K Number
K091310

Validate with FDA (Live)

Manufacturer
Date Cleared
2010-08-11

(464 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ablation Generator System and Ablation Instrument are indicated for palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.

Device Description

Ablation Generator System is a microprocessor controlled radiofrequency (RF) generator designed for use with the hand held, bipolar, sterile, single use Ablation Instrument for percutaneous delivery of low power bi-polar RF energy by the physician to the tissue site in contact with the Electrode of the Ablation Instrument.

The Ablation Instrument is a hand held, bi-polar, sterile, single use device designed specifically for use with the Ablation Generator System. The user connects the Ablation Instrument to the Ablation Generator System using the AE Cable (for delivery of RF energy). The user positions the distal tip of the device containing the active and return electrodes at the tissue site where the specific treatment is to be performed.

AI/ML Overview

The provided text describes a 510(k) summary for the Ablation Generator System and Ablation Instrument. It focuses on demonstrating substantial equivalence to predicate devices rather than establishing de novo performance criteria against specific numerical acceptance values. Therefore, much of the requested information regarding acceptance criteria and performance metrics (e.g., effect size, specific accuracy metrics, training set details) is not explicitly present in the provided document.

Here's an analysis based on the available information:

Acceptance Criteria and Study Details

1. Table of Acceptance Criteria and Reported Device Performance

This information is not explicitly provided in the given text.
The filing relies on showing substantial equivalence to predicate devices. The "Performance" section within the "Test Results" states: "Results of in-vitro and in-vivo testing (performance, mechanical, electrical, software) demonstrate that the Ablation Generator System and the Ablation Instrument function as intended and the results of tests were as expected." This indicates that the device met internal performance benchmarks, but these specific benchmarks and their numerical acceptance criteria are not detailed.

2. Sample Size Used for the Test Set and Data Provenance

This information is not explicitly provided.
The text mentions "in-vitro and in-vivo testing" but does not specify the sample sizes (e.g., number of patients, animals, or in-vitro samples) used for these tests, nor the provenance of any data (e.g., country of origin, retrospective or prospective nature).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not applicable/provided.
The submission does not describe a study involving expert-established ground truth in the way one would for diagnostic or AI-assisted devices. The "testing" appears to be focused on functional performance, mechanical safety, electrical safety, and software functionality.

4. Adjudication Method for the Test Set

This information is not applicable/provided.
As there's no mention of expert consensus-based ground truth establishment, no adjudication method would have been used in this context.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, a multi-reader multi-case (MRMC) comparative effectiveness study was not conducted or described.
This type of study is typically performed to evaluate the impact of a device on human reader performance, often in AI-assisted diagnostic tools. The provided information relates to an ablation device, where the focus is on its functional safety and efficacy in delivering RF energy for palliative treatment.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

This information is not applicable/provided.
The device described (Ablation Generator System and Ablation Instrument) is a physical medical device for delivering RF energy. It does not appear to involve an "algorithm" in the sense of a diagnostic or image analysis algorithm that would have standalone performance. Its "performance" refers to its mechanical, electrical, and functional operation.

7. The Type of Ground Truth Used

The "ground truth" for this device's evaluation appears to be based on engineering specifications, safety standards (e.g., biocompatibility testing against ISO 10993-1), and expected functional outcomes during in-vitro and in-vivo performance tests. It's not about establishing a diagnostic "ground truth" for a condition.

8. The Sample Size for the Training Set

This information is not applicable/provided.
The device is not described as having a "training set" in the context of machine learning or AI. Performance testing was conducted on the physical device.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/provided.
As there is no mention of a training set, the method for establishing its ground truth is not relevant.


Summary of Device Evaluation in the Provided Document:

The 510(k) submission for the Ablation Generator System and Ablation Instrument primarily relies on demonstrating substantial equivalence to existing predicate devices (ArthroCare Cavity SpineWand and ArthroCare Orthopedic Electrosurgery System). The "Test Results" section indicates that:

  • Performance: In-vitro and in-vivo tests showed the device "function[s] as intended and the results of tests were as expected." This implies the device met its internal design specifications and safety requirements.
  • Biocompatibility: The materials met ISO 10993-1 requirements.

The core argument for clearance is that "The minor technological differences between the subject devices... and their predicate devices... raise no new issues of safety or effectiveness. Performance data demonstrate that Ablation Generator System and the Ablation Instrument are as safe and effective as the predicate devices for their intended use."

Therefore, the "acceptance criteria" were implicitly tied to meeting internal design specifications, safety standards, and demonstrating performance comparable to the predicate devices, rather than a specific numerical performance metric in a clinical study against a defined ground truth.

{0}------------------------------------------------

510(K) SUMMARY 3.

This summary of 510(k)-safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 878.4400.

Date Prepared: 07/19/10

510(k) number:K091310
-------------------------

Applicant Information:

DFine Inc. 3047 Orchard Parkway San Jose, CA 95134, USA AUG 1 1 2010

Contact Person:

Sandeep Saboo
Vice-President, Quality Assurance and Regulatory Affairs
Phone Number:408-321-9999 ext 248
Fax Number:408-321-9401
e-mail:sandeeps@dfineinc.com

Device Information:

Trade Name:Ablation Generator System, and Ablation Instrument
Classification:Class II (for both, Ablation Generator System and Ablation Instrument)
Classification Name:Electrosurgical Cutting and Coagulation and Accessories
Product Code:GEI
Regulation No.:21CFR 888.4400 878.4400

Physical Description:

Ablation Generator System is a microprocessor controlled radiofrequency (RF) generator designed for use with the hand held, bipolar, sterile, single use Ablation Instrument for percutaneous delivery of low power bi-polar RF energy by the physician to the tissue site in contact with the Electrode of the Ablation Instrument.

The Ablation Instrument is a hand held, bi-polar, sterile, single use device designed specifically for use with the Ablation Generator System. The user connects the Ablation Instrument to the Ablation Generator System using the AE Cable (for delivery of RF energy). The user positions the distal tip of the device containing the active and return electrodes at the tissue site where the specific treatment is to be performed.

Indications For Use:

The Ablation Generator System and Ablation Instrument are indicated for palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.

{1}------------------------------------------------

Test Results:

Performance

Results of in-vitro and in-vivo testing (performance, mechanical, electrical, software) demonstrate that the Ablation Generator System and the Ablation Instrument function as intended and the results of tests were as expected.

Biocompatibility

The materials used in the Ablation Generator System and Ablation Instrument meet the requirements of ISO 10993-1.

Substantial Equivalence and Summary:

The subject device:

  • · Ablation Instrument has the same technological characteristics and principles of operation as its predicate, ArthroCare Cavity SpineWand [K063172, manufactured by ArthroCare Corporation]. The intended use and indications for the Ablation Instrument is a subset of that for its predicate device.
  • · Ablation Generator System (used only with the Ablation Instrument) has the same technological characteristics and principles of operation as its predicate, ArthroCare Orthopedic Electrosurgery System [K992581] when used with ArthroCare Cavity SpineWand [K063172, manufactured by ArthroCare Corporation]. The intended use and indications for the Ablation Generator is a subset of that for its predicate device.

The minor technological differences between the subject devices (Ablation Generator System and Ablation Instrument) and their predicate devices (ArthroCare Orthopedic Electrosurgery System and ArthroCare Cavity SpineWand) raise no new issues of safety or effectiveness. Performance data demonstrate that Ablation Generator System and the Ablation Instrument are as safe and effective as the predicate devices for their intended use. Thus, the Ablation Generator System and Ablation Instrument are substantially equivalent to their identified predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal is circular and features the department's name around the perimeter. In the center of the seal is a stylized emblem that resembles an eagle or a bird in flight, composed of three curved lines above a wavy line.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

DFine, Inc. % Mr. Sandeep Saboo Vice President, Quality Assurance and Regulatory Affairs 3047 Orchard Parkway San Jose, California 95134

AUG 1 1 2010

Re: K091310

Trade/Device Name: Ablation Instrument and Ablation Generator System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 19, 2010 Received: July 20, 2010

Dear Mr. Saboo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indice includions for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Dr w and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract libility warranties. We remind you; however, that device labeling must be truthful and not misleding,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Sandeep Saboo

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm1118809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance Alexanding to note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Pred. 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DFine, Inc.

K091310

INDICATIONS FOR USE STATEMENT 2.

AUG 1 1 2010

510(k) Number (if known):K091310
Device Name:Ablation Instrument and Ablation Generator System

Indications for Use:

Ablation Instrument:

The Ablation Instrument is indicated for palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.

Ablation Generator System:

The Ablation Generator is indicated for palliative treatment in spinal procedures by ablation of metastatic malignant lesions in a vertebral body.

Prescription Use (Part 21 CFR 801 Subpart D)X
-------------------------------------------------

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 'of '

(Division Sign-Off)

Division of Surgical. Orthopedic, and Restorative Devices

510(k) Number K 091310

.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.