K Number
K091263

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-11-19

(203 days)

Product Code
Regulation Number
888.3560
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Zimmer Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting. The Zimmer Patient Specific Instruments System is to be used with Zimmer NEXGEN CR-FLEX fixed bearing and Zimmer NEXGEN LPS-FLEX fixed bearing prostheses families only.

The Zimmer Patient Specific Instruments are intended for single use only.

Device Description

The Zimmer Patient Specific Instruments System consists of a software component, Zimmer Patient Specific Instruments Planner and a hardware component, Zimmer Patient Specific Instruments and is designed to assist the surgeon in the placement of total knee replacement components for Zimmer NEXGEN CR-FLEX fixed bearing and Zimmer NEXGEN LPS-FLEX fixed bearing prostheses families.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the Zimmer Patient Specific Instruments System (K091263):

1. Table of Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state specific acceptance criteria in a quantitative or tabular format for the device's performance. Instead, it makes a general statement about the non-clinical tests.

Acceptance Criteria (Not explicitly stated in the document)Reported Device Performance
(Presumed: Sufficient accuracy and performance for intended use)"Testing verified that the accuracy and performance of the system is adequate to perform as intended."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size for the Test Set: Not specified in the provided document.
  • Data Provenance: Not specified in the provided document (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Experts

This information is not provided in the document as a "test set" and "ground truth" for the non-clinical tests are not detailed in this submission summary.

4. Adjudication Method for the Test Set

This information is not provided in the document.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Was an MRMC study done? No.
  • Effect size of human readers improvement with AI vs. without AI assistance: Not applicable, as no clinical data or MRMC study was performed or reported in this summary.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

The document mentions "The Zimmer Patient Specific Instruments Planner (software component)" which generates a pre-surgical plan. This plan is "then used pre-operatively by a qualified surgeon to inspect, fine-tune and approve the pre-surgical plan." This indicates that the system is designed with a human-in-the-loop for approval, so a purely standalone (algorithm-only) performance without human review is not explicitly stated or implied as the basis for evaluation.

7. Type of Ground Truth Used

The document states "Non-clinical tests were performed to assess the safety and effectiveness of the device. Testing verified that the accuracy and performance of the system is adequate to perform as intended."

Given the nature of the device (surgical planning and instrumentation), the ground truth for these non-clinical tests would likely involve:

  • Benchtop measurements: Comparing the planned cuts/drilling locations to actual surgical guide dimensions and accuracy.
  • Engineering specifications/tolerances: Conformance to predefined precision and accuracy requirements for the guides.
  • Simulated surgical scenarios: While not a clinical study, these might involve phantom models or cadavers where known anatomical landmarks could serve as a "truth."

However, the specific "type of ground truth" (e.g., expert consensus, pathology, outcomes data) is not detailed in the provided 510(k) summary for the non-clinical tests.

8. Sample Size for the Training Set

The document does not provide any information regarding a training set or its sample size. This type of detail is often associated with machine learning or AI algorithm development, which is not extensively detailed for this device in this submission.

9. How the Ground Truth for the Training Set Was Established

Since no training set information is provided, how its ground truth was established is also not applicable/not provided.

{0}------------------------------------------------

K091263
P 1/2

11,

510(k) Summary

..

NOV 1 9 2009

Submitter information

Company nameMaterialise N.V.
Establishment registration number3003998208
Street AddressTechnologielaan 15
CityLeuven
Postal code3001
CountryBelgium
Phone number+32 16 39 66 75
Fax number+32 16 39 66 06
Contact nameKarl vom Berge
Contact titleProduct Manager
Contact e-mail addresskarl.vomberge@materialise.be

Submission date

The date of the Traditional 510(k) submission is April 29th, 2009. The date of the Additional Information request submission is August 4th 2009.

Submission information

Trade NameZimmer Patient Specific InstrumentsZimmer Patient Specific Instruments Planner
Common NameKnee prosthesis
Classification NameKnee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesisKnee joint patellofemorotibial.metal/polymer porous-coated uncemented prosthesis
Product codeJWH (21 CFR § 888.3560) and MBH (21 CFR § 888.3565)

Predicate device

Predicate Device
Trade or proprietary or modelnamePatient Matched Cutting Blocks
510(k) numberK082358
Decision date11/25/2008
Product codeJWH, MBH
ManufacturerSmith and Nephew, Inc.

Device Information

Description of the device

The Zimmer Patient Specific Instruments System consists of a software component, Zimmer Patient Specific Instruments Planner and a hardware component, Zimmer Patient Specific Instruments and is designed to assist the surgeon in the placement of total knee

. 4

{1}------------------------------------------------

replacement components for Zimmer NEXGEN CR-FLEX fixed bearing and Zimmer NEXGEN LPS-FLEX fixed bearing prostheses families.

Functioning of the device

The Zimmer Patient Specific Instruments System generates a pre-surgical plan based on MRI imaging data using the Zimmer Patient Specific Instruments Planner (software component). The software is then used pre-operatively by a qualified surgeon to inspect, fine-tune and approve the pre-surgical plan. Next, Zimmer Patient Specific Instruments are designed and manufactured based on the approved pre-surgical plan. Zimmer Patient Specific Instruments are patient specific templates that transfer the pre-operatively determined positioning of the Total Knee Replacement components to the patient intra-operatively, assisting the surgeon in positioning and aligning the actual Total Knee Replacement components by guiding and marking drill locations.

Intended use

The Zimmer Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting. The Zimmer Patient Specific Instruments System is to be used with Zimmer NEXGEN CR-FLEX fixed bearing and Zimmer NEXGEN LPS-FLEX fixed bearing prostheses families only.

The Zimmer Patient Specific Instruments are intended for single use only.

Summary of technological characteristics

Device comparison showed that the proposed device is substantially equivalent in intended use, materials and performance characteristics to the predicate device.

Performance data

Non-clinical tests

Non-clinical tests were performed to assess the safety and effectiveness of the device. Testing verified that the accuracy and performance of the system is adequate to perform as intended.

Clinical data

Not applicable.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR 1 5 2011

Materialise N.V. % Mr. Karl vom Berge Technologielaan 15 3001 Leuven Belgium

Re: K091263

Trade/Device Name: Zimmer Patient Specific Instruments System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: OOG, JWH Dated: October 22, 2009 Received: October 26, 2009

Dear Mr. vom Berge:

This letter corrects our substantially equivalent letter of November 19, 2009.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Karl vom Berge

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Hy B. Rake

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K091263

Device Name: Zimmer Patient Specific Instruments System (Zimmer Patient Specific Instruments, Zimmer Patient Specific Instruments Planner)

Indications For Use:

The Zimmer Patient Specific Instruments System is intended to be used as a surgical instrument to assist in the positioning of Total Knee Replacement components intraoperatively and in guiding the marking of bone before cutting. The Zimmer Patient Specific Instruments System is to be used with Zimmer NEXGEN CR-FLEX fixed bearing and Zimmer NEXGEN LPS-FLEX fixed bearing prostheses families only.

The Zimmer Patient Specific Instruments are intended for single use only.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nil R.P. Ozhe for mxm

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

Page 1 of 1

510(k) Number: K091263

§ 888.3560 Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis.

(a)
Identification. A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).(b)
Classification. Class II.