K Number
K091241

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-11-06

(193 days)

Product Code
Regulation Number
870.2700
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RR). The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.

Device Description

The Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories (Rad 87) include the MX board with Masimo Rainbow SET technology. The Rad 87 provides noninvasive monitoring of arterial oxygen saturation (%SpO₂), pulse rate, carboxyhemoglobin saturation (%SpCO), methemoglobin saturation (%SpMet), total hemoglobin concentration (g/dl SpHb), and/or respiratory rate (RR). Other information displayed by the Rad 87 includes: Low Signal IQ (Low SIQ), Perfusion Index (PI), Pleth Variability Index (PVI), Total Arterial Oxygen Content (SpOC), Respiratory Signal Quality (RSQ), alarm status, alarm silence, battery life, sensor status, and trends. The Rad 87 has output interfaces include: Nurse Call analog output, and RS-232 serial output. The Rad 87 in this filing are the Rad-87 in the K080238 filing, but with the addition of respiratory rate monitoring.

AI/ML Overview

The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and Accessories are indicated for the continuous noninvasive monitoring of several physiological parameters. The acceptance criteria and device performance are outlined below, along with details of the supporting studies.

1. Table of Acceptance Criteria and Reported Device Performance:

FeatureAcceptance Criteria (Specifications)Reported Device Performance (Accuracy)
SpO2 (Oxygen Saturation)
No MotionDisplay Range: 0-100%60-80% ± 3% (adults/pediatrics/infants)
Motion70-100% ± 2% (adults/pediatrics/infants); ± 3% (neonates)
Low Perfusion70-100% ± 2% (adults/pediatrics/infants/neonates)
Pulse Rate
No MotionDisplay Range: 25-240 bpm25-240 bpm ± 3 bpm (adults/pediatrics/infants/neonates)
Motion25-240 bpm ± 5 bpm (adults/pediatrics/infants/neonates)
Low Perfusion25-240 bpm ± 3 bpm (adults/pediatrics/infants/neonates)
SpCO (Carboxyhemoglobin Saturation)Display Range: 0-99%1-40% ± 3% (adults/pediatrics/infants)
SpMet (Methemoglobin Saturation)Display Range: 0-99.9%1-15% ± 1% (adults/pediatrics/infants/neonates)
SpHb (Total Hemoglobin)Display Range: 0-25 g/dL8-17 g/dl ± 1 g/dl (arterial or venous, adults/pediatrics/infants). (Accuracy not validated with motion or low perfusion.)
RR (Respiratory Rate)Display Range: 0-150 breaths per minute4-70 breaths per minute ± 1 breath per minute (adults > 30kg)

2. Sample Size Used for the Test Set and Data Provenance:

The document provides specific details for different parameters:

  • SpO2, SpCO, SpMet (Adults): "healthy adult volunteers" were tested.
  • SpO2, SpMet (Neonates): 16 neonatal NICU patients (age 7-135 days, weighing 0.5-4.25 kg). 79 data samples were collected.
  • SpO2 (No Motion): "healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies."
  • SpO2 (Motion): "healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and tapping motions."
  • SpO2 (Low Perfusion): Bench top testing using a Biotek Index 2 simulator and Masimo's simulator.
  • Pulse Rate: Bench top testing using a Biotek Index 2 simulator.
  • SpHb: "healthy adult male and female volunteers and on surgical patients with light to dark skin pigmentation."

Data Provenance:
The studies appear to be prospective human blood studies on healthy volunteers and surgical patients for SpO2, SpCO, SpMet, and SpHb. Neonatal data was also collected prospectively in a NICU setting. Benchtop testing was used for low perfusion SpO2 and Pulse Rate.

Country of Origin: The document does not explicitly state the country of origin for the human volunteer studies; however, the submission is from Masimo Corporation based in Irvine, CA, USA, and approved by the FDA in the USA, suggesting the studies were likely conducted in the US or under US regulatory standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

The document does not specify the number of experts used or their qualifications for establishing ground truth. The ground truth for the physiological parameters was established using:

  • A laboratory CO-Oximeter for SpO2, SpCO, SpMet, and SpHb.
  • An ECG monitor for SpO2 (no motion and motion studies).
  • A Biotek Index 2 simulator and Masimo's simulator for low perfusion SpO2 and Pulse Rate.

4. Adjudication Method for the Test Set:

The document does not describe any adjudication method involving human experts for the test set. Ground truth was established by reference devices (laboratory CO-Oximeter, ECG monitor) or simulators.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

No MRMC comparative effectiveness study is mentioned in the provided text. The device is a direct physiological measurement device, not an image interpretation or diagnostic aid that typically involves human readers. Therefore, the concept of human readers improving with AI vs without AI assistance does not directly apply here.

6. Standalone (Algorithm Only) Performance Study:

Yes, the accuracy statements are for the device's standalone performance (algorithm only without human-in-the-loop performance). The accuracies are reported as deviations from concurrently measured reference values (e.g., laboratory CO-Oximeter, ECG monitor, or simulators). For example:

  • "SpO2, SpCO and SpMet accuracy was determined by testing on healthy adult volunteers... against a laboratory CO-Oximeter."
  • "The Masimo sensors have been validated for no motion accuracy... against a laboratory CO-Oximeter and ECG monitor."

7. Type of Ground Truth Used:

The ground truth used for the accuracy studies consists of:

  • Reference device measurements: A laboratory CO-Oximeter and an ECG monitor.
  • Simulator data: Biotek Index 2 simulator and Masimo's simulator for specific conditions (low perfusion SpO2 and pulse rate).

8. Sample Size for the Training Set:

The document does not provide information regarding the sample size for any training sets. Given the nature of a pulse CO-Oximeter which uses established physiological principles and signal processing, it's possible that the "training" (development and refinement) of the algorithms was not conducted in the same manner as for AI/machine learning algorithms that typically require large, distinct training datasets. It likely involved extensive engineering, signal processing, and iterative testing rather than a formalized "training set" in the context of deep learning.

9. How the Ground Truth for the Training Set Was Established:

As no specific training set is described, there is no information on how ground truth for such a set would have been established. Any internal development and testing would have likely relied on similar reference methods as described for the performance studies.

{0}------------------------------------------------

KO91241

Submitted by:Masimo Corporation40 ParkerIrvine, CA 92618Phone: (949) 297-7000FAX: (949) 297-7001NOV - 6 2009
Contact:Marguerite Thomlinson, Manager of Regulatory Affairs
Date Summary Prepared:August 14, 2009
Trade NameMasimo Rainbow SET® Rad 87 CO-Oximeter and Accessories
Common NamePulse Oximeter and Sensor
Regulation Number:21 CFR 870.2700
Regulation Name:Oximeter
Regulation Class:Class II
Product CodeDQA, BZQ, DPZ, JKS
Substantially Equivalent DevicesMasimo Rainbow SET® Rad 87 Pulse CO-Oximeters and Accessories, 510(k) Number – K080238Oridion Capnography Inc., Capnostream 20 with IntegratedPulmonary Index, 510(k) Number – K082268

Andromed Inc., Biological Sound Monitor (BSM) Sensor 510(k) Number - K021389

Description of the Device

The Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories (Rad 87) include the MX board with Masimo Rainbow SET technology.

The Rad 87 provides noninvasive monitoring of arterial oxygen saturation (%SpO₂), pulse rate, carboxyhemoglobin saturation (%SpCO), methemoglobin saturation (%SpMet), total hemoglobin concentration (g/dl SpHb), and/or respiratory rate (RR). Other information displayed by the Rad 87 includes: Low Signal IQ (Low SIQ), Perfusion Index (PI), Pleth Variability Index (PVI), Total Arterial Oxygen Content (SpOC), Respiratory Signal Quality (RSQ), alarm status, alarm silence, battery life, sensor status, and trends. The Rad 87 has output interfaces include: Nurse Call analog output, and RS-232 serial output.

The Rad 87 in this filing are the Rad-87 in the K080238 filing, but with the addition of respiratory rate monitoring.

Intended Use/Indications for Use

The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO₂), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RR). The Masimo Rainbow SET® Rad 87 Pulse CO-

{1}------------------------------------------------

Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.

Principles of Operation

SpO2 and Pulse Rate

Pulse oximetry is governed by the principles that oxyhemoglobin (oxygenated blood), deoxyhemoglobin (non-oxygenated blood), carboxyhemoglobin (blood with carbon monoxide content), and methemoglobin (blood with oxidized hemoglobin content) species differ in their absorption of visible and infrared light. The amount of arterial blood in tissue changes with the pulse (photoplethysography). Therefore the amount of light, absorbed by the varying quantities of arterial blood, changes accordingly.

SpCO. SpMet, and SpHb General Description

The Rad-87 include the Masimo Rainbow SET technology for SpCO, SpMet and SpHb measurements, based on the same principles of pulse oximetry. The Masimo Rainbow SET technology uses a multiwavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, blood with oxidized hemoglobin and blood plasma.

Once the Masimo Rainbow SET technology receives the signal from the sensor, it calculates the patient's functional oxygen saturation (SpO2), fractional concentration of carboxyhemoglobin (SpCO), fractional concentration of methemoglobin (SpMet), total hemoglobin concentration (SpHb) and pulse rate.

Respiratory Rate General Description

The Masimo Rainbow SET technology also provides respiratory rate measurements, based on vibratory signals from respiratory sounds.

Method of Operation

SpO2, SpCO, SpMet, and SpHb

The instrument (Rad 87) is turned on. A sensor is attached to a patient's finger. The other end of the sensor is attached to a patient cable. The other end of the patient cable is connected to the Dual Channel cable or directly to the instrument. The connection of the patient cable to the Dual Channel cable is only needed when pulse CO-oximetry monitoring is concurrent with respiratory rate monitoring.

The instrument will begin continuously displaying the patient's pulse rate and SpO2 value. Depending on the type and/or configuration of the instrument, monitoring information would also include SpCO, SpMet, SpHb, PVI, and/or SpOC. The practitioner can then use the information to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner.

Once the practitioner determines the patient no longer requires monitoring, the cable is disconnected from the sensor, the sensor is removed (and disposed of if it is a single use device), and the power to the instrument is turned off.

Respiratory Rate

The instrument (Rad 87) is turned on. A sensor is attached to a patient's neck. The other end of the sensor is connected to a patient cable. The other end of the cable is connected to the Dual Channel cable. The Dual Channel cable is then connected to the instrument.

The instrument will begin continuously display the patient's respiratory rate. The practitioner can then use the information to help assess the condition of the patient and as an aide in determining if any intervention is required by the practitioner. Once the practitioner determines the patient no longer requires monitoring, the patient cable is disconnected from the sensor, the sensor is disposed and the power to the instrument is turned off.

{2}------------------------------------------------

Specifications

The specifications for the Rad 87 are the following:

FEATURESSPECIFICATIONS
Display RangesOxygen Saturation (SpO2): 0-100%Pulse Rate (beat per minute or bpm): 25-240 bpmCarboxyhemoglobin Saturation (SpCO): 0-99%Methemoglobin Saturation (SpMet): 0-99.9%Total Hemoglobin (SpHb): 0-25 g/dLRespiratory Rate (RR): 0-150 breaths per minuteTotal Oxygen Concentration (SpOC): 0-35 ml/dlPerfusion Index: 0.02-20%Pleth Variability Index: 0-100%
AccuracySee Footnotes 1, 2, 3, 4, 5, 6, and 7
SpO2, No Motion60-80 ± 3%, adults/pediatrics/infants
SpO2, Motion70-100 ± 2%, adults/pediatrics/infants; ± 3%, neonates
SpO2, Low Perfusion70-100 ± 2%, adults/pediatrics/infants/neonates
Pulse Rate, No Motion25-240 ± 3 bpm, adults/pediatrics/infants/neonates
Pulse Rate, Motion25-240 ± 5 bpm, adults/pediatrics/infants/neonates
Pulse Rate, Low Perfusion25-240 ± 3 bpm, adults/pediatrics/infants/neonates
SpCO1-40 ± 3%, adults/pediatrics/infants
SpMet1-15 ± 1%, adults/pediatrics/infants/neonates
SpHb8-17 ±1 g/dl (arterial or venous), adults/pediatrics/infants
RR4-70 ± 1 breath per minute, adults (> 30kg)
General
ResolutionSpO2: 1%Pulse Rate: 1 bpmSpCO: 1%SpMet: 0.1%SpHb: 0.1 g/dlRR: 1 breath per minute
MeasurementsLow Signal IQPerfusion Index (PI)Total Oxygen Concentration (SpOC)Pleth Variability Index (PVI)Respiratory Signal Quality (RSQ)
Electrical
Power (AC)Voltage Input Range: 100-240 VAC, 47-63 Hz
BatteriesRechargeable
CircuitryMicroprocessor controlledAutomatic self-test of pulse CO-oximeter when powered onAutomatic setting of default parametersAutomatic alarm messagesTrend data output
FirmwareRainbow SET technology, MX Board/Circuitry
Mechanical
MaterialPolycarbonate/ABS Blend
FEATURESSPECIFICATIONS
Environmental
Operating Temperature32 to 122°F (0 to 50°C)
Storage Temperature-40 to 158°F (-40 to 70°C)
Relative Storage Humidity10 to 95% non-condensing
Operating AltitudePressure: 500-1,060 mbarAltitude: -1,000-18,000 ft (-304-5,486m)
Mode & Sensitivity
SpO2 Averaging Mode2, 4, 6, 8, 10, 12 and 16 seconds; FastSat
SpO2 SensitivityAPOD, Normal, Maximum
Alarms
Volume Level Adjustment: Pulse/ToneOFF; 25% to 100% in 4 increments
Alarm Silence120 seconds delay; All mute: continuous silence
Out of Limit Alarms: SpO2, Pulse Rate, SpCO, SpMet, SpHb, RR, PI, PVIHigh/low alarms
Sensor Condition AlarmNo Sensor; Sensor Off; Sensor Defect
SystemSystem failure
Battery AlarmLow battery
Display and Indicators
Data DisplaySpO2 (%)Pulse rate (bpm)SpCO (%)SpMet (%)SpHb (g/dl)SpHbv(g/dl)RRSpOC(ml/dl)Perfusion Index-PI (%)Pleth Variability Index-PVI (%)Signal IQRespiratory Signal Quality (RSQ)Sensitivity indicatorSensor statusSensor timeStatus messagesAlarm statusBattery status
Output Interface
Analog outputNurse Call
Serial Port(RS-232 connector)PC/printer connectionPhilips VuelinkRadNetPatient Safety NetTrends
Compliance
EMC ComplianceEN 60601-1-2, Class B
Electrical SafetyIEC 60601-1, 2nd Edition; UL 60601-1
Radio802.11 a/b/g
Type of Protection (AC Power)Class 1
Type of Protection (battery power)Internally Powered
Degree of Protection-Patient CableType BF-Applied Part
Liquid Ingress-Degree of ProtectionIPX1
Mode of OperationContinuous

:

{3}------------------------------------------------

{4}------------------------------------------------

Footnotes

  • SpO2, SpCO and SpMet accuracy was determined by testing on healthy adult volunteers in the 1 range of 60-100% SpO2, 0-40% SpCO, and 0-15% SpMet against a laboratory CO-Oximeter. SpO2 and SpMet accuracy was determined on 16 neonatal NICU patients ranging in age from 7-135 days old and weighing between 0.5-4.25 kg. Seventy-nine (79) data samples were collected over a range of 70-100% SaO2 and 0.5-2.5% MetHb with a resultant accuracy of 2.9% SpO2 and 0.9% SpMet.
  • ನ The Masimo sensors have been validated for no motion accuracy in human blood studies on healthy adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-Oximeter and ECG monitor. This variation equals plus or minus one standard deviation. Plus or minus one standard deviation encompasses 68% of the population.
  • The Masimo sensors have been validated for motion accuracy in human blood studies on healthy ന adult male and female volunteers with light to dark skin pigmentation in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an amplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO- against a laboratory CO-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
  • The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing য against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population:
  • The Masimo sensors have been validated for pulse rate accuracy for the range of 25-240 bpm in ട് bench top testing against a Biotek Index 2 simulator. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
  • SpHb accuracy has been validated on healthy adult male and female volunteers and on surgical റ patients with light to dark skin pigmentation in the range of 8-17 g/dl SpHb against a laboratory CO-oximeter. This variation equals plus or minus one standard deviation which encompasses 68% of the population. The SpHb accuracy has not been validated with motion or low perfusion.
  • The following substances may interfere with pulse CO-oximetry measurements: 7
    • · Elevated levels of Methemoglobin (MetHb) may lead to inaccurate SpO2 and SpCO measurements
    • · Elevated levels of Carboxyhemoglobin (COHb) may lead to inaccurate SpO₂ measurements.
    • · Very low arterial Oxygen Saturation (SpO2) levels may cause inaccurate SpCO and SpMet measurements
    • · Severe anemia may cause erroneous SpO2 readings.
    • · Dyes, or any substance containing dyes, that change usual blood pigmentation may cause erroneous readings.
    • · Elevated levels of total bilirubin may lead to inaccurate SpO2, SpMet, SpCO and SpHb readings

{5}------------------------------------------------

Test Summary

The Rad 87 comply with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the Rad 87:

  • . Risk Analysis
  • Design Reviews .
  • Biocompatibility Testing .
  • . Performance Testing
  • . Safety Testing
  • Environmental Testing .
  • Clinical Testing .

Conclusions

The information in this 510(k) submission demonstrates that the Rad 87 are substantially equivalent to the predicate devices, with respect to safety, effectiveness, and performance.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo features the department's name arranged in a circular fashion around an emblem. The emblem is a stylized representation of an eagle.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-066-0609 Silver Spring, MD 20993-0002

Ms. Marguerite Thomlinson Manager of Regulatory Affairs Masimo Corporation 40 Parker Irvine, California 92618

M-M-M

Re: K091241

Trade/Device Name: Masimo Rainbow SET Rad 87 Pulse CO-Oximeter and Accessories Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA, BZQ,DPZ, JKS Dated: October 29, 2009 Received: October 30, 2009

NOV - 6 2009

Dear Ms. Thomlinson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{7}------------------------------------------------

Page 2- Ms. Thomlinson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to

http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours

Susan Rose

Susan Runner, D.D.S., M.A. Acting Division Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

Section 4 - Indications for Use

510(k) Number (if known):

Masimo Rainbow SET Rad 87 Pulse CO-Oximeter and Accessories Device Name:

Indications For Use:

The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), methemoglobin saturation (SpMet), total hemoglobin concentration (SpHb), and/or respiratory rate (RR). The Masimo Rainbow SET® Rad 87 Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (Per 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CD/RH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number: K091241

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

510(k) Number:

K091241-Rad 87 w/ ARRM Masimo Response, 8/14/09 Page 7 of 180

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).