K Number
K091368

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-06-04

(27 days)

Product Code
Regulation Number
892.2050
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

R2 DigitalNow HD is a software application intended to process digitized screen-film mammograms for comparison purposes only. The software processes digitized prior film images to produce lossy-compressed DICOM images that more closely resemble digital mammography images. R2 DigitalNow HD images are intended for comparison purposes only and cannot be used for primary diagnosis.

DigitalNow HD is a software application which runs on the Hologic Cenova server (Class I exempt per 21 CFR § 892.2010 and 21 CFR § 892.2020).

Device Description

R2 DigitalNow HD is a software application intended to process digitized screen-film mammograms.

AI/ML Overview

The provided text does not contain detailed information about specific acceptance criteria or the study that rigorously proves the device meets those criteria. The submission is a 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device rather than comprehensive performance validation against defined acceptance criteria.

However, based on the implied acceptance criteria for a 510(k) submission and the information provided, we can infer some aspects:

The primary "acceptance criterion" for this 510(k) submission appears to be demonstrating substantial equivalence to the predicate device (DexTop Mammography Workstation K080351 for certain software functions) in terms of intended use, technological characteristics, and safety and effectiveness.

Here's an attempt to structure an answer based on the given information, highlighting what is present and what is absent:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criterion (Inferred from 510(k) Goal)Reported Device Performance (Summary from Document)
Substantial Equivalence to Predicate DeviceThe FDA reviewed the 510(k) and determined the device is substantially equivalent to legally marketed predicate devices for the stated indications for use.
Intended Use Equivalence"R2 DigitalNow HD is a software application intended to process digitized screen-film mammograms for comparison purposes only." This matches the general scope of such software.
Technological Characteristics EquivalenceDigitalNow HD is described as "a software application intended to process digitized screen-film mammograms." The submission would have detailed its algorithms and processing.
Safety and Effectiveness"The 510(k) Pre-Market Notification for DigitalNow HD contains adequate information and data to enable FDA - CDRH to determine substantial equivalence..." "General Safety and Effectiveness Concerns: The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management is ensured via a risk analysis, which is used to identify potential hazards."
Compliance with StandardsDesigned and manufactured in accordance with ISO 13485, ISO 14971, ANSI/AAMI SW68:2001, 21 CFR § 820.

Note: The document explicitly states, "The submission contains the results of a hazard analysis and the 'Level of Concern' for potential hazards has been classified as 'Moderate'." This indicates risk management was performed and reviewed as part of establishing safety.

2. Sample Size Used for the Test Set and Data Provenance

The provided text does not specify a sample size for a test set (e.g., number of images, number of patients). It also does not mention the data provenance (e.g., country of origin, retrospective or prospective).

For a 510(k) submission, particularly one involving image processing for comparison only and not primary diagnosis, the "test set" might not be a clinical image dataset for performance metrics like sensitivity/specificity. Instead, testing might focus on technical validation, such as image quality assessment (e.g., preservation of detail, compression artifacts, adherence to DICOM standards) and functional testing against specified requirements. The document mentions "The performance of the software is also tested in accordance with Hologic's SOPs and testing procedures to demonstrate adequate performance," which implies internal validation.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

This information is not provided in the given text. Since the device is for "comparison purposes only" and "cannot be used for primary diagnosis," a ground truth established by experts for diagnostic performance (e.g., disease presence/absence) would likely not be the primary focus of the performance testing presented in this summary. If image quality was assessed by experts, their qualifications are not mentioned.

4. Adjudication Method for the Test Set

This information is not provided in the given text.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

An MRMC study is not mentioned in the provided text. Given the device's intended use ("for comparison purposes only" and "cannot be used for primary diagnosis"), a comparative effectiveness study showing human readers' improvement with AI assistance would not be required or relevant for this specific 510(k) pathway.

6. Standalone (Algorithm Only) Performance Study

The text does not explicitly describe a standalone performance study in terms of specific clinical metrics like sensitivity or specificity. The "performance" mentioned likely refers to technical performance, such as:

  • Ability to process digitized screen-film mammograms.
  • Production of lossy-compressed DICOM images.
  • Resemblance to digital mammography images (qualitative assessment).

The submission implies internal testing ("The performance of the software is also tested in accordance with Hologic's SOPs and testing procedures to demonstrate adequate performance"), which would include standalone functional and technical validation.

7. Type of Ground Truth Used

The text does not specify the type of ground truth used. For this type of device, ground truth would likely relate to objective measurements of image fidelity to the original film, proper DICOM formatting, and the visual similarity ("more closely resemble digital mammography images") which could be a qualitative assessment by an internal team rather than a clinical ground truth like pathology or outcomes data.

8. Sample Size for the Training Set

The text does not provide any information regarding a training set sample size. Given the stated function ("process digitized screen-film mammograms to produce lossy-compressed DICOM images that more closely resemble digital mammography images"), it's possible the algorithms are based on image processing techniques that don't rely on a "training set" in the machine learning sense (i.e., for classification or detection). Instead, they might use fixed, rule-based, or adaptive algorithms for image manipulation. If machine learning was involved for the "resemblance" aspect, the training data is not discussed.

9. How the Ground Truth for the Training Set Was Established

This information is not provided in the given text, as no training set or its ground truth is mentioned.

{0}------------------------------------------------

Hologic, Inc.

5091368

DigitalNow™ HD 510(k) Submission

B.1 510(k) Summary of Safety & Effectiveness

(as required by 21 CFR § 807.92c)

Date Prepared: April 14, 2009

JUN - 4 2009

Submitted by:

Hologic, Inc. 35 Crosby Drive Bedford, MA 01730, USA

Name, Title and Phone Number of Contact:

Michaela Mahl Senior Regulatory Affairs Specialist Phone: (408) 745-5148 FAX: (408) 744-1905 Email: michaela.mahl(@hologic.com

Trade Name and Common Name:

Trade Name:R2 DigitalNow™ HD (DigitalNow™ HD)
Software Version:1.0
Common Name:Picture Archiving and Communications System

Classification:

Regulatory Class:II
Classification Panel:Radiology

Image Processing System 21 CFR § 892.2050 Product Code 90-LLZ

Predicate Devices:

The predicate device for DigitalNow HD software is certain software functions contained in the following device:

K080351, Apr 22, 2008

DexTop Mammography Workstation [Dexela Limited]

{1}------------------------------------------------

Device Description:

R2 DigitalNow HD is a software application intended to process digitized screen-film mammograms.

DigitalNow HD has been designed and will be manufactured in accordance with the following standards:

· ISO 13485Medical Devices - Quality Management Systems
· ISO 14971Medical Devices – Application of Risk Management to Medical Devices
· ANSI/AAMISW68:2001Medical Device Software – Software Life Cycle Processes
· 21 CFR § 820US Food and Drug Administration, Quality System Regulation
· N/AGeneral Principles of Software Validation; Final Guidance for Industry andFDA

The performance of the software is also tested in accordance with Hologic's SOPs and testing procedures to demonstrate adequate performance.

Intended Use:

R2 DigitalNow HD is a software application intended to process digitized screen-film mammograms for comparison purposes only. The software processes digitized prior film images to produce lossycompressed DICOM images that more closely resemble digital mammography images. R2 DigitalNow HD images are intended for comparison purposes only and cannot be used for primary diagnosis.

DigitalNow HD is a software application which runs on the Hologic Cenova server (Class I exempt per 21 CFR § 892.2010 and 21 CFR § 892.2020).

Technological Characteristics:

DigitalNow HD is a software application intended to process digitized screen-film mammograms. The device does not contact the patient, nor does it control any life-sustaining devices.

General Safety and Effectiveness Concerns:

The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of this device. Risk management is ensured via a risk analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing.

Conclusion:

The 510(k) Pre-Market Notification for DigitalNow HD contains adequate information and data to enable FDA - CDRH to determine substantial equivalence to the predicate device. The submission contains the results of a hazard analysis and the "Level of Concern" for potential hazards has been classified as "Moderate".

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular pattern around the symbol. The text is in all caps and is evenly spaced around the circle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUN - 4 2009

Hologic, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K091368

Trade/Device Name: R2 DigitalNowTM HD (DigitalNow™ HD) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system

Regulatory Class: II Product Code: LLZ Dated: May 23, 2009 Received: May 26, 2009

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

21 CFR 876.xxx(Gastroenterology/Renal/Urology)(240) 276-0115
21 CFR 884.xxx(Obstetrics/Gynecology)(240) 276-0115
21 CFR 892.xxx(Radiology)(240) 276-0120
Other(240) 276-0100

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please contact the CDRH/Office of Surveillance and Biometrics/Division of Postmarket Surveillance at 240-276-3464. For more information regarding the reporting of adverse events, please go to http://www.fda.gov/cdrh/mdr/.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

Page 2

{4}------------------------------------------------

B.2 Indication(s) for Use Statement

510(k) Number (if known):

K09/368

Device Name:

R2 DigitalNow™ HD (DigitalNow™ HD)

Indications for Use:

R2 DigitalNow HD is a software application intended to process digitized screen-film mammograms for comparison purposes only. The software processes digitized prior film images to produce lossy-compressed DICOM images that more closely resemble digital mammography images. R2 DigitalNow HD images are intended for comparison purposes only and cannot be used for primary diagnosis.

DigitalNow HD is a software application which runs on the Hologic Cenova server (Class I exempt per 21 CFR § 892.2010 and 21 CFR § 892.2020). 1

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED),

Concurrence of CDRH, Office of Device Evaluation (ODE)

Hele Lesan

(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Device 510(k) Number

§ 892.2050 Medical image management and processing system.

(a)
Identification. A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.(b)
Classification. Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).