K Number
K091159

Validate with FDA (Live)

Manufacturer
Date Cleared
2009-07-10

(80 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Synthes Oracle Plate is indicated for use, via the lateral or anterolateral surgical approach above the bifurcation of the great vessels, or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal stenosis, or failed previous fusion.

Device Description

The Synthes Oracle Plate is an addition to Synthes' existing anterior/anterolateral/lateral thoracolumbar spine systems. The Oracle Plate implant consists of a plate and 4 bone screws, all manufactured from Titanium Aluminum Niobium (Ti-6Al-7Nb) ASTM F1295, the same material as the predicates.

AI/ML Overview

The acceptance criteria and study proving the device meets them are described below:

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
Non-Clinical: Substantial equivalence to predicate devices based on design, function, material, and intended use.Bench testing results demonstrate that the Synthes Oracle Plate System is substantially equivalent to the predicate devices.
Clinical: Not required.Clinical data and conclusions were not needed for this device.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not applicable. The submission states "Clinical data and conclusions were not needed for this device." The non-clinical performance relied on bench testing, for which sample sizes are not provided in this summary.
  • Data Provenance: Not applicable. Clinical data was not required or provided. For non-clinical bench testing, the provenance of materials or components used in testing is not detailed, but it would typically be from the manufacturer's internal processes.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

Not applicable. Clinical ground truth was not established as clinical data was not required. For non-clinical bench testing, the "ground truth" would be engineering specifications and standards, verified by engineers or technicians. The number and qualifications of individuals involved in the bench testing are not provided.

4. Adjudication method for the test set:

Not applicable. No human adjudication method for a test set is mentioned, as clinical data was not required.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

Not applicable. This device is a spinal implant system, not an AI-assisted diagnostic or imaging device. Therefore, an MRMC study is not relevant, nor is the concept of human readers improving with AI assistance.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

Not applicable. This device is a physical spinal implant system and does not involve an algorithm or AI.

7. The type of ground truth used:

  • For Non-Clinical Performance: The ground truth for the non-clinical performance was based on engineering specifications, material properties (Titanium Aluminum Niobium (Ti-6Al-7Nb) ASTM F1295), and the functional performance criteria established for substantial equivalence to predicate devices. This is assessed via bench testing.
  • For Clinical Performance: Not applicable, as clinical data was not required.

8. The sample size for the training set:

Not applicable. This device is a physical spinal implant system and does not involve AI or machine learning models that require a training set.

9. How the ground truth for the training set was established:

Not applicable. As noted above, this device does not involve AI or machine learning and therefore does not have a "training set."

{0}------------------------------------------------

KO91159

JUL 1-0 2009

510(k) Summary

:

510(k) Summary
Name of Firm:Synthes Spine1302 Wrights Lane EastWest Chester, PA 19380
510(k) Contact:Stacey BonnellRegulatory Affairs SpecialistTelephone: 610-719-5895 Facsimile: 610-719-5102Email: bonnell.stacey@synthes.com
Date Prepared:April 17, 2009
Trade Name:Synthes Oracle Plate System
Classification:21 CFR 888.3060 - Spinal Intervertebral Body Fixation OrthosisClass IIOrthopaedic and Rehabilitation Devices PanelProduct Code: KWQ
Predicates:Synthes Oracle Plate System is substantially equivalent to similar previously cleared predicate devices.
Device Description:The Synthes Oracle Plate is an addition to Synthes' existing anterior/anterolateral/lateral thoracolumbar spine systems. The Oracle Plate implant consists of a plate and 4 bone screws, all manufactured from Titanium Aluminum Niobium (Ti-6Al-7Nb) ASTM F1295, the same material as the predicates.
Intended Use/ Indications for Use:The Synthes Oracle Plate is indicated for use via the lateral or anterolateral surgical approach above the bifurcation of the great vessels, or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (L1-S1) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal stenosis, or failed previous fusion.Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.
Comparison of the device to predicate device(s):The Synthes Oracle Plate System is a result of design modifications to the predicate devices. It is substantially equivalent to the predicates in design, function, material, and intended use.
Performance Date (Non-Clinical and/or Clinical):Non-Clinical Performance and Conclusions:Bench testing results demonstrate that the Synthes Oracle Plate System is substantially equivalent to the predicate devices.Clinical Performance and Conclusions:Clinical data and conclusions were not needed for this device.

ની જુ.
જિલ્લાના ગુજરાત lof l

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird in flight.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Synthes Spine % Ms. Stacey Bonnell Regulatory Affairs Specialist 1302 Wrights Lane East West Chester, Pennsylvania 19380

2009

Re: K091159

Trade/Device Name: Synthese Oracle Plate Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervetebral body fixation orthosis Regulatory Class: II Product Code: KWQ Dated: April 17, 2009 Received: April 21, 2009

Dear Ms. Bonnell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Stacey Bonnell

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrh/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbara Buellius

Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

@ SYNTHES® Spine

Indications for Use Statement 4

510(k) Number: (if known)

K 691159

Device Name:

Synthes Oracle Plate

The Synthes Oracle Plate is indicated for use, via the lateral or anterolateral surgical approach above the bifurcation of the great vessels, or via the anterior surgical approach, below the bifurcation of the great vessels as an adjunct to fusion. This system is indicated in the treatment of lumbar and lumbosacral (LI-SI) spine instability as a result of fracture (including dislocation and subluxation), tumor, degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by patient history and radiographic studies), pseudoarthrosis, spondylolysis, spondylolisthesis, scoliosis, lordotic deformities of the spine, spinal stenosis, or failed previous fusion.

Warning: This device is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic, or lumbar spine.

X Prescription Use (21 CFR 801 Subpart D) AND / OR

Over-the-Counter Use (21 CFR 801 Subpart C)

Page 7 of 24

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

E Z (EXT. For MXM)

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K091159

Traditional 510(k) - Synthes Oracle Plate System

N/A