITAL RADIOGRAPHY CXDI-55C is a solid state x-ray imager which has 35 x 43 cm imaging area. The DIGITAL RADIOGRAPHY CXDI-55C intercepts x-ray photons and the scintillator of the CXDI-55C emits visible spectrum photons that illuminate an array of photo-detectors that create electrical signals. After the electrical signals are generated, the images are converted to digital and displayed on a monitor.

AI/ML Overview

The provided text is related to a 510(k) submission for a digital radiography device (CXDI-55C). However, it focuses on demonstrating substantial equivalence to a predicate device (CXDI-50C) rather than providing detailed acceptance criteria and a standalone study proving the device meets those criteria with specific performance metrics.

Here's a breakdown of what can be extracted and what is NOT available based on the provided text:

1. A table of acceptance criteria and the reported device performance:

This information is not present in the provided text. The document states "The Performance Data demonstrates that CXDI-55C is safe and effective just as the CXDI-50C," but it does not provide specific performance metrics or acceptance criteria in a table format.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

This information is not present in the provided text. The document refers to "Performance Data" and "non-clinical testing" but does not detail the nature of these tests, the sample sizes used, or the data provenance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

This information is not present in the provided text. There is no mention of expert involvement in establishing ground truth for any test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

This information is not present in the provided text.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

This information is not present in the provided text. The device described is a digital X-ray imager, not an AI-assisted diagnostic tool. Its comparison is to a previous model of a similar imager, not to human readers with or without AI.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

This information is not present in the provided text. The device is a hardware component for capturing X-ray images, not an algorithm that performs a diagnostic task independently.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

This information is not present in the provided text.

8. The sample size for the training set:

This information is not present in the provided text. The device is a digital radiography imager, not an AI model that requires a training set in the conventional sense.

9. How the ground truth for the training set was established:

This information is not present in the provided text.


Summary of available information related to acceptance criteria and study:

The provided document is a 510(k) summary for the Canon CXDI-55C, a solid-state x-ray imager. The primary method for gaining market clearance through this submission is demonstrating substantial equivalence to a predicate device, the Canon CXDI-50C.

  • Acceptance Criteria & Reported Performance:

    • Acceptance Criteria (Implied): Substantial equivalence to the predicate device, CXDI-50C, in terms of safety and effectiveness. This implies meeting similar performance standards to the predicate device.
    • Reported Device Performance: "The Performance Data demonstrates that CXDI-55C is safe and effective just as the CXDI-50C." However, no specific quantitative metrics (e.g., DQE, MTF, SNR) are provided to exemplify this safety and effectiveness.
    • Basis for Equivalence: The primary differences highlighted are physical characteristics:
      • External Dimensions: CXDI-55C changed from 491 x 477 x 23 mm (CXDI-50C) to 480 x 481 x 15 mm.
      • Weight: CXDI-55C changed from 4.8kg (CXDI-50C) to 3.4kg.
    • Intended Use: Identical to the predicate device: "digital image capture for conventional film/screen radiographic examinations... to replace radiographic film/screen systems in all general purpose diagnostic procedures." (Not for mammography).
  • Study Details (Limited):

    • The document refers to "design control activities and non-clinical testing" that led to the conclusion of substantial equivalence. However, no specific details about these tests are provided, such as sample sizes, methodologies, or outcomes in measurable terms.
    • There is no evidence of studies involving human readers, AI assistance, expert review for ground truth, or adjudication methods for diagnostic performance in this document. The focus is on the equivalence of the hardware's imaging capabilities.

In conclusion, this document demonstrates substantial equivalence to a predicate device rather than providing detailed acceptance criteria and a standalone study for the device's diagnostic performance. The "performance data" mentioned is not quantified or elaborated upon in this summary.

{0}------------------------------------------------

KC91436

Section 10: Summary

MAY 2 9 2009

510(k) Summary

April 6, 2009

Prepared:

Submitter:

Company Name: Company Address:

Contact Person: Phone Number: Fax Number:

Canon USA, Inc. (U.S. agent for Canon Inc.) One Canon Plaza Lake Success, NY 11042 Ms. Sheila Driscoll (516) 328-5602 (216) 328-2169

Proposed Device:

Reason for 510(k): Manufacturer: Trade Name: Model Name: Classification Name: FDA 510(k) #:

New Model Canon Inc. Canon CXDI-55C MQB, Solid State X-ray Imager To be assigned

Predicate Device:

Manufacturer: Trade Name: Model Name: Classification Name: FDA 510(k) #:

Canon Inc. Canon CXDI-50C 90MQB, Solid State X-ray Imager K060433

Description of Device:

The DIGITAL RADIOGRAPHY CXDI-55C is a solid state x-ray imager which has 35 x 43 cm imaging area.

The DIGITAL RADIOGRAPHY CXDI-55C intercepts x-ray photons and the scintillator of the CXDI-55C emits visible spectrum photons that illuminate an array of photo-detectors that create electrical signals. After the electrical signals are generated, the images are converted to digital and displayed on a monitor.

Intended Use:

The DIGITAL RADIOGRAPHY CXDI-55C provides digital image capture for conventional film/screen radiographic examinations. The device is intended to replace radiographic film/screen systems in all general purpose diagnostic procedures.

This device is not intended for mammography applications.

{1}------------------------------------------------

Section 10: Summary

Comparison to Predicate:

CXDI-55C's intended use is the same as that of CXDI-50C. The differences are the external dimensions and the weight which are:

The external dimensions of CXDI-55C are changed from 491 x 477 x 23 mm to 480 x 481 x 15 mm.

The weight of CXDI-55C is changed from 4.8kg to 3.4kg.

Conclusion:

The Performance Data demonstrates that CXDI-55C is safe and effective just as the CXDI-50C.

Based on the information in this submission, similarity to the predicate device (Digital Radiography CXDI-50C), and the results of our design control activities and non-clinical testing, it is the opinion of Canon Inc. that the DIGITAL RADIOGRAPHY CXDI-55C described in this submission is substantially equivalent to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo of the Department of Health & Human Services. The logo consists of a stylized graphic of human figures in blue, along with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in blue font. The text is positioned to the right of the graphic.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

Canon USA, Inc. % Mr. Jeff D. Rongero Third Party Reviewer-Senior Project Engineer Underwriters Laboratories, Inc. 12 Laboratory Drive Research Triangle Park, NC 27709

AUG 2 3 2013

Re: K091436 Trade/Device Name: CXDI-55C Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: May 13, 2009 Received: May 14. 2009

Dear Mr. Rongero:

This letter corrects our substantially equivalent letter of May 29, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

Page 2

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Janine M. Morri Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications Statement

510(K)Number(if known):K091436
----------------------------------

Device Name: CXDI-55C

· Indications for Use:

DIGITAL RADIOQRAPHY CXDI-55C provides digital image capture for

conventional film/screan radiographic examinations.

The device is intended to replace radiographio film/screen systems in all general purpose diagnostic procedures. "

This device is not intended for mammography applications.

Presoription Uso OR (Part 21 OFR 801 Subpert D)

Over-The-Counter Use (Part 21 OFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LIKE-CONTINUE ON ANOTHERT PAGE IF NEEDED).

Ooncurrence of CDRH, Office of Device Evaluation(ODE)

:

.

(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number